| | Class 2 Device Recall Revolution Ascend |  |
| Date Initiated by Firm | March 26, 2026 |
| Date Posted | May 08, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2149-2026 |
| Recall Event ID |
98738 |
| 510(K)Number | K213938 |
| Product Classification |
System, x-ray, tomography, computed - Product Code JAK
|
| Product | Revolution Ascend, computed tomography, Model Numbers 6969000-100 and 6969000-300 |
| Code Information |
UDI-DI: 00840682146173;
Serial Numbers: AS10G2200136YC
AS10G2200101YC
AS10G2200068YC
AS10G2200035YC
AS10G2300093YC
AS10G2400094YC
AS10G2400079YC
AS10G2100008YC
AS10G2500002YC
CBDWG2400019HM
CBDWG2300081HM
AS10G2400056YC
AS10G2400081YC
AS10G2400087YC
CBDWG2400117HM
CBDWG2400085HM
CBDWG2500003HM
CBDWG2400140HM
CBDWG2400126HM
AS10G2200085YC
AS10G2300057YC
CBDWG2500011HM
CBDWG2300056HM
CBDWG2300041HM
AS10G2300041YC |
| FEI Number |
2126677
|
Recalling Firm/ Manufacturer |
GE Medical Systems, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
| For Additional Information Contact | GE HealthCare Service 800-437-1171 |
Manufacturer Reason for Recall | GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems. |
FDA Determined Cause 2 | Software design |
| Action | An URGENT MEDICAL DEVICE CORRECTION notification letter dated 3/26/26 was sent to customers.
Actions to be taken by Customer/User Pending corrections by GE HealthCare, you can continue to use your system by following security and cybersecurity best practices. Before the interim solution (see Product Correction section below) is applied, please follow the instructions below as identified in the respective CT system User Manuals: CAUTION: 3D or slab reconstructions provide additional supplemental information, complementing diagnosis that should be based on classical techniques. WARNING: Do not use 3D or slab views only to perform any measurements (distance, angle, region of interest, report cursor, area, volume, etc.). Always check measurement points position and refer to 2D baseline views (acquisition images or reformatted images of minimal thickness) to confirm measurements.
Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions.
Please retain this document for your records.
Please complete and return the attached acknowledgement form electronically via (https://gehealthcare-svc.my.site.com/publicForm/s/?formId=aGjUr000002jnD7) or print, fill out manually, scan and email to recall.23015@gehealthcare.com
Product Correction
GE HealthCare will correct all affected products at no cost to you in two steps. A GE HealthCare representative will contact you to arrange for an interim solution to deactivate AW Server authentication through the CT system on the EHL-based Smart Subscription, and later, to install a CT software update. Note: With the interim solution, all alternative authentication mechanisms (direct application launch from CT console user interface, login using users defined in AW Server user database, federation with hospital user directory through LDAP) will remain available.
Contact Information
If you have any questions or concerns regarding this notification, please contact GE HealthCare |
| Quantity in Commerce | 25 units |
| Distribution | Worldwide - US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = JAK
|
|
|
|