| | Class 2 Device Recall Olympus 4K UHD LCD Monitor |  |
| Date Initiated by Firm | July 07, 2026 |
| Date Posted | July 20, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2761-2026 |
| Recall Event ID |
99288 |
| 510(K)Number | K220069 |
| Product Classification |
Laparoscope, general & plastic surgery - Product Code GCJ
|
| Product | Olympus 4K UHD LCD Monitor. Model Number: OEV321UH.
An electrically powered visual display unit designed for displaying graphical information in color. |
| Code Information |
Model Number: OEV321UH. UDI-DI: 04953170418723. Serial Numbers: 7503474, 7503477, 7503479, 7503263, 7503264, 7503262, 7503261, 7503260, 7503259, 7503335.
|
| FEI Number |
2429304
|
Recalling Firm/ Manufacturer |
Olympus Corporation of the Americas 3500 Corporate Pkwy Center Valley PA 18034-8229
|
| For Additional Information Contact | Ashley Troiani 484-5531029 |
Manufacturer Reason for Recall | Potential for the HDMI and DisplayPort (DP) connectors can become faulty resulting in image loss or image abnormalities. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | On July 7, 2026, the firm issued "Urgent: Medical Device Correction" letters to consignees.
Customers were informed of the product issue.
While the identified 4K UHD LCD monitor may be used with VISERA S, VISERA ELITE III, EVIS X1, and EU-ME3 systems, the firm notes that the affected HDMI or DP connectors are more commonly used with the VISERA S video system center. Its use with the three other systems does not typically involve the affected HDMI or DP connectors.
Customers are requested to examine their equipment, looking for the identified monitor model number and serial numbers listed in the customer letter. They should then acknowledge receipt of the letter through the firm's recall web portal. The firm will coordinate repair of the monitors.
If customer needs to use the monitor prior to its repair, the firm recommends that a spare unit be available in case issues are encountered with the primary unit during use. |
| Quantity in Commerce | 10 units |
| Distribution | Worldwide - US Nationwide distribution in the states of OH, CA, NJ, FL and the country of Japan.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|