• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Philips EPIQ Elite Diagnostic Ultrasound Circular Edition System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Philips EPIQ Elite Diagnostic Ultrasound Circular Edition Systemsee related information
Date Initiated by FirmOctober 01, 2026
Date PostedNovember 17, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0470-2027
Recall Event ID 100039
510(K)NumberK260123 K260680 
Product Classification System, imaging, pulsed doppler, ultrasonic - Product Code IYN
ProductPhilips EPIQ Elite Diagnostic Ultrasound System, Model: 795125
Code Information All Philips EPIQ Elite Ultrasound Systems running software version 14.0, UDI-DI: 00884838114234
FEI Number 3019216
Recalling Firm/
Manufacturer
Philips Ultrasound, LLC
22100 Bothell Everett Hwy
Bothell WA 98021-8431
For Additional Information ContactPhilips Customer Care Solutions Center
800-722-9377
Manufacturer Reason
for Recall
Due to software issue with ultrasound system, doppler sample is displayed shallower than its actual measurement location, potentially causing inaccurate blood flow velocity measurements and misdiagnosis.
FDA Determined
Cause 2
Software Design Change
ActionOn 10/1/2026, Philips issued an Urgent Medical Device Correction Notification to affected consignees. Philips requested consignees take the following actions: 1. Due to possibility of a weak PW Doppler Signal on devices with affected software, customers are advised not to use C5-1 Abd Vasc Tissue Specific Preset (TSP). Please use a different TSP (e.g., Abd Gen, Abd Renal) within C5-1. 2. Otherwise, you may continue to use your device in accordance with the IFU. 3.Circulate this notice to all users of the device so that they are aware of the product issue and associated hazard / harm. 4. Post this notice near the affected unit(s) for ease of reference. 5. Complete and return the response form via email to alps.ssd.c-r.gbs@philips.com
Quantity in Commerce1
DistributionInternational distribution to the country of Switzerland.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = IYN
-
-