| Date Initiated by Firm | October 05, 2026 |
| Date Posted | November 05, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0359-2027 |
| Recall Event ID |
100036 |
| 510(K)Number | K243013 |
| Product Classification |
Instrumentation for clinical multiplex test systems - Product Code NSU
|
| Product | LIAISON PLEX Gram-Negative Blood Culture Assay
Catalog No. 20-005-303 |
| Code Information |
All lots
UDI-DI pending |
| FEI Number |
3006028115
|
Recalling Firm/ Manufacturer |
Luminex Corporation
4088 Commercial Ave
Northbrook
IL
60062-1829
|
| For Additional Information Contact | Technical Support 877-785-2323 Ext. 9029 |
Manufacturer Reason for Recall | Instructions for Use updated to include explicit guidance that repeat testing is recommended whenever a positive family-level only result is obtained. Issue may contribute to incorrect or delayed diagnosis of bloodstream infection. |
FDA Determined Cause 2 | Error in labeling |
| Action | On October 5, 2026, the firm began notifying customers via "Urgent Field Safety Notice - Medical Device Recall" letters.
Consignees were informed that the current Instructions for Use for the Gram-Negative Blood Culture Assay do not adequately describe interpretation of certain positive family-level results. Specifically, when a positive family-level only result is obtained ("family-level only result" refers to a positive organism family-level detection without an accompanying genus- or species-level identification), repeat testing with the LIAISON PLEX Gram-Negative Blood Culture Assay is recommended. Absence of this instruction may increase the likelihood that certain reportable detections are not appropriately recognized by the user and may contribute to an incorrect interpretation of test results.
Customers were instructed to implement the following interpretation guidance:
When a positive family-level only result is obtained, repeat testing with the LIAISON PLEX BCN assay is recommended.
DiaSorin is updating the Instructions for Use to include the revised guidance. Revised labeling will be provided following implementation of the labeling revision.
Please contact Technical Support Services with any questions or concerns at the following phone numbers:1-877-785-2323 (U.S. and Canada); +1-512-381-4397 (Outside U.S. and Canada); 1-512-219-5114 (Fax) and email:support@luminexcorp.com. |
| Quantity in Commerce | 30,168 test cartridges (2,514 Kits) |
| Distribution | US distribution nationwide. No OUS distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|