| | Class 2 Device Recall HeartSine |  |
| Date Initiated by Firm | July 27, 2026 |
| Date Posted | September 04, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-3120-2026 |
| Recall Event ID |
99568 |
| PMA Number | P160008 |
| Product Classification |
Automated external defibrillators (non-wearable) - Product Code MKJ
|
| Product | HeartSine SAM 350P
350P US Device, US English - 2 Min CPR
UDI-DI codes:
0506016712067
350P US Device, US English (American Airlines) - 2 Min CPR
UDI-DI code:
0506016712207
350P US Device, US English with Gateway - 2 Min CPR
UDI-DI code:
0506016712804
HeartSine SAM 360P
360P US Device, US English - 2 Min CPR
UDI-DI code:
0506016712068
HeartSine SAM 450P
SAM 450P US Device, US English (2 min CPR)
UDI-DI code:
0506016712069
The HeartSine samaritan PAD (Public Access Defibrillator) Models SAM 350P, SAM 360P and SAM 450P are indicated for use on victims of cardiac arrest who are exhibiting the following signs:
" Unconscious
" Not breathing
" Without circulation (without a pulse)
The devices are indicated for use on patients greater than 8 years old or over 55 lbs. (25 kg) when used with the adult Pad-Pak (Pad-Pak-01 or Pad-Pak-07). They are indicated for use on children between 1 and 8 years of age or up to 55 lbs. (25 kg) when used with the Pediatric-Pad (Pad-Pak-02). |
| Code Information |
350P US Device, US English - 2 Min CPR
Item code: 350-STR-US-10
UDI-DI codes: 0506016712067
Serial Numbers:
22D91164357L22
22D91164989L22
22D91180808K23
23D98000415K23
23D98000597K23
23D98001134L23
23D98001634I23
23D98001722F23
23D98002004K23
23D98002054K23
23D98002059I23
23D98002086L23
23D98002225K23
23D98002639K23
23D98002649K23
23D98002771L23
23D98002897K23
23D98003187A24
23D98003188L23
23D98003198F23
23D98003221K23
23D98003523K23
23D98004710K23
23D98004956L23
23D98005238E23
23D98005242K23
23D98008495L23
23D98008520K23
23D98008591K23
23D98008832K23
23D98008910E23
23D98009571F23
23D98009597F23
23D98009629I23
23D98009835L23
23D98009905F23
23D98010830K23
23D98012539K23
23D98012748G23
23D98013452K23
23D98013684F23
23D98014324L23
23D98014750K23
23D98014988K23
23D98015098K23
23D98015120K23
23D98015207K23
23D98015221K23
23D98015222K23
23D98015231L23
23D98015571F23
23D98016139F23
23D98016168K23
23D98016187F23
23D98016436K23
23D98016463F23
23D98016490L23
23D98016718K23
23D98016725K23
23D98017989L23
23D98018269K23
23D98018271K23
23D98018272J23
23D98018274J23
23D98018277J23
23D98018280K23
23D98018284J23
23D98018285K23
23D98018286J23
23D98018287J23
23D98018290K23
23D98018291K23
23D98018292J23
23D98018294F23
23D98020005K23
23D98023298L23
23D98024058K23
23D98024075K23
23D98100344K23
23D98100567I23
23D98100620L23
23D98100627I23
23D98102225K23
23D98102340K23
23D98102422K23
23D98102666L23
23D98103547K23
23D98103822J23
23D98103971J23
23D98103972K23
23D98103973J23
23D98103974J23
23D98103975J23
23D98103976J23
23D98103977J23
23D98103979J23
23D98103980J23
23D98103981J23
23D98103982J23
23D98103983J23
23D98103984J23
23D98103985J23
23D98103988J23
23D98103989J23
23D98103990J23
23D98103991L23
23D98103992J23
23D98103993J23
23D98103994J23
23D98103995J23
23D98104017L23
23D98104195J23
23D98104196J23
23D98104197K23
23D98104198J23
23D98104199J23
23D98104231J23
23D98104232K23
23D98104234J23
23D98104235J23
23D98104236K23
23D98104237J23
23D98104238J23
23D98104239J23
23D98104240J23
23D98104243J23
23D98104244J23
23D98104245J23
23D98104246J23
23D98104247J23
23D98104249J23
23D98104250J23
23D98104251J23
23D98104252J23
23D98104253J23
23D98104254J23
23D98104256J23
23D98104257J23
23D98104258J23
23D98104259J23
23D98104260J23
23D98104261J23
23D98104262J23
23D98104263J23
23D98104264J23
23D98104265J23
23D98104266J23
23D98104267J23
23D98104268J23
23D98104269J23
23D98104270J23
23D98104272J23
23D98104273J23
23D98104274J23
23D98104275J23
23D98104276J23
23D98104277J23
23D98104278J23
23D98104279J23
23D98104286J23
23D98104292L23
23D98104310K23
23D98104315J23
23D98104327K23
23D98104543J23
23D98104544J23
23D98104545J23
23D98104546J23
23D98104547J23
23D98104548J23
23D98104549J23
23D98104551J23
23D98104553J23
23D98104555J23
23D98104556J23
23D98104557J23
23D98104558J23
23D98104561J23
23D98104562J23
23D98104631K23
23D98104659L23
23D98104705J23
23D98104706J23
23D98104707J23
23D98104708J23
23D98104709J23
23D98104710J23
23D98104717J23
23D98104719J23
23D98104720J23
23D98104721J23
23D98104722J23
23D98104723J23
23D98104724J23
23D98104725J23
23D98104726J23
23D98104728J23
23D98104729J23
23D98104730J23
23D98104731J23
23D98104732J23
23D98104733J23
23D98104734J23
23D98104735J23
23D98104736J23
23D98104738J23
23D98104739J23
23D98104740J23
23D98104742J23
23D98104743J23
23D98104744J23
23D98104745J23
23D98104746J23
23D98104747J23
23D98104748J23
23D98104749J23
23D98104751J23
23D98104754J23
23D98104755J23
23D98104756J23
23D98104757J23
23D98104762J23
23D98104763J23
23D98104772J23
23D98104773J23
23D98104776J23
23D98104777J23
23D98104778J23
23D98104779J23
23D98104781J23
23D98104782J23
23D98104783J23
23D98104784J23
23D98104785J23
23D98104786J23
23D98104787J23
23D98104788J23
23D98104789J23
23D98104790J23
23D98105005K23
23D98105009K23
23D98105076J23
23D98105164J23
23D98105195J23
23D98105196J23
23D98105197J23
23D98105198J23
23D98105199J23
23D98105200J23
23D98105201J23
23D98105202J23
23D98105215J23
23D98105218J23
23D98105219J23
23D98105275J23
23D98105481L23
23D98105845J23
23D98105873J23
23D98105983J23
23D98105984J23
23D98105986J23
23D98105988J23
23D98106003J23
23D98106247J23
23D98106274K23
23D98107424K23
23D98108605K23
23D98108621I23
23D98108660K23
23D98108719K23
23D98108961I23
23D98109140I23
23D98109142I23
23D98109144I23
23D98109147I23
23D98109153I23
23D98109163L23
23D98109181K23
23D98109216I23
23D98109253K23
23D98109395I23
23D98109398I23
23D98109402I23
23D98109408I23
23D98109409I23
23D98109415I23
23D98109420I23
23D98109423I23
23D98109425I23
23D98109426I23
23D98109453K23
23D98109554L23
23D98109561K23
23D98109631I23
23D98109632I23
23D98109633I23
23D98109634I23
23D98109635I23
23D98109636I23
23D98109637I23
23D98109686I23
23D98109712I23
23D98109713I23
23D98109716I23
23D98109734K23
23D98109800L23
23D98109824I23
23D98109968K23
23D98110031I23
23D98110035I23
23D98110036I23
23D98110039I23
23D98110040I23
23D98110041I23
23D98110042I23
23D98110128I23
23D98110131I23
23D98110137I23
23D98110140I23
23D98110146I23
23D98110208I23
23D98110210I23
23D98110211I23
23D98110213I23
23D98110217I23
23D98110219I23
23D98110221I23
23D98110223I23
23D98110224I23
23D98110236I23
23D98110239I23
23D98110255I23
23D98110270I23
23D98110271I23
23D98110273I23
23D98110274I23
23D98110307K23
23D98110405I23
23D98110406I23
23D98110407I23
23D98110409I23
23D98110410I23
23D98110411I23
23D98110412I23
23D98110413I23
23D98110414I23
23D98110416I23
23D98110417I23
23D98110418I23
23D98110419I23
23D98110420I23
23D98110421I23
23D98110422I23
23D98110423I23
23D98110424I23
23D98110425I23
23D98110426I23
23D98110427I23
23D98110428I23
23D98110429I23
23D98110430I23
23D98110431I23
23D98110432I23
23D98110433I23
23D98110434I23
23D98110435I23
23D98110436I23
23D98110437I23
23D98110438I23
23D98110439I23
23D98110440I23
23D98110441I23
23D98110442I23
23D98110443I23
23D98110444I23
23D98110445I23
23D98110446I23
23D98110447I23
23D98110448I23
23D98110449I23
23D98110450I23
23D98110452I23
23D98110453I23
23D98110454I23
23D98110455I23
23D98110456I23
23D98110457I23
23D98110458I23
23D98110459I23
23D98110460I23
23D98110461I23
23D98110462I23
23D98110463I23
23D98110464I23
23D98110465I23
23D98110466I23
23D98110468I23
23D98110469I23
23D98110470I23
23D98110471I23
23D98110472I23
23D98110473I23
23D98110474I23
23D98110475I23
23D98110476I23
23D98110477I23
23D98110478I23
23D98110479I23
23D98110480I23
23D98110481I23
23D98110482I23
23D98110483I23
23D98110484I23
23D98110485I23
23D98110486I23
23D98110488I23
23D98110489I23
23D98110490I23
23D98110491I23
23D98110492I23
23D98110493I23
23D98110494I23
23D98110495I23
23D98110496I23
23D98110497I23
23D98110510H23
23D98110513H23
23D98110519H23
23D98110647H23
23D98110702K23
23D98110731H23
23D98110746K23
23D98111340K23
23D98111416L23
23D98111689I23
23D98111690I23
23D98111691I23
23D98111693I23
23D98111694I23
23D98111695I23
23D98111697I23
23D98111698I23
23D98111699I23
23D98111700I23
23D98111701K23
23D98111702I23
23D98111703I23
23D98111704I23
23D98111705I23
23D98111706I23
23D98111707I23
23D98111708I23
23D98111709I23
23D98111710I23
23D98111711I23
23D98111712I23
23D98111713I23
23D98111714I23
23D98111715I23
23D98111716I23
23D98111717K23
23D98111718I23
23D98111719I23
23D98111720I23
23D98111721I23
23D98111722I23
23D98111723I23
23D98111724I23
23D98111725I23
23D98111726I23
23D98111727I23
23D98111728I23
23D98111729I23
23D98111730I23
23D98111731I23
23D98111732I23
23D98111733I23
23D98111734I23
23D98111735I23
23D98111736I23
23D98111811I23
23D98111812I23
23D98111813I23
23D98111814I23
23D98111815I23
23D98111816I23
23D98111817I23
23D98111818I23
23D98111819I23
23D98111820I23
23D98111821I23
23D98111822I23
23D98111823I23
23D98111824I23
23D98111825I23
23D98111826I23
23D98111827I23
23D98111828I23
23D98111829I23
23D98111830I23
23D98111831I23
23D98111832I23
23D98111833I23
23D98111834I23
23D98111835I23
23D98111836I23
23D98111837I23
23D98111838I23
23D98111839I23
23D98111840I23
23D98111841I23
23D98111842I23
23D98111843I23
23D98111844I23
23D98111845I23
23D98111846I23
23D98111847I23
23D98111848I23
23D98111849I23
23D98111850I23
23D98111851I23
23D98111852I23
23D98111853I23
23D98111854I23
23D98111855I23
23D98111856I23
23D98111858I23
23D98111860I23
23D98111891I23
23D98111905I23
23D98111913I23
23D98111928I23
23D98112342J23
23D98112343J23
23D98112344J23
23D98112345J23
23D98112346J23
23D98112347J23
23D98112348J23
23D98112349J23
23D98112350J23
23D98112351J23
23D98112353J23
23D98112354J23
23D98112364J23
23D98112365J23
23D98112366J23
23D98112367J23
23D98112369J23
23D98112370J23
23D98112371J23
23D98112372J23
23D98112373J23
23D98112374J23
23D98112375J23
23D98112376J23
23D98112377J23
23D98112378J23
23D98112379J23
23D98112380J23
23D98112381J23
23D98112523K23
23D98112541J23
23D98112690I23
23D98112699I23
23D98112883J23
23D98114373J23
23D98114956K23
23D98114957J23
23D98114958J23
23D98114961J23
23D98114962J23
23D98114965J23
23D98114966J23
23D98114969J23
23D98114970J23
23D98114973J23
23D98114975J23
23D98115598H23
23D98115711K23
23D98115718K23
23D98115829L23
23D98115907K23
23D98115956K23
23D98116225G23
23D98117758L23
23D98117759L23
23D98117767L23
23D98117768L23
23D98117770L23
23D98117922I23
23D98117924I23
23D98117925I23
23D98117926I23
23D98117927I23
23D98117928I23
23D98117929I23
23D98117930I23
23D98117931I23
23D98118483L23
23D98118517J23
23D98118673I23
23D98118936J23
23D98119204L23
23D98119848K23
23D98119850K23
23D98119851K23
23D98119852K23
23D98119927L23
23D98119928L23
23D98120303I23
23D98120305I23
23D98120306I23
23D98120307I23
23D98120308I23
23D98120311I23
23D98120439I23
23D98120447I23
23D98120462I23
23D98120466I23
23D98120471I23
23D98120474I23
23D98120477I23
23D98120482I23
23D98120488I23
23D98120489I23
23D98120496I23
23D98120497I23
23D98120498I23
23D98120499I23
23D98120500I23
23D98120511I23
23D98120512I23
23D98120552I23
23D98120553I23
23D98120566K23
23D98120660I23
23D98120661I23
23D98120665I23
23D98120667I23
23D98120671I23
23D98120672I23
23D98120673I23
23D98120674I23
23D98120846I23
23D98120872K23
23D98120874I23
23D98120875I23
23D98120877I23
23D98120879I23
23D98120880I23
23D98120881I23
23D98120889I23
23D98120898I23
23D98120918K23
23D98121054J23
23D98121058J23
23D98121059J23
23D98121062J23
23D98121065K23
23D98121096J23
23D98121097J23
23D98121098J23
23D98121099J23
23D98121202J23
23D98121203J23
23D98121204J23
23D98121205J23
23D98121206J23
23D98121207J23
23D98121208J23
23D98121209J23
23D98121210J23
23D98121211J23
23D98121212J23
23D98121213J23
23D98121214J23
23D98121215J23
23D98121216J23
23D98121217J23
23D98121218J23
23D98121219J23
23D98121220J23
23D98121221J23
23D98121222J23
23D98121223J23
23D98121224J23
23D98121225J23
23D98121226J23
23D98121227J23
23D98121228J23
23D98121229J23
23D98121230J23
23D98121231J23
23D98121232J23
23D98121234J23
23D98121235J23
23D98121236J23
23D98121238J23
23D98121239J23
23D98121240J23
23D98121241J23
23D98121242J23
23D98121243J23
23D98121244J23
23D98121245J23
23D98121246J23
23D98121247L23
23D98121248J23
23D98121249J23
23D98121251J23
23D98121305J23
23D98121306J23
23D98121308J23
23D98121311J23
23D98121314J23
23D98121317J23
23D98121320J23
23D98121321J23
23D98121322J23
23D98121323J23
23D98121433J23
23D98121435J23
23D98121437J23
23D98121447J23
23D98121448J23
23D98121450J23
23D98121452J23
23D98121454J23
23D98121462J23
23D98121465J23
23D98121467J23
23D98121468J23
23D98121470J23
23D98121472J23
23D98121473J23
23D98121475J23
23D98121476J23
23D98121741J23
23D98121776J23
23D98121787J23
23D98121796J23
23D98121802L23
23D98121803J23
23D98121804J23
23D98121806J23
23D98121809J23
23D98121847J23
23D98121975K23
23D98121976K23
23D98123159K23
23D98123496K23
23D98123945I23
23D98124382J23
23D98124383J23
23D98124384J23
23D98124385J23
23D98124386J23
23D98124387J23
23D98124388J23
23D98124389J23
23D98124783J23
23D98124784J23
23D98124785J23
23D98124786J23
23D98124788J23
23D98124789J23
23D98124790J23
23D98124791J23
23D98124796J23
23D98124806J23
23D98124807J23
23D98124808J23
23D98124809J23
23D98124810J23
23D98124811J23
23D98124812J23
23D98124813J23
23D98124814J23
23D98124815J23
23D98124816J23
23D98124817J23
23D98124818J23
23D98124819K23
23D98124820J23
23D98124822J23
23D98124823J23
23D98124824J23
23D98124825J23
23D98124826J23
23D98124827K23
23D98124828J23
23D98124829J23
23D98124830J23
23D98124831J23
23D98124832J23
23D98124835J23
23D98124836J23
23D98124837J23
23D98124838J23
23D98124839J23
23D98124840J23
23D98124841J23
23D98124842J23
23D98124843J23
23D98124844J23
23D98124845J23
23D98124846J23
23D98124847J23
23D98124848J23
23D98124849J23
23D98124850J23
23D98124851J23
23D98124854J23
23D98124856J23
23D98124857J23
23D98124858J23
23D98124859J23
23D98124860J23
23D98124861J23
23D98124862J23
23D98124864J23
23D98124865J23
23D98124866J23
23D98124867J23
23D98124868J23
23D98124869J23
23D98124870J23
23D98124871J23
23D98124872J23
23D98124873J23
23D98124874J23
23D98124875J23
23D98124876J23
23D98124877J23
23D98124878J23
23D98124879J23
23D98124880J23
23D98124881J23
23D98124882J23
23D98124883J23
23D98124884J23
23D98124885J23
23D98124970J23
23D98124971J23
23D98124991J23
23D98124994J23
23D98124995J23
23D98124997J23
23D98124998J23
23D98124999J23
23D98125000J23
23D98125023L23
23D98125025G23
23D98125026L23
23D98125027L23
23D98125216L23
23D98125237L23
23D98125340H23
23D98125341H23
23D98125343H23
23D98125396H23
23D98125398H23
23D98125400H23
23D98125434H23
23D98125435H23
23D98125436H23
23D98125438H23
23D98125441H23
23D98125443H23
23D98125462I23
23D98125463I23
23D98125466I23
23D98125467I23
23D98125469I23
23D98125471I23
23D98125475I23
23D98125476I23
23D98125478I23
23D98125479I23
23D98125480I23
23D98125481H23
23D98125489I23
23D98125490I23
23D98125491I23
23D98125492I23
23D98125493I23
23D98125495I23
23D98125496I23
23D98125497I23
23D98125498I23
23D98125499I23
23D98125826K23
23D98125828L23
23D98126091L23
23D98126104I23
23D98126174K23
23D98126224K23
23D98126297I23
23D98126372J23
23D98126405J23
23D98126500I23
23D98126550J23
23D98126755I23
23D98126756I23
23D98126757J23
23D98126758J23
23D98126759J23
23D98126760J23
23D98126761J23
23D98126762J23
23D98126763J23
23D98126764J23
23D98126765J23
23D98126766J23
23D98126768J23
23D98126769J23
23D98126771J23
23D98126772J23
23D98126773J23
23D98126775J23
23D98126784J23
23D98126787J23
23D98126814L23
23D98126816J23
23D98126817J23
23D98126818J23
23D98126819J23
23D98126820J23
23D98126822J23
23D98126823J23
23D98126824J23
23D98126825J23
23D98126826J23
23D98126827J23
23D98126828J23
23D98126829J23
23D98126919J23
23D98126956J23
23D98126957J23
23D98126958J23
23D98126959J23
23D98126960J23
23D98126961J23
23D98126978K23
23D98126990J23
23D98126993J23
23D98126999J23
23D98127005J23
23D98127010J23
23D98127028J23
23D98127029J23
23D98127057J23
23D98127058J23
23D98127060J23
23D98127061J23
23D98127085J23
23D98127138J23
23D98127148J23
23D98127149J23
23D98127150J23
23D98127151J23
23D98127152J23
23D98127153J23
23D98127154J23
23D98127155J23
23D98127156J23
23D98127157J23
23D98127158J23
23D98127159J23
23D98127160J23
23D98127161J23
23D98127162J23
23D98127163J23
23D98127164J23
23D98127165J23
23D98127166J23
23D98127167J23
23D98127211J23
23D98127213J23
23D98127527K23
23D98127528L23
23D98127653H23
23D98127673K23
23D98127699K23
23D98127719K23
23D98127756K23
23D98128030K23
23D98128047H23
23D98128048H23
23D98128169H23
23D98128951H23
23D98128952H23
23D98128961H23
23D98128962H23
23D98128964H23
23D98129394I23
23D98129395I23
23D98129396I23
23D98129397I23
23D98129413H23
23D98129414H23
23D98129416H23
23D98129417H23
23D98129421H23
23D98129431H23
23D98129437H23
23D98129441H23
23D98129443H23
23D98129447H23
23D98129467H23
23D98129733H23
23D98129928K23
23D98129965L23
23D98130050K23
23D98130218H23
23D98130230H23
23D98130234H23
23D98130235H23
23D98130236H23
23D98130237H23
23D98130276H23
23D98130472K23
23D98130580A24
23D98130670H23
23D98130672H23
23D98130674H23
23D98130676H23
23D98130684H23
23D98130685H23
23D98130689H23
23D98130695H23
23D98130698H23
23D98130701H23
23D98130704H23
23D98130709H23
23D98130741H23
23D98130753H23
23D98130757H23
23D98130760H23
23D98130773H23
23D98130875H23
23D98130885H23
23D98130890H23
23D98130891H23
23D98130893H23
23D98130900H23
23D98130901H23
23D98130902H23
23D98130905H23
23D98130907H23
23D98130909H23
23D98130910H23
23D98130911H23
23D98130912H23
23D98130913H23
23D98130915H23
23D98130916H23
23D98130917H23
23D98130918H23
23D98130941H23
23D98130968H23
23D98131033K23
23D98131117H23
23D98131121H23
23D98131154H23
23D98131769K23
23D98131817H23
23D98132006K23
23D98132106H23
23D98132159K23
23D98132204L23
23D98132205L23
23D98132206L23
23D98132208L23
23D98132209L23
23D98132210L23
23D98132212L23
23D98132213L23
23D98132215L23
23D98132217L23
23D98132777L23
23D98132829L23
23D98132954J23
23D98132959G23
23D98133372H23
23D98133373H23
23D98133374H23
23D98133375H23
23D98133376H23
23D98133377H23
23D98133378H23
23D98133389K23
23D98133440H23
23D98133441H23
23D98133444H23
23D98133445H23
23D98133446H23
23D98134555K23
23D98134560H23
23D98134973H23
23D98135234K23
23D98135311H23
23D98135330K23
23D98135944H23
23D98135947H23
23D98135949H23
23D98135970K23
23D98135976H23
23D98135982A24
23D98135986H23
23D98135988H23
23D98135992H23
23D98136592K23
23D98136730L23
23D98136950K23
23D98136951H23
23D98136952H23
23D98136953H23
23D98136954H23
23D98136956H23
23D98136957H23
23D98136960H23
23D98136961H23
23D98136963H23
23D98136964H23
23D98136966H23
23D98136967K23
23D98136968H23
23D98136969H23
23D98136970H23
23D98137174A24
23D98137218H23
23D98137248H23
23D98137253H23
23D98137266H23
23D98137331L23
23D98137451H23
23D98139012K23
23D98139015I23
23D98139021I23
23D98139128I23
23D98139130I23
23D98139319J23
23D98139420I23
23D98139422I23
23D98139423I23
23D98139424I23
23D98139425I23
23D98139426I23
23D98139427I23
23D98139428I23
23D98139429I23
23D98139430I23
23D98139431I23
23D98139432I23
23D98139433I23
23D98139434I23
23D98139435I23
23D98139436I23
23D98139437I23
23D98139438I23
23D98139439I23
23D98139440I23
23D98139441I23
23D98139442I23
23D98139443I23
23D98139444I23
23D98139445I23
23D98139446I23
23D98139448I23
23D98139449I23
23D98139450I23
23D98139451I23
23D98139452I23
23D98139453I23
23D98139460K23
23D98139468L23
23D98139470K23
23D98139502J23
23D98139505J23
23D98139506J23
23D98139699I23
23D98139700I23
23D98139701I23
23D98139702I23
23D98139703I23
23D98139704I23
23D98139707I23
23D98139708I23
23D98139710I23
23D98139711I23
23D98139712I23
23D98139713I23
23D98139714I23
23D98139715I23
23D98139716I23
23D98139717I23
23D98139718I23
23D98139719I23
23D98139720I23
23D98139741I23
23D98139742I23
23D98139743I23
23D98139744I23
23D98139753I23
23D98139754I23
23D98139755I23
23D98139756I23
23D98139757I23
23D98139758I23
23D98139759I23
23D98139760I23
23D98139763I23
23D98139764I23
23D98139765I23
23D98139767I23
23D98139768I23
23D98139851I23
23D98140913K23
23D98141558K23
23D98141686L23
23D98141688L23
23D98143016L23
23D98145053J23
23D98145057J23
23D98145058J23
23D98145059J23
23D98145062J23
23D98145063J23
23D98145064J23
23D98145383J23
23D98146003K23
23D98146973L23
23D98146988L23
23D98147006L23
23D98147484K23
23D98147602K23
23D98147860L23
23D98147861L23
23D98147862L23
23D98147863L23
23D98147864L23
23D98147866L23
23D98147867L23
23D98147868L23
23D98147881L23
23D98147972L23
23D98147973L23
23D98147980L23
23D98147982L23
23D98147984L23
23D98147985L23
23D98147991L23
23D98147993L23
23D98147995L23
23D98147996L23
23D98147998L23
23D98148020K23
23D98148151J23
23D98148168J23
23D98148280K23
23D98148284J23
23D98148289J23
23D98148293J23
23D98148294J23
23D98148383J23
23D98148499J23
23D98148513J23
23D98148518J23
23D98148522J23
23D98148525J23
23D98148536J23
23D98148539J23
23D98148540J23
23D98149629K23
23D98149630K23
23D98149637K23
23D98149639K23
23D98149654K23
23D98149787K23
23D98149795K23
23D98149797K23
23D98149825K23
23D98149864K23
23D98149866K23
24D98029184A24
24D98029186A24
24D98029359A24
24D98029370A24
24D98029373A24
24D98029383A24
24D98029977A24
24D98029979A24
24D98029980A24
24D98030065A24
24D98030158A24
24D98030159A24
24D98030160A24
24D98030161A24
24D98030164A24
24D98030165A24
24D98030168A24
24D98030169A24
24D98030170A24
24D98030171A24
24D98030174A24
24D98030175A24
24D98030176A24
24D98030177A24
24D98030178A24
24D98030180A24
24D98030182A24
24D98030183A24
24D98030232A24
24D98030525A24
24D98030534A24
24D98030700A24
24D98030702A24
24D98030703A24
24D98030705A24
24D98030934A24
24D98030936A24
24D98030938A24
24D98030939A24
24D98034006A24
24D98034008A24
24D98034420A24
24D98034421A24
24D98034422A24
24D98034423A24
24D98037331A24
24D98037546A24
24D98037560A24
24D98037625A24
24D98039525B24
24D98039538B24
24D98039541B24
24D98039548B24
24D98039555B24
24D98039609B24
24D98041069B24
24D98041275B24
24D98041279B24
24D98041282B24
24D98041286B24
24D98041288B24
24D98041838B24
24D98041839B24
24D98042012A24
24D98043079A24
24D98043151A24
24D98043351A24
24D98046152A24
24D98046226A24
24D98046265A24
24D98046267A24
24D98046269A24
24D98046274A24
24D98046276A24
24D98046383A24
24D98046385A24
24D98046387A24
24D98046394A24
24D98046484A24
24D98046503A24
24D98046505A24
24D98046506A24
24D98046507A24
24D98046513A24
24D98046516A24
24D98046517A24
24D98046520A24
24D98046521A24
24D98046708A24
24D98046715A24
24D98046724A24
24D98070023F24
24D98072553B24
24D98056051D24
24D98057467D24
24D98057468D24
24D98057469D24
24D98068091E24
24D98068097E24
24D98070022F24
24D98070024F24
24D98070025F24
24D98070026F24
24D98070027F24
24D98070028F24
24D98070029F24
24D98070030F24
24D98070031F24
24D98070032F24
24D98070033F24
24D98070034F24
24D98070035F24
24D98070036F24
24D98070037F24
24D98070038F24
24D98070039F24
24D98070040F24
24D98070041F24
24D98070042F24
24D98070043F24
24D98070044F24
24D98070045F24
24D98070046F24
24D98070047F24
24D98070048F24
24D98070049F24
24D98070050F24
24D98070051F24
24D98070052F24
24D98070053F24
24D98070054F24
24D98070055F24
24D98070056F24
24D98070057F24
24D98070058F24
24D98070059F24
24D98070060F24
24D98070061F24
24D98070062F24
24D98070063F24
24D98070064F24
24D98070065F24
24D98070066F24
24D98070067F24
24D98070068F24
24D98070069F24
24D98070071F24
24D98070072F24
24D98070073F24
24D98070074F24
24D98070075F24
24D98070076F24
24D98070077F24
24D98070078F24
24D98070079F24
24D98070080F24
24D98070081F24
24D98070082F24
24D98070083F24
24D98070084F24
24D98070085F24
24D98070087F24
24D98070088F24
24D98070089F24
24D98070090F24
24D98070091F24
24D98070092F24
24D98070093F24
24D98070094F24
24D98070095F24
24D98070096F24
24D98070097F24
24D98070098F24
24D98070099F24
24D98070100F24
24D98070101F24
24D98070102F24
24D98070103F24
24D98070104F24
24D98070105F24
24D98070107F24
24D98070108F24
24D98070109F24
24D98070110F24
24D98070111F24
24D98070112F24
24D98070113F24
24D98070114F24
24D98070115F24
24D98070117F24
24D98070118F24
24D98070119F24
24D98070120F24
24D98070121F24
24D98070122F24
24D98070123F24
24D98070124F24
24D98070125F24
24D98070126F24
24D98070127F24
24D98070128F24
24D98070129F24
24D98070130F24
24D98070131F24
24D98070132F24
24D98070133F24
24D98070134F24
24D98070135F24
24D98070136F24
24D98070138F24
24D98070139F24
24D98070140F24
24D98070141F24
24D98070142F24
24D98070143F24
24D98070144F24
24D98070145F24
24D98070146F24
24D98070147F24
24D98070148F24
24D98070149F24
24D98070150F24
24D98070151F24
24D98070157F24
24D98070160F24
24D98070161F24
24D98070162F24
24D98070163F24
24D98070166F24
22D91165150L22
24D98026868F24
24D98026936F24
24D98030913A24
24D98030924A24
24D98030925A24
24D98030927A24
24D98048388F24
24D98048554F24
24D98048555F24
24D98048556F24
24D98048566F24
24D98048595F24
24D98048604F24
24D98048605F24
24D98048606F24
24D98048607F24
24D98048609F24
24D98048610F24
24D98048611F24
24D98048614F24
24D98048615F24
24D98048616F24
24D98048617F24
24D98048618F24
24D98048619F24
24D98029982A24
24D98057648D24
24D98053900E24
24D98054034E24
24D98054543E24
24D98054669E24
24D98050678E24
24D98050945E24
24D98055733E24
24D98070116F24
24D98026619F24
24D98026634F24
24D98026855F24
24D98026897F24
24D98048208F24
24D98026834F24
24D98026836F24
24D98026838F24
24D98026869F24
24D98026921F24
24D98026922F24
24D98026928F24
24D98026933F24
24D98026934F24
24D98026935F24
24D98048491F24
24D98048492F24
24D98048495F24
24D98048496F24
24D98048497F24
24D98048498F24
24D98048499F24
24D98048500F24
24D98048501F24
24D98048502F24
24D98048503F24
24D98048504F24
24D98048557F24
24D98048558F24
24D98048559F24
24D98048560F24
24D98048561F24
24D98048562F24
24D98048563F24
24D98048564F24
24D98048565F24
24D98048567F24
24D98048568F24
24D98048569F24
24D98048570F24
24D98048571F24
24D98048572F24
24D98048573F24
24D98048574F24
24D98048575F24
24D98048576F24
24D98048577F24
24D98048578F24
24D98048579F24
24D98048580F24
24D98048581F24
24D98048582F24
24D98048583F24
24D98048584F24
24D98048585F24
24D98048586F24
24D98048587F24
24D98048588F24
24D98048589F24
24D98048590F24
24D98048591F24
24D98048592F24
24D98048593F24
24D98048594F24
24D98048596F24
24D98048597F24
24D98048598F24
24D98048599F24
24D98048600F24
24D98048601F24
24D98048602F24
24D98048603F24
24D98048608F24
24D98048625F24
24D98048709F24
24D98048710F24
24D98048711F24
24D98048712F24
24D98028430G24
24D98062685G24
24D98048955G24
24D98071082G24
24D98038091G24
24D98071057G24
24D98038133G24
24D98074570G24
24D98074572G24
24D98074606G24
24D98074541G24
24D98024808H24
24D98024818H24
24D98024650H24
24D98019809H24
350P US Device, US English (American Airlines) - 2 Min CPR
Item code: 350-STR-AA-AV
UDI-DI code: 0506016712207
Serial Numbers:
23D98109359I23
23D98109384I23
23D98110320I23
23D98110321I23
23D98119022J23
23D98119023J23
23D98119024J23
23D98119025J23
23D98119026J23
23D98119027J23
23D98119028J23
23D98119029J23
23D98119030J23
23D98119031J23
23D98119032J23
23D98119033J23
23D98119034J23
23D98119035J23
23D98119036J23
23D98119037J23
23D98119039J23
23D98119040J23
23D98119041J23
23D98119042J23
24D98041472B24
24D98041569B24
24D98041710B24
24D98041712B24
24D98041724B24
24D98041726B24
24D98041729B24
24D98041734B24
24D98041738B24
24D98041746B24
24D98041749B24
24D98041755B24
24D98041758B24
24D98041759B24
24D98041760B24
24D98041765B24
24D98041768B24
24D98041774B24
24D98041781B24
24D98041785B24
24D98041786B24
24D98041801B24
24D98041802B24
24D98041804B24
24D98067144D24
24D98067171D24
24D98067175D24
24D98067179D24
24D98067180D24
24D98067181D24
24D98067182D24
24D98067177D24
24D98067178D24
24D98067183D24
24D98067184D24
24D98067185D24
24D98067186D24
24D98067187D24
24D98067188D24
24D98067189D24
24D98067190D24
24D98067191D24
24D98067192D24
24D98067193D24
24D98067194D24
24D98067195D24
24D98067196D24
24D98067197D24
24D98067198D24
24D98067200D24
24D98067201D24
24D98067202D24
24D98067203D24
24D98067204D24
24D98067205D24
24D98067206D24
24D98067207D24
24D98067208D24
24D98067209D24
24D98067210D24
24D98067211D24
24D98067212D24
24D98067213D24
24D98067214D24
24D98067215D24
24D98067216D24
24D98067217D24
24D98067218D24
24D98067219D24
24D98067220D24
24D98067221D24
24D98067222D24
24D98067223D24
24D98067224D24
24D98067225D24
24D98067226D24
24D98067227D24
24D98067228D24
24D98067229D24
24D98067230D24
24D98067231D24
24D98067232D24
24D98067233D24
24D98067234D24
24D98067235D24
24D98067236D24
24D98067237D24
24D98067238D24
24D98067239D24
24D98067240D24
24D98067241D24
24D98067242D24
24D98067243D24
24D98067244D24
24D98067245D24
24D98067246D24
24D98067247D24
24D98067248D24
24D98067249D24
24D98067250D24
24D98067251D24
24D98067252D24
24D98067253D24
24D98056938D24
24D98067153D24
24D98067155D24
24D98067156D24
24D98067157D24
24D98067158D24
24D98067159D24
24D98067160D24
24D98067161D24
24D98067162D24
24D98067163D24
24D98067164D24
24D98067165D24
24D98067166D24
24D98067167D24
24D98067168D24
24D98067169D24
24D98067170D24
24D98067172D24
24D98067173D24
24D98067174D24
24D98067176D24
24D98067121D24
24D98067131D24
350P US Device, US English with Gateway - 2 Min CPR
Item code: 350-STR-US-GW
UDI-DI code: 0506016712804
Serial Numbers:
24D98033695E24
24D98033696E24
24D98033711E24
24D98073174E24
24D98033694E24
24D98033688E24
HeartSine SAM 360P
360P US Device, US English - 2 Min CPR
Item code: 360-STR-US-10
Serial Numbers:
23E90000585K23
23E90003566K23
23E90004161K23
23E90006888K23
23E90013295K23
23E90013878K23
23E90017687K23
23E90102264G23
23E90102269G23
23E90102273G23
23E90102274G23
23E90102275G23
23E90102279G23
23E90102280G23
23E90102281G23
23E90102282G23
23E90102283G23
23E90102284G23
23E90102285G23
23E90102286G23
23E90102287G23
23E90102288G23
23E90102289G23
23E90102290G23
23E90102291G23
23E90102292G23
23E90102293G23
23E90102294G23
23E90102295G23
23E90102296G23
23E90102297G23
23E90102298G23
23E90102299G23
23E90102300G23
23E90102303G23
23E90102304G23
23E90102305G23
23E90102306G23
23E90102307G23
23E90102308G23
23E90102309G23
23E90102312G23
23E90102313G23
23E90102314G23
23E90102315G23
23E90102316G23
23E90102317G23
23E90102318G23
23E90102319G23
23E90102320G23
23E90102321G23
23E90102406G23
23E90102411G23
23E90102539G23
23E90102540G23
23E90102542G23
23E90102543G23
23E90102586G23
23E90102587G23
23E90102601G23
23E90102605G23
23E90102607G23
23E90102609G23
23E90102611G23
23E90102615G23
23E90102616G23
23E90102617G23
23E90102619G23
23E90102620G23
23E90102704G23
23E90102721G23
23E90102722G23
23E90102809G23
23E90102811G23
23E90102953G23
23E90105179I23
23E90105321I23
23E90105891I23
23E90107011G23
23E90107016G23
23E90107017G23
23E90107018G23
23E90107033G23
23E90107085G23
23E90107101G23
23E90107110G23
23E90107111G23
23E90107258G23
23E90107259G23
23E90107260G23
23E90107261G23
23E90107262G23
23E90107263G23
23E90107265G23
23E90107267G23
23E90107268G23
23E90107269G23
23E90107270G23
23E90107271G23
23E90107272G23
23E90107273G23
23E90107274G23
23E90107275G23
23E90107317G23
23E90107318G23
23E90107320G23
23E90107321G23
23E90107322G23
23E90107323G23
23E90107324G23
23E90107325G23
23E90107326G23
23E90107327G23
23E90107328G23
23E90107329G23
23E90107330G23
23E90107331G23
23E90107332G23
23E90107333G23
23E90107334G23
23E90107335G23
23E90107336G23
23E90108023K23
23E90113889K23
23E90113890K23
23E90113891K23
23E90113896K23
23E90113897K23
23E90113898K23
23E90116036K23
23E90117111I23
23E90117117I23
23E90117327I23
23E90117331K23
23E90117631I23
23E90117635I23
23E90119047K23
23E90130028J23
23E90130029J23
23E90130030J23
23E90130031J23
23E90130033J23
23E90130038J23
23E90130039J23
23E90130040J23
23E90130048J23
23E90130053J23
23E90130054J23
23E90130317J23
23E90130320J23
23E90130321K23
23E90130323J23
23E90130367J23
23E90130398J23
23E90130417J23
23E90130418J23
23E90130419J23
23E90130421J23
23E90130422J23
23E90130423J23
23E90130425J23
23E90130426J23
23E90130427J23
23E90130428J23
23E90130429J23
23E90130431J23
23E90130432J23
23E90130433J23
23E90130434J23
23E90130436J23
23E90130437J23
23E90130438J23
23E90130439J23
23E90130440J23
23E90130444J23
23E90130451J23
23E90130453J23
23E90130459J23
23E90130460J23
23E90130465J23
23E90130470J23
23E90130487J23
23E90131455K23
23E90133283I23
23E90133290I23
23E90133291I23
23E90134127I23
23E90135280J23
23E90135501J23
23E90135511J23
23E90135521J23
23E90135526J23
23E90135533J23
23E90135536J23
23E90135554J23
23E90135555J23
23E90135556J23
23E90135557J23
23E90135559J23
23E90135560J23
23E90135561J23
23E90135562J23
23E90135563J23
23E90135564J23
23E90135565J23
23E90135566J23
23E90135567J23
23E90135568J23
23E90135569J23
23E90135570J23
23E90135571J23
23E90135572J23
23E90135573J23
23E90135574J23
23E90135575J23
23E90135576J23
23E90135590J23
23E90135592J23
23E90135739J23
23E90135741J23
23E90135750J23
23E90135752J23
23E90135754J23
23E90135755J23
23E90135756J23
23E90135757J23
23E90135758J23
23E90135765J23
23E90135766J23
23E90135769J23
23E90135772J23
23E90135774J23
23E90135776J23
23E90135777J23
23E90135806J23
23E90135810J23
23E90135823J23
23E90135888J23
23E90135889J23
23E90135890J23
23E90135892J23
23E90135893J23
23E90135894J23
23E90135895J23
23E90135898J23
23E90135899J23
23E90135901J23
23E90135902J23
23E90135904J23
23E90135905J23
23E90135907J23
23E90135908J23
23E90135909J23
23E90135910J23
23E90135911J23
23E90135912J23
23E90135913J23
23E90135914J23
23E90135915J23
23E90135916J23
23E90135917J23
23E90135918J23
23E90135919J23
23E90135921J23
23E90135922J23
23E90135923J23
23E90135925J23
23E90135926J23
23E90135927J23
23E90135930J23
23E90135931J23
23E90135932J23
23E90135933J23
23E90137003I23
23E90137059I23
23E90137406I23
23E90137519I23
23E90137611I23
23E90137645I23
23E90137673I23
23E90137686I23
23E90137694I23
23E90137696I23
23E90137701I23
23E90137704I23
23E90137706I23
23E90137708I23
23E90137714K23
23E90137717I23
23E90137721I23
23E90137722I23
23E90137723I23
23E90137726I23
23E90137727I23
23E90137736I23
23E90137739I23
23E90137741I23
23E90137743I23
23E90137745I23
23E90137758I23
23E90139453I23
23E90139461I23
23E90139480I23
23E90139493I23
23E90139538I23
23E90143202K23
23E90143213K23
23E90143270K23
23E90143284K23
23E90143300K23
23E90143793K23
23E90143828K23
23E90145795K23
23E90146023J23
23E90146176J23
23E90146177J23
23E90146179J23
23E90146181J23
23E90146182J23
23E90146183J23
23E90146184J23
23E90146185J23
23E90146186J23
23E90146188J23
23E90146189J23
23E90146191J23
23E90146192J23
23E90146319J23
23E90146333J23
23E90146377J23
23E90146426J23
23E90146428J23
23E90146430J23
23E90146431J23
23E90146432J23
23E90146433J23
23E90146435J23
23E90146436J23
23E90146437J23
23E90146438J23
23E90146439J23
23E90146440J23
23E90146441J23
23E90146442J23
23E90146443J23
23E90146445J23
23E90146446J23
23E90146450J23
23E90146451J23
23E90146452J23
23E90146453J23
23E90146454J23
23E90146455J23
23E90146566J23
23E90148086K23
23E90148149K23
23E90148170K23
23E90148173K23
23E90148291K23
23E90148293K23
23E90148297K23
23E90148301K23
24E90001961B24
24E90002370B24
24E90003112B24
24E90003215B24
24E90003216B24
24E90003254B24
24E90003312B24
24E90003493B24
24E90003494B24
24E90003739B24
24E90003744B24
24E90003745B24
24E90003747B24
24E90003760B24
24E90003787B24
24E90003791B24
24E90004170B24
24E90004612B24
24E90004613B24
24E90004617B24
24E90004619B24
24E90004621B24
24E90004622B24
24E90004624B24
24E90004625B24
24E90004633B24
24E90004634B24
24E90004635B24
24E90004638B24
24E90004640B24
24E90004641B24
24E90004695B24
24E90004696B24
24E90004697B24
24E90004698B24
24E90004699B24
24E90004704B24
24E90004705B24
24E90004706B24
24E90004726B24
24E90004732B24
24E90005978B24
24E90013330B24
24E90014785A24
24E90015664B24
24E90015665B24
24E90016471A24
24E90016891A24
24E90016893A24
24E90016894A24
24E90016895A24
24E90016896A24
24E90016897A24
24E90016900A24
24E90016901A24
24E90016902A24
24E90016903A24
24E90016904A24
24E90017087A24
24E90017112A24
24E90017277A24
24E90017409A24
24E90017410A24
24E90017417A24
24E90017418A24
24E90017421A24
24E90017422A24
24E90017423A24
24E90017424A24
24E90017426A24
24E90017427A24
24E90017428A24
24E90017429A24
24E90017430A24
24E90017431A24
24E90017432A24
24E90017433A24
24E90017439A24
24E90017440A24
24E90017441A24
24E90017442A24
24E90017443A24
24E90017444A24
24E90017445A24
24E90017446A24
24E90017447A24
24E90017448A24
24E90017449A24
24E90017450A24
24E90017451A24
24E90017452A24
24E90017454A24
24E90017455A24
24E90017456A24
24E90017457A24
24E90017458A24
24E90017459A24
24E90017460A24
24E90017461A24
24E90017462A24
24E90017463A24
24E90017464A24
24E90017465A24
24E90017466A24
24E90017467A24
24E90017471A24
24E90017472A24
24E90017473A24
24E90017474A24
24E90017475A24
24E90017476A24
24E90017477A24
24E90017478A24
24E90017629A24
24E90017787A24
24E90017925A24
24E90017930A24
24E90017933A24
24E90017936A24
24E90018647A24
24E90020472B24
24E90020541B24
23E90139374B24
24E90000739B24
24E90000958B24
24E90001072B24
24E90001288B24
24E90001501B24
24E90001548B24
24E90001638B24
24E90001785B24
24E90002115B24
24E90002912B24
24E90003807B24
24E90003808B24
24E90003809B24
24E90003818B24
24E90003835B24
24E90004172B24
24E90004386B24
24E90004387B24
24E90004388B24
24E90004389B24
24E90004390B24
24E90004391B24
24E90004392B24
24E90004393B24
24E90004396B24
24E90004397B24
24E90004398B24
24E90004399B24
24E90004400B24
24E90004448B24
24E90004449B24
24E90004450B24
24E90004451B24
24E90004452B24
24E90004453B24
24E90004454B24
24E90004455B24
24E90004456B24
24E90004457B24
24E90004458B24
24E90004459B24
24E90004460B24
24E90004461B24
24E90004462B24
24E90004463B24
24E90004464B24
24E90004465B24
24E90004466B24
24E90004467B24
24E90004468B24
24E90004469B24
24E90004470B24
24E90004471B24
24E90004472B24
24E90004473B24
24E90004474B24
24E90004476B24
24E90004477B24
24E90004478B24
24E90004530B24
24E90004531B24
24E90004532B24
24E90004593B24
24E90004594B24
24E90004598B24
24E90004599B24
24E90004600B24
24E90004601B24
24E90004602B24
24E90004605B24
24E90004607B24
24E90004609B24
24E90004610B24
24E90004615B24
24E90004616B24
24E90004618B24
24E90004623B24
24E90004626B24
24E90004629B24
24E90004631B24
24E90004632B24
24E90004636B24
24E90004637B24
24E90004639B24
24E90004642B24
24E90004693B24
24E90004694B24
24E90004700B24
24E90004701B24
24E90004702B24
24E90004707B24
24E90004708B24
24E90004710B24
24E90004711B24
24E90004715B24
24E90004716B24
24E90004717B24
24E90004718B24
24E90004719B24
24E90004721B24
24E90004722B24
24E90004723B24
24E90004724B24
24E90004727B24
24E90004728B24
24E90004730B24
24E90004731B24
24E90004733B24
24E90004734B24
24E90004735B24
24E90004749B24
24E90004750B24
24E90004751B24
24E90004752B24
24E90004755B24
24E90005781B24
24E90005783B24
24E90005784B24
24E90005785B24
24E90005788B24
24E90005790B24
24E90005791B24
24E90005792B24
24E90005793B24
24E90005794B24
24E90005795B24
24E90005797B24
24E90005880B24
24E90005881B24
24E90005882B24
24E90005883B24
24E90005884B24
24E90005885B24
24E90005886B24
24E90005888B24
24E90005889B24
24E90005890B24
24E90005891B24
24E90005892B24
24E90005893B24
24E90005894B24
24E90005895B24
24E90005896B24
24E90005897B24
24E90005898B24
24E90005899B24
24E90005900B24
24E90005901B24
24E90005902B24
24E90005903B24
24E90005904B24
24E90005905B24
24E90005906B24
24E90005907B24
24E90005910B24
24E90005913B24
24E90005915B24
24E90005925B24
24E90005937B24
24E90005976B24
24E90005977B24
24E90006143B24
24E90006144B24
24E90006145B24
24E90006146B24
24E90006148B24
24E90006149B24
24E90006150B24
24E90006151B24
24E90006152B24
24E90006153B24
24E90006154B24
24E90006155B24
24E90006156B24
24E90006157B24
24E90006158B24
24E90006159B24
24E90006160B24
24E90006161B24
24E90006162B24
24E90006163B24
24E90006164B24
24E90006165B24
24E90006167B24
24E90006168B24
24E90006169B24
24E90006172B24
24E90006173B24
24E90006174B24
24E90006175B24
24E90006176B24
24E90006177B24
24E90006178B24
24E90006179B24
24E90006180B24
24E90006181B24
24E90006182B24
24E90006183B24
24E90006184B24
24E90006185B24
24E90006186B24
24E90006188B24
24E90006190B24
24E90006192B24
24E90006193B24
24E90006197B24
24E90006198B24
24E90006200B24
24E90012555A24
24E90012556A24
24E90012561A24
24E90012577A24
24E90012586A24
24E90012588A24
24E90012589A24
24E90012590A24
24E90012591A24
24E90012592A24
24E90012594A24
24E90012599A24
24E90012606A24
24E90012608A24
24E90012613A24
24E90014632B24
24E90014657B24
24E90014938B24
24E90015411B24
24E90015420B24
24E90015530B24
24E90015823A24
24E90015921A24
24E90015923A24
24E90015933A24
24E90015934A24
24E90015935A24
24E90015936A24
24E90015937A24
24E90015940A24
24E90016700A24
24E90016942A24
24E90016944A24
24E90016945A24
24E90016948A24
24E90016951A24
24E90016952A24
24E90016953A24
24E90016955A24
24E90017713A24
24E90017714A24
24E90017715A24
24E90017716A24
24E90017725A24
24E90017735A24
24E90017736A24
24E90017737A24
24E90017738A24
24E90017739A24
24E90017740A24
24E90017741A24
24E90017742A24
24E90017743A24
24E90017747A24
24E90017760A24
24E90017761A24
24E90017762A24
24E90017764A24
24E90017766A24
24E90017923B24
24E90017926A24
24E90017927A24
24E90017928A24
24E90017931A24
24E90017932A24
24E90017934A24
24E90017935A24
24E90018028A24
24E90018029A24
24E90018030A24
24E90018031A24
24E90018032A24
24E90018033A24
24E90018034A24
24E90018035A24
24E90018036A24
24E90018037A24
24E90018038A24
24E90018039A24
24E90018040A24
24E90018041A24
24E90018042A24
24E90018043A24
24E90018044A24
24E90018045A24
24E90018046A24
24E90018047A24
24E90018048A24
24E90018049A24
24E90018054A24
24E90018062A24
24E90018126A24
24E90018128A24
24E90018129A24
24E90018130A24
24E90018132A24
24E90018133A24
24E90018134A24
24E90018135A24
24E90018144A24
24E90018155A24
24E90018156A24
24E90018159A24
24E90018160A24
24E90018161A24
24E90018162A24
24E90018163A24
24E90018164A24
24E90018165A24
24E90018167A24
24E90018168A24
24E90018169A24
24E90018170A24
24E90018171A24
24E90018172A24
24E90018178A24
24E90018179A24
24E90018183A24
24E90018184A24
24E90018185A24
24E90018186A24
24E90018187A24
24E90018188A24
24E90018189A24
24E90018192A24
24E90018194A24
24E90018201A24
24E90018208A24
24E90018218A24
24E90018223A24
24E90018225A24
24E90018226A24
24E90018227A24
24E90018228A24
24E90018229A24
24E90018230A24
24E90018232A24
24E90018233A24
24E90018236A24
24E90018237A24
24E90018238A24
24E90018239A24
24E90018240A24
24E90018241A24
24E90018242A24
24E90018243A24
24E90018244A24
24E90018245A24
24E90018246A24
24E90018247A24
24E90018248A24
24E90018250A24
24E90018251A24
24E90018346A24
24E90018347A24
24E90018348A24
24E90018349A24
24E90018350A24
24E90018352A24
24E90018354A24
24E90018355A24
24E90018356A24
24E90018358A24
24E90018359A24
24E90018360A24
24E90018367A24
24E90018368A24
24E90018370A24
24E90018374A24
24E90018378A24
24E90018383A24
24E90018384A24
24E90018387A24
24E90018389A24
24E90018391A24
24E90018393A24
24E90018408A24
24E90018544A24
24E90018545A24
24E90018549A24
24E90018553A24
24E90018556A24
24E90018557A24
24E90018558A24
24E90018562A24
24E90018594A24
24E90018595A24
24E90018596A24
24E90018597A24
24E90018598A24
24E90018599A24
24E90018600A24
24E90018602A24
24E90018603A24
24E90018604A24
24E90018605A24
24E90018606A24
24E90018607A24
24E90018608A24
24E90018609A24
24E90018610A24
24E90018611A24
24E90018612A24
24E90018613A24
24E90018614A24
24E90018615A24
24E90018616A24
24E90018617A24
24E90018618A24
24E90018619A24
24E90018620A24
24E90018621A24
24E90018622A24
24E90018623A24
24E90018624A24
24E90018625A24
24E90018626A24
24E90018627A24
24E90018628A24
24E90018629A24
24E90018630A24
24E90018631A24
24E90018632A24
24E90018633A24
24E90018634A24
24E90018635A24
24E90018636A24
24E90018637A24
24E90018638A24
24E90018639A24
24E90018640A24
24E90018641A24
24E90018642A24
24E90018643A24
24E90018644A24
24E90018645A24
24E90018646A24
24E90018657B24
23E90144949A24
24E90000277A24
24E90000279A24
24E90000280A24
24E90000281A24
24E90000282A24
24E90000283A24
24E90000284A24
24E90000296A24
24E90000301A24
24E90000302A24
24E90000368A24
24E90000374A24
24E90000377A24
24E90000378A24
24E90000381A24
24E90000390A24
24E90000391A24
24E90000392A24
24E90000393A24
24E90000394A24
24E90000395A24
24E90000451A24
24E90000452A24
24E90000453A24
24E90000456A24
24E90000457A24
24E90000458A24
24E90000459A24
24E90000508A24
24E90000509A24
24E90000513A24
24E90000557A24
24E90000593A24
24E90000598A24
24E90000600A24
24E90000601A24
24E90000602A24
24E90000603A24
24E90000604A24
24E90000605A24
24E90000606A24
24E90000614A24
24E90000615A24
24E90000616A24
24E90000617A24
24E90000620A24
24E90000621A24
24E90000622A24
24E90000627A24
24E90000628A24
24E90000629A24
24E90000630A24
24E90000631A24
24E90000632A24
24E90000633A24
24E90000634A24
24E90000635A24
24E90000636A24
24E90000637A24
24E90000638A24
24E90000639A24
24E90000640A24
24E90000641A24
24E90000642A24
24E90000643A24
24E90000644A24
24E90000645A24
24E90000646A24
24E90000648A24
24E90000664A24
24E90000677A24
24E90000678A24
24E90000682A24
24E90000683A24
24E90000689A24
24E90000690A24
24E90000692A24
24E90001251A24
24E90001253A24
24E90001254A24
24E90001256A24
24E90001267A24
24E90001269A24
24E90001278A24
24E90001280A24
24E90001281A24
24E90001283A24
24E90001284A24
24E90001285A24
24E90001286A24
24E90001287A24
24E90001289A24
24E90001381A24
24E90001962B24
24E90001963B24
24E90001964B24
24E90001965B24
24E90001967B24
24E90001968B24
24E90001969B24
24E90001970B24
24E90001972B24
24E90001974B24
24E90001975B24
24E90001977B24
24E90001978B24
24E90001985B24
24E90001986B24
24E90001987B24
24E90001988B24
24E90001989B24
24E90001992B24
24E90001993B24
24E90001995B24
24E90001996B24
24E90001997B24
24E90002141B24
24E90002142B24
24E90002143B24
24E90002148B24
24E90002149B24
24E90002150B24
24E90002151B24
24E90002164B24
24E90002166B24
24E90002167B24
24E90002168B24
24E90002169B24
24E90002170B24
24E90002172B24
24E90002173B24
24E90002301B24
24E90002302B24
24E90002303B24
24E90002304B24
24E90002305B24
24E90002306B24
24E90002307B24
24E90002308B24
24E90002309B24
24E90002310B24
24E90002311B24
24E90002312B24
24E90002315B24
24E90002316B24
24E90002318B24
24E90002320B24
24E90002321B24
24E90002322B24
24E90002323B24
24E90002324B24
24E90002326B24
24E90002327B24
24E90002328B24
24E90002329B24
24E90002330B24
24E90002331B24
24E90002332B24
24E90002333B24
24E90002334B24
24E90002335B24
24E90002336B24
24E90002337B24
24E90002338B24
24E90002339B24
24E90002340B24
24E90002341B24
24E90002342B24
24E90002343B24
24E90002344B24
24E90002345B24
24E90002346B24
24E90002348B24
24E90003077B24
24E90003078B24
24E90003083B24
24E90003084B24
24E90003085B24
24E90003086B24
24E90003087B24
24E90003088B24
24E90003090B24
24E90003091B24
24E90003092B24
24E90003093B24
24E90003096B24
24E90003098B24
24E90003099B24
24E90003100B24
24E90003102B24
24E90003103B24
24E90003105B24
24E90003106B24
24E90003110B24
24E90003111B24
24E90003115B24
24E90003116B24
24E90003117B24
24E90003118B24
24E90003119B24
24E90003120B24
24E90003121B24
24E90003122B24
24E90003123B24
24E90003125B24
24E90003126B24
24E90003127B24
24E90003129B24
24E90003130B24
24E90003131B24
24E90003132B24
24E90003217B24
24E90003223B24
24E90003225B24
24E90003226B24
24E90003227B24
24E90003228B24
24E90003229B24
24E90003230B24
24E90003231B24
24E90003232B24
24E90003314B24
24E90003317B24
24E90003506B24
24E90003555B24
24E90003565B24
24E90003657B24
24E90003658B24
24E90003659B24
24E90003660B24
24E90003661B24
24E90003663B24
24E90003665B24
24E90003671B24
24E90003672B24
24E90003680B24
24E90003681B24
24E90003682B24
24E90003684B24
24E90003738B24
24E90003740B24
24E90003742B24
24E90003743B24
24E90003757B24
24E90003758B24
24E90003759B24
24E90003761B24
24E90003762B24
24E90003763B24
24E90003764B24
24E90003765B24
24E90003766B24
24E90003767B24
24E90003768B24
24E90003769B24
24E90003770B24
24E90003771B24
24E90003772B24
24E90003773B24
24E90003774B24
24E90003775B24
24E90003776B24
24E90003777B24
24E90003778B24
24E90003779B24
24E90003780B24
24E90003781B24
24E90003782B24
24E90003783B24
24E90003784B24
24E90003785B24
24E90003786B24
24E90003788B24
24E90003789B24
24E90003792B24
24E90003794B24
24E90003987B24
24E90003988B24
24E90003989B24
24E90003994B24
24E90003995B24
24E90003996B24
24E90003998B24
24E90003999B24
24E90004000B24
24E90004001B24
24E90004002B24
24E90004003B24
24E90004004B24
24E90004005B24
24E90004006B24
24E90004007B24
24E90004008B24
24E90004009B24
24E90004010B24
24E90004011B24
24E90004012B24
24E90004013B24
24E90004015B24
24E90004016B24
24E90004017B24
24E90004089B24
24E90004091B24
24E90004092B24
24E90004093B24
24E90004094B24
24E90004095B24
24E90004097B24
24E90004098B24
24E90004099B24
24E90004100B24
24E90004101B24
24E90004102B24
24E90004104B24
24E90004105B24
24E90004106B24
24E90004107B24
24E90004108B24
24E90004109B24
24E90004110B24
24E90004120B24
24E90004127B24
24E90004129B24
24E90004130B24
24E90004131B24
24E90004159B24
24E90004160B24
24E90004164B24
24E90004165B24
24E90004168B24
24E90004169B24
24E90012610A24
24E90012680A24
24E90013058A24
24E90013059A24
24E90013060A24
24E90013062A24
24E90013063A24
24E90013064A24
24E90013066A24
24E90013067A24
24E90013309A24
24E90013313A24
24E90013327A24
24E90014622A24
24E90014764A24
24E90014765A24
24E90014767A24
24E90014768A24
24E90014769A24
24E90014770A24
24E90014771A24
24E90014772A24
24E90014773A24
24E90014777A24
24E90014780A24
24E90014783A24
24E90014784A24
24E90014789A24
24E90014791A24
24E90014792A24
24E90014793A24
24E90014794A24
24E90014796A24
24E90014797A24
24E90014800A24
24E90014801A24
24E90014802A24
24E90014803A24
24E90014804A24
24E90014814A24
24E90014816A24
24E90014818A24
24E90014819A24
24E90014820A24
24E90014823A24
24E90014829A24
24E90014831A24
24E90014832A24
24E90014833A24
24E90014834A24
24E90014835A24
24E90014836A24
24E90014837A24
24E90014848A24
24E90014961A24
24E90014962A24
24E90014963A24
24E90014964A24
24E90014965A24
24E90015011A24
24E90015201A24
24E90015215A24
24E90015217A24
24E90015219A24
24E90015221A24
24E90015222A24
24E90015224A24
24E90015225A24
24E90015226A24
24E90015229A24
24E90015230A24
24E90015231A24
24E90015232A24
24E90015233A24
24E90015235A24
24E90015236A24
24E90015275A24
24E90015277A24
24E90015279A24
24E90015280A24
24E90015281A24
24E90015282A24
24E90015283A24
24E90015284A24
24E90015285A24
24E90015286A24
24E90015287A24
24E90015288A24
24E90015289A24
24E90015290A24
24E90015291A24
24E90015296A24
24E90015302A24
24E90015310A24
24E90015373A24
24E90015382A24
24E90015383A24
24E90015386A24
24E90015389A24
24E90015391A24
24E90015394A24
24E90015468A24
24E90015471A24
24E90015474A24
24E90015475A24
24E90015480A24
24E90015490A24
24E90015491A24
24E90015492A24
24E90015493A24
24E90015496A24
24E90015497A24
24E90015498A24
24E90015499A24
24E90015500A24
24E90015501A24
24E90015502A24
24E90015507A24
24E90015532A24
24E90015535A24
24E90015536A24
24E90015544A24
24E90015545A24
24E90015546A24
24E90015555A24
24E90015556A24
24E90015557A24
24E90015558A24
24E90015559A24
24E90015560A24
24E90015561A24
24E90015562A24
24E90015563A24
24E90015570A24
24E90015571A24
24E90015572A24
24E90015573A24
24E90015902A24
24E90015986A24
24E90016241A24
24E90016439A24
24E90016443A24
24E90016454A24
24E90016457A24
24E90016460A24
24E90016467A24
24E90016474A24
24E90016475A24
24E90016476A24
24E90016477A24
24E90016508A24
24E90016512A24
24E90016570A24
24E90016571A24
24E90016585A24
24E90016698A24
24E90016703A24
24E90016721A24
24E90016745A24
24E90016747A24
24E90016805A24
24E90016810A24
24E90016837A24
24E90016840A24
24E90016842A24
24E90016862A24
24E90016978A24
24E90016979A24
24E90016980A24
24E90016981A24
24E90016988A24
24E90016989A24
24E90016990A24
24E90016991A24
24E90017057A24
24E90017066A24
24E90017075A24
24E90017078A24
24E90017079A24
24E90017080A24
24E90017081A24
24E90017082A24
24E90017083A24
24E90017085A24
24E90017086A24
24E90017088A24
24E90017089A24
24E90017090A24
24E90017091A24
24E90017092A24
24E90017094A24
24E90017095A24
24E90017096A24
24E90017097A24
24E90017098A24
24E90017099A24
24E90017100A24
24E90017101A24
24E90017103A24
24E90017105A24
24E90017106A24
24E90017107A24
24E90017109A24
24E90017110A24
24E90017111A24
24E90017113A24
24E90017114A24
24E90017115A24
24E90017116A24
24E90017117A24
24E90017118A24
24E90017119A24
24E90017120A24
24E90017121A24
24E90017122A24
24E90017123A24
24E90017124A24
24E90017125A24
24E90017126A24
24E90017127A24
24E90017128A24
24E90017129A24
24E90017130A24
24E90017131A24
24E90017132A24
24E90017133A24
24E90017134A24
24E90017136A24
24E90017138A24
24E90017139A24
24E90017140A24
24E90017141A24
24E90017142A24
24E90017143A24
24E90017160A24
24E90017179A24
24E90017180A24
24E90017181A24
24E90017182A24
24E90017183A24
24E90017184A24
24E90017185A24
24E90017186A24
24E90017187A24
24E90017188A24
24E90017189A24
24E90017191A24
24E90017192A24
24E90017193A24
24E90017194A24
24E90017195A24
24E90017196A24
24E90017197A24
24E90017198A24
24E90017199A24
24E90017200A24
24E90017201A24
24E90017202A24
24E90017203A24
24E90017204A24
24E90017205A24
24E90017206A24
24E90017207A24
24E90017208A24
24E90017209A24
24E90017210A24
24E90017211A24
24E90017212A24
24E90017213A24
24E90017214A24
24E90017215A24
24E90017216A24
24E90017217A24
24E90017218A24
24E90017219A24
24E90017220A24
24E90017221A24
24E90017222A24
24E90017223A24
24E90017224A24
24E90017225A24
24E90017226A24
24E90017227A24
24E90017228A24
24E90017229A24
24E90017230A24
24E90017231A24
24E90017232A24
24E90017233A24
24E90017234A24
24E90017235A24
24E90017236A24
24E90017237A24
24E90017238A24
24E90017239A24
24E90017240A24
24E90017241A24
24E90017242A24
24E90017243A24
24E90017244A24
24E90017245A24
24E90017246A24
24E90017247A24
24E90017248A24
24E90017250A24
24E90017251A24
24E90017252A24
24E90017253A24
24E90017254A24
24E90017255A24
24E90017256A24
24E90017257A24
24E90017258A24
24E90017259A24
24E90017260A24
24E90017261A24
24E90017262A24
24E90017263A24
24E90017264A24
24E90017265A24
24E90017266A24
24E90017267A24
24E90017268A24
24E90017269A24
24E90017271A24
24E90017272A24
24E90017273A24
24E90017274A24
24E90017275A24
24E90017279A24
24E90017280A24
24E90017281A24
24E90017282A24
24E90017283A24
24E90017285A24
24E90017286A24
24E90017288A24
24E90017289A24
24E90017290A24
24E90017291A24
24E90017292A24
24E90017293A24
24E90017294A24
24E90017295A24
24E90017296A24
24E90017297A24
24E90017298A24
24E90017299A24
24E90017300A24
24E90017301A24
24E90017302A24
24E90017303A24
24E90017304A24
24E90017305A24
24E90017306A24
24E90017307A24
24E90017308A24
24E90017309A24
24E90017310A24
24E90017312A24
24E90017314A24
24E90017315A24
24E90017318A24
24E90017319A24
24E90017320A24
24E90017322A24
24E90017323A24
24E90017324A24
24E90017325A24
24E90017326A24
24E90017328A24
24E90017329A24
24E90017330A24
24E90017331A24
24E90017332A24
24E90017333A24
24E90017334A24
24E90017335A24
24E90017343A24
24E90017352A24
24E90017353A24
24E90017354A24
24E90017355A24
24E90017359A24
24E90017360A24
24E90017361A24
24E90017362A24
24E90017363A24
24E90017364A24
24E90017365A24
24E90017366A24
24E90017367A24
24E90017368A24
24E90017369A24
24E90017370A24
24E90017371A24
24E90017372A24
24E90017373A24
24E90017374A24
24E90017375A24
24E90017376A24
24E90017377A24
24E90017378A24
24E90017402A24
24E90017403A24
24E90017404A24
24E90017405A24
24E90017406A24
24E90017408A24
24E90017479A24
24E90017480A24
24E90017481A24
24E90017482A24
24E90017484A24
24E90017485A24
24E90017486A24
24E90017487A24
24E90017488A24
24E90017489A24
24E90017490A24
24E90017492A24
24E90017493A24
24E90017494A24
24E90017495A24
24E90017497A24
24E90017498A24
24E90017499A24
24E90017500A24
24E90017501A24
24E90017502A24
24E90017503A24
24E90017504A24
24E90017505A24
24E90017506A24
24E90017507A24
24E90017508A24
24E90017509A24
24E90017510A24
24E90017511A24
24E90017512A24
24E90017513A24
24E90017514A24
24E90017515A24
24E90017516A24
24E90017517A24
24E90017518A24
24E90017519A24
24E90017520A24
24E90017521A24
24E90017522A24
24E90017523A24
24E90017524A24
24E90017525A24
24E90017526A24
24E90017527A24
24E90017528A24
24E90017529A24
24E90017532A24
24E90017533A24
24E90017534A24
24E90017535A24
24E90017536A24
24E90017537A24
24E90017538A24
24E90017539A24
24E90017541A24
24E90017563A24
24E90017617A24
24E90017618A24
24E90017619A24
24E90017620A24
24E90017621A24
24E90017622A24
24E90017623A24
24E90017624A24
24E90017625A24
24E90017626A24
24E90017627A24
24E90017628A24
24E90017630A24
24E90017631A24
24E90017632A24
24E90017633A24
24E90017634A24
24E90017635A24
24E90017656A24
24E90017658A24
24E90017659A24
24E90017661A24
24E90017662A24
24E90017663A24
24E90017664A24
24E90017665A24
24E90017707A24
24E90017782A24
24E90017785A24
24E90017788A24
24E90017789A24
24E90017790A24
24E90017791A24
24E90017792A24
24E90017793A24
24E90017794A24
24E90017795A24
24E90017798A24
24E90017811A24
24E90017812A24
24E90017815A24
24E90017867A24
24E90017869A24
24E90017870A24
24E90017871A24
24E90017872A24
24E90017874A24
24E90017879A24
24E90017883A24
24E90017884A24
24E90017885A24
24E90017886A24
24E90017887A24
24E90017888A24
24E90017889A24
24E90017892A24
24E90017893A24
24E90017894A24
24E90017896A24
24E90017937A24
24E90017938A24
24E90017939A24
24E90017940A24
24E90017941A24
24E90017942A24
24E90017943A24
24E90017946A24
24E90017947A24
24E90017948A24
24E90017949A24
24E90017950A24
24E90017954A24
24E90017955A24
24E90017956A24
24E90017976A24
24E90017977A24
24E90017978A24
24E90017979A24
24E90017980A24
24E90017981A24
24E90017982A24
24E90017984A24
24E90017985A24
24E90017986A24
24E90017987A24
24E90017988A24
24E90017989A24
24E90017990A24
24E90017991A24
24E90017992A24
24E90017993A24
24E90017994A24
24E90017995A24
24E90017996A24
24E90017998A24
24E90017999A24
24E90018000A24
24E90018001A24
24E90018004A24
24E90018005A24
24E90018006A24
24E90018007A24
24E90018008A24
24E90018009A24
24E90018010A24
24E90018011A24
24E90018012A24
24E90018013A24
24E90018014A24
24E90018015A24
24E90018016A24
24E90018017A24
24E90018018A24
24E90018022A24
24E90018023A24
24E90018091A24
24E90018224A24
24E90018274A24
24E90018275A24
24E90018277A24
24E90018279A24
24E90018280A24
24E90018281A24
24E90018282A24
24E90018283A24
24E90018284A24
24E90018285A24
24E90018286A24
24E90018287A24
24E90018288A24
24E90018289A24
24E90018290A24
24E90018291A24
24E90018292A24
24E90018293A24
24E90018294A24
24E90018295A24
24E90018296A24
24E90018297A24
24E90018298A24
24E90018300A24
24E90018302A24
24E90018303A24
24E90018304A24
24E90018305A24
24E90018306A24
24E90018307A24
24E90018308A24
24E90018309A24
24E90018310A24
24E90018311A24
24E90018312A24
24E90018314A24
24E90018315A24
24E90018316A24
24E90018317A24
24E90018318A24
24E90018319A24
24E90018320A24
24E90018321A24
24E90018322A24
24E90018323A24
24E90018324A24
24E90018325A24
24E90018326A24
24E90018327A24
24E90018328A24
24E90018329A24
24E90018330A24
24E90018331A24
24E90018332A24
24E90018333A24
24E90018334A24
24E90018335A24
24E90018336A24
24E90018337A24
24E90018338A24
24E90018339A24
24E90018340A24
24E90018341A24
24E90018342A24
24E90018343A24
24E90018344A24
24E90018345A24
24E90018351A24
24E90018357A24
24E90018410A24
24E90018413A24
24E90018420A24
24E90018424A24
24E90018425A24
24E90018426A24
24E90018427A24
24E90018428A24
24E90018429A24
24E90018431A24
24E90018432A24
24E90018433A24
24E90018434A24
24E90018435A24
24E90018436A24
24E90018439A24
24E90018440A24
24E90018441A24
24E90018442A24
24E90018443A24
24E90018481A24
24E90018482A24
24E90018483A24
24E90018484A24
24E90018487A24
24E90018488A24
24E90018523A24
24E90018526A24
24E90018527A24
24E90018529A24
24E90018530A24
24E90018532A24
24E90018533A24
24E90018534A24
24E90018535A24
24E90018536A24
24E90018538A24
24E90018540A24
24E90018541A24
24E90018542A24
24E90018543A24
24E90018546A24
24E90018547A24
24E90018548A24
24E90018550A24
24E90018551A24
24E90018552A24
24E90018554A24
24E90018555A24
24E90018559A24
24E90018560A24
24E90018561A24
24E90018563A24
24E90018564A24
24E90018565A24
24E90018566A24
24E90018567A24
24E90018568A24
24E90018569A24
24E90018570A24
24E90018571A24
24E90018572A24
24E90018573A24
24E90018574A24
24E90018575A24
24E90018576A24
24E90018577A24
24E90018580A24
24E90018581A24
24E90018583A24
24E90018584A24
24E90018585A24
24E90018586A24
24E90018587A24
24E90018588A24
24E90018589A24
24E90018590A24
24E90018591A24
24E90018658A24
23E90116066K23
23E90117536A24
23E90117546A24
23E90117549A24
23E90139221K23
24E90000285A24
24E90000533A24
24E90000538A24
24E90000539A24
24E90000541A24
24E90000543A24
24E90000544A24
24E90000545A24
24E90000547A24
24E90000551A24
24E90000552A24
24E90000553A24
24E90000554A24
24E90000555A24
24E90000556A24
24E90000563A24
24E90000564A24
24E90000568A24
24E90000569A24
24E90000570A24
24E90000571A24
24E90000572A24
24E90000573A24
24E90000574A24
24E90000575A24
24E90000576A24
24E90000577A24
24E90000578A24
24E90000579A24
24E90000580A24
24E90000581A24
24E90001064A24
24E90003064B24
24E90003067B24
24E90003068B24
24E90003069B24
24E90003071B24
24E90003072B24
24E90003073B24
24E90003074B24
24E90003075B24
24E90003076B24
24E90003080B24
24E90003081B24
24E90003082B24
24E90003194B24
24E90003195B24
24E90003203B24
24E90003233B24
24E90003234B24
24E90003235B24
24E90003251B24
24E90003253B24
24E90003255B24
24E90003256B24
24E90003257B24
24E90003258B24
24E90003259B24
24E90003260B24
24E90003261B24
24E90003262B24
24E90003263B24
24E90003264B24
24E90003265B24
24E90003266B24
24E90003267B24
24E90003268B24
24E90003269B24
24E90003270B24
24E90003271B24
24E90003272B24
24E90003273B24
24E90003274B24
24E90003275B24
24E90003276B24
24E90003277B24
24E90003278B24
24E90003279B24
24E90003282B24
24E90003283B24
24E90003284B24
24E90003285B24
24E90003286B24
24E90003288B24
24E90003289B24
24E90003290B24
24E90003291B24
24E90003293B24
24E90003294B24
24E90003295B24
24E90003296B24
24E90003298B24
24E90003299B24
24E90003300B24
24E90003306B24
24E90003309B24
24E90003313B24
24E90003544B24
24E90003556B24
24E90003887B24
24E90003888B24
24E90003889B24
24E90003890B24
24E90003894B24
24E90003896B24
24E90003897B24
24E90003900B24
24E90003902B24
24E90003905B24
24E90003911B24
24E90003912B24
24E90003914B24
24E90003915B24
24E90003917B24
24E90003921B24
24E90003922B24
24E90003923B24
24E90003927B24
24E90003929B24
24E90003930B24
24E90003931B24
24E90003935B24
24E90003936B24
24E90003940B24
24E90003941B24
24E90003942B24
24E90003943B24
24E90003944B24
24E90003945B24
24E90003946B24
24E90003947B24
24E90003948B24
24E90003949B24
24E90003950B24
24E90003951B24
24E90003952B24
24E90003953B24
24E90003954B24
24E90003957B24
24E90003958B24
24E90003964B24
24E90003984B24
24E90004090B24
24E90004144B24
24E90004153B24
24E90004163B24
24E90004166B24
24E90012523A24
24E90012525A24
24E90012526A24
24E90012527A24
24E90012528A24
24E90012531A24
24E90012532A24
24E90012533A24
24E90012668A24
24E90012669A24
24E90012670A24
24E90012671A24
24E90012672A24
24E90012673A24
24E90012674A24
24E90012675A24
24E90012676A24
24E90012677A24
24E90012678A24
24E90012679A24
24E90012681A24
24E90012682A24
24E90012683A24
24E90012684A24
24E90015211A24
24E90015318A24
24E90015334A24
24E90015337A24
24E90015338A24
24E90015339A24
24E90015340A24
24E90015341A24
24E90015342A24
24E90015365A24
24E90015366A24
24E90015367A24
24E90015368A24
24E90015369A24
24E90015370A24
24E90015371A24
24E90015372A24
24E90015375A24
24E90015376A24
24E90015378A24
24E90015379A24
24E90015380A24
24E90015381A24
24E90015384A24
24E90015385A24
24E90015387A24
24E90015388A24
24E90015390A24
24E90015392A24
24E90015393A24
24E90015395A24
24E90015396A24
24E90015397A24
24E90015398A24
24E90015399A24
24E90015400A24
24E90015402A24
24E90015403A24
24E90015404A24
24E90015406A24
24E90015407A24
24E90015408A24
24E90015409A24
24E90015413A24
24E90015415A24
24E90015416A24
24E90015417A24
24E90015418A24
24E90015419A24
24E90015423A24
24E90015426A24
24E90015427A24
24E90015428A24
24E90015429A24
24E90015430A24
24E90015431A24
24E90015433A24
24E90015434A24
24E90015435A24
24E90015436A24
24E90015437A24
24E90015438A24
24E90015439A24
24E90015440A24
24E90015441A24
24E90015444A24
24E90015445A24
24E90015451A24
24E90015459A24
24E90015463A24
24E90015465A24
24E90015531A24
24E90015533A24
24E90015537A24
24E90015538A24
24E90015540A24
24E90015541A24
24E90015542A24
24E90015543A24
24E90015547A24
24E90015548A24
24E90015549A24
24E90015550A24
24E90015551A24
24E90015552A24
24E90015553A24
24E90015554A24
24E90015838A24
24E90015841A24
24E90015846A24
24E90015871A24
24E90015875A24
24E90015876A24
24E90015877A24
24E90015878A24
24E90015879A24
24E90015880A24
24E90015882A24
24E90015884A24
24E90015886A24
24E90015888A24
24E90015889A24
24E90015891A24
24E90015892A24
24E90015895A24
24E90015896A24
24E90015897A24
24E90015898A24
24E90015899A24
24E90015900A24
24E90015901A24
24E90015951A24
24E90015955A24
24E90015956A24
24E90015957A24
24E90015958A24
24E90015962A24
24E90015967A24
24E90015968A24
24E90015970A24
24E90015971A24
24E90015972A24
24E90015973A24
24E90015974A24
24E90015975A24
24E90015977A24
24E90015978A24
24E90015979A24
24E90015980A24
24E90015981A24
24E90015982A24
24E90015983A24
24E90015984A24
24E90015985A24
24E90015987A24
24E90015988A24
24E90015989A24
24E90015994A24
24E90016342A24
24E90016347A24
24E90016348A24
24E90016349A24
24E90016350A24
24E90016351A24
24E90016352A24
24E90016353A24
24E90016354A24
24E90016355A24
24E90016358A24
24E90016359A24
24E90016360A24
24E90016361A24
24E90016362A24
24E90016364A24
24E90016365A24
24E90016366A24
24E90016367A24
24E90016368A24
24E90016370A24
24E90016371A24
24E90016372A24
24E90016373A24
24E90016374A24
24E90016376A24
24E90016379A24
24E90016380A24
24E90016381A24
24E90016382A24
24E90016383A24
24E90016384A24
24E90016385A24
24E90016386A24
24E90016387A24
24E90016766A24
24E90016767A24
24E90016768A24
24E90016769A24
24E90016770A24
24E90016772A24
24E90016773A24
24E90016774A24
24E90017997A24
23E90135542J23
24E90017983A24
24E90003919B24
HeartSine SAM 450P
SAM 450P US Device, US English (2 min CPR)
Item code: 450-STR-US-10
UDI-DI code: 0506016712069
Serial Numbers:
22G91037109D22
22G91037113D22
22G91037115D22
22G91037234D22
22G91037235D22
22G91037238D22
22G91037239D22
23G90000012H23
23G90000015H23
23G90000182H23
23G90000188H23
23G90000198H23
23G90000201H23
23G90000218K23
23G90000252H23
23G90000273H23
23G90000300L23
23G90000457J23
23G90000458J23
23G90000459J23
23G90000460J23
23G90000461J23
23G90000462J23
23G90000463J23
23G90000465J23
23G90000466J23
23G90000467J23
23G90000468J23
23G90000469J23
23G90000470J23
23G90000471J23
23G90000472J23
23G90000474J23
23G90000475J23
23G90000476J23
23G90000477J23
23G90000479J23
23G90000481J23
23G90000482J23
23G90000483J23
23G90000485J23
23G90000486J23
23G90000487J23
23G90000488J23
23G90000489J23
23G90000490J23
23G90000491J23
23G90000493J23
23G90000494J23
23G90000495J23
23G90000496J23
23G90000497J23
23G90000498J23
23G90000499J23
23G90001076L23
23G90001157L23
23G90001296H23
23G90001297H23
23G90001298H23
23G90001299H23
23G90001300H23
23G90001306H23
23G90001307H23
23G90001309H23
23G90001310H23
23G90001311H23
23G90001312H23
23G90001315H23
23G90001316H23
23G90001317H23
23G90001319H23
23G90001320H23
23G90001321H23
23G90001322H23
23G90001323H23
23G90001325H23
23G90001326H23
23G90001327H23
23G90001328H23
23G90001329L23
23G90001330H23
23G90001331H23
23G90001332H23
23G90001334H23
23G90001336H23
23G90001337H23
23G90001413H23
23G90001414H23
23G90001425H23
23G90001426H23
23G90001427H23
23G90001428H23
23G90001430H23
23G90001432H23
23G90001433H23
23G90001579L23
23G90002485J23
23G90002486J23
23G90002487J23
23G90002488J23
23G90002489J23
23G90002491J23
23G90002492J23
23G90002493J23
23G90002494J23
23G90002496J23
23G90002497J23
23G90002498J23
23G90002499J23
23G90002500J23
23G90003117H23
23G90003118A23
23G90003626L23
23G90003849J23
23G90003850J23
23G90003851J23
23G90003852J23
23G90003853J23
23G90003854J23
23G90003863J23
23G90003864J23
23G90003865J23
23G90003866J23
23G90003867J23
23G90003868J23
23G90003869J23
23G90003870J23
23G90003871J23
23G90003895H23
23G90003898H23
23G90003900H23
23G90003905H23
23G90003914H23
23G90003915L23
23G90003916H23
23G90003917H23
23G90003919H23
23G90003920H23
23G90003921H23
23G90003922H23
23G90003923H23
23G90003924H23
23G90003925H23
23G90003926H23
23G90003929H23
23G90003930H23
23G90003931H23
23G90003932H23
23G90003933H23
23G90003934H23
23G90003935H23
23G90003936H23
23G90003937H23
23G90003939H23
23G90003940H23
23G90003941H23
23G90003943H23
23G90003944H23
23G90003946H23
23G90003947H23
23G90003948H23
23G90003949H23
23G90003950H23
23G90003951H23
23G90003953H23
23G90003954H23
23G90003955H23
23G90003956L23
23G90003957H23
23G90003958H23
23G90003959H23
23G90003960H23
23G90003961H23
23G90003962H23
23G90003963H23
23G90003964H23
23G90003965H23
23G90003966H23
23G90003967H23
23G90003969H23
23G90003970H23
23G90004442L23
23G90004637H23
23G90004943I23
23G90004994I23
23G90004997I23
23G90006350L23
23G90006617L23
23G90006671J23
23G90006672J23
23G90006673J23
23G90006674J23
23G90006675J23
23G90006676J23
23G90006677J23
23G90006678J23
23G90006679J23
23G90006680J23
23G90006682J23
23G90006683J23
23G90006684J23
23G90006685J23
23G90006686J23
23G90006687J23
23G90006688J23
23G90006689J23
23G90006690J23
23G90006691J23
23G90006694J23
23G90006695J23
23G90006697J23
23G90006698J23
23G90006699J23
23G90006700J23
23G90006701J23
23G90007992J23
23G90007993J23
23G90007994J23
23G90007995J23
23G90007996J23
23G90007997J23
23G90007998J23
23G90007999J23
23G90008101L23
23G90008663L23
23G90008706L23
23G90008751G23
23G90009002L23
23G90009006H23
23G90009007L23
23G90009008H23
23G90009012H23
23G90009017H23
23G90009019H23
23G90009024H23
23G90009029H23
23G90009067H23
23G90009072H23
23G90009074H23
23G90009075H23
23G90009076H23
23G90009080H23
23G90009085H23
23G90009086H23
23G90009093H23
23G90009296H23
23G90009305H23
23G90009311H23
23G90009330K23
23G90009339K23
23G90009564L23
23G90009591L23
23G90009605H23
23G90009690H23
23G90009692H23
23G90009695H23
23G90009696H23
23G90009697H23
23G90009707H23
23G90009710H23
23G90009718H23
23G90009738H23
23G90009739H23
23G90009742H23
23G90009778H23
23G90009830K23
23G90009993L23
23G90009997L23
23G90100110K23
23G90100277L23
23G90100312K23
23G90100327K23
23G90100389K23
23G90100442K23
23G90100801L23
23G90100913L23
23G90100915L23
23G90100917L23
23G90100918L23
23G90100920L23
23G90100921L23
23G90100922L23
23G90100923L23
23G90100924L23
23G90100926L23
23G90100927L23
23G90100929L23
23G90100935L23
23G90100936L23
23G90100937L23
23G90100942L23
23G90100943L23
23G90100944L23
23G90100945L23
23G90100947L23
23G90100948L23
23G90100949L23
23G90100950L23
23G90100953L23
23G90100954L23
23G90100955L23
23G90100956L23
23G90100957L23
23G90100958L23
23G90100959L23
23G90100961L23
23G90102084K23
23G90102157K23
23G90102159K23
23G90102176K23
23G90102245L23
23G90102285K23
23G90102318L23
23G90102420K23
23G90102482K23
23G90102488L23
23G90102589J23
23G90102595L23
23G90102605J23
23G90102608J23
23G90102610L23
23G90102617J23
23G90102619L23
23G90102621L23
23G90102715J23
23G90102716J23
23G90102717J23
23G90102718J23
23G90102719J23
23G90102720J23
23G90102721J23
23G90102834L23
23G90103243L23
23G90103373L23
23G90103527J23
23G90103540J23
23G90103564J23
23G90103700L23
23G90103809J23
23G90105001L23
23G90105010K23
23G90105132K23
23G90105412K23
23G90105415K23
23G90105416K23
23G90105417K23
23G90105429K23
23G90105431K23
23G90105433K23
23G90105435K23
23G90105436K23
23G90105438K23
23G90105439K23
23G90105441L23
23G90105452K23
23G90105459K23
23G90105463K23
23G90105533K23
23G90105538K23
23G90105539K23
23G90105542K23
23G90105552K23
23G90105557K23
23G90105602K23
23G90105603K23
23G90105739L23
23G90105841L23
23G90105847L23
23G90105851L23
23G90105863L23
23G90105865L23
23G90106002K23
23G90106004K23
23G90106006K23
23G90106008K23
23G90106009K23
23G90106013K23
23G90106014K23
23G90106015K23
23G90106017K23
23G90106018K23
23G90106019K23
23G90106023K23
23G90106025K23
23G90106026K23
23G90106027K23
23G90106030K23
23G90106031K23
23G90106034K23
23G90106037K23
23G90106039K23
23G90106040K23
23G90106041K23
23G90106042K23
23G90106043K23
23G90106044K23
23G90106047K23
23G90106048K23
23G90106050K23
23G90106055K23
23G90106057K23
23G90106059K23
23G90106062L23
23G90106063K23
23G90106064K23
23G90106065K23
23G90106066K23
23G90106069K23
23G90106071K23
23G90106072K23
23G90106075K23
23G90106076K23
23G90106077K23
23G90106078K23
23G90106079K23
23G90106080K23
23G90106081K23
23G90106085K23
23G90106086K23
23G90106087K23
23G90106088K23
23G90106089K23
23G90106091K23
23G90106094K23
23G90106096K23
23G90106097K23
23G90106099K23
23G90106101K23
23G90106102K23
23G90106207K23
23G90106220K23
23G90106221K23
23G90106240K23
23G90106241K23
23G90106243K23
23G90106247K23
23G90106254K23
23G90106257K23
23G90106263K23
23G90106269K23
23G90106275K23
23G90106277K23
23G90106292K23
23G90106293K23
23G90106294K23
23G90106296K23
23G90106298K23
23G90106301K23
23G90106305K23
23G90106310K23
23G90106313K23
23G90106316K23
23G90106317K23
23G90106318K23
23G90106319K23
23G90106321K23
23G90106322K23
23G90106356K23
23G90106357K23
23G90106361K23
23G90106367K23
23G90106369K23
23G90106370K23
23G90106371L23
23G90106519L23
23G90106524K23
23G90106559L23
23G90106594L23
23G90106608L23
23G90106844K23
23G90107038J23
23G90107044J23
23G90107050J23
23G90107052J23
23G90107152J23
23G90107231J23
23G90107242J23
23G90107250J23
23G90107251J23
23G90107279L23
23G90107291J23
23G90107300J23
23G90107484J23
23G90107533J23
23G90107534J23
23G90107535J23
23G90107536J23
23G90107561K23
23G90107563K23
23G90107665J23
23G90107717K23
23G90107720K23
23G90107723K23
23G90107724K23
23G90107732L23
23G90107969K23
23G90107970K23
23G90108098L23
23G90108197L23
23G90108248L23
23G90108249L23
23G90108250L23
23G90108252L23
23G90108253L23
23G90108254L23
23G90108255L23
23G90108256L23
23G90108257L23
23G90108258L23
23G90108259L23
23G90108260L23
23G90108261L23
23G90108262L23
23G90108263L23
23G90108264L23
23G90108265L23
23G90108266L23
23G90108267L23
23G90108268L23
23G90108269L23
23G90108270L23
23G90108271L23
23G90108272L23
23G90108273L23
23G90108274L23
23G90108275L23
23G90108276L23
23G90108278L23
23G90108279L23
23G90108280L23
23G90108281L23
23G90108282L23
23G90108283L23
23G90108284L23
23G90108285L23
23G90108286L23
23G90108287L23
23G90108288L23
23G90108289L23
23G90108290L23
23G90108291L23
23G90108292L23
23G90108293L23
23G90108294L23
23G90108295L23
23G90108296L23
23G90108313L23
23G90108319L23
23G90108454L23
23G90108455L23
23G90108456L23
23G90108458L23
23G90108459L23
23G90108461L23
23G90108462L23
23G90108463L23
23G90108464L23
23G90108465L23
23G90108466L23
23G90108467L23
23G90108468L23
23G90108469L23
23G90108470L23
23G90108471L23
23G90108472L23
23G90108474L23
23G90108475L23
23G90108476L23
23G90108478L23
23G90108480L23
23G90108481L23
23G90108482L23
23G90108483L23
23G90108484L23
23G90108485L23
23G90108486L23
23G90108487L23
23G90108488L23
23G90108489L23
23G90108490L23
23G90108491L23
23G90108493L23
23G90108494L23
23G90108496L23
23G90108498L23
23G90108499L23
23G90108500L23
23G90108501L23
23G90108502L23
23G90108505L23
23G90108506L23
23G90108507L23
23G90108508L23
23G90108509L23
23G90108510L23
23G90108511L23
23G90108512L23
23G90108513L23
23G90108514L23
23G90108515L23
23G90108517L23
23G90108518L23
23G90108519L23
23G90108521L23
23G90108522L23
23G90108523L23
23G90108524L23
23G90108525L23
23G90108526L23
23G90108527L23
23G90108528L23
23G90108529L23
23G90108530L23
23G90108531L23
23G90108532L23
23G90108533L23
23G90108534L23
23G90108535L23
23G90108536L23
23G90108537L23
23G90108538L23
23G90108540L23
23G90108541L23
23G90108542L23
23G90108544L23
23G90108545L23
23G90108546L23
23G90108547L23
23G90108548L23
23G90108549L23
23G90108550L23
23G90108551L23
23G90108552L23
23G90108553L23
23G90108564L23
23G90108592L23
23G90108604L23
23G90108605L23
23G90108606L23
23G90108607L23
23G90108608L23
23G90108609L23
23G90108610L23
23G90108611L23
23G90108612L23
23G90108613L23
23G90108614L23
23G90108615L23
23G90108616L23
23G90108617L23
23G90108618L23
23G90108619L23
23G90108620L23
23G90108621L23
23G90108623L23
23G90108624L23
23G90108625L23
23G90108626L23
23G90108628L23
23G90108630L23
23G90108631L23
23G90108633L23
23G90108634L23
23G90108635L23
23G90108636L23
23G90108639L23
23G90108640L23
23G90108641L23
23G90108642L23
23G90108643L23
23G90108644L23
23G90108645L23
23G90108648L23
23G90108654L23
23G90108655L23
23G90108656L23
23G90108657L23
23G90108658L23
23G90108659L23
23G90108660L23
23G90108661L23
23G90108662L23
23G90108663L23
23G90108664L23
23G90108665L23
23G90108666L23
23G90108667L23
23G90108668L23
23G90108670L23
23G90108671L23
23G90108672L23
23G90108673L23
23G90108674L23
23G90108675L23
23G90108676L23
23G90108677L23
23G90108678L23
23G90108679L23
23G90108680L23
23G90108683L23
23G90108684L23
23G90108686L23
23G90108687L23
23G90108688L23
23G90108689L23
23G90108690L23
23G90108691L23
23G90108692L23
23G90108693L23
23G90108694L23
23G90108695L23
23G90108696L23
23G90108697L23
23G90108698L23
23G90108699L23
23G90108700L23
23G90108702L23
23G90108703L23
23G90108704L23
23G90108705L23
23G90108706L23
23G90108707L23
23G90108708L23
23G90108709L23
23G90109058J23
23G90109059J23
23G90109060J23
23G90109061J23
23G90109062J23
23G90109063J23
23G90109064J23
23G90109066J23
23G90109068J23
23G90109069J23
23G90109070J23
23G90109071J23
23G90109072J23
23G90109073J23
23G90109074J23
23G90109075J23
23G90109263J23
23G90109313J23
23G90109314J23
23G90109317J23
23G90109319J23
23G90109320J23
23G90109323J23
23G90109324J23
23G90109325J23
23G90109327J23
23G90109332J23
23G90109336J23
23G90109337J23
23G90109345J23
23G90109346J23
23G90109347J23
23G90109351J23
23G90109355J23
23G90109357J23
23G90109358J23
23G90109359J23
23G90109360J23
23G90109363J23
23G90109364J23
23G90109368J23
23G90109370J23
23G90109372J23
23G90109375J23
23G90109385J23
23G90109388J23
23G90109389J23
23G90109391J23
23G90109393J23
23G90109398J23
23G90109401J23
23G90109403J23
23G90109407J23
23G90109408J23
23G90109409J23
23G90109411J23
23G90109415J23
23G90109421J23
23G90109424J23
23G90109431J23
23G90109436J23
23G90109437J23
23G90109446J23
23G90109447J23
23G90109448J23
23G90109451J23
23G90109452J23
23G90109457J23
23G90109458J23
23G90109460J23
23G90109461J23
23G90109472J23
23G90109477J23
23G90109537J23
23G90109632J23
23G90109639J23
23G90109646J23
23G90109662J23
23G90109676J23
23G90109778J23
23G90109779J23
23G90110022I23
23G90110026I23
23G90110027I23
23G90110028I23
23G90110030I23
23G90110031I23
23G90110032I23
23G90110033I23
23G90110035I23
23G90110036I23
23G90110037I23
23G90110040I23
23G90110074I23
23G90110087I23
23G90110151I23
23G90110152I23
23G90110269I23
23G90110270I23
23G90110276I23
23G90110277L23
23G90110279I23
23G90110294L23
23G90110362I23
23G90110370I23
23G90110374I23
23G90110389I23
23G90112532I23
23G90112553L23
23G90112691L23
23G90112701L23
23G90112708L23
23G90112709L23
23G90112711L23
23G90112712L23
23G90112714L23
23G90112715L23
23G90112719L23
23G90112720L23
23G90112721L23
23G90112722L23
23G90112724L23
23G90112727L23
23G90112728L23
23G90112731L23
23G90112732L23
23G90112735L23
23G90112737L23
23G90112738L23
23G90112893I23
24G90000085B24
24G90000086B24
24G90000089B24
24G90000099B24
24G90000102B24
24G90000103B24
24G90000104B24
24G90000105B24
24G90000106B24
24G90000109B24
24G90000112B24
24G90000114B24
24G90000115B24
24G90000116B24
24G90000117B24
24G90000118B24
24G90000119B24
24G90000120B24
24G90000122B24
24G90000123B24
24G90000129B24
24G90000130B24
24G90000132B24
24G90001563D24
24G90005260B24
24G90005262B24
24G90005267B24
24G90005298B24
24G90005299B24
24G90005304B24
24G90005375B24
24G90005376B24
24G90005400B24
24G90005409B24
24G90005431B24
24G90005433B24
24G90005435B24
24G90005452B24
24G90005454B24
24G90005458B24
24G90005460B24
24G90005477B24
24G90007696D24
24G90009459D24
24G90010794A24
24G90010843A24
24G90011010B24
24G90011011B24
24G90011012B24
24G90011016B24
24G90011017B24
24G90011018B24
24G90011021B24
24G90011024B24
24G90011025B24
24G90011336B24
24G90012517B24
24G90012519B24
24G90012528B24
24G90012531B24
24G90012533B24
24G90012589B24
24G90012591B24
24G90012592B24
24G90012594B24
24G90012595B24
24G90012596B24
24G90012597B24
24G90012598B24
24G90012600B24
24G90012749D24
24G90015765A24
24G90015869A24
24G90016018A24
24G90016021A24
24G90016030A24
24G90016079A24
24G90016088A24
24G90016324A24
24G90016325A24
24G90016326A24
24G90016327A24
24G90016329A24
24G90016330A24
24G90016331A24
24G90016332A24
24G90016334A24
24G90016335A24
24G90016337A24
24G90016338A24
24G90016339A24
24G90016341A24
24G90016350A24
24G90016352A24
24G90016373A24
24G90016374A24
24G90016375A24
24G90016376A24
24G90016377A24
24G90016378A24
24G90016379A24
24G90016380A24
24G90016381A24
24G90016382A24
24G90016383A24
24G90016384A24
24G90016385A24
24G90016386A24
24G90016387A24
24G90016388A24
24G90016389A24
24G90016390A24
24G90016391A24
24G90016392A24
24G90016393A24
24G90016394A24
24G90016437A24
24G90016438A24
24G90016439A24
24G90016440A24
24G90016441A24
24G90016443A24
24G90016473A24
24G90016474A24
24G90016475A24
24G90016476A24
24G90016478A24
24G90016479A24
24G90016480A24
24G90016481A24
24G90016482A24
24G90016483A24
24G90016484A24
24G90016485A24
24G90016486A24
24G90016487A24
24G90016488A24
24G90016489A24
24G90016490A24
24G90016491A24
24G90016492A24
24G90016493A24
24G90016494A24
24G90016495A24
24G90016496A24
24G90016497A24
24G90016498A24
24G90016499A24
24G90016500A24
24G90016501A24
24G90016502A24
24G90016503A24
24G90016504A24
24G90016505A24
24G90016506A24
24G90016507A24
24G90016508A24
24G90016509A24
24G90016510A24
24G90016511A24
24G90016512A24
24G90016513A24
24G90016515A24
24G90016516A24
24G90016517A24
24G90016518A24
24G90016519A24
24G90016520A24
24G90016521A24
24G90016582A24
24G90016586A24
24G90016627A24
24G90016713A24
24G90016715A24
24G90016717A24
24G90016723A24
24G90016727A24
24G90016757A24
24G90016915B24
24G90016927B24
24G90016930B24
24G90017061B24
24G90017062B24
24G90017065B24
24G90017066B24
24G90017067B24
24G90017069B24
24G90017071B24
24G90017263B24
24G90017270B24
24G90017271B24
24G90017278B24
24G90017446B24
24G90017449B24
24G90017450B24
24G90017482B24
24G90017483B24
24G90017486B24
24G90017487B24
24G90017488B24
24G90017489B24
24G90017490B24
24G90017491B24
24G90017492B24
24G90017493B24
24G90017494B24
24G90017497B24
24G90017498B24
24G90017501B24
24G90017502B24
24G90017590B24
24G90017591B24
24G90017592B24
24G90017593B24
24G90017596B24
24G90017597B24
24G90017599B24
24G90017600B24
24G90017601B24
24G90017602B24
24G90018091H24
23G90004172G23
23G90008119L23
23G90009047L23
23G90100150B24
23G90102061J23
23G90102062J23
23G90102063J23
23G90102086J23
23G90102087J23
23G90102088J23
23G90102095J23
23G90102096J23
23G90102097J23
23G90102098J23
23G90102103J23
23G90102166K23
23G90102182K23
23G90102184K23
23G90102212K23
23G90102213K23
23G90102221K23
23G90102222K23
23G90102223K23
23G90102224K23
23G90102225K23
23G90102226K23
23G90102227K23
23G90102228K23
23G90102229K23
23G90102230K23
23G90102231K23
23G90102232K23
23G90102233K23
23G90102234K23
23G90102235K23
23G90102236K23
23G90102237K23
23G90102238K23
23G90102239K23
23G90102241K23
23G90102243K23
23G90102244K23
23G90102246K23
23G90102247K23
23G90102248K23
23G90102249K23
23G90102250K23
23G90102255K23
23G90102274K23
23G90102276K23
23G90102277K23
23G90102279K23
23G90102280K23
23G90102281K23
23G90102295K23
23G90102351K23
23G90102354K23
23G90102363K23
23G90102365K23
23G90102368K23
23G90102375K23
23G90102376K23
23G90102377K23
23G90102378K23
23G90102380K23
23G90102381K23
23G90102383K23
23G90102385K23
23G90102395K23
23G90102398K23
23G90102438K23
23G90102440K23
23G90102445K23
23G90102447K23
23G90102448K23
23G90102449K23
23G90102455K23
23G90102458K23
23G90102460K23
23G90102461K23
23G90102462K23
23G90102463K23
23G90102464K23
23G90102465K23
23G90102466K23
23G90102467K23
23G90102469K23
23G90102473K23
23G90102477K23
23G90102633J23
23G90102822J23
23G90102825J23
23G90102831J23
23G90102832K23
23G90102836K23
23G90102838K23
23G90102843K23
23G90102844K23
23G90102846K23
23G90102848K23
23G90102851K23
23G90102853K23
23G90102856K23
23G90102858K23
23G90102860K23
23G90102862K23
23G90102864K23
23G90103012J23
23G90103015J23
23G90103066J23
23G90103084J23
23G90103087J23
23G90103095J23
23G90103100J23
23G90103214J23
23G90103215J23
23G90103216J23
23G90103222J23
23G90103234J23
23G90103307J23
23G90103308J23
23G90103309J23
23G90103310J23
23G90103316J23
23G90103339J23
23G90103341J23
23G90103364J23
23G90103369J23
23G90103511J23
23G90103518J23
23G90103536J23
23G90103537J23
23G90103538J23
23G90103539J23
23G90103541J23
23G90103542J23
23G90103543J23
23G90103544J23
23G90103545J23
23G90103549J23
23G90103550J23
23G90103552J23
23G90103553J23
23G90103554J23
23G90103555J23
23G90103556J23
23G90103557J23
23G90103558J23
23G90103560J23
23G90103562J23
23G90103570J23
23G90103572J23
23G90103574J23
23G90103576J23
23G90103581J23
23G90103588J23
23G90103592J23
23G90103594J23
23G90103597J23
23G90103605J23
23G90103606J23
23G90103607J23
23G90103608J23
23G90103609J23
23G90103649J23
23G90103652J23
23G90103656J23
23G90103658J23
23G90103659J23
23G90103660J23
23G90103668J23
23G90103677J23
23G90103678J23
23G90103680J23
23G90103681J23
23G90103683J23
23G90103684J23
23G90103686J23
23G90103687J23
23G90103688J23
23G90103689J23
23G90103690J23
23G90103691J23
23G90103692J23
23G90103694J23
23G90103696J23
23G90103697J23
23G90103699J23
23G90103702J23
23G90103703J23
23G90103705J23
23G90103707J23
23G90103712J23
23G90103799J23
23G90103800J23
23G90103801J23
23G90103802J23
23G90103803J23
23G90103804J23
23G90103844J23
23G90103848J23
23G90103850J23
23G90103852J23
23G90103856J23
23G90103858J23
23G90103953J23
23G90103954J23
23G90103955J23
23G90103956J23
23G90103957J23
23G90103958J23
23G90103959J23
23G90103983J23
23G90103991J23
23G90103996J23
23G90103997J23
23G90103998J23
23G90103999J23
23G90104000J23
23G90105394K23
23G90105644K23
23G90107046J23
23G90107047J23
23G90107053J23
23G90107054J23
23G90107056J23
23G90107057J23
23G90107058J23
23G90107060J23
23G90107061J23
23G90107062J23
23G90107447J23
23G90107448J23
23G90107662J23
23G90107710J23
23G90107712J23
23G90107713J23
23G90108198L23
23G90108199L23
23G90108200L23
23G90108201L23
23G90108202L23
23G90108203L23
23G90108204L23
23G90108205L23
23G90108207L23
23G90108208L23
23G90108209L23
23G90108210L23
23G90108211L23
23G90108212L23
23G90108213L23
23G90108215L23
23G90108218L23
23G90108219L23
23G90108221L23
23G90108222L23
23G90108223L23
23G90108225L23
23G90108226L23
23G90108227L23
23G90108235L23
23G90108236L23
23G90108237L23
23G90108238L23
23G90108239L23
23G90108240L23
23G90108241L23
23G90108242L23
23G90108244L23
23G90108246L23
23G90108247L23
23G90108322L23
23G90108326L23
23G90108339L23
23G90108341L23
23G90108342L23
23G90108344L23
23G90108345L23
23G90108347L23
23G90108363L23
23G90108365L23
23G90108368L23
23G90108398L23
23G90108399L23
23G90108401L23
23G90108402L23
23G90108403L23
23G90108404L23
23G90108405L23
23G90108408L23
23G90108409L23
23G90108410L23
23G90108411L23
23G90108412L23
23G90108415L23
23G90108416L23
23G90108417L23
23G90108418L23
23G90108419L23
23G90108428L23
23G90108437L23
23G90108440L23
23G90108441L23
23G90108442L23
23G90108457L23
23G90108473L23
23G90108477L23
23G90108479L23
23G90108492L23
23G90108495L23
23G90108497L23
23G90109799J23
23G90109816J23
23G90109861J23
23G90109878J23
23G90109879J23
23G90109880J23
23G90109884J23
23G90109885J23
23G90109886J23
23G90109887J23
23G90109893J23
23G90109896J23
23G90109897J23
23G90109899J23
24G90000014B24
24G90000032B24
24G90000034B24
24G90000035B24
24G90000049B24
24G90000050B24
24G90000051B24
24G90000084B24
24G90000087B24
24G90000090B24
24G90000091B24
24G90000098B24
24G90000101B24
24G90000107B24
24G90000108B24
24G90000111B24
24G90000113B24
24G90000121B24
24G90000128B24
24G90000133B24
24G90000134B24
24G90000135B24
24G90000136B24
24G90000137B24
24G90000138B24
24G90000139B24
24G90000141B24
24G90000142B24
24G90000143B24
24G90000144B24
24G90000145B24
24G90000146B24
24G90000147B24
24G90000148B24
24G90000149B24
24G90000150B24
24G90000151B24
24G90000152B24
24G90001466D24
24G90001545D24
24G90001546D24
24G90001549D24
24G90001550D24
24G90001552D24
24G90001553D24
24G90001558D24
24G90001564D24
24G90001567D24
24G90001568D24
24G90001584D24
24G90001585D24
24G90001586D24
24G90001587D24
24G90001588D24
24G90001589D24
24G90001590D24
24G90001591D24
24G90001592D24
24G90001593D24
24G90001594D24
24G90001595D24
24G90001597D24
24G90001598D24
24G90001599D24
24G90001600D24
24G90001601D24
24G90001603D24
24G90001604D24
24G90001605D24
24G90001606D24
24G90001609D24
24G90001610D24
24G90001611D24
24G90001613D24
24G90001614D24
24G90001616D24
24G90001617D24
24G90001618D24
24G90001619D24
24G90001620D24
24G90001624D24
24G90001626D24
24G90001628D24
24G90001629D24
24G90001630D24
24G90001631D24
24G90001632D24
24G90001633D24
24G90001634D24
24G90001635D24
24G90001636D24
24G90001637D24
24G90001638D24
24G90001639D24
24G90001640D24
24G90001641D24
24G90001642D24
24G90001643D24
24G90001644D24
24G90001646D24
24G90001647D24
24G90001648D24
24G90001649D24
24G90001650D24
24G90001651D24
24G90001652D24
24G90001653D24
24G90001654D24
24G90001656D24
24G90001657D24
24G90001658D24
24G90001659D24
24G90001661D24
24G90001663D24
24G90001664D24
24G90001666D24
24G90001667D24
24G90001668D24
24G90001669D24
24G90001670D24
24G90005261B24
24G90005265B24
24G90005266B24
24G90005268B24
24G90005269B24
24G90005273B24
24G90005274B24
24G90005275B24
24G90005276B24
24G90005277B24
24G90005278B24
24G90005279B24
24G90005281B24
24G90005282B24
24G90005300B24
24G90005301B24
24G90005303B24
24G90005305B24
24G90005307B24
24G90005308B24
24G90005311B24
24G90005491B24
24G90005492B24
24G90005493B24
24G90005494B24
24G90005646B24
24G90005647B24
24G90005648B24
24G90005649B24
24G90005650B24
24G90005653B24
24G90005655B24
24G90005656B24
24G90005657B24
24G90005658B24
24G90005659B24
24G90005660B24
24G90005661B24
24G90005662B24
24G90005663B24
24G90005664B24
24G90005665B24
24G90005666B24
24G90005667B24
24G90005668B24
24G90005669B24
24G90005670B24
24G90005671B24
24G90005672B24
24G90005673B24
24G90005674B24
24G90005675B24
24G90005676B24
24G90005677B24
24G90005678B24
24G90005679B24
24G90005680B24
24G90005681B24
24G90005682B24
24G90005683B24
24G90005684B24
24G90005685B24
24G90005686B24
24G90005687B24
24G90005688B24
24G90005689B24
24G90005690B24
24G90005691B24
24G90005692B24
24G90005693B24
24G90005694B24
24G90005695B24
24G90005696B24
24G90005697B24
24G90005699B24
24G90005700B24
24G90005701B24
24G90005702B24
24G90005703B24
24G90005704B24
24G90005705B24
24G90005706B24
24G90005707B24
24G90005708B24
24G90007313H24
24G90008358D24
24G90008359D24
24G90008360D24
24G90008361D24
24G90008363D24
24G90008364D24
24G90008378D24
24G90008380D24
24G90008383D24
24G90008387D24
24G90008404D24
24G90010612A24
24G90010613A24
24G90010622A24
24G90010623A24
24G90010624A24
24G90010626A24
24G90010631A24
24G90010640A24
24G90010641A24
24G90010642A24
24G90010643A24
24G90010644A24
24G90010645A24
24G90010646A24
24G90010647A24
24G90010649A24
24G90010650A24
24G90010651A24
24G90010658A24
24G90010659A24
24G90010661A24
24G90010683A24
24G90010684A24
24G90010685A24
24G90010686A24
24G90010687A24
24G90010689A24
24G90010691A24
24G90010692A24
24G90010693A24
24G90010694A24
24G90010695A24
24G90010696A24
24G90010697A24
24G90010698A24
24G90010699A24
24G90010700A24
24G90010701A24
24G90010702A24
24G90010703A24
24G90010704A24
24G90010705A24
24G90010706A24
24G90010707A24
24G90010708A24
24G90010710A24
24G90010711A24
24G90010712A24
24G90010713A24
24G90010715A24
24G90010716A24
24G90010717A24
24G90010718A24
24G90010719A24
24G90010720A24
24G90010721A24
24G90010722A24
24G90010723A24
24G90010724A24
24G90010725A24
24G90010726A24
24G90010727A24
24G90010728A24
24G90010729A24
24G90010730A24
24G90010731A24
24G90010732A24
24G90010733A24
24G90010734A24
24G90010735A24
24G90010736A24
24G90010737A24
24G90010749A24
24G90010750A24
24G90010752A24
24G90010755A24
24G90010758A24
24G90010761A24
24G90010771A24
24G90010772A24
24G90010780A24
24G90010781A24
24G90010785A24
24G90010788A24
24G90010791A24
24G90010792A24
24G90010793A24
24G90010796A24
24G90010800A24
24G90010803A24
24G90010804A24
24G90010805A24
24G90010807A24
24G90010808A24
24G90010809A24
24G90010810A24
24G90010811A24
24G90010813A24
24G90010814A24
24G90010815A24
24G90010816A24
24G90010817A24
24G90010818A24
24G90010819A24
24G90010820A24
24G90010821A24
24G90010822A24
24G90010823A24
24G90010824A24
24G90010826A24
24G90010827A24
24G90010828A24
24G90010830A24
24G90010833A24
24G90010834A24
24G90010835A24
24G90010836A24
24G90010837A24
24G90010839A24
24G90010841A24
24G90010842A24
24G90010844A24
24G90010845A24
24G90010978B24
24G90010979B24
24G90010980B24
24G90010984B24
24G90010985B24
24G90010987B24
24G90010988B24
24G90010989B24
24G90010990B24
24G90010991B24
24G90010992B24
24G90010993B24
24G90010994B24
24G90010995B24
24G90010997B24
24G90010999B24
24G90011004B24
24G90011005B24
24G90011006B24
24G90011009B24
24G90011013B24
24G90011029B24
24G90011030B24
24G90011031B24
24G90011032B24
24G90011033B24
24G90011034B24
24G90011035B24
24G90011036B24
24G90011037B24
24G90011038B24
24G90011039B24
24G90011044B24
24G90011045B24
24G90011046B24
24G90011047B24
24G90011048B24
24G90011049B24
24G90011050B24
24G90011051B24
24G90011054B24
24G90011055B24
24G90011056B24
24G90011058B24
24G90011059B24
24G90011060B24
24G90011061B24
24G90011062B24
24G90011063B24
24G90011066B24
24G90011067B24
24G90011068B24
24G90011069B24
24G90011070B24
24G90011074B24
24G90011075B24
24G90011076B24
24G90011077B24
24G90011179B24
24G90011183B24
24G90011186B24
24G90011187B24
24G90011188B24
24G90011189B24
24G90011191B24
24G90011192B24
24G90011194B24
24G90011196B24
24G90011197B24
24G90011198B24
24G90011199B24
24G90011200B24
24G90011201B24
24G90011202B24
24G90011203B24
24G90011204B24
24G90011205B24
24G90011206B24
24G90011207B24
24G90011208B24
24G90011209B24
24G90011211B24
24G90011212B24
24G90011213B24
24G90011214B24
24G90011216B24
24G90011220B24
24G90011221B24
24G90011225B24
24G90011226B24
24G90011286B24
24G90011290B24
24G90011292B24
24G90011293B24
24G90011330B24
24G90011340B24
24G90011341B24
24G90011344B24
24G90011346B24
24G90011351B24
24G90011352B24
24G90011354B24
24G90011356B24
24G90011357B24
24G90011358B24
24G90011359B24
24G90011360B24
24G90011501B24
24G90011502B24
24G90011503B24
24G90011504B24
24G90011505B24
24G90011506B24
24G90011507B24
24G90011508B24
24G90011509B24
24G90011510B24
24G90011511B24
24G90011512B24
24G90011513B24
24G90011514B24
24G90011515B24
24G90011516B24
24G90011517B24
24G90011518B24
24G90011519B24
24G90011520B24
24G90011521B24
24G90011522B24
24G90011524B24
24G90011525B24
24G90011526B24
24G90011528B24
24G90011529B24
24G90011530B24
24G90011531B24
24G90011533B24
24G90011534B24
24G90011535B24
24G90011536B24
24G90011539B24
24G90011541B24
24G90011542B24
24G90011543B24
24G90011545B24
24G90011546B24
24G90011547B24
24G90011548B24
24G90011550B24
24G90011588B24
24G90011589B24
24G90011590B24
24G90011591B24
24G90011592B24
24G90011593B24
24G90011594B24
24G90011595B24
24G90011596B24
24G90011598B24
24G90011599B24
24G90011600B24
24G90011651B24
24G90011652B24
24G90011653B24
24G90011654B24
24G90011655B24
24G90011656B24
24G90011657B24
24G90011658B24
24G90011659B24
24G90011660B24
24G90011661B24
24G90011662B24
24G90011663B24
24G90011664B24
24G90011665B24
24G90011666B24
24G90011667B24
24G90011668B24
24G90011669B24
24G90011670B24
24G90011671B24
24G90011672B24
24G90011673B24
24G90011674B24
24G90011675B24
24G90011676B24
24G90011677B24
24G90011678B24
24G90011679B24
24G90011680B24
24G90011681B24
24G90011683B24
24G90011684B24
24G90011685B24
24G90011686B24
24G90011687B24
24G90011688B24
24G90011689B24
24G90011690B24
24G90011691B24
24G90011692B24
24G90011693B24
24G90011694B24
24G90011695B24
24G90011696B24
24G90011698B24
24G90011699B24
24G90011700B24
24G90011701B24
24G90011702B24
24G90011703B24
24G90011704B24
24G90011705B24
24G90011706B24
24G90011707B24
24G90011708B24
24G90011709B24
24G90011710B24
24G90011711B24
24G90011712B24
24G90011713B24
24G90011714B24
24G90011733B24
24G90011736B24
24G90011737B24
24G90011739B24
24G90011740B24
24G90011741B24
24G90011742B24
24G90011743B24
24G90011744B24
24G90011745B24
24G90011746B24
24G90011747B24
24G90011748B24
24G90011749B24
24G90011750B24
24G90011751B24
24G90011752B24
24G90011753B24
24G90011754B24
24G90011755B24
24G90011757B24
24G90011758B24
24G90011759B24
24G90011760B24
24G90011761B24
24G90011762B24
24G90011763B24
24G90011764B24
24G90011765B24
24G90011766B24
24G90011767B24
24G90011768B24
24G90011769B24
24G90011771B24
24G90011773B24
24G90011774B24
24G90011775B24
24G90011776B24
24G90011777B24
24G90011778B24
24G90011779B24
24G90011780B24
24G90011781B24
24G90011782B24
24G90011783B24
24G90011784B24
24G90011785B24
24G90011786B24
24G90011787B24
24G90011788B24
24G90011789B24
24G90011790B24
24G90011792B24
24G90011793B24
24G90011794B24
24G90011795B24
24G90011797B24
24G90011799B24
24G90011801B24
24G90011802B24
24G90011803B24
24G90011804B24
24G90011806B24
24G90011808B24
24G90011809B24
24G90011810B24
24G90011811B24
24G90011814B24
24G90011816B24
24G90011817B24
24G90011818B24
24G90011820B24
24G90011821B24
24G90011822B24
24G90011903B24
24G90011904B24
24G90011910B24
24G90011912B24
24G90011913B24
24G90011920B24
24G90011921B24
24G90011922B24
24G90011923B24
24G90011925B24
24G90011926B24
24G90011930B24
24G90011931B24
24G90011932B24
24G90011933B24
24G90011934B24
24G90011935B24
24G90011936B24
24G90011937B24
24G90011938B24
24G90011939B24
24G90011940B24
24G90011942B24
24G90011943B24
24G90011944B24
24G90011945B24
24G90011946B24
24G90011947B24
24G90011948B24
24G90011949B24
24G90011950B24
24G90011951B24
24G90011952B24
24G90011953B24
24G90011954B24
24G90011955B24
24G90011956B24
24G90011957B24
24G90011958B24
24G90011960B24
24G90011964B24
24G90011965B24
24G90011966B24
24G90011967B24
24G90011968B24
24G90011969B24
24G90011970B24
24G90011971B24
24G90011973B24
24G90011974B24
24G90011975B24
24G90011976B24
24G90011977B24
24G90011978B24
24G90011979B24
24G90011980B24
24G90012063B24
24G90012065B24
24G90012066B24
24G90012067B24
24G90012069B24
24G90012071B24
24G90012091B24
24G90012093B24
24G90012095B24
24G90012097B24
24G90012098B24
24G90012099B24
24G90012105B24
24G90012229B24
24G90012235B24
24G90012501B24
24G90012502B24
24G90012503B24
24G90012504B24
24G90012505B24
24G90012506B24
24G90012507B24
24G90012508B24
24G90012509B24
24G90012510B24
24G90012511B24
24G90012512B24
24G90012513B24
24G90012514B24
24G90012515B24
24G90012516B24
24G90012520B24
24G90012521B24
24G90012522B24
24G90012523B24
24G90012524B24
24G90012525B24
24G90012526B24
24G90012527B24
24G90012530B24
24G90012537B24
24G90012551B24
24G90012552B24
24G90012553B24
24G90012554B24
24G90012555B24
24G90012556B24
24G90012557B24
24G90012558B24
24G90012559B24
24G90012560B24
24G90012561B24
24G90012562B24
24G90012563B24
24G90012564B24
24G90012567B24
24G90012568B24
24G90012569B24
24G90012570B24
24G90012571B24
24G90012572B24
24G90012573B24
24G90012575B24
24G90012576B24
24G90012577B24
24G90012582B24
24G90012593B24
24G90012599B24
24G90012869H24
24G90012872H24
24G90012873H24
24G90012876H24
24G90012879H24
24G90012885H24
24G90012886H24
24G90012888H24
24G90012890H24
24G90012891H24
24G90012894H24
24G90012895H24
24G90012897H24
24G90012898H24
24G90012899H24
24G90012901H24
24G90012903H24
24G90012904H24
24G90012905H24
24G90012906H24
24G90012910H24
24G90012912H24
24G90012914H24
24G90012915H24
24G90012916H24
24G90012917H24
24G90012918H24
24G90012919H24
24G90012921H24
24G90012922H24
24G90012925H24
24G90012927H24
24G90012929H24
24G90012930H24
24G90012931H24
24G90012932H24
24G90012934H24
24G90012935H24
24G90012936H24
24G90012937H24
24G90012939H24
24G90012940H24
24G90012952H24
24G90012953H24
24G90012960H24
24G90012961H24
24G90012963H24
24G90012964H24
24G90012966H24
24G90013174H24
24G90013175H24
24G90013178H24
24G90013179H24
24G90013183H24
24G90013209H24
24G90013210H24
24G90013211H24
24G90013213H24
24G90013215H24
24G90013216H24
24G90013217H24
24G90013218H24
24G90013220H24
24G90013222H24
24G90013225H24
24G90013226H24
24G90013227H24
24G90013229H24
24G90013230H24
24G90013234H24
24G90013248H24
24G90013249H24
24G90013250H24
24G90013253H24
24G90013254H24
24G90013255H24
24G90013256H24
24G90013257H24
24G90013258H24
24G90013259H24
24G90013260H24
24G90013265H24
24G90013269H24
24G90013270H24
24G90013271H24
24G90013281H24
24G90013282H24
24G90013283H24
24G90013284H24
24G90013285H24
24G90013286H24
24G90013287H24
24G90013288H24
24G90013289H24
24G90013290H24
24G90013291H24
24G90013292H24
24G90013293H24
24G90013294H24
24G90013295H24
24G90013296H24
24G90013297H24
24G90013311H24
24G90013316H24
24G90013317H24
24G90013318H24
24G90013319H24
24G90013320H24
24G90013321H24
24G90013322H24
24G90013323H24
24G90013324H24
24G90013325H24
24G90013326H24
24G90013327H24
24G90013328H24
24G90013329H24
24G90013331H24
24G90013357H24
24G90013364H24
24G90013365H24
24G90013366H24
24G90013367H24
24G90013368H24
24G90013369H24
24G90013370H24
24G90013371H24
24G90013372H24
24G90013373H24
24G90013374H24
24G90013375H24
24G90015796A24
24G90015801A24
24G90015851A24
24G90015852A24
24G90015853A24
24G90015854A24
24G90015855A24
24G90015856A24
24G90015857A24
24G90015858A24
24G90015859A24
24G90015860A24
24G90015861A24
24G90015862A24
24G90015863A24
24G90015864A24
24G90015865A24
24G90015866A24
24G90015867A24
24G90015870A24
24G90015871A24
24G90015872A24
24G90015875A24
24G90015876A24
24G90015877A24
24G90015879A24
24G90015880A24
24G90015881A24
24G90015904A24
24G90015905A24
24G90015908A24
24G90015912A24
24G90015915A24
24G90015922A24
24G90015926A24
24G90015928A24
24G90015930A24
24G90015943A24
24G90015944A24
24G90015948A24
24G90016000A24
24G90016001A24
24G90016002A24
24G90016003A24
24G90016004A24
24G90016005A24
24G90016006A24
24G90016017A24
24G90016019A24
24G90016020A24
24G90016022A24
24G90016023A24
24G90016024A24
24G90016025A24
24G90016026A24
24G90016027A24
24G90016028A24
24G90016029A24
24G90016031A24
24G90016032A24
24G90016033A24
24G90016034A24
24G90016036A24
24G90016037A24
24G90016038A24
24G90016039A24
24G90016040A24
24G90016041A24
24G90016042A24
24G90016044A24
24G90016046A24
24G90016068A24
24G90016069A24
24G90016070A24
24G90016072A24
24G90016075A24
24G90016076A24
24G90016077A24
24G90016078A24
24G90016080A24
24G90016082A24
24G90016084A24
24G90016086A24
24G90016087A24
24G90016579A24
24G90016580A24
24G90016583A24
24G90016584A24
24G90016585A24
24G90016679A24
24G90016681A24
24G90016683A24
24G90016684A24
24G90016685A24
24G90016686A24
24G90016687A24
24G90016689A24
24G90016692A24
24G90016694A24
24G90016695A24
24G90016696A24
24G90016699A24
24G90016702A24
24G90016703A24
24G90016704A24
24G90016709A24
24G90016710A24
24G90016712A24
24G90016716A24
24G90016718A24
24G90016720A24
24G90016721A24
24G90016722A24
24G90016724A24
24G90016726A24
24G90016728A24
24G90016758A24
24G90016760A24
24G90016761A24
24G90016762A24
24G90016763A24
24G90016764A24
24G90016765A24
24G90016766A24
24G90016767A24
24G90016768A24
24G90016769A24
24G90016770A24
24G90016771A24
24G90016773A24
24G90016774A24
24G90016775A24
24G90016776A24
24G90016777A24
24G90016778A24
24G90016780A24
24G90016781A24
24G90016782A24
24G90017011B24
24G90017012B24
24G90017013B24
24G90017014B24
24G90017015B24
24G90017017B24
24G90017018B24
24G90017019B24
24G90017020B24
24G90017021B24
24G90017022B24
24G90017024B24
24G90017025B24
24G90017026B24
24G90017027B24
24G90017028B24
24G90017029B24
24G90017030B24
24G90017031B24
24G90017032B24
24G90017033B24
24G90017034B24
24G90017035B24
24G90017036B24
24G90017037B24
24G90017038B24
24G90017039B24
24G90017040B24
24G90017041B24
24G90017042B24
24G90017043B24
24G90017044B24
24G90017045B24
24G90017046B24
24G90017047B24
24G90017048B24
24G90017050B24
24G90017051B24
24G90017052B24
24G90017053B24
24G90017055B24
24G90017056B24
24G90017057B24
24G90017058B24
24G90017059B24
24G90017060B24
24G90017064B24
24G90017068B24
24G90017072B24
24G90017073B24
24G90017074B24
24G90017075B24
24G90017076B24
24G90017077B24
24G90017078B24
24G90017079B24
24G90017080B24
24G90017081B24
24G90017082B24
24G90017083B24
24G90017084B24
24G90017085B24
24G90017087B24
24G90017090B24
24G90017091B24
24G90017094B24
24G90017095B24
24G90017097B24
24G90017098B24
24G90017099B24
24G90017100B24
24G90017101B24
24G90017102B24
24G90017103B24
24G90017104B24
24G90017105B24
24G90017107B24
24G90017108B24
24G90017109B24
24G90017110B24
24G90017233B24
24G90017236B24
24G90017237B24
24G90017252B24
24G90017264B24
24G90017265B24
24G90017268B24
24G90017279B24
24G90017280B24
24G90017448B24
24G90017481B24
24G90017484B24
24G90017495B24
24G90017496B24
24G90017499B24
24G90017500B24
24G90017503B24
24G90017504B24
24G90017505B24
24G90017506B24
24G90017507B24
24G90017508B24
24G90017509B24
24G90017510B24
24G90017511B24
24G90017512B24
24G90017513B24
24G90017514B24
24G90017515B24
24G90017516B24
24G90017518B24
24G90017520B24
24G90017521B24
24G90017522B24
24G90017523B24
24G90017524B24
24G90017525B24
24G90017526B24
24G90017527B24
24G90017528B24
24G90017529B24
24G90017530B24
24G90017531B24
24G90017532B24
24G90017534B24
24G90017535B24
24G90017536B24
24G90017537B24
24G90017538B24
24G90017539B24
24G90017540B24
24G90017541B24
24G90017542B24
24G90017543B24
24G90017544B24
24G90017545B24
24G90017546B24
24G90017547B24
24G90017548B24
24G90017549B24
24G90017550B24
24G90017551B24
24G90017552B24
24G90017553B24
24G90017554B24
24G90017555B24
24G90017556B24
24G90017557B24
24G90017558B24
24G90017559B24
24G90017560B24
24G90017561B24
24G90017562B24
24G90017563B24
24G90017564B24
24G90017565B24
24G90017566B24
24G90017567B24
24G90017568B24
24G90017569B24
24G90017571B24
24G90017572B24
24G90017573B24
24G90017574B24
24G90017575B24
24G90017576B24
24G90017577B24
24G90017578B24
24G90017579B24
24G90017580B24
24G90017581B24
24G90017582B24
24G90017583B24
24G90017584B24
24G90017585B24
24G90017586B24
24G90017587B24
24G90017588B24
24G90017589B24
24G90017594B24
24G90017595B24
24G90017598B24
24G90017603B24
24G90017604B24
24G90017605B24
24G90017606B24
24G90017607B24
24G90017608B24
24G90017609B24
24G90017610B24
24G90017611B24
24G90017612B24
24G90017613B24
24G90017616B24
24G90017618B24
24G90017619B24
24G90017620B24
24G90017621B24
24G90017623B24
24G90017624B24
24G90017628B24
24G90017629B24
24G90017630B24
24G90017631B24
24G90017632B24
24G90017633B24
24G90017634B24
24G90017635B24
24G90017636B24
24G90017637B24
24G90017638B24
24G90017639B24
24G90017640B24
24G90017641B24
24G90017642B24
24G90017643B24
24G90017644B24
24G90017645B24
24G90017646B24
24G90017647B24
24G90017648B24
24G90017649B24
24G90017650B24
24G90017651B24
24G90001536D24
24G90001537D24
24G90001538D24
24G90001542D24
24G90001543D24
24G90001544D24
24G90001547D24
24G90001548D24
24G90001551D24
24G90001555D24
24G90001556D24
24G90001557D24
24G90001559D24
24G90001560D24
24G90001561D24
24G90001562D24
24G90001565D24
24G90001655D24
24G90001671D24
24G90001672D24
24G90001673D24
24G90001674D24
24G90001675D24
24G90001676D24
24G90001677D24
24G90001678D24
24G90001679D24
24G90001681D24
24G90001682D24
24G90001683D24
24G90001684D24
24G90001686D24
24G90001687D24
24G90001883D24
24G90001884D24
24G90001885D24
24G90001886D24
24G90001887D24
24G90001888D24
24G90001891D24
24G90006836D24
24G90006837D24
24G90006838D24
24G90006839D24
24G90006847D24
24G90006848D24
24G90006854D24
24G90007698D24
24G90007701D24
24G90007704D24
24G90007706D24
24G90008491D24
24G90008493D24
24G90008494D24
24G90008495D24
24G90008496D24
24G90010344D24
24G90010345D24
24G90010346D24
24G90010347D24
24G90010348D24
24G90010349D24
24G90010350D24
24G90010351D24
24G90010488H24
24G90011426H24
24G90011427H24
24G90011428H24
24G90011429H24
24G90011430H24
24G90011431H24
24G90012868H24
24G90012870H24
24G90012871H24
24G90012874H24
24G90012875H24
24G90012887H24
24G90012892H24
24G90012893H24
24G90012896H24
24G90012900H24
24G90012902H24
24G90012907H24
24G90012908H24
24G90012909H24
24G90012913H24
24G90012923H24
24G90012924H24
24G90012926H24
24G90012928H24
24G90012944H24
24G90012945H24
24G90012955H24
24G90012956H24
24G90012957H24
24G90013192H24
24G90013193H24
24G90013195H24
24G90013197H24
24G90013199H24
24G90013201H24
24G90013221H24
24G90013223H24
24G90013224H24
24G90013378H24
24G90013381H24
24G90013382H24
24G90013383H24
24G90013384H24
24G90013385H24
24G90013386H24
24G90013387H24
24G90013388H24
24G90013389H24
24G90013390H24
24G90013391H24
24G90013392H24
24G90013395H24
24G90013396H24
24G90013397H24
24G90013399H24
24G90013400H24
24G90013401H24
24G90013402H24
24G90013403H24
24G90013404H24
24G90013612H24
24G90013617H24
24G90013620H24
24G90013622H24
24G90013623H24
24G90013625H24
24G90013626H24
24G90013627H24
24G90013628H24
24G90013629H24
24G90013630H24
24G90013631H24
24G90013632H24
24G90013633H24
24G90013634H24
24G90013635H24
24G90013636H24
24G90013637H24
24G90013648H24
24G90013650H24
24G90013652H24
24G90013655H24
24G90013656H24
24G90013661H24
24G90013665H24
24G90013672H24
24G90013678H24
24G90017726C24
24G90017996H24
24G90017999H24
24G90018000H24
24G90018002H24
24G90018009H24
24G90018017H24
24G90018018H24
24G90018021H24
24G90018024H24
24G90018025H24
24G90018028H24
24G90018030H24
24G90018031H24
24G90018032H24
24G90018033H24
24G90018034H24
24G90012942H24
24G90012949H24
24G90013068H24
24G90013069H24
24G90013071H24
24G90018086H24
24G90018089H24
24G90018106H24
24G90018108H24
24G90018109H24
24G90007690C24
24G90008245D24
24G90008246D24
24G90008250D24
24G90008251D24
24G90008256D24
24G90008276D24
24G90008301D24
24G90008302D24
24G90008303D24
24G90008304D24
24G90008306D24
24G90008307D24
24G90008310D24
24G90008311D24
24G90008313D24
24G90008314D24
24G90008316D24
24G90008318D24
24G90008337D24
24G90008497D24
24G90008499D24
24G90009383D24
24G90009387D24
24G90009426D24
24G90009430D24
24G90009431D24
24G90009432D24
24G90009433D24
24G90009434D24
24G90009435D24
24G90009436D24
24G90009437D24
24G90009438D24
24G90009439D24
24G90009440D24
24G90009441D24
24G90009442D24
24G90009443D24
24G90009444D24
24G90009445D24
24G90009446D24
24G90009447D24
24G90009448D24
24G90009449D24
24G90009450D24
24G90009451D24
24G90009452D24
24G90009453D24
24G90009454D24
24G90009455D24
24G90009456D24
24G90009457D24
24G90009458D24
24G90009460D24
24G90009461D24
24G90009462D24
24G90009463D24
24G90009464D24
24G90009465D24
24G90009466D24
24G90009467D24
24G90009468D24
24G90009471D24
24G90009472D24
24G90009473D24
24G90009474D24
24G90009475D24
24G90009476D24
24G90009477D24
24G90009478D24
24G90009479D24
24G90009480D24
24G90009481D24
24G90009482D24
24G90009483D24
24G90009484D24
24G90009544D24
24G90009550D24
24G90009571D24
24G90009572D24
24G90009573D24
24G90009574D24
24G90009575D24
24G90009576D24
24G90009577D24
24G90009578D24
24G90009579D24
24G90009580D24
24G90009581D24
24G90009582D24
24G90009583D24
24G90009584D24
24G90012735D24
24G90012744D24
24G90012747D24
24G90017736C24
23G90003968A24
23G90009117A24
23G90009633A24
23G90103211A24
23G90112713A24
24G90001825F24
24G90001830F24
24G90004244F24
24G90004246F24
24G90004247F24
24G90004248F24
24G90007430F24
24G90007432F24
24G90007490F24
24G90007494F24
24G90007499F24
24G90007500F24
24G90016007A24
24G90016009A24
24G90016010A24
24G90016615A24
24G90001822F24
24G90001823F24
24G90001824F24
24G90001826F24
24G90001827F24
24G90001828F24
24G90001829F24
24G90001831F24
24G90001833F24
24G90001834F24
24G90004242F24
24G90004243F24
24G90004245F24
24G90004249F24
24G90007378F24
24G90007419F24
24G90007422F24
24G90007423F24
24G90007424F24
24G90007425F24
24G90007426F24
24G90007427F24
24G90007428F24
24G90007429F24
24G90007431F24
24G90007433F24
24G90007493F24
24G90007495F24
24G90007496F24
24G90007497F24
24G90009499F24
24G90009500F24
24G90010355F24
24G90001608D24
|
| FEI Number |
3004123209
|
Recalling Firm/ Manufacturer |
HeartSine Technologies Ltd
207 Airport Road West
Belfast
United Kingdom
|
| For Additional Information Contact | Joyce Staggs 425-867-4597 |
Manufacturer Reason for Recall | Due to a damaged (over stretched) membrane tail that causes an intermittent fault. |
FDA Determined Cause 2 | Process control |
| Action | On 07/28/2026, several versions of an "Urgent: Medical device recall" letter that outlines multiple issues:
1) Stryker identified an issue with the membrane tail, an internal component, which could cause an intermittent fault. This failure could occur at any point when the device is holding a charge in preparation to deliver therapy, while delivering a shock, or after shock delivery. The device becomes inoperable after the failure occurs.
2) Stryker has also identified that a limited number of HeartSine samaritan PAD devices were distributed without the required maintenance insert. This insert contains periodic maintenance instructions implemented as an additional risk control measure. While most devices contain the maintenance insert, this voluntary correction is being initiated to ensure all end users have access to the supplemental maintenance information. This issue was identified through a customer complaint.
For the membrane tail issue (issue 1), Customers are instructed to:
|
| Quantity in Commerce | 11,912 |
| Distribution | Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AZ, CA, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME MI, MN, NC, NJ, NM, NV, NY, OH, OK, OR, PA, SD,TN, TX, UT, VA, WA, WI, WV and WY. The countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Hong Kong, Indonesia, ISRAEL, JAPAN, Malaysia, Netherlands, Philippines, POLAND, SINGAPORE, South Korea, Taiwan, Thailand and UNITED KINGDOM. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|