| | Class 2 Device Recall GORE VIABAHN Endoprosthesis with Heparin Bioactive Surface, GORE VIABAHN Endoprosthesis with PRO |  |
| Date Initiated by Firm | August 28, 2026 |
| Date Posted | November 06, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0360-2027 |
| Recall Event ID |
99797 |
| PMA Number | P040037 P130006 |
| Product Classification |
Stent, superficial femoral artery - Product Code NIP
|
| Product | Brand Name: GORE VIABAHN Endoprosthesis with Heparin Bioactive Surface, GORE VIABAHN Endoprosthesis with PROPATEN Bioactive Surface
Product Name: GORE VIABAHN Endoprosthesis with Heparin Bioactive Surface, GORE VIABAHN Endoprosthesis with PROPATEN Bioactive Surface
Model/Catalog Number: VBJR081002A, VBJR081502A, VBJR060702A, VBJR071002A, PAJR081002B, VBJR080702A, VBJR072502A, VBJR082502A, PAHR131002B, PAJR071002B, VBJR080502A, VBJR061502A, VBJR051502A, VBJR062502A, VBHR111002A, VBJR052502A, VBJR051002A, VBHR110502A,
VBHR131002A, VBHR131002H, VBHR110502W, VBHR131002W, JHJR060702J, JHJR071002J, PAJR081002E, PAHR131002E, PAJR061502E, PAJR070502E, PAJR071002E, PAJR062502E,
PAJR061002E, PAJR051502E, PAJR070702E, PAHR110502E, PAHR130502E
Reference Attachment 6 for an explanation of the part numbers.
Product Description: The GORE VIABAHN Endoprosthesis with Heparin/PROPATEN Bioactive Surface is intended for use as an endovascular prosthesis. The endoprosthesis functions as a new blood conduit, with inlet and outlet to the endoprosthesis for blood flow. The self-expanding stent frame serves to maintain an open vessel lumen while the expanded polytetrafluoroethylene (ePTFE) graft material serves as a blood flow conduit. Surface bound heparin on the device is used as an anticoagulant to provide thromboresistance. The GORE VIABAHN Endoprosthesis with Heparin/PROPATEN Bioactive Surface is a flexible, self-expanding endoluminal endoprosthesis consisting of an ePTFE lining with an external nitinol (NiTi = Nickel:Titanium) support extending along its entire length. Four radiopaque markers are present on each end of the endoprosthesis to facilitate accurate placement and improve visibility of the device post-deployment. The CBAS Heparin Surface on the GORE VIABAHN Endoprosthesis with Heparin/PROPATEN Bioactive Surface consists of stable, covalent, end-point attached heparin of porcine origin. The endoprosthesis is compressed and attached to a dual lumen delivery catheter. The larger central catheter lumen is used for flushing and guidewire introduction. The smaller lumen contains elements of the deployment mechanism. The delivery catheter hub assembly has one port for the deployment system and one port for flushing and guidewire insertion. To facilitate accurate endoprosthesis placement, two radiopaque metallic bands are attached to the catheter shaft, marking the ends of the compressed endoprosthesis.
Component: N/A |
| Code Information |
Model Number / UDI-DI code: JHJR060702J 04993024010482
Lot Numbers:
31674434
31674435
31674436
Model Number / UDI-DI code: JHJR071002J 04993024010239
Lot Numbers:
31750021
31724571
Model Number / UDI-DI code: PAHR110502E 00733132650118
Lot Numbers:
31822842
31822830
31822838
31846601
31822836
31846583
31846582
31822843
31822819
31822817
31822831
31846595
31846596
31846592
31822839
31822824
31822825
31822840
31822841
31846591
31846593
31822844
31846585
31846586
31846587
31846600
31846590
31822833
31846597
31822822
31822816
31822826
31822821
31822835
31846589
31846599
31822818
31822820
31846598
31822834
31846594
31822832
31846588
31846584
Model Number / UDI-DI code: PAHR130502E 00733132650132
Lot Numbers:
31832526
31832532
31832529
31832527
31832525
31832528
31832530
Model Number / UDI-DI code: PAHR131002B 00733132650248
Lot Numbers:
31849227
31720620
Model Number / UDI-DI code: PAHR131002E 00733132650149
Lot Numbers:
31822989
31822990
31822986
31704734
31822965
31704735
31822988
31822963
31704732
31822993
31822987
31822967
31822964
31822991
31822992
31822966
31822968
Model Number / UDI-DI code: PAJR051502E 00733132623846
Lot Numbers:
31775425
31775427
31775426
31775428
Model Number / UDI-DI code: PAJR061002E 00733132623983
Lot Numbers:
31761002
31760999
31760997
31760998
31761000
31761003
31761007
31761008
31761001
31760996
31761006
Model Number / UDI-DI code: PAJR061502E 00733132623990
Lot Numbers:
31723358
31723357
31723360
Model Number / UDI-DI code: PAJR062502E 00733132624003
Lot Numbers:
31749884
31775388
31775385
31775386
31775387
Model Number / UDI-DI code: PAJR070502E 00733132627431
Lot Numbers:
31726726
31726709
31726731
31726736
31726700
31726735
31726705
31726708
31726729
31726704
31726732
31726706
31726727
31726728
31726730
31726707
31726698
31726701
31726702
31726734
31726703
31726699
Model Number / UDI-DI code: PAJR070702E 00733132627844
Lot Numbers:
31790107
31790108
31790105
31790106
Model Number / UDI-DI code: PAJR071002B 00733132639618
Lot Numbers:
31723720
31723721
31723718
31723719
Model Number / UDI-DI code: PAJR071002E 00733132627448
Lot Numbers:
31733204
31733211
31733207
31733212
31733213
31733203
31733208
31733206
31733210
31733214
31733209
31733205
Model Number / UDI-DI code: PAJR081002B 00733132639663
Lot Numbers:
31704840
31714266
31704842
31714265
31714264
31714268
31704844
31704841
31704843
31714267
Model Number / UDI-DI code: PAJR081002E 00733132627493
Lot Numbers:
31735027
31658986
31658987
31658985
31735029
31607103
31607104
31607105
31607106
31735026
31735024
Model Number / UDI-DI code: VBHR110502A 00733132646081
Lot Numbers:
31831734
31831738
31831735
31820787
31831730
31831739
Model Number / UDI-DI code: VBHR110502W 00733132651139
Lot Numbers:
31823157
31846809
31823158
31823160
31823161
31846811
31823156
31823159
31846810
31823162
31823163
Model Number / UDI-DI code: VBHR111002A 00733132646104
Lot Numbers:
31831566
31766110
31752926
31766113
31766108
31831567
31831569
31766115
31831564
31831565
Model Number / UDI-DI code: VBHR131002A 00733132646142
Lot Numbers:
31844911
31844913
31844914
31844916
31844922
31844917
31820789
31844919
31844912
31820790
31831663
31831664
Model Number / UDI-DI code: VBHR131002H 00733132651955
Lot Number: 31828735
Model Number / UDI-DI code: VBHR131002W 00733132644797
Lot Numbers:
31828718
31828717
31828719
31828720
Model Number / UDI-DI code: VBJR051002A 00733132623914
Lot Numbers:
31802822
31802821
31802827
31802820
31802823
31802826
31802824
31802818
Model Number / UDI-DI code: VBJR051502A 00733132623921
Lot Numbers:
31742391
31742392
31742393
31742394
31742390
31742395
Model Number / UDI-DI code: VBJR052502A 00733132623938
Lot Numbers:
31753593
31753596
31753594
31753595
Model Number / UDI-DI code: VBJR060702A 00733132624096
Lot Numbers:
31747775
31682473
31682479
31747776
31682475
31682476
31682480
31726170
31682474
31726167
31682477
31747777
31682478
31726166
31747774
31726169
31726171
Model Number / UDI-DI code: VBJR061502A 00733132623747
Lot Numbers:
31729956
31729957
31815491
31729946
31729954
31729960
31729955
31729950
31729951
31729958
31729952
31815492
31729948
31729959
31729945
31732530
31729947
31732531
31732532
31729949
Model Number / UDI-DI code: VBJR062502A 00733132623976
Lot Numbers:
31766195
31747773
31766191
31766192
31766225
31766220
31747756
31747762
31766218
31766226
31766197
31747757
31747755
31766223
31766190
31747764
31747758
31747767
31747765
31766202
31766203
31747759
31747763
31766221
31747761
31747772
31766219
31766193
31766224
31747771
31747766
31766198
31747754
31747768
31747769
31747770
Model Number / UDI-DI code: VBJR071002A 00733132623778
Lot Numbers:
31732851
31687507
31732846
31718036
31732852
31732847
31687503
31687504
31687501
31732848
31687510
31732849
31687502
31687505
31732835
31687506
31687509
31732854
31687508
31687511
Model Number / UDI-DI code: VBJR072502A 00733132623792
Lot Numbers:
31713035
31713041
31713040
31713042
31713036
31725945
31725938
31725939
31757523
31725942
31725940
31713037
31713044
31725941
31757522
31713039
Model Number / UDI-DI code: VBJR080502A 00733132623815
Lot Numbers:
31725819
31725820
Model Number / UDI-DI code: VBJR080702A 00733132624119
Lot Numbers:
31706824
31718137
31808025
31706828
31706825
31718128
31808028
31706829
31747554
31706830
31808027
31718119
31718121
31718136
31808026
31718131
31808023
31706832
31808024
31718132
31706831
31718127
31732552
31718135
31718133
31718123
31718125
31718129
31706852
31706853
31718130
Model Number / UDI-DI code: VBJR081002A 00733132623822
Lot Numbers:
31657868
31726064
31701883
31726060
31615834
31807953
31701882
31687933
31726032
31615830
31726066
31668958
31726061
31615829
31747616
31657869
31657863
31615832
31807954
31687915
31687916
31726033
31687912
31687913
31807956
31747614
31657859
31747618
31701887
31726063
31687914
31668959
31747617
31687917
31726065
31701885
31687910
31807955
31657862
31807957
31615833
31657870
31657861
31687934
31657867
31668960
31687911
31657860
31726067
31615831
31701886
31701884
Model Number / UDI-DI code: VBJR081502A 00733132623839
Lot Numbers:
31706875
31717918
31706872
31706877
31652771
31706874
Model Number / UDI-DI code: VBJR082502A 00733132623693
Lot Number: 31717732
Lot Code: Reference Attachment 7
Note to "Product Code" above - secondary product code of FPV, in addition to NPI
Note to "Expected Life" below - The selected device lifetime used by Gore is limited to 1-10 years (depending on the indication) to meet regulatory requirements due to factors such as: " differing regulatory guidance and standards " post market clinical study follow-up time periods " disease progression " duration of clinical use " primary function of the endoprosthesis and endoprosthesis characteristics that deliver the function " indication/product literature assessment " mechanical integrity. The selected device lifetime as used by Gore to meet regulatory requirements is not meant to be a recommendation or indication for reintervention or removal of the endoprosthesis except as otherwise determined by the physician and/or per the applicable labeling. The selected device lifetime is not meant to be a warranty or guarantee as the actual duration of clinical use or implantation of the endoprosthesis for a particular patient may be shorter or longer than the selected device lifetime based on the patient s medical conditions, anatomy, or other medical treatment as well as procedure and clinical factors and decisions.
|
| FEI Number |
2017233
|
Recalling Firm/ Manufacturer |
W L Gore & Associates, Inc.
1505 N 4th St
Flagstaff
AZ
86004-5867
|
| For Additional Information Contact | 800-5288763 |
Manufacturer Reason for Recall | Endovascular prosthesis were manufactured on equipment that was not operating with valid parameters and may cause the distal tip to separate. |
FDA Determined Cause 2 | Process change control |
| Action | On 09/03/2026, an "URGENT MEDICAL DEVICE RECALL/FIELD SAFETY NOTICE" was sent via UPS 2nd Day Air by a Third-Party Service (Sedgwick) to U.S. and Candian customers informing them that, Gore has identified through quality control testing that some VIABAHN devices were processed using equipment that was functioning outside of the validated range. This could potentially result in an incorrect bonding location of the distal tip onto the delivery catheter.
To Healthcare providers:
For patients already implanted with an affected device, Gore does not recommend prophylactic removal, reintervention, or additional monitoring solely as a result of this recall as the risk of surgical removal or reintervention are expected to outweigh the theoretical risks associated with this issue. Patient management should continue according to routine clinical care and the treating physician s judgment.
Instructions for customers:
1. Discontinue the use of listed device(s).
2. Identify and return any unused devices within the scope of this recall for replacement.
3. Complete and sign the enclosed CUSTOMER RESPONSE FORM and return to
WLGore6907@sedgwick.com within 2 weeks of receipt of this notification.
4. This notice needs to be shared with those who need to be aware within your institution or to any organization where potentially affected devices have been transferred (as appropriate). Transfer this notice to other organization(s) on which this action has an impact (as appropriate).
5. If a listed device has been implanted, indicate the used device(s) on the CUSTOMER RESPONSE FORM and return to WLGore6907@sedgwick.com. |
| Quantity in Commerce | 424 |
| Distribution | Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Argentina, Austria, Belgium, Canada, France, Germany, Israel, Italy, Japan, Netherlands, Poland, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, and United Kingdom. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| PMA Database | PMAs with Product Code = NIP
|
|
|
|