 Class 2 Device Recall ETHILON" Nylon Suturesee related information
Date Initiated by FirmSeptember 23, 2026
Date PostedOctober 23, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0284-2027
Recall Event ID 99947
510(K)NumberK946173 
Product Classification Suture, nonabsorbable, steel, monofilament and multifilament, sterile - Product Code GAQ
ProductETHILON Nylon Suture, Product Code 697G
Code Information GTIN Box/Unit: 30705031008568/10705031008564. Lot: 10B3QS
FEI Number 2648650
Recalling Firm/
Manufacturer
Ethicon, LLC
Highway 183 Km 8.3
San Lorenzo PR 00754
For Additional Information Contact
787-783-7070
Manufacturer Reason
for Recall
Nylon suture indicated for use in general soft tissue approximation and/or ligation may have needles with distorted curvature, that may break during use, or could cause additional soft tissue or vessel injury, including tearing or bleeding intraoperatively.
FDA Determined
Cause 2
Process control
ActionOn 9/3/2026, field safety notices were mailed to customers who were asked to do the following: 1) Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). If you have product subject to this recall, please maintain a copy of this notice with the quarantined product and keep a copy for your records. 2. Return affected product to firm. 3. Communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return and forward them this recall notice. 4. Complete and return the Business Reply Form (BRF) via email to RA-ITSUS-JJMCanada@ITS.JNJ.com
Quantity in Commerce4,003
DistributionInternational distribution to the countries of JP, CA, AU.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = GAQ
-
-