• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall Medline

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Medlinesee related information
Date Initiated by FirmApril 10, 2026
Date PostedMay 28, 2026
Recall Status1 Open3, Classified
Recall NumberZ-2237-2026
Recall Event ID 98837
Product Classification Orthopedic tray - Product Code OJH
ProductMedline kits containing Huons Co. Bupivacaine HCL in Dextrose Injection, USP, labeled as: 1) SPINAL ANESTHESIA TRAY-LF, Medline Kit SKU DYNJRA0102A; 2) SPINAL TRAY-LF, Medline Kit SKU DYNJRA0143B; 3) SPINAL PREP TRAY-LF, Medline Kit SKU DYNJRA0205B; 4) SPINAL TRAY, Medline Kit SKU DYNJRA0255B; 5) SPINAL TRAY, Medline Kit SKU DYNJRA0320; 6) SPINAL TRAY, Medline Kit SKU DYNJRA0320A; 7) SPINAL BLOCK TRAY- LF, Medline Kit SKU DYNJRA0413C; 8) SPINAL TRAY, Medline Kit SKU DYNJRA0461D; 9) 25G SPINAL BLOCK TRAY, Medline Kit SKU DYNJRA0945; 10) SPINAL TRAY, Medline Kit SKU DYNJRA1140B; 11) SPINAL TRAY, Medline Kit SKU DYNJRA1192; 12) SPINAL TRAY 25G, Medline Kit SKU DYNJRA1240; 13) TRAY, SPINAL PENCIL POINT 25G, Medline Kit SKU DYNJRA1288; 14) SPINAL 25G WHITACRE, Medline Kit SKU DYNJRA1301; 15) SPINAL BLOCK TRAY WHITACRE 24G, Medline Kit SKU DYNJRA1304; 16) AHN SPINAL TRAY, Medline Kit SKU DYNJRA1376; 17) SPINAL TRAY, Medline Kit SKU DYNJRA1477A; 18) ADULT SPINAL KIT, Medline Kit SKU DYNJRA1538; 19) SPINAL TRAY, Medline Kit SKU DYNJRA1558A; 20) SPINAL TRAY, Medline Kit SKU DYNJRA1560A; 21) SPINAL TRAY, Medline Kit SKU DYNJRA1565; 22) SPINAL ANESTHESIA TRAY, Medline Kit SKU DYNJRA1577C; 23) SPINAL NEEDLE TRAY, Medline Kit SKU DYNJRA1584; 24) 25G WHITACRE SPINAL TRAY, Medline Kit SKU DYNJRA1740A; 25) SPINAL TRAY, Medline Kit SKU DYNJRA1744A; 26) SPINAL TRAY, Medline Kit SKU DYNJRA1750A; 27) SPINAL TRAY, Medline Kit SKU DYNJRA1784; 28) SPINAL TRAY, Medline Kit SKU DYNJRA1790; 29) SPINAL TRAY, Medline Kit SKU DYNJRA1796; 30) MSSC SPINAL TRAY, Medline Kit SKU DYNJRA1802B; 31) SPINAL TRAY, Medline Kit SKU DYNJRA1818; 32) SPINAL TRAY, Medline Kit SKU DYNJRA1839; 33) SPINAL TRAY, Medline Kit SKU DYNJRA1884; 34) SPINAL TRAY, Medline Kit SKU DYNJRA1884A; 35) 24G SPINAL KIT, Medline Kit SKU DYNJRA1889; 36) SPINAL TRAY, Medline Kit SKU DYNJRA1929; 37) SPINAL TRAY, Medline Kit SKU DYNJRA1969; 38) SPINAL TRAY, Medline Kit SKU DYNJRA1982; 39) SPINAL TRAY/PENCAN, Medline Kit SKU DYNJRA1983A; 40) SPINAL NEEDLE TRAY, Medline Kit SKU DYNJRA1987; 41) SPINAL TRAY - 3 ML CHLORAPREP, Medline Kit SKU DYNJRA2005; 42) SPINAL TRAY, Medline Kit SKU DYNJRA2023; 43) SPINAL TRAY, Medline Kit SKU DYNJRA2068A; 44) SPINAL BLOCK TRAY, Medline Kit SKU DYNJRA2088; 45) 25G WHITACRE SPINAL TRAY, Medline Kit SKU DYNJRA2115; 46) SPINAL BLOCK 25G WHTCRE 5S BU, Medline Kit SKU DYNJRA2117; 47) SPINAL TRAY, Medline Kit SKU DYNJRA2140; 48) SPINAL TRAY, Medline Kit SKU DYNJRA2153; 49) SPINAL BLOCK TRAY WHITACRE 24G, Medline Kit SKU DYNJRA2213; 50) SPINAL TRAY, Medline Kit SKU DYNJRA2225; 51) OR SPINAL BLOCK KIT, Medline Kit SKU DYNJRA2349; 52) SPINAL TRAY, Medline Kit SKU DYNJRA2360; 53) L&D SPINAL TRAY, Medline Kit SKU DYNJRA2371; 54) TRAY, 25G WHIT SPINAL W/PHARM, Medline Kit SKU DYNJRA2381; 55) SPINAL TRAY, Medline Kit SKU DYNJRA2432A; 56) OB SPINAL TRAY, Medline Kit SKU DYNJRA2532; 57) SPINAL TRAY, Medline Kit SKU DYNJRA2534; 58) SPINAL TRAY, Medline Kit SKU DYNJRA2556; 59) SPINAL TRAY, Medline Kit SKU DYNJRA2566A; 60) SPINAL TRAY BRAUN, Medline Kit SKU DYNJRA2575; 61) SPINAL TRAY, Medline Kit SKU DYNJRA2592; 62) SPINAL TRAY, Medline Kit SKU DYNJRA2597; 63) SPINAL BLOCK, Medline Kit SKU DYNJRA2605; 64) SPINAL TRAY, Medline Kit SKU DYNJRA2606; 65) SPINAL TRAY, Medline Kit SKU DYNJRA2607; 66) SPINAL TRAY, Medline Kit SKU DYNJRA2607A; 67) SPINAL TRAY, Medline Kit SKU DYNJRA2611; 68) SPINAL BLOCK TRAY, Medline Kit SKU DYNJRA2613; 69) SPINAL TRAY, Medline Kit SKU DYNJRA2618A; 70) SPINAL TRAY, Medline Kit SKU DYNJRA2634; 71) SPINAL TRAY, Medline Kit SKU DYNJRA2638; 72) SPINAL TRAY, Medline Kit SKU DYNJRA2639; 73) JOI SPINAL BLOCK TRAY, Medline Kit SKU DYNJRA2641; 74) SPINAL TRAY, Medline Kit SKU DYNJRA2653; 75) SPINAL BLOCK 25G WHTCRE 5S BU, Medline Kit SKU DYNJRA2669; 76) SPINAL TRAY, Medline Kit SKU DYNJRA2670; 77) SPINAL BLOCK TRAY 18G INTRO, Medline Kit SKU DYNJRA2672; 78) SPINAL TRAY 24G, Medline Kit
Code Information Medline Kit SKU DYNJRA0102A: UDI/DI 10884389459003 each, 40884389459004 case, Lot Number 26ABG838; Medline Kit SKU DYNJRA0102A: UDI/DI 10884389459003 each, 40884389459004 case, Lot Number 25JBR222; Medline Kit SKU DYNJRA0102A: UDI/DI 10884389459003 each, 40884389459004 case, Lot Number 25IBJ307; Medline Kit SKU DYNJRA0102A: UDI/DI 10884389459003 each, 40884389459004 case, Lot Number 25EBR346; Medline Kit SKU DYNJRA0102A: UDI/DI 10884389459003 each, 40884389459004 case, Lot Number 24LDB145; Medline Kit SKU DYNJRA0102A: UDI/DI 10884389459003 each, 40884389459004 case, Lot Number 24FDB896; Medline Kit SKU DYNJRA0143B: UDI/DI 10195327125820 each, 40195327125821 case, Lot Number 24ABQ670; Medline Kit SKU DYNJRA0205B: UDI/DI 10195327472344 each, 40195327472345 case, Lot Number 24KMD212; Medline Kit SKU DYNJRA0255B: UDI/DI 10195327282042 each, 40195327282043 case, Lot Number 26CMG585; Medline Kit SKU DYNJRA0255B: UDI/DI 10195327282042 each, 40195327282043 case, Lot Number 26BMK211; Medline Kit SKU DYNJRA0255B: UDI/DI 10195327282042 each, 40195327282043 case, Lot Number 25JMC385; Medline Kit SKU DYNJRA0255B: UDI/DI 10195327282042 each, 40195327282043 case, Lot Number 25IMB105; Medline Kit SKU DYNJRA0255B: UDI/DI 10195327282042 each, 40195327282043 case, Lot Number 25EMC687; Medline Kit SKU DYNJRA0255B: UDI/DI 10195327282042 each, 40195327282043 case, Lot Number 24HMD675; Medline Kit SKU DYNJRA0255B: UDI/DI 10195327282042 each, 40195327282043 case, Lot Number 24HMB249; Medline Kit SKU DYNJRA0255B: UDI/DI 10195327282042 each, 40195327282043 case, Lot Number 24GMH436; Medline Kit SKU DYNJRA0320: UDI/DI 10884389656778 each, 40884389656779 case, Lot Number 24EDB314; Medline Kit SKU DYNJRA0320: UDI/DI 10884389656778 each, 40884389656779 case, Lot Number 24BDA550; Medline Kit SKU DYNJRA0320A: UDI/DI 10198459027796 each, 40198459027797 case, Lot Number 26ABJ309; Medline Kit SKU DYNJRA0320A: UDI/DI 10198459027796 each, 40198459027797 case, Lot Number 25HBT750; Medline Kit SKU DYNJRA0320A: UDI/DI 10198459027796 each, 40198459027797 case, Lot Number 25GBW261; Medline Kit SKU DYNJRA0320A: UDI/DI 10198459027796 each, 40198459027797 case, Lot Number 25FBP025; Medline Kit SKU DYNJRA0413C: UDI/DI 10195327253363 each, 40195327253364 case, Lot Number 26AMK323; Medline Kit SKU DYNJRA0413C: UDI/DI 10195327253363 each, 40195327253364 case, Lot Number 25HMB655; Medline Kit SKU DYNJRA0413C: UDI/DI 10195327253363 each, 40195327253364 case, Lot Number 25FMF763; Medline Kit SKU DYNJRA0413C: UDI/DI 10195327253363 each, 40195327253364 case, Lot Number 24KMF380; Medline Kit SKU DYNJRA0413C: UDI/DI 10195327253363 each, 40195327253364 case, Lot Number 24KMA760; Medline Kit SKU DYNJRA0413C: UDI/DI 10195327253363 each, 40195327253364 case, Lot Number 23IBH412; Medline Kit SKU DYNJRA0461D: UDI/DI 10195327120689 each, 40195327120680 case, Lot Number 24KME483; Medline Kit SKU DYNJRA0461D: UDI/DI 10195327120689 each, 40195327120680 case, Lot Number 23KMD757; Medline Kit SKU DYNJRA0945: UDI/DI 10889942598267 each, 40889942598268 case, Lot Number 26BBP431; Medline Kit SKU DYNJRA0945: UDI/DI 10889942598267 each, 40889942598268 case, Lot Number 25KBN750; Medline Kit SKU DYNJRA0945: UDI/DI 10889942598267 each, 40889942598268 case, Lot Number 25IBK141; Medline Kit SKU DYNJRA0945: UDI/DI 10889942598267 each, 40889942598268 case, Lot Number 25EBG343; Medline Kit SKU DYNJRA0945: UDI/DI 10889942598267 each, 40889942598268 case, Lot Number 25ADA288; Medline Kit SKU DYNJRA0945: UDI/DI 10889942598267 each, 40889942598268 case, Lot Number 24FDA636; Medline Kit SKU DYNJRA1140B: UDI/DI 10193489432015 each, 40193489432016 case, Lot Number 26BMI341; Medline Kit SKU DYNJRA1140B: UDI/DI 10193489432015 each, 40193489432016 case, Lot Number 25LMF246; Medline Kit SKU DYNJRA1140B: UDI/DI 10193489432015 each, 40193489432016 case, Lot Number 25JMA359; Medline Kit SKU DYNJRA1140B: UDI/DI 10193489432015 each, 40193489432016 case, Lot Number 25GMB242; Medline Kit SKU DYNJRA1140B: UDI/DI 10193489432015 each, 40193489432016 case, Lot Number 25DMI678; Medline Kit SKU DYNJRA1140B: UDI/DI 10193489432015 each, 40193489432016 case, Lot Number 24JMA139; Medline Kit SKU DYNJRA1140B: UDI/DI 10193489432015 each, 40193489432016 case, Lot Number 24BME491; Medline Kit SKU DYNJRA1140B: UDI/DI 10193489432015 each, 40193489432016 case, Lot Number 23KMB687; Medline Kit SKU DYNJRA1140B: UDI/DI 10193489432015 each, 40193489432016 case, Lot Number 23HMD531; Medline Kit SKU DYNJRA1192: UDI/DI 10888277762947 each, 40888277762948 case, Lot Number 24JDA083; Medline Kit SKU DYNJRA1192: UDI/DI 10888277762947 each, 40888277762948 case, Lot Number 24FDA375; Medline Kit SKU DYNJRA1192: UDI/DI 10888277762947 each, 40888277762948 case, Lot Number 24CDA048; Medline Kit SKU DYNJRA1192: UDI/DI 10888277762947 each, 40888277762948 case, Lot Number 23KDC201; Medline Kit SKU DYNJRA1192: UDI/DI 10888277762947 each, 40888277762948 case, Lot Number 23JDA946; Medline Kit SKU DYNJRA1240: UDI/DI 10889942965144 each, 40889942965145 case, Lot Number 26AMG752; Medline Kit SKU DYNJRA1240: UDI/DI 10889942965144 each, 40889942965145 case, Lot Number 25LME370; Medline Kit SKU DYNJRA1240: UDI/DI 10889942965144 each, 40889942965145 case, Lot Number 25HMC381; Medline Kit SKU DYNJRA1240: UDI/DI 10889942965144 each, 40889942965145 case, Lot Number 25GMG557; Medline Kit SKU DYNJRA1240: UDI/DI 10889942965144 each, 40889942965145 case, Lot Number 25FMD116; Medline Kit SKU DYNJRA1240: UDI/DI 10889942965144 each, 40889942965145 case, Lot Number 24JMA140; Medline Kit SKU DYNJRA1240: UDI/DI 10889942965144 each, 40889942965145 case, Lot Number 24IMH937; Medline Kit SKU DYNJRA1240: UDI/DI 10889942965144 each, 40889942965145 case, Lot Number 24BMH596; Medline Kit SKU DYNJRA1240: UDI/DI 10889942965144 each, 40889942965145 case, Lot Number 23JMF215; Medline Kit SKU DYNJRA1240: UDI/DI 10889942965144 each, 40889942965145 case, Lot Number 23HMI647; Medline Kit SKU DYNJRA1288: UDI/DI 10193489566291 each, 40193489566292 case, Lot Number 25IMI489; Medline Kit SKU DYNJRA1288: UDI/DI 10193489566291 each, 40193489566292 case, Lot Number 25HMB361; Medline Kit SKU DYNJRA1288: UDI/DI 10193489566291 each, 40193489566292 case, Lot Number 25DMG334; Medline Kit SKU DYNJRA1288: UDI/DI 10193489566291 each, 40193489566292 case, Lot Number 25DMD623; Medline Kit SKU DYNJRA1288: UDI/DI 10193489566291 each, 40193489566292 case, Lot Number 25DMC001; Medline Kit SKU DYNJRA1288: UDI/DI 10193489566291 each, 40193489566292 case, Lot Number 24JMA144; Medline Kit SKU DYNJRA1288: UDI/DI 10193489566291 each, 40193489566292 case, Lot Number 24GMI472; Medline Kit SKU DYNJRA1288: UDI/DI 10193489566291 each, 40193489566292 case, Lot Number 24GMG962; Medline Kit SKU DYNJRA1288: UDI/DI 10193489566291 each, 40193489566292 case, Lot Number 24GMA452; Medline Kit SKU DYNJRA1288: UDI/DI 10193489566291 each, 40193489566292 case, Lot Number 24DMF012; Medline Kit SKU DYNJRA1288: UDI/DI 10193489566291 each, 40193489566292 case, Lot Number 23IBL331; Medline Kit SKU DYNJRA1301: UDI/DI 10193489586732 each, 40193489586733 case, Lot Number 26BBQ480; Medline Kit SKU DYNJRA1301: UDI/DI 10193489586732 each, 40193489586733 case, Lot Number 25IBR270; Medline Kit SKU DYNJRA1301: UDI/DI 10193489586732 each, 40193489586733 case, Lot Number 25GBE596; Medline Kit SKU DYNJRA1301: UDI/DI 10193489586732 each, 40193489586733 case, Lot Number 24KDB789; Medline Kit SKU DYNJRA1301: UDI/DI 10193489586732 each, 40193489586733 case, Lot Number 24FDA635; Medline Kit SKU DYNJRA1304: UDI/DI 10193489590135 each, 40193489590136 case, Lot Number 25JMA655; Medline Kit SKU DYNJRA1304: UDI/DI 10193489590135 each, 40193489590136 case, Lot Number 25IMF958; Medline Kit SKU DYNJRA1304: UDI/DI 10193489590135 each, 40193489590136 case, Lot Number 24JMB108; Medline Kit SKU DYNJRA1304: UDI/DI 10193489590135 each, 40193489590136 case, Lot Number 24DMG675; Medline Kit SKU DYNJRA1304: UDI/DI 10193489590135 each, 40193489590136 case, Lot Number 23IBM788; Medline Kit SKU DYNJRA1376: UDI/DI 10193489658941 each, 40193489658942 case, Lot Number 26BBQ529; Medline Kit SKU DYNJRA1376: UDI/DI 10193489658941 each, 40193489658942 case, Lot Number 26ABT406; Medline Kit SKU DYNJRA1376: UDI/DI 10193489658941 each, 40193489658942 case, Lot Number 25LBT592; Medline Kit SKU DYNJRA1376: UDI/DI 10193489658941 each, 40193489658942 case, Lot Number 25KBR352; Medline Kit SKU DYNJRA1376: UDI/DI 10193489658941 each, 40193489658942 case, Lot Number 25JBO649; Medline Kit SKU DYNJRA1376: UDI/DI 10193489658941 each, 40193489658942 case, Lot Number 25JBB377; Medline Kit SKU DYNJRA1376: UDI/DI 10193489658941 each, 40193489658942 case, Lot Number 25HBP557; Medline Kit SKU DYNJRA1376: UDI/DI 10193489658941 each, 40193489658942 case, Lot Number 25FBU697; Medline Kit SKU DYNJRA1376: UDI/DI 10193489658941 each, 40193489658942 case, Lot Number 25DBT832; Medline Kit SKU DYNJRA1376: UDI/DI 10193489658941 each, 40193489658942 case, Lot Number 25CBC755; Medline Kit SKU DYNJRA1376: UDI/DI 10193489658941 each, 40193489658942 case, Lot Number 23JBI433; Medline Kit SKU DYNJRA1376: UDI/DI 10193489658941 each, 40193489658942 case, Lot Number 23IBL601; Medline Kit SKU DYNJRA1376: UDI/DI 10193489658941 each, 40193489658942 case, Lot Number 23GBW707; Medline Kit SKU DYNJRA1376: UDI/DI 10193489658941 each, 40193489658942 case, Lot Number 23GBE466; Medline Kit SKU DYNJRA1477A: UDI/DI 10195327298692 each, 40195327298693 case, Lot Number 23JBC917; Medline Kit SKU DYNJRA1538: UDI/DI 10193489215366 each, 40193489215367 case, Lot Number 26ABT377; Medline Kit SKU DYNJRA1538: UDI/DI 10193489215366 each, 40193489215367 case, Lot Number 26ABD865; Medline Kit SKU DYNJRA1538: UDI/DI 10193489215366 each, 40193489215367 case, Lot Number 25KBB315; Medline Kit SKU DYNJRA1538: UDI/DI 10193489215366 each, 40193489215367 case, Lot Number 25HBT787; Medline Kit SKU DYNJRA1538: UDI/DI 10193489215366 each, 40193489215367 case, Lot Number 25HBJ292; Medline Kit SKU DYNJRA1538: UDI/DI 10193489215366 each, 40193489215367 case, Lot Number 25FBN804; Medline Kit SKU DYNJRA1538: UDI/DI 10193489215366 each, 40193489215367 case, Lot Number 25EBM784; Medline Kit SKU DYNJRA1538: UDI/DI 10193489215366 each, 40193489215367 case, Lot Number 25ALA898; Medline Kit SKU DYNJRA1538: UDI/DI 10193489215366 each, 40193489215367 case, Lot Number 24FLA415; Medline Kit SKU DYNJRA1538: UDI/DI 10193489215366 each, 40193489215367 case, Lot Number 24DLB151; Medline Kit SKU DYNJRA1538: UDI/DI 10193489215366 each, 40193489215367 case, Lot Number 24BLA472; Medline Kit SKU DYNJRA1538: UDI/DI 10193489215366 each, 40193489215367 case, Lot Number 23JLB282; Medline Kit SKU DYNJRA1538: UDI/DI 10193489215366 each, 40193489215367 case, Lot Number 23ILA430; Medline Kit SKU DYNJRA1558A: UDI/DI 10193489422450 each, 40193489422451 case, Lot Number 26ABM540; Medline Kit SKU DYNJRA1558A: UDI/DI 10193489422450 each, 40193489422451 case, Lot Number 25IBO446; Medline Kit SKU DYNJRA1558A: UDI/DI 10193489422450 each, 40193489422451 case, Lot Number 25GBR046; Medline Kit SKU DYNJRA1558A: UDI/DI 10193489422450 each, 40193489422451 case, Lot Number 25GBP595; Medline Kit SKU DYNJRA1558A: UDI/DI 10193489422450 each, 40193489422451 case, Lot Number 25DBM632; Medline Kit SKU DYNJRA1558A: UDI/DI 10193489422450 each, 40193489422451 case, Lot Number 23ILA580; Medline Kit SKU DYNJRA1560A: UDI/DI 10193489992427 each, 40193489992428 case, Lot Number 26CLA918; Medline Kit SKU DYNJRA1560A: UDI/DI 10193489992427 each, 40193489992428 case, Lot Number 25LLA085; Medline Kit SKU DYNJRA1560A: UDI/DI 10193489992427 each, 40193489992428 case, Lot Number 25HLA940; Medline Kit SKU DYNJRA1560A: UDI/DI 10193489992427 each, 40193489992428 case, Lot Number 25DLA809; Medline Kit SKU DYNJRA1560A: UDI/DI 10193489992427 each, 40193489992428 case, Lot Number 25ALA743; Medline Kit SKU DYNJRA1560A: UDI/DI 10193489992427 each, 40193489992428 case, Lot Number 24ALA606; Medline Kit SKU DYNJRA1565: UDI/DI 10193489251654 each, 40193489251655 case, Lot Number 26BBJ338; Medline Kit SKU DYNJRA1565: UDI/DI 10193489251654 each, 40193489251655 case, Lot Number 25LBJ822; Medline Kit SKU DYNJRA1565: UDI/DI 10193489251654 each, 40193489251655 case, Lot Number 25HBT749; Medline Kit SKU DYNJRA1565: UDI/DI 10193489251654 each, 40193489251655 case, Lot Number 25FBB360; Medline Kit SKU DYNJRA1565: UDI/DI 10193489251654 each, 40193489251655 case, Lot Number 25ALA674; Medline Kit SKU DYNJRA1577C: UDI/DI 10198459030956 each, 40198459030957 case, Lot Number 26CBF846; Medline Kit SKU DYNJRA1577C: UDI/DI 10198459030956 each, 40198459030957 case, Lot Number 26ABC756; Medline Kit SKU DYNJRA1584: UDI/DI 10193489278408 each, 40193489278409 case, Lot Number 24LLA814; Medline Kit SKU DYNJRA1584: UDI/DI 10193489278408 each, 40193489278409 case, Lot Number 23KLA590; Medline Kit SKU DYNJRA1740A: UDI/DI 10195327088934 each, 40195327088935 case, Lot Number 26BMI370; Medline Kit SKU DYNJRA1740A: UDI/DI 10195327088934 each, 40195327088935 case, Lot Number 26AMD201; Medline Kit SKU DYNJRA1740A: UDI/DI 10195327088934 each, 40195327088935 case, Lot Number 25JMJ116; Medline Kit SKU DYNJRA1740A: UDI/DI 10195327088934 each, 40195327088935 case, Lot Number 25IMI411; Medline Kit SKU DYNJRA1740A: UDI/DI 10195327088934 each, 40195327088935 case, Lot Number 25GMB253; Medline Kit SKU DYNJRA1740A: UDI/DI 10195327088934 each, 40195327088935 case, Lot Number 25DMI661; Medline Kit SKU DYNJRA1740A: UDI/DI 10195327088934 each, 40195327088935 case, Lot Number 25DMD627; Medline Kit SKU DYNJRA1740A: UDI/DI 10195327088934 each, 40195327088935 case, Lot Number 25CMJ576; Medline Kit SKU DYNJRA1740A: UDI/DI 10195327088934 each, 40195327088935 case, Lot Number 25AMI062; Medline Kit SKU DYNJRA1740A: UDI/DI 10195327088934 each, 40195327088935 case, Lot Number 24JMI045; Medline Kit SKU DYNJRA1740A: UDI/DI 10195327088934 each, 40195327088935 case, Lot Number 24JMA122; Medline Kit SKU DYNJRA1740A: UDI/DI 10195327088934 each, 40195327088935 case, Lot Number 24GMJ821; Medline Kit SKU DYNJRA1740A: UDI/DI 10195327088934 each, 40195327088935 case, Lot Number 24CMF337; Medline Kit SKU DYNJRA1740A: UDI/DI 10195327088934 each, 40195327088935 case, Lot Number 24BBP807; Medline Kit SKU DYNJRA1740A: UDI/DI 10195327088934 each, 40195327088935 case, Lot Number 23JBA530; Medline Kit SKU DYNJRA1740A: UDI/DI 10195327088934 each, 40195327088935 case, Lot Number 23HBJ527; Medline Kit SKU DYNJRA1744A: UDI/DI 10195327297749 each, 40195327297740 case, Lot Number 26BBH656; Medline Kit SKU DYNJRA1750A: UDI/DI 10198459270192 each, 40198459270193 case, Lot Number 25CBF201; Medline Kit SKU DYNJRA1784: UDI/DI 10193489875126 each, 40193489875127 case, Lot Number 24CMA243; Medline Kit SKU DYNJRA1784: UDI/DI 10193489875126 each, 40193489875127 case, Lot Number 23KMD775; Medline Kit SKU DYNJRA1790: UDI/DI 10193489880441 each, 40193489880442 case, Lot Number 24GLA034; Medline Kit SKU DYNJRA1790: UDI/DI 10193489880441 each, 40193489880442 case, Lot Number 24ALA809; Medline Kit SKU DYNJRA1796: UDI/DI 10193489898477 each, 40193489898478 case, Lot Number 25GMB228; Medline Kit SKU DYNJRA1796: UDI/DI 10193489898477 each, 40193489898478 case, Lot Number 25BMC173; Medline Kit SKU DYNJRA1796: UDI/DI 10193489898477 each, 40193489898478 case, Lot Number 24KME293; Medline Kit SKU DYNJRA1796: UDI/DI 10193489898477 each, 40193489898478 case, Lot Number 24KMC186; Medline Kit SKU DYNJRA1796: UDI/DI 10193489898477 each, 40193489898478 case, Lot Number 24JMI040; Medline Kit SKU DYNJRA1796: UDI/DI 10193489898477 each, 40193489898478 case, Lot Number 23IBC188; Medline Kit SKU DYNJRA1796: UDI/DI 10193489898477 each, 40193489898478 case, Lot Number 23HBH576; Medline Kit SKU DYNJRA1802B: UDI/DI 10195327343897 each, 40195327343898 case, Lot Number 24BBA726; Medline Kit SKU DYNJRA1818: UDI/DI 10193489946888 each, 40193489946889 case, Lot Number 26BBT339; Medline Kit SKU DYNJRA1818: UDI/DI 10193489946888 each, 40193489946889 case, Lot Number 26ABW298; Medline Kit SKU DYNJRA1818: UDI/DI 10193489946888 each, 40193489946889 case, Lot Number 25LBJ837; Medline Kit SKU DYNJRA1818: UDI/DI 10193489946888 each, 40193489946889 case, Lot Number 25JBO685; Medline Kit SKU DYNJRA1818: UDI/DI 10193489946888 each, 40193489946889 case, Lot Number 25IBA272; Medline Kit SKU DYNJRA1818: UDI/DI 10193489946888 each, 40193489946889 case, Lot Number 25GBK749; Medline Kit SKU DYNJRA1818: UDI/DI 10193489946888 each, 40193489946889 case, Lot Number 25FBH676; Medline Kit SKU DYNJRA1818: UDI/DI 10193489946888 each, 40193489946889 case, Lot Number 25CBD837; Medline Kit SKU DYNJRA1818: UDI/DI 10193489946888 each, 40193489946889 case, Lot Number 25ALA358; Medline Kit SKU DYNJRA1818: UDI/DI 10193489946888 each, 40193489946889 case, Lot Number 25ALA119; Medline Kit SKU DYNJRA1818: UDI/DI 10193489946888 each, 40193489946889 case, Lot Number 24KLA542; Medline Kit SKU DYNJRA1818: UDI/DI 10193489946888 each, 40193489946889 case, Lot Number 24FLA929; Medline Kit SKU DYNJRA1818: UDI/DI 10193489946888 each, 40193489946889 case, Lot Number 24DLA123; Medline Kit SKU DYNJRA1818: UDI/DI 10193489946888 each, 40193489946889 case, Lot Number 24ALA052; Medline Kit SKU DYNJRA1818: UDI/DI 10193489946888 each, 40193489946889 case, Lot Number 23KLA873; Medline Kit SKU DYNJRA1818: UDI/DI 10193489946888 each, 40193489946889 case, Lot Number 23JLA294; Medline Kit SKU DYNJRA1818: UDI/DI 10193489946888 each, 40193489946889 case, Lot Number 23ILA293; Medline Kit SKU DYNJRA1839: UDI/DI 10193489981605 each, 40193489981606 case, Lot Number 26BBS481; Medline Kit SKU DYNJRA1839: UDI/DI 10193489981605 each, 40193489981606 case, Lot Number 26BBB042; Medline Kit SKU DYNJRA1839: UDI/DI 10193489981605 each, 40193489981606 case, Lot Number 26ABQ774; Medline Kit SKU DYNJRA1839: UDI/DI 10193489981605 each, 40193489981606 case, Lot Number 25HBB827; Medline Kit SKU DYNJRA1839: UDI/DI 10193489981605 each, 40193489981606 case, Lot Number 25GBF892; Medline Kit SKU DYNJRA1839: UDI/DI 10193489981605 each, 40193489981606 case, Lot Number 24FBT400; Medline Kit SKU DYNJRA1839: UDI/DI 10193489981605 each, 40193489981606 case, Lot Number 23HBG227; Medline Kit SKU DYNJRA1884: UDI/DI 10195327044442 each, 40195327044443 case, Lot Number 24ALA860; Medline Kit SKU DYNJRA1884: UDI/DI 10195327044442 each, 40195327044443 case, Lot Number 23LLA854; Medline Kit SKU DYNJRA1884: UDI/DI 10195327044442 each, 40195327044443 case, Lot Number 23JLA222; Medline Kit SKU DYNJRA1884: UDI/DI 10195327044442 each, 40195327044443 case, Lot Number 23ILA091; Medline Kit SKU DYNJRA1884A: UDI/DI 10195327630461 each, 40195327630462 case, Lot Number 26BBQ460; Medline Kit SKU DYNJRA1884A: UDI/DI 10195327630461 each, 40195327630462 case, Lot Number 26ABS301; Medline Kit SKU DYNJRA1884A: UDI/DI 10195327630461 each, 40195327630462 case, Lot Number 25JBT564; Medline Kit SKU DYNJRA1884A: UDI/DI 10195327630461 each, 40195327630462 case, Lot Number 25JBD791; Medline Kit SKU DYNJRA1884A: UDI/DI 10195327630461 each, 40195327630462 case, Lot Number 25IBJ312; Medline Kit SKU DYNJRA1884A: UDI/DI 10195327630461 each, 40195327630462 case, Lot Number 25GBA592; Medline Kit SKU DYNJRA1884A: UDI/DI 10195327630461 each, 40195327630462 case, Lot Number 25EBO914; Medline Kit SKU DYNJRA1884A: UDI/DI 10195327630461 each, 40195327630462 case, Lot Number 24GBI534; Medline Kit SKU DYNJRA1889: UDI/DI 10195327052409 each, 40195327052400 case, Lot Number 25LBL251; Medline Kit SKU DYNJRA1889: UDI/DI 10195327052409 each, 40195327052400 case, Lot Number 25LBG746; Medline Kit SKU DYNJRA1889: UDI/DI 10195327052409 each, 40195327052400 case, Lot Number 25KBN759; Medline Kit SKU DYNJRA1889: UDI/DI 10195327052409 each, 40195327052400 case, Lot Number 25HBG277; Medline Kit SKU DYNJRA1889: UDI/DI 10195327052409 each, 40195327052400 case, Lot Number 25FBL050; Medline Kit SKU DYNJRA1889: UDI/DI 10195327052409 each, 40195327052400 case, Lot Number 24BBR617; Medline Kit SKU DYNJRA1889: UDI/DI 10195327052409 each, 40195327052400 case, Lot Number 23JBE887; Medline Kit SKU DYNJRA1889: UDI/DI 10195327052409 each, 40195327052400 case, Lot Number 23GBP040; Medline Kit SKU DYNJRA1929: UDI/DI 10195327117436 each, 40195327117437 case, Lot Number 25JMH654; Medline Kit SKU DYNJRA1929: UDI/DI 10195327117436 each, 40195327117437 case, Lot Number 25GMB229; Medline Kit SKU DYNJRA1929: UDI/DI 10195327117436 each, 40195327117437 case, Lot Number 25DMF106; Medline Kit SKU DYNJRA1929: UDI/DI 10195327117436 each, 40195327117437 case, Lot Number 24FMI985; Medline Kit SKU DYNJRA1969: UDI/DI 10195327155759 each, 40195327155750 case, Lot Number 26BBJ366; Medline Kit SKU DYNJRA1969: UDI/DI 10195327155759 each, 40195327155750 case, Lot Number 26ABJ212; Medline Kit SKU DYNJRA1969: UDI/DI 10195327155759 each, 40195327155750 case, Lot Number 25KBR357; Medline Kit SKU DYNJRA1969: UDI/DI 10195327155759 each, 40195327155750 case, Lot Number 25IBI578; Medline Kit SKU DYNJRA1969: UDI/DI 10195327155759 each, 40195327155750 case, Lot Number 25GBS433; Medline Kit SKU DYNJRA1969: UDI/DI 10195327155759 each, 40195327155750 case, Lot Number 25DBS842; Medline Kit SKU DYNJRA1969: UDI/DI 10195327155759 each, 40195327155750 case, Lot Number 25BBJ513; Medline Kit SKU DYNJRA1969: UDI/DI 10195327155759 each, 40195327155750 case, Lot Number 24BBR619; Medline Kit SKU DYNJRA1969: UDI/DI 10195327155759 each, 40195327155750 case, Lot Number 23JBI474; Medline Kit SKU DYNJRA1969: UDI/DI 10195327155759 each, 40195327155750 case, Lot Number 23IBG943; Medline Kit SKU DYNJRA1969: UDI/DI 10195327155759 each, 40195327155750 case, Lot Number 23GBW701; Medline Kit SKU DYNJRA1969: UDI/DI 10195327155759 each, 40195327155750 case, Lot Number 23GBG665; Medline Kit SKU DYNJRA1982: UDI/DI 10195327172220 each, 40195327172221 case, Lot Number 25KBH337; Medline Kit SKU DYNJRA1982: UDI/DI 10195327172220 each, 40195327172221 case, Lot Number 25FBJ700; Medline Kit SKU DYNJRA1982: UDI/DI 10195327172220 each, 40195327172221 case, Lot Number 25DBP048; Medline Kit SKU DYNJRA1982: UDI/DI 10195327172220 each, 40195327172221 case, Lot Number 23IBU749; Medline Kit SKU DYNJRA1982: UDI/DI 10195327172220 each, 40195327172221 case, Lot Number 23HBN821; Medline Kit SKU DYNJRA1983A: UDI/DI 10195327328559 each, 40195327328550 case, Lot Number 24HMA882; Medline Kit SKU DYNJRA1983A: UDI/DI 10195327328559 each, 40195327328550 case, Lot Number 24FMI986; Medline Kit SKU DYNJRA1983A: UDI/DI 10195327328559 each, 40195327328550 case, Lot Number 24DMB103; Medline Kit SKU DYNJRA1983A: UDI/DI 10195327328559 each, 40195327328550 case, Lot Number 24CMH742; Medline Kit SKU DYNJRA1983A: UDI/DI 10195327328559 each, 40195327328550 case, Lot Number 24BBP805; Medline Kit SKU DYNJRA1983A: UDI/DI 10195327328559 each, 40195327328550 case, Lot Number 23JBI335; Medline Kit SKU DYNJRA1983A: UDI/DI 10195327328559 each, 40195327328550 case, Lot Number 23JBB539; Medline Kit SKU DYNJRA1983A: UDI/DI 10195327328559 each, 40195327328550 case, Lot Number 23IBM793; Medline Kit SKU DYNJRA1983A: UDI/DI 10195327328559 each, 40195327328550 case, Lot Number 23HBX589; Medline Kit SKU DYNJRA1983A: UDI/DI 10195327328559 each, 40195327328550 case, Lot Number 23GBO102; Medline Kit SKU DYNJRA1987: UDI/DI 10195327178055 each, 40195327178056 case, Lot Number 25DMI991; Medline Kit SKU DYNJRA1987: UDI/DI 10195327178055 each, 40195327178056 case, Lot Number 24IMG184; Medline Kit SKU DYNJRA1987: UDI/DI 10195327178055 each, 40195327178056 case, Lot Number 24CMJ744; Medline Kit SKU DYNJRA1987: UDI/DI 10195327178055 each, 40195327178056 case, Lot Number 24CME769; Medline Kit SKU DYNJRA2005: UDI/DI 10195327211134 each, 40195327211135 case, Lot Number 26BMI302; Medline Kit SKU DYNJRA2005: UDI/DI 10195327211134 each, 40195327211135 case, Lot Number 25HME970; Medline Kit SKU DYNJRA2005: UDI/DI 10195327211134 each, 40195327211135 case, Lot Number 25EMI731; Medline Kit SKU DYNJRA2005: UDI/DI 10195327211134 each, 40195327211135 case, Lot Number 24KME368; Medline Kit SKU DYNJRA2005: UDI/DI 10195327211134 each, 40195327211135 case, Lot Number 24KMD667; Medline Kit SKU DYNJRA2005: UDI/DI 10195327211134 each, 40195327211135 case, Lot Number 24GMI469; Medline Kit SKU DYNJRA2023: UDI/DI 10195327218140 each, 40195327218141 case, Lot Number 26BMH421; Medline Kit SKU DYNJRA2023: UDI/DI 10195327218140 each, 40195327218141 case, Lot Number 25IMI450; Medline Kit SKU DYNJRA2023: UDI/DI 10195327218140 each, 40195327218141 case, Lot Number 25HMC853; Medline Kit SKU DYNJRA2023: UDI/DI 10195327218140 each, 40195327218141 case, Lot Number 25CMI825; Medline Kit SKU DYNJRA2023: UDI/DI 10195327218140 each, 40195327218141 case, Lot Number 25BMH953; Medline Kit SKU DYNJRA2023: UDI/DI 10195327218140 each, 40195327218141 case, Lot Number 24HMD118; Medline Kit SKU DYNJRA2023: UDI/DI 10195327218140 each, 40195327218141 case, Lot Number 24BBO065; Medline Kit SKU DYNJRA2023: UDI/DI 10195327218140 each, 40195327218141 case, Lot Number 23IBL654; Medline Kit SKU DYNJRA2023: UDI/DI 10195327218140 each, 40195327218141 case, Lot Number 23HBR220; Medline Kit SKU DYNJRA2068A: UDI/DI 10195327427870 each, 40195327427871 case, Lot Number 26AMK410; Medline Kit SKU DYNJRA2068A: UDI/DI 10195327427870 each, 40195327427871 case, Lot Number 26AMH169; Medline Kit SKU DYNJRA2068A: UDI/DI 10195327427870 each, 40195327427871 case, Lot Number 25KMJ048; Medline Kit SKU DYNJRA2068A: UDI/DI 10195327427870 each, 40195327427871 case, Lot Number 25IMA352; Medline Kit SKU DYNJRA2068A: UDI/DI 10195327427870 each, 40195327427871 case, Lot Number 25DMH258; Medline Kit SKU DYNJRA2068A: UDI/DI 10195327427870 each, 40195327427871 case, Lot Number 25CMA052; Medline Kit SKU DYNJRA2068A: UDI/DI 10195327427870 each, 40195327427871 case, Lot Number 24JMA141; Medline Kit SKU DYNJRA2068A: UDI/DI 10195327427870 each, 40195327427871 case, Lot Number 24CMH183; Medline Kit SKU DYNJRA2068A: UDI/DI 10195327427870 each, 40195327427871 case, Lot Number 24BMJ537; Medline Kit SKU DYNJRA2068A: UDI/DI 10195327427870 each, 40195327427871 case, Lot Number 23JBB546; Medline Kit SKU DYNJRA2088: UDI/DI 10195327288495 each, 40195327288496 case, Lot Number 25LBO166; Medline Kit SKU DYNJRA2088: UDI/DI 10195327288495 each, 40195327288496 case, Lot Number 25IBP107; Medline Kit SKU DYNJRA2088: UDI/DI 10195327288495 each, 40195327288496 case, Lot Number 25DBI023; Medline Kit SKU DYNJRA2115: UDI/DI 10195327327101 each, 40195327327102 case, Lot Number 26BBR460; Medline Kit SKU DYNJRA2115: UDI/DI 10195327327101 each, 40195327327102 case, Lot Number 26ABL341; Medline Kit SKU DYNJRA2115: UDI/DI 10195327327101 each, 40195327327102 case, Lot Number 25KBT165; Medline Kit SKU DYNJRA2115: UDI/DI 10195327327101 each, 40195327327102 case, Lot Number 25FBJ911; Medline Kit SKU DYNJRA2115: UDI/DI 10195327327101 each, 40195327327102 case, Lot Number 25EBM786; Medline Kit SKU DYNJRA2115: UDI/DI 10195327327101 each, 40195327327102 case, Lot Number 24BBQ039; Medline Kit SKU DYNJRA2115: UDI/DI 10195327327101 each, 40195327327102 case, Lot Number 23IBK437; Medline Kit SKU DYNJRA2115: UDI/DI 10195327327101 each, 40195327327102 case, Lot Number 23GBE482; Medline Kit SKU DYNJRA2117: UDI/DI 10195327328801 each, 40195327328802 case, Lot Number 23HBB082; Medline Kit SKU DYNJRA2140: UDI/DI 10195327351816 each, 40195327351817 case, Lot Number 26BBB083; Medline Kit SKU DYNJRA2140: UDI/DI 10195327351816 each, 40195327351817 case, Lot Number 25KBT830; Medline Kit SKU DYNJRA2140: UDI/DI 10195327351816 each, 40195327351817 case, Lot Number 25JBJ002; Medline Kit SKU DYNJRA2140: UDI/DI 10195327351816 each, 40195327351817 case, Lot Number 25FBR550; Medline Kit SKU DYNJRA2140: UDI/DI 10195327351816 each, 40195327351817 case, Lot Number 25DBM220; Medline Kit SKU DYNJRA2140: UDI/DI 10195327351816 each, 40195327351817 case, Lot Number 24GBD223; Medline Kit SKU DYNJRA2140: UDI/DI 10195327351816 each, 40195327351817 case, Lot Number 23IBP024; Medline Kit SKU DYNJRA2153: UDI/DI 10195327366391 each, 40195327366392 case, Lot Number 26BBJ361; Medline Kit SKU DYNJRA2153: UDI/DI 10195327366391 each, 40195327366392 case, Lot Number 26ABL239; Medline Kit SKU DYNJRA2153: UDI/DI 10195327366391 each, 40195327366392 case, Lot Number 25LBC432; Medline Kit SKU DYNJRA2153: UDI/DI 10195327366391 each, 40195327366392 case, Lot Number 25JBO638; Medline Kit SKU DYNJRA2153: UDI/DI 10195327366391 each, 40195327366392 case, Lot Number 25HBK428; Medline Kit SKU DYNJRA2153: UDI/DI 10195327366391 each, 40195327366392 case, Lot Number 25EBR104; Medline Kit SKU DYNJRA2153: UDI/DI 10195327366391 each, 40195327366392 case, Lot Number 25DBU420; Medline Kit SKU DYNJRA2153: UDI/DI 10195327366391 each, 40195327366392 case, Lot Number 24GBI551; Medline Kit SKU DYNJRA2153: UDI/DI 10195327366391 each, 40195327366392 case, Lot Number 23GBH424; Medline Kit SKU DYNJRA2213: UDI/DI 10195327425876 each, 40195327425877 case, Lot Number 26CBA281; Medline Kit SKU DYNJRA2213: UDI/DI 10195327425876 each, 40195327425877 case, Lot Number 26ABL340; Medline Kit SKU DYNJRA2213: UDI/DI 10195327425876 each, 40195327425877 case, Lot Number 25KBI543; Medline Kit SKU DYNJRA2213: UDI/DI 10195327425876 each, 40195327425877 case, Lot Number 25IBR271; Medline Kit SKU DYNJRA2213: UDI/DI 10195327425876 each, 40195327425877 case, Lot Number 25FBN745; Medline Kit SKU DYNJRA2213: UDI/DI 10195327425876 each, 40195327425877 case, Lot Number 25FBJ701; Medline Kit SKU DYNJRA2213: UDI/DI 10195327425876 each, 40195327425877 case, Lot Number 25DBS268; Medline Kit SKU DYNJRA2213: UDI/DI 10195327425876 each, 40195327425877 case, Lot Number 23HBV950; Medline Kit SKU DYNJRA2225: UDI/DI 10195327432683 each, 40195327432684 case, Lot Number 23HBT397; Medline Kit SKU DYNJRA2349: UDI/DI 10195327607135 each, 40195327607136 case, Lot Number 26BBI277; Medline Kit SKU DYNJRA2349: UDI/DI 10195327607135 each, 40195327607136 case, Lot Number 25JBO671; Medline Kit SKU DYNJRA2349: UDI/DI 10195327607135 each, 40195327607136 case, Lot Number 25JBG136; Medline Kit SKU DYNJRA2349: UDI/DI 10195327607135 each, 40195327607136 case, Lot Number 25FBH739; Medline Kit SKU DYNJRA2360: UDI/DI 10195327629489 each, 40195327629480 case, Lot Number 26CBD944; Medline Kit SKU DYNJRA2360: UDI/DI 10195327629489 each, 40195327629480 case, Lot Number 26BBP631; Medline Kit SKU DYNJRA2360: UDI/DI 10195327629489 each, 40195327629480 case, Lot Number 26BBK190; Medline Kit SKU DYNJRA2360: UDI/DI 10195327629489 each, 40195327629480 case, Lot Number 26ABW305; Medline Kit SKU DYNJRA2360: UDI/DI 10195327629489 each, 40195327629480 case, Lot Number 26ABG715; Medline Kit SKU DYNJRA2371: UDI/DI 10195327644895 each, 40195327644896 case, Lot Number 26CBF837; Medline Kit SKU DYNJRA2371: UDI/DI 10195327644895 each, 40195327644896 case, Lot Number 25IBB962; Medline Kit SKU DYNJRA2371: UDI/DI 10195327644895 each, 40195327644896 case, Lot Number 25IBB964; Medline Kit SKU DYNJRA2371: UDI/DI 10195327644895 each, 40195327644896 case, Lot Number 25HBK417; Medline Kit SKU DYNJRA2371: UDI/DI 10195327644895 each, 40195327644896 case, Lot Number 24GBD226; Medline Kit SKU DYNJRA2381: UDI/DI 10195327667146 each, 40195327667147 case, Lot Number 25JBW800; Medline Kit SKU DYNJRA2381: UDI/DI 10195327667146 each, 40195327667147 case, Lot Number 25JBC388; Medline Kit SKU DYNJRA2381: UDI/DI 10195327667146 each, 40195327667147 case, Lot Number 24GBD228; Medline Kit SKU DYNJRA2432A: UDI/DI 10198459382727 each, 40198459382728 case, Lot Number 26CBF847; Medline Kit SKU DYNJRA2432A: UDI/DI 10198459382727 each, 40198459382728 case, Lot Number 26BBM310; Medline Kit SKU DYNJRA2432A: UDI/DI 10198459382727 each, 40198459382728 case, Lot Number 25GBC739; Medline Kit SKU DYNJRA2532: UDI/DI 10198459188497 each, 40198459188498 case, Lot Number 25LBF411; Medline Kit SKU DYNJRA2532: UDI/DI 10198459188497 each, 40198459188498 case, Lot Number 25HBU760; Medline Kit SKU DYNJRA2534: UDI/DI 10198459191251 each, 40198459191252 case, Lot Number 25GBW196; Medline Kit SKU DYNJRA2534: UDI/DI 10198459191251 each, 40198459191252 case, Lot Number 25FBT598; Medline Kit SKU DYNJRA2556: UDI/DI 10198459247514 each, 40198459247515 case, Lot Number 26BBE200; Medline Kit SKU DYNJRA2556: UDI/DI 10198459247514 each, 40198459247515 case, Lot Number 25LBN290; Medline Kit SKU DYNJRA2556: UDI/DI 10198459247514 each, 40198459247515 case, Lot Number 25KBQ920; Medline Kit SKU DYNJRA2556: UDI/DI 10198459247514 each, 40198459247515 case, Lot Number 25JBU726; Medline Kit SKU DYNJRA2556: UDI/DI 10198459247514 each, 40198459247515 case, Lot Number 25JBC448; Medline Kit SKU DYNJRA2556: UDI/DI 10198459247514 each, 40198459247515 case, Lot Number 25HBS669; Medline Kit SKU DYNJRA2556: UDI/DI 10198459247514 each, 40198459247515 case, Lot Number 25FBR642; Medline Kit SKU DYNJRA2556: UDI/DI 10198459247514 each, 40198459247515 case, Lot Number 25DBI029; Medline Kit SKU DYNJRA2566A: UDI/DI 10198459421822 each, 40198459421823 case, Lot Number 26CBD938; Medline Kit SKU DYNJRA2566A: UDI/DI 10198459421822 each, 40198459421823 case, Lot Number 25HBP480; Medline Kit SKU DYNJRA2566A: UDI/DI 10198459421822 each, 40198459421823 case, Lot Number 25HBK616; Medline Kit SKU DYNJRA2575: UDI/DI 10198459271830 each, 40198459271831 case, Lot Number 25LBC427; Medline Kit SKU DYNJRA2575: UDI/DI 10198459271830 each, 40198459271831 case, Lot Number 25JBX992; Medline Kit SKU DYNJRA2575: UDI/DI 10198459271830 each, 40198459271831 case, Lot Number 25JBC445; Medline Kit SKU DYNJRA2575: UDI/DI 10198459271830 each, 40198459271831 case, Lot Number 25GBO568; Medline Kit SKU DYNJRA2575: UDI/DI 10198459271830 each, 40198459271831 case, Lot Number 25GBB711; Medline Kit SKU DYNJRA2575: UDI/DI 10198459271830 each, 40198459271831 case, Lot Number 25GBG218; Medline Kit SKU DYNJRA2575: UDI/DI 10198459271830 each, 40198459271831 case, Lot Number 25BBH733; Medline Kit SKU DYNJRA2592: UDI/DI 10198459300417 each, 40198459300418 case, Lot Number 26ABL364; Medline Kit SKU DYNJRA2592: UDI/DI 10198459300417 each, 40198459300418 case, Lot Number 25KBT846; Medline Kit SKU DYNJRA2592: UDI/DI 10198459300417 each, 40198459300418 case, Lot Number 25FBC729; Medline Kit SKU DYNJRA2597: UDI/DI 10198459304835 each, 40198459304836 case, Lot Number 25GBN789; Medline Kit SKU DYNJRA2597: UDI/DI 10198459304835 each, 40198459304836 case, Lot Number 25EBA564; Medline Kit SKU DYNJRA2605: UDI/DI 10198459326370 each, 40198459326371 case, Lot Number 26ABQ772; Medline Kit SKU DYNJRA2605: UDI/DI 10198459326370 each, 40198459326371 case, Lot Number 26ABK647; Medline Kit SKU DYNJRA2605: UDI/DI 10198459326370 each, 40198459326371 case, Lot Number 25KBT820; Medline Kit SKU DYNJRA2605: UDI/DI 10198459326370 each, 40198459326371 case, Lot Number 25FBG507; Medline Kit SKU DYNJRA2606: UDI/DI 10198459328657 each, 40198459328658 case, Lot Number 25LBF412; Medline Kit SKU DYNJRA2606: UDI/DI 10198459328657 each, 40198459328658 case, Lot Number 25JBM797; Medline Kit SKU DYNJRA2606: UDI/DI 10198459328657 each, 40198459328658 case, Lot Number 25FBT691; Medline Kit SKU DYNJRA2606: UDI/DI 10198459328657 each, 40198459328658 case, Lot Number 25EBC231; Medline Kit SKU DYNJRA2607: UDI/DI 10198459331657 each, 40198459331658 case, Lot Number 25EBK954; Medline Kit SKU DYNJRA2607A: UDI/DI 10198459421839 each, 40198459421830 case, Lot Number 26CBI128; Medline Kit SKU DYNJRA2607A: UDI/DI 10198459421839 each, 40198459421830 case, Lot Number 26ABV387; Medline Kit SKU DYNJRA2611: UDI/DI 10198459336256 each, 40198459336257 case, Lot Number 26BBB084; Medline Kit SKU DYNJRA2611: UDI/DI 10198459336256 each, 40198459336257 case, Lot Number 25LBN247; Medline Kit SKU DYNJRA2611: UDI/DI 10198459336256 each, 40198459336257 case, Lot Number 25IBB903; Medline Kit SKU DYNJRA2611: UDI/DI 10198459336256 each, 40198459336257 case, Lot Number 25EBF537; Medline Kit SKU DYNJRA2613: UDI/DI 10198459339356 each, 40198459339357 case, Lot Number 25EBR108; Medline Kit SKU DYNJRA2618A: UDI/DI 10198459537998 each, 40198459537999 case, Lot Number 26CBE369; Medline Kit SKU DYNJRA2618A: UDI/DI 10198459537998 each, 40198459537999 case, Lot Number 26ABJ331; Medline Kit SKU DYNJRA2634: UDI/DI 10198459349355 each, 40198459349356 case, Lot Number 26BBL356; Medline Kit SKU DYNJRA2634: UDI/DI 10198459349355 each, 40198459349356 case, Lot Number 25JBA669; Medline Kit SKU DYNJRA2634: UDI/DI 10198459349355 each, 40198459349356 case, Lot Number 25FBP070; Medline Kit SKU DYNJRA2638: UDI/DI 10198459349591 each, 40198459349592 case, Lot Number 25FBG845; Medline Kit SKU DYNJRA2639: UDI/DI 10198459352218 each, 40198459352219 case, Lot Number 25FBU637; Medline Kit SKU DYNJRA2641: UDI/DI 10198459352546 each, 40198459352547 case, Lot Number 26BBE022; Medline Kit SKU DYNJRA2641: UDI/DI 10198459352546 each, 40198459352547 case, Lot Number 25EBO148; Medline Kit SKU DYNJRA2653: UDI/DI 10198459411021 each, 40198459411022 case, Lot Number 25IBL292; Medline Kit SKU DYNJRA2669: UDI/DI 10198459466489 each, 40198459466480 case, Lot Number 26CBI134; Medline Kit SKU DYNJRA2669: UDI/DI 10198459466489 each, 40198459466480 case, Lot Number 26BBJ402; Medline Kit SKU DYNJRA2669: UDI/DI 10198459466489 each, 40198459466480 case, Lot Number 25LBF213; Medline Kit SKU DYNJRA2669: UDI/DI 10198459466489 each, 40198459466480 case, Lot Number 25IBL293; Medline Kit SKU DYNJRA2670: UDI/DI 10198459467974 each, 40198459467975 case, Lot Number 25HBQ154; Medline Kit SKU DYNJRA2672: UDI/DI 10198459475382 each, 40198459475383 case, Lot Number 26CBD940; Medline Kit SKU DYNJRA2672: UDI/DI 10198459475382 each, 40198459475383 case, Lot Number 25KBT183; Medline Kit SKU DYNJRA2672: UDI/DI 10198459475382 each, 40198459475383 case, Lot Number 25HBT583; Medline Kit SKU DYNJRA2675A: UDI/DI 10198459616471 each, 40198459616472 case, Lot Number 26ABV388; Medline Kit SKU DYNJRA2700: UDI/DI 10198459540615 each, 40198459540616 case, Lot Number 26BBC162; Medline Kit SKU DYNJRA2700: UDI/DI 10198459540615 each, 40198459540616 case, Lot Number 25LBA445; Medline Kit SKU DYNJRA2700: UDI/DI 10198459540615 each, 40198459540616 case, Lot Number 25JBO627; Medline Kit SKU DYNJRA2708: UDI/DI 10198459555756 each, 40198459555757 case, Lot Number 26BBN582; Medline Kit SKU DYNJRA2708: UDI/DI 10198459555756 each, 40198459555757 case, Lot Number 26BBL108; Medline Kit SKU DYNJRA2708: UDI/DI 10198459555756 each, 40198459555757 case, Lot Number 25KBT823; Medline Kit SKU DYNJRA2710: UDI/DI 10198459559075 each, 40198459559076 case, Lot Number 25LBL119; Medline Kit SKU DYNJRA2710: UDI/DI 10198459559075 each, 40198459559076 case, Lot Number 25KBM981; Medline Kit SKU DYNJRA2710: UDI/DI 10198459559075 each, 40198459559076 case, Lot Number 25JBD673; Medline Kit SKU DYNJRA9029: UDI/DI 10889942466436 each, 40889942466437 case, Lot Number 25KBD330; Medline Kit SKU DYNJRA9029: UDI/DI 10889942466436 each, 40889942466437 case, Lot Number 25KBB936; Medline Kit SKU DYNJRA9029: UDI/DI 10889942466436 each, 40889942466437 case, Lot Number 25JBB447; Medline Kit SKU DYNJRA9029: UDI/DI 10889942466436 each, 40889942466437 case, Lot Number 24FBJ029; Medline Kit SKU DYNJRA9029: UDI/DI 10889942466436 each, 40889942466437 case, Lot Number 24BBR611; Medline Kit SKU DYNJRA9029: UDI/DI 10889942466436 each, 40889942466437 case, Lot Number 23JBM198; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 26ABR020; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 26ABG927; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 25LBH190; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 25KBR348; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 25JBV917; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 25JBL130; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 25JBH184; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 25IBW244; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 25DBE496; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 25DBE497; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 25DBD657; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 25DBE019; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 25CBK276; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 25CBH117; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 25CBE763; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 25CBD665; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 24GBD232; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 24FBJ504; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 24FBJ509; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 24BBR622; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 24BBR604; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 23JBH420; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 23JBH419; Medline Kit SKU DYNJRA9034: UDI/DI 10193489467574 each, 40193489467575 case, Lot Number 23FBR086; Medline Kit SKU DYNJRA9038: UDI/DI 10193489830477 each, 40193489830478 case, Lot Number 26BBQ555; Medline Kit SKU DYNJRA9038: UDI/DI 10193489830477 each, 40193489830478 case, Lot Number 26BBM438; Medline Kit SKU DYNJRA9038: UDI/DI 10193489830477 each, 40193489830478 case, Lot Number 26ABM602; Medline Kit SKU DYNJRA9038: UDI/DI 10193489830477 each, 40193489830478 case, Lot Number 25KBS235; Medline Kit SKU DYNJRA9038: UDI/DI 10193489830477 each, 40193489830478 case, Lot Number 25GBO785; Medline Kit SKU DYNJRA9038: UDI/DI 10193489830477 each, 40193489830478 case, Lot Number 25FBM871; Medline Kit SKU DYNJRA9038: UDI/DI 10193489830477 each, 40193489830478 case, Lot Number 25ABC406; Medline Kit SKU DYNJRA9038: UDI/DI 10193489830477 each, 40193489830478 case, Lot Number 24FBJ512; Medline Kit SKU DYNJRA9038: UDI/DI 10193489830477 each, 40193489830478 case, Lot Number 24FBJ126; Medline Kit SKU DYNJRA9038: UDI/DI 10193489830477 each, 40193489830478 case, Lot Number 24BBR602; Medline Kit SKU PAIN1279A: UDI/DI 10193489302509 each, 40193489302500 case, Lot Number 26BBC168; Medline Kit SKU PAIN1279A: UDI/DI 10193489302509 each, 40193489302500 case, Lot Number 25LBF437; Medline Kit SKU PAIN1279A: UDI/DI 10193489302509 each, 40193489302500 case, Lot Number 25JBC449; Medline Kit SKU PAIN1279A: UDI/DI 10193489302509 each, 40193489302500 case, Lot Number 25HBU729; Medline Kit SKU PAIN1279A: UDI/DI 10193489302509 each, 40193489302500 case, Lot Number 25HBM736; Medline Kit SKU PAIN1279A: UDI/DI 10193489302509 each, 40193489302500 case, Lot Number 25GBW272; Medline Kit SKU PAIN1279A: UDI/DI 10193489302509 each, 40193489302500 case, Lot Number 25BBM498; Medline Kit SKU PAIN1279A: UDI/DI 10193489302509 each, 40193489302500 case, Lot Number 25ALA737; Medline Kit SKU PAIN1279A: UDI/DI 10193489302509 each, 40193489302500 case, Lot Number 25ALA309; Medline Kit SKU PAIN1279A: UDI/DI 10193489302509 each, 40193489302500 case, Lot Number 24KLA204; Medline Kit SKU PAIN1279A: UDI/DI 10193489302509 each, 40193489302500 case, Lot Number 24FLA466; Medline Kit SKU PAIN1279A: UDI/DI 10193489302509 each, 40193489302500 case, Lot Number 24ALB132; Medline Kit SKU PAIN1279A: UDI/DI 10193489302509 each, 40193489302500 case, Lot Number 23ILA681; Medline Kit SKU PAIN1279A: UDI/DI 10193489302509 each, 40193489302500 case, Lot Number 23HLA822; Medline Kit SKU PAIN1280: UDI/DI 10888277747784 each, 40888277747785 case, Lot Number 26BMJ295; Medline Kit SKU PAIN1280: UDI/DI 10888277747784 each, 40888277747785 case, Lot Number 26AMH186; Medline Kit SKU PAIN1280: UDI/DI 10888277747784 each, 40888277747785 case, Lot Number 25KMG729; Medline Kit SKU PAIN1280: UDI/DI 10888277747784 each, 40888277747785 case, Lot Number 25JMA658; Medline Kit SKU PAIN1280: UDI/DI 10888277747784 each, 40888277747785 case, Lot Number 25GMB271; Medline Kit SKU PAIN1280: UDI/DI 10888277747784 each, 40888277747785 case, Lot Number 25DMK280; Medline Kit SKU PAIN1280: UDI/DI 10888277747784 each, 40888277747785 case, Lot Number 25DMI542; Medline Kit SKU PAIN1280: UDI/DI 10888277747784 each, 40888277747785 case, Lot Number 25DMC359; Medline Kit SKU PAIN1280: UDI/DI 10888277747784 each, 40888277747785 case, Lot Number 24JMA154; Medline Kit SKU PAIN1280: UDI/DI 10888277747784 each, 40888277747785 case, Lot Number 24HMD114; Medline Kit SKU PAIN1280: UDI/DI 10888277747784 each, 40888277747785 case, Lot Number 24EMJ390; Medline Kit SKU PAIN1280: UDI/DI 10888277747784 each, 40888277747785 case, Lot Number 24DMJ999; Medline Kit SKU PAIN1280: UDI/DI 10888277747784 each, 40888277747785 case, Lot Number 24CMD645; Medline Kit SKU PAIN1280: UDI/DI 10888277747784 each, 40888277747785 case, Lot Number 24BMA231; Medline Kit SKU PAIN1280: UDI/DI 10888277747784 each, 40888277747785 case, Lot Number 23KMD774; Medline Kit SKU PAIN1280: UDI/DI 10888277747784 each, 40888277747785 case, Lot Number 23JMF327; Medline Kit SKU PAIN1280: UDI/DI 10888277747784 each, 40888277747785 case, Lot Number 23JMB758; Medline Kit SKU PAIN1280: UDI/DI 10888277747784 each, 40888277747785 case, Lot Number 23GMG995; Medline Kit SKU PAIN1280: UDI/DI 10888277747784 each, 40888277747785 case, Lot Number 24DMJ999A; Medline Kit SKU PAIN1289A: UDI/DI 10193489495478 each, 40193489495479 case, Lot Number 26AMG738; Medline Kit SKU PAIN1289A: UDI/DI 10193489495478 each, 40193489495479 case, Lot Number 24GMH190; Medline Kit SKU PAIN1393: UDI/DI 10193489629965 each, 40193489629966 case, Lot Number 25LMH336; Medline Kit SKU PAIN1393: UDI/DI 10193489629965 each, 40193489629966 case, Lot Number 25LMF024; Medline Kit SKU PAIN1393: UDI/DI 10193489629965 each, 40193489629966 case, Lot Number 25HMB657; Medline Kit SKU PAIN1393: UDI/DI 10193489629965 each, 40193489629966 case, Lot Number 25GMI070; Medline Kit SKU PAIN1393: UDI/DI 10193489629965 each, 40193489629966 case, Lot Number 25FMA572; Medline Kit SKU PAIN1393: UDI/DI 10193489629965 each, 40193489629966 case, Lot Number 24JMA110; Medline Kit SKU PAIN1393: UDI/DI 10193489629965 each, 40193489629966 case, Lot Number 23HBM570; Medline Kit SKU PAIN1393: UDI/DI 10193489629965 each, 40193489629966 case, Lot Number 23HBE099; Medline Kit SKU PAIN1393: UDI/DI 10193489629965 each, 40193489629966 case, Lot Number 25HMB657B; Medline Kit SKU PAIN1405B: UDI/DI 10193489768206 each, 40193489768207 case, Lot Number 26BBN581; Medline Kit SKU PAIN1405B: UDI/DI 10193489768206 each, 40193489768207 case, Lot Number 26ABS329; Medline Kit SKU PAIN1405B: UDI/DI 10193489768206 each, 40193489768207 case, Lot Number 25IBE918; Medline Kit SKU PAIN1405B: UDI/DI 10193489768206 each, 40193489768207 case, Lot Number 25GBT026; Medline Kit SKU PAIN1405B: UDI/DI 10193489768206 each, 40193489768207 case, Lot Number 25FBL017; Medline Kit SKU PAIN1405B: UDI/DI 10193489768206 each, 40193489768207 case, Lot Number 25EBC429; Medline Kit SKU PAIN1405B: UDI/DI 10193489768206 each, 40193489768207 case, Lot Number 25ALA031; Medline Kit SKU PAIN1405B: UDI/DI 10193489768206 each, 40193489768207 case, Lot Number 24CLA245; Medline Kit SKU PAIN1405B: UDI/DI 10193489768206 each, 40193489768207 case, Lot Number 23ILA303; Medline Kit SKU SAMPA0124: UDI/DI 10195327691691 each, 40195327691692 case, Lot Number 24GBL109; Medline Kit SKU SAMPA0128: UDI/DI 10195327691813 each, 40195327691814 case, Lot Number 24GBF551; Medline Kit SKU SAMPA0139: UDI/DI 10198459042034 each, 40198459042035 case, Lot Number 24FBT389; Medline Kit SKU SAMPA0144: UDI/DI 10198459292194 each, 40198459292195 case, Lot Number 25CBD916; Medline Kit SKU SAMPA0146: UDI/DI 10198459421815 each, 40198459421816 case, Lot Number 25GBQ852.
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries, LP
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactHaley Barclay
800-359-1704
Manufacturer Reason
for Recall
Medline kits were distributed containing Huons Bupivacaine Hydrochloride in Dextrose Injection, USP, which was subsequently recalled due to quality issues, including microbiology testing data integrity concerns and reported efficacy complaints.
FDA Determined
Cause 2
Component design/selection
ActionMedline Industries, LP issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 4/10/2026 via USPS first class mail and email. The notice explained the issue with the component and requested the following: - Identify and quarantine the affected item and lot numbers (kit). - Respond to notice via the portal: Website link: https://recalls.medline.com, Recall Reference #: R-26-065-FGX1, using the Recall Code provided. - Once the response is received, over-labels will be provided to be placed on the affected inventory with instructions to remove and discard the affected component. - Distributors or those who have further distributed the product are directed to notify those to whom the affected kits were distributed and include the customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity in Commerce200702 kits
DistributionWorldwide - US Nationwide distribution in the US Territory of Virgin Islands and the countries of Bahamas, Panama, Barbados.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-