Class 2 Device Recall SIGNA Artist and SIGNA Artist Evo

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall SIGNA Artist and SIGNA Artist Evosee related information
Date Initiated by FirmAugust 27, 2026
Date PostedNovember 13, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0448-2027
Recall Event ID 99911
510(K)NumberK202238 K213603 
Product Classification System, nuclear magnetic resonance imaging - Product Code LNH
ProductSIGNA Artist SIGNA Artist Evo
Code Information SIGMA Artist Product - 1.5T SIGNA ARTIST EVO UPGRADE Serial Number / System ID / UDI code: IPGEV2400140SC 630833MRI (01)00195278117021(11)241000(21)IPGEV2400140SC IPGEV2300012SC 937641MR (01)00195278117021(11)230300(21)IPGEV2300012SC IPGEV2300013SC HPAD01MR01 (01)00195278117021(11)230300(21)IPEV2300013SC IPGEV2400170SC MR00290 (01)00195278117021(11)241200(21)IPGEV2400170SC Product - 1.5T SIGNA ARTIST FLO IPM Serial Number / System ID / UDI code: IPGAR2500009SC 13829180 (01)00195278117021(11)250300(21)IPGAR2500009SC IPGEV2400101SC 019022MR01 (01)00195278117021(11)240700(21)IPGEV2400101SC IPGEV2400098SC 019022MR07 (01)00195278117021(11)240700(21)IPGEV2400098SC PGRART2500008 0853270243 (01)00195278117021(11)250801(21)PGRART2500008 IPGEV2400174SC 13051973 (01)00195278117021(11)250100(21)IPGEV2400174SC PGRART2500014 205801TKCMR1 (01)00195278117021(11)250821(21)PGRART2500014 IPGAR2500017SC 208932MR1 (01)00195278117021(11)250500(21)IPGAR2500017SC IPGAR2500003SC 216476MR (01)00195278117021(11)250200(21)IPGAR2500003SC IPGAR2500026SC 314RAYUSMR (01)00195278117021(11)250611(21)IPGAR2500026SC IPGAR2500027SC 336527MR1 (01)00195278117021(11)250627(21)IPGAR2500027SC PGRART2500024 352333ERMRI (01)00195278117021(11)250908(21)PGRART2500024 IPGEV2400155SC 352401MR2 (01)00195278117021(11)241100(21)IPGEV2400155SC IPGEV2400159SC 405949IBMR2 (01)00195278117021(11)241200(21)IPGEV2400159SC PGRART2500011 423VAMRI (01)00195278117021(11)250806(21)PGRART2500011 PGRART2500002 480323SMCMR2 (01)00195278117021(11)250630(21)PGRART2500002 IPGEV2400164SC 502875ARTIST (01)00195278117021(11)241200(21)IPGEV2400164SC PGRART2500005 509765ART (01)00195278117021(11)250701(21)PGRART2500005 IPGAR2500012SC 516231MR (01)00195278117021(11)250400(21)IPGAR2500012SC PGRART2500054 539355MR1 (01)00195278117021(11)251119(21)PGRART2500054 IPGAR2500022SC 580223AMR2 (01)00195278117021(11)250528(21)IPGAR2500022SC PGRART2500009 605504MRART (01)00195278117021(11)250801(21)PGRART2500009 IPGAR2500008SC 606408MRW (01)00195278117021(11)250300(21)IPGAR2500008SC IPGEV2400119SC 608EMCMR3 (01)00195278117021(11)240800(21)IPGEV2400119SC IPGAR2500001SC 614DRMCMR (01)00195278117021(11)250200(21)IPGAR2500001SC IPGAR2500023SC 701775GEMR1 (01)00195278117021(11)250528(21)IPGAR2500023SC IPGAR2500021SC 703504MR1 (01)00195278117021(11)250500(21)IPGAR2500021SC IPGEV2400160SC 713572MR1 (01)00195278117021(11)241200(21)IPGEV2400160SC IPGAR2500193SC 717738MR (01)00195278117021(11)250200(21)IPGAR2500193SC IPGAR2500025SC 718780MR1 (01)00195278117021(11)250609(21)IPGAR2500025SC IPGAR2500019SC 770400MR (01)00195278117021(11)250500(21)IPGAR2500019SC IPGEV2400172SC 773270MR (01)00195278117021(11)250100(21)IPGEV2400172SC IPGAR2500002SC 80428915T (01)00195278117021(11)250200(21)IPGAR2500002SC IPGAR2500016SC 813685MR3 (01)00195278117021(11)250500(21)IPGAR2500016SC IPGEV2400135SC 813MCCMR1 (01)00195278117021(11)241000(21)IPGEV2400135SC IPGAR2500004SC 815968MR2 (01)00195278117021(11)250200(21)IPGAR2500004SC IPGAR2500011SC 832826MR6 (01)00195278117021(11)250300(21)IPGAR2500011SC IPGAR2500010SC 832912MR2 (01)00195278117021(11)250300(21)IPGAR2500010SC PGRART2500006 859MR000011 (01)00195278117021(11)250718(21)PGRART2500006 IPGEV2500179SC 860889MRART (01)00195278117021(11)250100(21)IPGEV2500179SC IPGAR2500020SC 901287ART2 (01)00195278117021(11)250500(21)IPGAR2500020SC IPGAR2500007SC 904627MR4 (01)00195278117021(11)250200(21)IPGAR2500007SC IPGEV2400151SC 904639MR2 (01)00195278117021(11)241100(21)IPGEV2400151SC IPGEV2400115SC 913588WPMR (01)00195278117021(11)240900(21)IPGEV2400115SC IPGEV2400167SC 913676LCCMR (01)00195278117021(11)241200(21)IPGEV2400167SC IPGAR2500024SC 916784MR1 (01)00195278117021(11)250528(21)IPGAR2500024SC PGRART2500019 916SUTARDENMR (01)00195278117021(11)250825(21)PGRART2500019 PGRART2500010 E416249MR1ART (01)00195278117021(11)250806(21)PGRART2500010 PGRART2500007 LVIPCMR (01)00195278117021(11)250718(21)PGRART2500007 IPGAR2500014SC LVITPMR (01)00195278117021(11)250400(21)IPGAR2500014SC PGRART2500018 MCCLOLMR1 (01)00195278117021(11)250825(21)PGRART2500018 IPGEV2400171SC MR00291 (01)00195278117021(11)250100(21)IPGEV2400171SC IPGEV2400152SC MR00296 (01)00195278117021(11)241100(21)IPGEV2400152SC PGRART2500017 MR00316 (01)00195278117021(11)250825(21)PGRART2500017 IPGAR2500028SC MR1045981 (01)00195278117021(11)250627(21)IPGAR2500028SC IPGEV2500180SC MR54443751 (01)00195278117021(11)250100(21)IPGEV2500180SC IPGEV2400145SC MRPM1634 (01)00195278117021(11)241000(21)IPGEV2400145SC PGRART2500003 MRPM1958 (01)00195278117021(11)250701(21)PGRART2500003 PGRART2500012 MRPM1991 (01)00195278117021(11)250812(21)PGRART2500012 IPGAR2500006SC NYPTO15TMR1 (01)00195278117021(11)250200(21)IPGAR2500006SC IPGAR2500013SC SHMBLMR (01)00195278117021(11)250400(21)IPGAR2500013SC PGRART2500016 VNMRI26563342 (01)00195278117021(11)250825(21)PGRART2500016 Product - 1-5T SIGNA ARTIST FLO IPM Serial Number / System ID / UDI code: IPGEV2400139SC Not Available (01)00195278117021(11)241000(21)IPGEV2400139SC IPGEV2400104SC MRPM1500 (01)00195278117021(11)240700(21)IPGEV2400104SC IPGEV2400109SC MRPM1458 (01)00195278117021(11)240700(21)IPGEV2400109SC IPGEV2400118SC MR5678PE (01)00195278117021(11)240800(21)IPGEV2400118SC PGRART2500013 0853270247 (01)00195278117021(11)250820(21)PGRART2500013 IPGEV2400163SC 0853270236 (01)00195278117021(11)241200(21)IPGEV2400163SC Product - 3.0T SIGNA PIONEER MR28 FLO Serial Number / System ID / UDI code: IPGAR2500018SC 508998MR (01)00195278117021(11)250500(21)IPGAR2500018SC Product - SIGNA Artist MDR IPGR with XG5, CTO Serial Number / System ID / UDI code: PGRART2500001 Not Available (01)00195278117021(11)250630(21)PGRART2500001 PGRART2500004 Not Available (01)00195278117021(11)250701(21)PGRART2500004 Product - UPG 1.5T LX UPGRADE Serial Number / System ID / UDI code: IPGAR2500015SC 209521SGMR2 (01)00195278117021(11)250400(21)IPGAR2500015SC SIGNA Artist Evo Product - 1.5T HDXT TO HD23 UPGRADE Serial Number / System ID / UDI code: IPGEV2400062SC 506452MR (01)00195278481382(11)240200(21)IPGEV2400062SC Product - 1.5T SIGNA ARTIST EVO CPO Serial Number / System ID / UDI code: IPGEV2500206SC 501552SVNMR (01)00195278481382(11)250400(21)IPGEV2500206SC IPGEVO2500011 203797MRCPO (01)00195278481382(11)250813(21)IPGEVO2500011 IPGEV2500212SC 402609EVO (01)00195278481382(11)250500(21)IPGEV2500212SC IPGEV2500196SC 402AMIMR15 (01)00195278481382(11)250200(21)IPGEV2500196SC IPGEV2500176SC 214420RDMR2 (01)00195278481382(11)250100(21)IPGEV2500176SC IPGEV2400082SC 262544MR2 (01)00195278481382(11)240500(21)IPGEV2400082SC Product - 1.5T SIGNA ARTIST EVO UPGRADE Serial Number / System ID / UDI code: IPGEV2400056SC 907796JMR (01)00195278481382(11)240100(21)IPGEV2400056SC IPGEV2400092SC 256ALOPMR (01)00195278481382(11)240600(21)IPGEV2400092SC IPGEV2400089SC 205934CVMR (01)00195278481382(11)240600(21)IPGEV2400089SC IPGEV2400169SC 256543MR (01)00195278481382(11)241200(21)IPGEV2400169SC IPGEV2500197SC 520381MR1 (01)00195278481382(11)250200(21)IPGEV2500197SC IPGEV2500189SC 623478MR1 (01)00195278481382(11)250200(21)IPGEV2500189SC IPGEV2200007SC 602406MR5 (01)00195278481382(11)221100(21)IPGEV2200007SC IPGEV2400173SC 602512MR3 (01)00195278481382(11)250100(21)IPGEV2400173SC IPGEVO2500006 916RASCMR (01)00195278481382(11)250718(21)IPGEVO2500006 IPGEV2500188SC 530894MR2 (01)00195278481382(11)250200(21)IPGEV2500188SC IPGEV2500215SC 707428FFMR (01)00195278481382(11)250500(21)IPGEV2500215SC IPGEV2400064SC UCDROCKMR (01)00195278481382(11)240200(21)IPGEV2400064SC IPGEV2400148SC 719587MR (01)00195278481382(11)241100(21)IPGEV2400148SC IPGEV2500185SC 303695MR (01)00195278481382(11)250100(21)IPGEV2500185SC IPGEV2400085SC 303938MR (01)00195278481382(11)240500(21)IPGEV2400085SC IPGEV2300010SC 970244AMPMR (01)00195278481382(11)230100(21)IPGEV2300010SC IPGEV2400141SC 970248MR (01)00195278481382(11)241000(21)IPGEV2400141SC IPGEV2300051SC 970RMOMR (01)00195278481382(11)231200(21)IPGEV2300051SC IPGEV2400121SC 203453MR1 (01)00195278481382(11)240900(21)IPGEV2400121SC IPGEV2400137SC 203407MREVO (01)00195278481382(11)241000(21)IPGEV2400137SC IPGEV2300020SC 860231MR (01)00195278481382(11)230600(21)IPGEV2300020SC IPGEV2400086SC 302744MR (01)00195278481382(11)240500(21)IPGEV2400086SC IPGEV2400088SC RADNETMR1725 (01)00195278481382(11)240600(21)IPGEV2400088SC IPGEV2400100SC 352333NFMR1 (01)00195278481382(11)240700(21)IPGEV2400100SC IPGEV2500198SC 904399MR2 (01)00195278481382(11)250200(21)IPGEV2500198SC IPGEV2400102SC 954978MR1 (01)00195278481382(11)240700(21)IPGEV2400102SC IPGEV2400127SC 850678MR (01)00195278481382(11)240900(21)IPGEV2400127SC IPGEV2300036SC 407303MRER (01)00195278481382(11)231000(21)IPGEV2300036SC IPGEV2400161SC 727521MRI (01)00195278481382(11)241200(21)IPGEV2400161SC IPGEV2500208SC 407CFRMR (01)00195278481382(11)250400(21)IPGEV2500208SC IPGEV2400166SC 954724MR1 (01)00195278481382(11)241200(21)IPGEV2400166SC IPGEV2500211SC 813751MR (01)00195278481382(11)250500(21)IPGEV2500211SC IPGEV2300046SC 561650MR2 (01)00195278481382(11)231100(21)IPGEV2300046SC IPGEV2300043SC 706651MR (01)00195278481382(11)231100(21)IPGEV2300043SC IPGEV2400077SC 706481MR (01)00195278481382(11)240400(21)IPGEV2400077SC IPGEV2500181SC 229276MR (01)00195278481382(11)250100(21)IPGEV2500181SC IPGEV2400162SC 770644WHMR (01)00195278481382(11)241200(21)IPGEV2400162SC IPGEV2300027SC 3012383 (01)00195278481382(11)230800(21)IPGEV2300027SC IPGEV2400150SC 1323566 (01)00195278481382(11)241100(21)IPGEV2400150SC IPGEV2500195SC 515247MR2 (01)00195278481382(11)250200(21)IPGEV2500195SC IPGEV2300042SC 208529MR2 (01)00195278481382(11)231100(21)IPGEV2300042SC IPGEV2400153SC 1000004646 (01)00195278481382(11)241100(21)IPGEV2400153SC IPGEV2400147SC 630275MR3 (01)00195278481382(11)241100(21)IPGEV2400147SC IPGEV2500225SC 1000004954 (01)00195278481382(11)250627(21)IPGEV2500225SC IPGEVO2500013 1000004074 (01)00195278481382(11)250825(21)IPGEVO2500013 IPGEV2300030SC 812376MR1 (01)00195278481382(11)230800(21)IPGEV2300030SC IPGEV2400065SC 574296MR (01)00195278481382(11)240200(21)IPGEV2400065SC IPGEVO2500018 260266MR (01)00195278481382(11)250825(21)IPGEVO2500018 IPGEV2400072SC 812482MR1 (01)00195278481382(11)240200(21)IPGEV2400072SC IPGEV2400059SC 913541OMR (01)00195278481382(11)240100(21)IPGEV2400059SC IPGEV2400081SC 270793HDE (01)00195278481382(11)240400(21)IPGEV2400081SC IPGEV2400112SC 502897BMR3 (01)00195278481382(11)240700(21)IPGEV2400112SC IPGEV2500177SC 318214MR (01)00195278481382(11)250100(21)IPGEV2500177SC IPGEV2500187SC 318675LMR2 (01)00195278481382(11)250200(21)IPGEV2500187SC IPGEV2500219SC RADNETMR1510 (01)00195278481382(11)250600(21)IPGEV2500219SC IPGEV2400124SC 410641MR1 (01)00195278481382(11)240900(21)IPGEV2400124SC IPGEV2400120SC RADNETMR1546 (01)00195278481382(11)240900(21)IPGEV2400120SC IPGEV2300016SC RADNETMR1548 (01)00195278481382(11)230500(21)IPGEV2300016SC IPGEV2500182SC RADNETMR1507 (01)00195278481382(11)250100(21)IPGEV2500182SC IPGEV2400128SC RADNETMR1505 (01)00195278481382(11)240900(21)IPGEV2400128SC IPGEV2500205SC RADNETMR1578 (01)00195278481382(11)250300(21)IPGEV2500205SC IPGEV2500200SC 734647MR (01)00195278481382(11)250200(21)IPGEV2500200SC IPGEV2500203SC 989583MRI (01)00195278481382(11)250300(21)IPGEV2500203SC IPGEV2400067SC 320693MR (01)00195278481382(11)240200(21)IPGEV2400067SC IPGEV2400116SC 660259LMR (01)00195278481382(11)240800(21)IPGEV2400116SC IPGEV2200004SC 406327AMR (01)00195278481382(11)221000(21)IPGEV2200004SC IPGEV2400084SC 704384BALL (01)00195278481382(11)240500(21)IPGEV2400084SC IPGEV2300018SC 402926NHI (01)00195278481382(11)230500(21)IPGEV2300018SC IPGEV2300031SC 908828MR (01)00195278481382(11)230800(21)IPGEV2300031SC IPGEV2300033SC 908890MR2 (01)00195278481382(11)230900(21)IPGEV2300033SC IPGEV2300047SC 732745MR (01)00195278481382(11)231100(21)IPGEV2300047SC IPGEV2400063SC 445MR2 (01)00195278481382(11)240200(21)IPGEV2400063SC IPGEVO2500001 505437MR1 (01)00195278481382(11)250630(21)IPGEVO2500001 IPGEV2200002SC 505888MR (01)00195278481382(11)220600(21)IPGEV2200002SC IPGEV2400142SC 775882MR (01)00195278481382(11)241000(21)IPGEV2400142SC IPGEV2400068SC 702731MR3 (01)00195278481382(11)240200(21)IPGEV2400068SC IPGEV2400057SC 914897MR (01)00195278481382(11)240100(21)IPGEV2400057SC IPGEV2200005SC 718226MR2 (01)00195278481382(11)221000(21)IPGEV2200005SC IPGEV2300034SC 330363MR (01)00195278481382(11)230900(21)IPGEV2300034SC IPGEV2500190SC 614234MR (01)00195278481382(11)250200(21)IPGEV2500190SC IPGEV2400144SC 937208MR3 (01)00195278481382(11)241000(21)IPGEV2400144SC IPGEV2500223SC 330945MR (01)00195278481382(11)250600(21)IPGEV2500223SC IPGEV2500184SC 614839MR (01)00195278481382(11)250100(21)IPGEV2500184SC IPGEV2300022SC 330746MR (01)00195278481382(11)230700(21)IPGEV2300022SC IPGEV2400093SC 740455MR (01)00195278481382(11)240600(21)IPGEV2400093SC IPGEV2400075SC 580421VMR (01)00195278481382(11)240400(21)IPGEV2400075SC IPGEV2500218SC 580931DMR (01)00195278481382(11)250600(21)IPGEV2500218SC IPGEVO2500017 405604DMR (01)00195278481382(11)250825(21)IPGEVO2500017 IPGEV2200008SC 580774WMR (01)00195278481382(11)221200(21)IPGEV2200008SC IPGEV2400130SC 541382SCMR (01)00195278481382(11)240900(21)IPGEV2400130SC IPGEVO2500014 503399MR1 (01)00195278481382(11)250825(21)IPGEVO2500014 IPGEV2400149SC 610740MR1 (01)00195278481382(11)241100(21)IPGEV2400149SC IPGEV2400080SC 610740MR2 (01)00195278481382(11)240400(21)IPGEV2400080SC IPGEV2400136SC 412647MR3 (01)00195278481382(11)241000(21)IPGEV2400136SC IPGEV2500199SC 412784MR2 (01)00195278481382(11)250200(21)IPGEV2500199SC IPGEV2400117SC 610568MR (01)00195278481382(11)240800(21)IPGEV2400117SC IPGEV2400113SC 215322MR2 (01)00195278481382(11)240800(21)IPGEV2400113SC IPGEV2300039SC 864487UPMR (01)00195278481382(11)231000(21)IPGEV2300039SC IPGEV2400126SC 843692MR (01)00195278481382(11)240900(21)IPGEV2400126SC IPGEV2400156SC 423855PEMR (01)00195278481382(11)241200(21)IPGEV2400156SC IPGEV2300035SC 615771MRMD (01)00195278481382(11)230900(21)IPGEV2300035SC IPGEV2400131SC 901259MR (01)00195278481382(11)241000(21)IPGEV2400131SC IPGEV2400073SC 615342MR3 (01)00195278481382(11)240200(21)IPGEV2400073SC IPGEVO2500003 512TSISAMR (01)00195278481382(11)250701(21)IPGEVO2500003 IPGEV2400099SC 936539MR1 (01)00195278481382(11)240700(21)IPGEV2400099SC IPGEV2400060SC 281758MR355 (01)00195278481382(11)240200(21)IPGEV2400060SC IPGEV2300019SC 972560AICMR (01)00195278481382(11)230500(21)IPGEV2300019SC IPGEV2500216SC 972566MR3 (01)00195278481382(11)250500(21)IPGEV2500216SC IPGEV2400129SC 214220WBCMR2 (01)00195278481382(11)240900(21)IPGEV2400129SC IPGEVO2500004 214559SRHMR1 (01)00195278481382(11)250701(21)IPGEVO2500004 IPGEV2400097SC 254897GRMR (01)00195278481382(11)240700(21)IPGEV2400097SC IPGEV2300040SC 830426MCMR (01)00195278481382(11)231000(21)IPGEV2300040SC IPGEV2200001SC 832522MR2 (01)00195278481382(11)220600(21)IPGEV2200001SC IPGEV2300049SC 214420MR1 (01)00195278481382(11)231200(21)IPGEV2300049SC IPGEV2400111SC 972434TSMR1 (01)00195278481382(11)240700(21)IPGEV2400111SC IPGEV2400066SC 469815BCMR1 (01)00195278481382(11)240200(21)IPGEV2400066SC IPGEV2300038SC 936568HDX (01)00195278481382(11)231000(21)IPGEV2300038SC IPGEV2300028SC 972617REDMR1 (01)00195278481382(11)230800(21)IPGEV2300028SC IPGEV2400158SC 210974NCBMR (01)00195278481382(11)241200(21)IPGEV2400158SC IPGEV2300026SC 903533HDXT (01)00195278481382(11)230800(21)IPGEV2300026SC IPGEV2300024SC 361572CICMR (01)00195278481382(11)230700(21)IPGEV2300024SC IPGEV2300044SC 817599WRMR1 (01)00195278481382(11)231100(21)IPGEV2300044SC IPGEV2200003SC 435HEBERMR1 (01)00195278481382(11)220700(21)IPGEV2200003SC IPGEV2200009SC 801373MR4 (01)00195278481382(11)221200(21)IPGEV2200009SC IPGEV2400091SC 801321LMR2 (01)00195278481382(11)240600(21)IPGEV2400091SC IPGEV2400168SC 435789MR1 (01)00195278481382(11)241200(21)IPGEV2400168SC IPGEV2500214SC 703893MR (01)00195278481382(11)250500(21)IPGEV2500214SC IPGEV2400061SC 540SRMCMR (01)00195278481382(11)240200(21)IPGEV2400061SC IPGEVO2500007 951276MR (01)00195278481382(11)250730(21)IPGEVO2500007 IPGEV2400071SC 804254LXMR (01)00195278481382(11)240200(21)IPGEV2400071SC IPGEVO2500010 804675HDMR (01)00195278481382(11)250806(21)IPGEVO2500010 IPGEV2300041SC 540776MR (01)00195278481382(11)231000(21)IPGEV2300041SC IPGEV2500224SC 802334MR360 (01)00195278481382(11)250600(21)IPGEV2500224SC IPGEV2300050SC 425486EVMR (01)00195278481382(11)231200(21)IPGEV2300050SC IPGEV2300023SC 509482MR2 (01)00195278481382(11)230700(21)IPGEV2300023SC IPGEV2500213SC 509891MR (01)00195278481382(11)250500(21)IPGEV2500213SC IPGEV2400123SC 253761CMMR (01)00195278481382(11)240900(21)IPGEV2400123SC IPGEV2300032SC 920406HDX (01)00195278481382(11)230800(21)IPGEV2300032SC IPGEV2400105SC 920720NSHD (01)00195278481382(11)240700(21)IPGEV2400105SC IPGEVO2500005 715847MR1 (01)00195278481382(11)250718(21)IPGEVO2500005 IPGEV2500186SC 0910274036 (01)00195278481382(11)250100(21)IPGEV2500186SC IPGEV2300021SC 833MR000035 (01)00195278481382(11)230600(21)IPGEV2300021SC IPGEV2300014SC 833MR000032 (01)00195278481382(11)230300(21)IPGEV2300014SC IPGEV2400143SC 2807634 (01)00195278481382(11)241000(21)IPGEV2400143SC IPGEV2400157SC 44172 (01)00195278481382(11)241200(21)IPGEV2400157SC IPGEV2500210SC 70537 (01)00195278481382(11)250400(21)IPGEV2500210SC IPGEV2400132SC 250717MR (01)00195278481382(11)241000(21)IPGEV2400132SC IPGEV2500222SC 613737MR2 (01)00195278481382(11)250600(21)IPGEV2500222SC IPGEVO2500008 306766MR (01)00195278481382(11)250730(21)IPGEVO2500008 IPGEV2400083SC 450778MRI (01)00195278481382(11)240500(21)IPGEV2400083SC IPGEV2400103SC Not Available (01)00195278481382(11)240700(21)IPGEV2400103SC IPGEV2400090SC 514431MRI (01)00195278481382(11)240600(21)IPGEV2400090SC IPGEV2400110SC 250370ARTIST (01)00195278481382(11)240700(21)IPGEV2400110SC IPGEV2500201SC 250727MR1 (01)00195278481382(11)250300(21)IPGEV2500201SC IPGEV2400108SC HVBMR01 (01)00195278481382(11)240700(21)IPGEV2400108SC IPGEV2500226SC CS1065MR01 (01)00195278481382(11)250627(21)IPGEV2500226SC IPGEVO2500002 Not Available (01)00195278481382(11)250701(21)IPGEVO2500002 IPGEV2400114SC DK1083MR06 (01)00195278481382(11)240800(21)IPGEV2400114SC IPGEV2200006SC DK1001MR01 (01)00195278481382(11)221100(21)IPGEV2200006SC IPGEV2300011SC DK1296MR01 (01)00195278481382(11)230300(21)IPGEV2300011SC IPGEV2400074SC 0920279030 (01)00195278481382(11)240300(21)IPGEV2400074SC IPGEV2500204SC HC0220MR02 (01)00195278481382(11)250300(21)IPGEV2500204SC IPGEV2400107SC H5912MR02 (01)00195278481382(11)240700(21)IPGEV2400107SC IPGEV2300029SC A5127236 (01)00195278481382(11)230800(21)IPGEV2300029SC IPGEV2500220SC A5624132 (01)00195278481382(11)250612(21)IPGEV2500220SC IPGEVO2500012 A5632405 (01)00195278481382(11)250820(21)IPGEVO2500012 IPGEV2500209SC A5632418 (01)00195278481382(11)250400(21)IPGEV2500209SC IPGEV2400078SC MR00248 (01)00195278481382(11)240400(21)IPGEV2400078SC IPGEV2500217SC MR00328 (01)00195278481382(11)250600(21)IPGEV2500217SC IPGEV2300052SC MR00217 (01)00195278481382(11)231200(21)IPGEV2300052SC IPGEV2500202SC MR00308 (01)00195278481382(11)250300(21)IPGEV2500202SC IPGEV2400125SC MR00269 (01)00195278481382(11)240900(21)IPGEV2400125SC IPGEV2300017SC MR00156 (01)00195278481382(11)230500(21)IPGEV2300017SC IPGEV2400146SC MR00287 (01)00195278481382(11)241100(21)IPGEV2400146SC IPGEV2500178SC MR00292 (01)00195278481382(11)250100(21)IPGEV2500178SC IPGEV2400070SC MR00242 (01)00195278481382(11)240200(21)IPGEV2400070SC IPGEV2300037SC MR00208 (01)00195278481382(11)231000(21)IPGEV2300037SC IPGEV2400058SC MR00221 (01)00195278481382(11)240100(21)IPGEV2400058SC IPGEV2400133SC MR00270 (01)00195278481382(11)241000(21)IPGEV2400133SC IPGEV2400165SC MR00306 (01)00195278481382(11)241200(21)IPGEV2400165SC IPGEV2400134SC MR00271 (01)00195278481382(11)241000(21)IPGEV2400134SC IPGEV2400154SC MR00293 (01)00195278481382(11)241100(21)IPGEV2400154SC IPGEV2300053SC MR00230 (01)00195278481382(11)231200(21)IPGEV2300053SC IPGEV2400069SC MR00237 (01)00195278481382(11)240200(21)IPGEV2400069SC IPGEV2400095SC MR00238 (01)00195278481382(11)240600(21)IPGEV2400095SC IPGEV2400076SC MR00245 (01)00195278481382(11)240400(21)IPGEV2400076SC IPGEVO2500048 525640SIGNA (01)00195278481382(11)251120(21)IPGEVO2500048 IPGEV2500192SC MR55222591 (01)00195278481382(11)250200(21)IPGEV2500192SC IPGEV2300045SC FPG103200 (01)00195278481382(11)231100(21)IPGEV2300045SC IPGEV2400055SC MPX82200 (01)00195278481382(11)240100(21)IPGEV2400055SC IPGEV2400122SC UP506A13 (01)00195278481382(11)240900(21)IPGEV2400122SC IPGEV2400138SC FPG25113 (01)00195278481382(11)241000(21)IPGEV2400138SC IPGEV2500221SC DPB15820 (01)00195278481382(11)250600(21)IPGEV2500221SC IPGEV2400087SC MEX68601 (01)00195278481382(11)240600(21)IPGEV2400087SC IPGEV2300048SC FPG46909 (01)00195278481382(11)231200(21)IPGEV2300048SC IPGEV2500183SC 70328MRS04 (01)00195278481382(11)250100(21)PGEV2500183SC Product - 1.5T SIGNA ARTIST FLO IPM Serial Number / System ID / UDI code: IPGEV2400096SC 617754MR9 (01)00195278481382(11)240600(21)IPGEV2400096SC Product - SIGNA Artist Evo IPGR with XG5, CTO Serial Number / System ID / UDI code: IPGEVO2600069 Not Available (01)00195278481382(11)260202(21)IPGEVO2600069
FEI Number 3005326970
Recalling Firm/
Manufacturer
GE Healthcare LLC
3001 W Radio Dr
Florence SC 29501-7806
For Additional Information ContactTracey Holevas
843-664-1488
Manufacturer Reason
for Recall
Due to an exposed sharp edge inside electronic cabinet that could damage power cables which could potentially pose an electrical fault such as an arc flash
FDA Determined
Cause 2
Process control
ActionOn 08/27/2026, an "URGENT MEDICAL DEVICE CORRECTION" Letter was sent to customers informing them that, GE HealthCare has become aware of an exposed sharp edge inside the SIGNA Artist Evo and SIGNA Artist electronics cabinet that could damage power cables, which may potentially cause an electrical fault. If the electrical fault is present and the cabinet door is opened while the system is powered, this could potentially result in an arc flash that could lead to serious injury to service personnel in immediate proximity. The issue would only present itself during servicing activities involving the electronics cabinet. The electronics cabinet is located within the equipment room and should only be accessed by qualified service personnel. Actions to be taken by Customers: Pending correction by GE HealthCare, you can continue to use the system. The electronics cabinet door should be closed any time there is power to the system. In the event the electronics cabinet needs to be serviced, your local GE HealthCare service representative or authorized third-party service providers are specifically trained and qualified to safely service your equipment, including the electronics cabinet, in accordance with standard GE HealthCare procedures and safety requirements detailed in the applicable GE HealthCare service manuals (Artist --- 5841150-2EN and Artist EVO --- 5884918-8EN). Please retain this document for your records. Please complete and return the attached acknowledgement form electronically via FMI 67959 Digital Response Form or print, fill out manually, scan, and email to recall.67959@gehealthcare.com. GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. For questions - contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Quantity in Commerce280 (200 US; 80 OUS)
DistributionWorldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI and the countries of Austria, Belgium, Brazil, Canada, Chile, Costa Rica, Czech Republic, Denmark, Estonia, Fiji, Germany, Greece, Italy Japan, Mexico, New Zealand, Panama, Peru, Poland, Serbia, Spain, Taiwan, United Kingdom and Vietnam.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.