| | Class 2 Device Recall GEM Premier 5000 |  |
| Date Initiated by Firm | May 08, 2026 |
| Date Posted | July 29, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2823-2026 |
| Recall Event ID |
99115 |
| 510(K)Number | K203790 |
| Product Classification |
Electrode measurement, blood-gases (pco2, po2) and blood ph - Product Code CHL
|
| Product | Brand Name: GEM Premier 5000
Product Name: GEM Premier 5000
Model/Catalog Number: 00024019255
portable blood analyzer system |
| Code Information |
GEM Premier 5000:
Part number: 00024019255
UDI: 08426950870227
Units distributed in Hong Kong:
GEM Premier 5000 Serial Nos.: 20106073, 20106039, 19052933, 20116415, 20064792, 22062065, 22062067, 19103490, 22092716, 22102881, 18112339, 19012518, 20126582, 20075256, 20064794, 20106036, 24016057, 21088758, 21088759, 18061910, 24016081, 24016108, 24108036, 24108044, 24067370, 24067371 and 24077468.
Affected Accessory:
Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026.
Part No. 00024015859; UDI No. 08426950869634
|
| FEI Number |
1217183
|
Recalling Firm/ Manufacturer |
Instrumentation Laboratory
180 Hartwell Rd
Bedford
MA
01730-2443
|
| For Additional Information Contact | 781-861-4467 |
Manufacturer Reason for Recall | Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID. |
FDA Determined Cause 2 | Software Design Change |
| Action | Werfen dba Instrumentation laboratory notified consignees on about 05/08/2026 via letter and/or email. Consignees were instructed to discontinue use of the affected BCR wands immediately, manually enter patient IDs when performing testing on the GEM Premier 5000 and/or 7000 until the BCR wands have been replaced, contact Werfen Service at 1-800-955-9525 to arrange for prompt replacement of the affected BCR wands, post this notification on or near affected GEM Premier 5000 and/or 7000 instruments, and inform all laboratory staff of this notification. Consignees were also instructed to forward the notification to all affected locations within each facility, retain a copy of the notification for records, and complete and return the Mandatory Response Tracking Form within 10 days of receipt.
In addition to the tracked customer notification, a Service Information (SI) 2026-09 was issued on May 11, 2026, for service personnel to track the destruction of the affected BCR wands. The Service Information includes a destruction form for service personnel to document the destruction of the affected BCR wands and to return the form to Regulatory for tracking purposes. |
| Quantity in Commerce | 27 |
| Distribution | US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|