 Class 2 Device Recall CryoPatch SynerGraftsee related information
Date Initiated by FirmSeptember 18, 2026
Date PostedOctober 30, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0325-2027
Recall Event ID 100012
510(K)NumberK101866 
Product Classification Patch, pledget and intracardiac, petp, ptfe, polypropylene - Product Code DXZ
ProductCryoPatch SynerGraft (SG) Pulmonary Human Cardiac Patch, Model: SGP010
Code Information Serial/Donor Number: 12617554/177871, UDI-DI: 00877234000478, Shipped: 22May2024, Implanted: 05June2024
FEI Number 3001451326
Recalling Firm/
Manufacturer
Artivion, Inc
1655 Roberts Blvd Nw
Kennesaw GA 30144-3632
For Additional Information ContactRochelle Maney
678-290-4531
Manufacturer Reason
for Recall
A cryopreserved pulmonary human cardiac patch, taken from a 64-year-old human donor, was outside the cardiac tissue donor age specification of newborn to 60 years. The patch was implanted, and due to limitations of long-term in vivo data, the possibility of premature graft dysfunction cannot be excluded.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionOn 9/18/2026, the notification was delivered via mail to customer who was advised of the following: Available biomechanical and histological evidence does not indicate an increased risk that would warrant additional patient testing, enhanced clinical monitoring, or other clinical intervention beyond the patient's routine standard of care. Accordingly, prophylactic removal or surgical replacement of the already implanted tissue solely because of the donor-age nonconformance is not clinically warranted. Complete and return the response form via email to fieldassurance@artivion.com Firm requests that any observed deficiency, performance concern, or other issue potentially associated with the tissue be reported to them vial email: fieldassurance@artivion.com or phone: +1-770-419-3355
Quantity in Commerce1
DistributionUS Nationwide distribution in the state of IL.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.