| Product | Philips EPIQ Elite Diagnostic Ultrasound System, Model: 795125 |
| Code Information |
All Philips EPIQ Elite Ultrasound Systems running software version 14.0, UDI-DI: 00884838114234 |
| FEI Number |
3019216
|
Recalling Firm/ Manufacturer |
Philips Ultrasound, LLC
22100 Bothell Everett Hwy
Bothell
WA
98021-8431
|
| For Additional Information Contact | Philips Customer Care Solutions Center 800-722-9377 |
Manufacturer Reason for Recall | Due to software issue with ultrasound system, doppler sample is displayed shallower than its actual measurement location, potentially causing inaccurate blood flow velocity measurements and misdiagnosis. |
FDA Determined Cause 2 | Software Design Change |
| Action | On 10/1/2026, Philips issued an Urgent Medical Device Correction Notification to affected consignees. Philips requested consignees take the following actions:
1. Due to possibility of a weak PW Doppler Signal on devices with affected software, customers are advised not to use C5-1 Abd Vasc Tissue Specific Preset (TSP). Please use a different TSP (e.g., Abd Gen, Abd Renal) within C5-1.
2. Otherwise, you may continue to use your device in accordance with the IFU.
3.Circulate this notice to all users of the device so that they are aware of the product issue and associated hazard / harm.
4. Post this notice near the affected unit(s) for ease of reference.
5. Complete and return the response form via email to alps.ssd.c-r.gbs@philips.com |
| Quantity in Commerce | 1 |
| Distribution | International distribution to the country of Switzerland.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|