| | Class 2 Device Recall RAPICIDE PA Part A HighLevel Disinfectant |  |
| Date Initiated by Firm | September 21, 2026 |
| Date Posted | October 14, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0155-2027 |
| Recall Event ID |
99804 |
| 510(K)Number | K082988 |
| Product Classification |
Accessories, cleaning, for endoscope - Product Code FEB
|
| Product | RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020116 |
| Code Information |
UDI: 00724995245955
Lot numbers with expiration dates:
689171 5/4/2027
689981 5/12/2027
690993 5/26/2027
691739 6/2/2027
692570 6/11/2027
694357 7/1/2027
|
| FEI Number |
1527821
|
Recalling Firm/ Manufacturer |
Steris Corporation
5960 Heisley Rd
Mentor
OH
44060-1834
|
| For Additional Information Contact | Michelle LaVan 440-2597417 |
Manufacturer Reason for Recall | If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity. |
FDA Determined Cause 2 | Process change control |
| Action | An URGENT MEDICAL DEVICE RECALL notification letter dated 9/8/26 was sent to customers.
User Action Users can continue using the RAPICIDE product through the remainder of its shelf life. Please immediately inspect your on-hand inventory for product affected by this recall. If bottle bulging is observed, follow the instructions below to release any accumulated pressure and dispose of the product:
1. Users MUST wear personal protective equipment (PPE) before handling the bottle, including:
a. Safety eyewear
b. Chemical-resistant gloves
c. A solvent-resistant apron
d. Protective footwear appropriate for chemical handling
2. Move the product to an appropriate, well-ventilated area. Respiratory protection is also recommended in accordance with the RAPICIDE PA Part A Safety Data Sheet (SDS). Refer to Attachment C to this letter for an excerpt of the SDS.
3. Carefully loosen the cap. Slowly begin untwisting the cap while keeping the threads engaged to allow any accumulated pressure to release gradually.
4. Allow the pressure to vent. Continue loosening the cap slowly over approximately 1 to 2 minutes until the pressure has been released and the cap can be fully removed.
5. Dispose of the bottle and contents through your facility's hazardous or special waste program in accordance with applicable local, regional, national, and/or international regulations. Do not replace the cap on a bottle that has exhibited bulging.
6. Please complete the Recall Response Form included with this letter, and return via email to Regulatory_Compliance@STERIS.com, or via fax to 440-392-8963. Replacement product will be provided for any product that had to be disposed.
Users are also reminded to report any adverse events via FDA s MedWatch Database or local competent authority should an event occur. We apologize for any inconvenience this matter may cause, and as always, STERIS is dedicated to supporting our products and valued Customers. If you have questions regarding this matter, please co |
| Quantity in Commerce | 11806 units |
| Distribution | Worldwide - US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = FEB
|
|
|
|