• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Medline

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Medlinesee related information
Date Initiated by FirmJuly 20, 2026
Date PostedAugust 31, 2026
Recall Status1 Open3, Classified
Recall NumberZ-3079-2026
Recall Event ID 99561
Product Classification Orthopedic tray - Product Code OJH
ProductMedline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators ORTHO ASPIRATION KIT DYKM1636A
Code Information UDI-DI 10889942925858 Lot 26EBI940
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries, LP
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactHaley Barclay
800-633-5463
Manufacturer Reason
for Recall
Downstream recall - manufacturer recalled specific lots of BD ChloraPrep Clear - 1mL Applicators. Affected product may exhibit an open or incomplete seal on the packaging of the applicator. Use of potentially contaminated applicator may lead to localized skin or soft tissue infection, or systemic infection or infectious complications.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionFirm began notifying consignees on 7/20/2026 via First-Class Mail and email. Medline Industries LP instructed customers to request stickers to over-label the affected convenience kits with a label indicating to remove the affected component and replace it with product from supply. No kits are to be returned.
Quantity in Commerce222,653 total
DistributionUS Nationwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-