| | Class 2 Device Recall Veritas Reverse Total Shoulder Replacement |  |
| Date Initiated by Firm | October 01, 2026 |
| Date Posted | November 03, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0348-2027 |
| Recall Event ID |
100047 |
| 510(K)Number | K243643 |
| Product Classification |
shoulder prosthesis, reverse configuration - Product Code PHX
|
| Product | Veritas Reverse Total Shoulder Replacement,
Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer Liner |
| Code Information |
All Lot codes/ UDIs:Product Code: 3500-140C3 UDI/GTIN:00840097513492;
Product Code: 3500-140U3 UDI/GTIN:00840097513461;
Product Code: 3500-140C1 UDI/GTIN:00840097513430;
Product Code: 3500-140U1 UDI/GTIN:00840097513409
|
| FEI Number |
3014833750
|
Recalling Firm/ Manufacturer |
restor3d Inc.
4001 E Nc 54 Hwy
Suite 3160
Durham
NC
27709
|
| For Additional Information Contact | Brianna Gabbard 1-786-5210501 |
Manufacturer Reason for Recall | (Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery.
|
FDA Determined Cause 2 | Device Design |
| Action | On 1, 2026 Restor3d reached out to affected consignees via phone call to make them aware of the recall. A Urgent; Medical Device Recall was issued to affected consignees on October 7, 2026 via email. Restor3d ask consignees to take the following actions:
1. Do not implant or otherwise use any affected units identified above.
2. Complete the attached Field Action Response Form (Attachment A) and return it to restor3d within 10 business days, so that restor3d can confirm effective distribution of this notice.
3. Promptly report to restor3d any intraoperative issue, patient adverse event, suspected device related complication, loosening, migration, revision surgery, or other clinical concern potentially associated with the affected product. |
| Quantity in Commerce | 240 units |
| Distribution | US Nationwide distribution in the states of Texas, Massachusetts, Oklahoma, Florida, Colorado, North Carolina. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|