Class 2 Device Recall Blue Rhino G2Multi Percutaneous Tracheostomy Introducer Tray

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Blue Rhino G2Multi Percutaneous Tracheostomy Introducer Traysee related information
Date Initiated by FirmMarch 05, 2026
Date PostedApril 06, 2026
Recall Status1 Open3, Classified
Recall NumberZ-1780-2026
Recall Event ID 98495
510(K)NumberK193133 
Product Classification Tube tracheostomy and tube cuff - Product Code JOH
ProductCOOK MEDICAL Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Tray: Reference Part Number C-PTISY-100-HC-G-NA, Order Number G57703; Reference Part Number C-PTISY-100-UNL-HC-G-NA, Order Number G57706; Reference Part Number C-PTISY-100-UNS-HC-G-NA, Order Number G57704.
Code Information Reference Part Number C-PTISY-100-HC-G-NA, UDI (01)00827002577039(17)270730(10)16183353X, Lot Number 16183353X; Reference Part Number C-PTISY-100-HC-G-NA, UDI (01)00827002577039(17)270524(10)16076948, Lot Number 16076948; Reference Part Number C-PTISY-100-HC-G-NA, UDI (01)00827002577039(17)260222(10)15473203X, Lot Number 15473203X; Reference Part Number C-PTISY-100-HC-G-NA, UDI (01)00827002577039(17)260201(10)15476984X, Lot Number 15476984X; Reference Part Number C-PTISY-100-HC-G-NA, UDI (01)00827002577039(17)261020(10)15719188X, Lot Number 15719188X; Reference Part Number C-PTISY-100-UNL-HC-G-NA, UDI (01)00827002577060(17)270814(10)16205108, Lot Number 16205108; Reference Part Number C-PTISY-100-UNS-HC-G-NA, UDI (01)00827002577046(17)270528(10)16080161X, Lot Number 16080161X; Reference Part Number C-PTISY-100-UNS-HC-G-NA, UDI (01)00827002577046(17)270901(10)16435112, Lot Number 16435112.
FEI Number 1820334
Recalling Firm/
Manufacturer
Cook Incorporated
750 N Daniels Way
Bloomington IN 47404-9120
For Additional Information ContactCook Medical Customer Relations Department
812-457-4500
Manufacturer Reason
for Recall
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
FDA Determined
Cause 2
Incorrect or no expiration date
ActionCook Medical issued an URGENT: MEDICAL DEVICE RECALL to its consignee on 3/5 2026 via courier or email. The notice explained the issue, potential risk to health, and requested the following: Actions to be Taken by the Customer 1. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all further distribution and use of the affected products. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter regardless of whether you have affected product(s) on hand. You may return the Form via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the affected devices have been transferred. 5. Please immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.
Quantity in Commerce441 units
DistributionWorldwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = JOH
-
-