| Date Initiated by Firm | September 30, 2026 |
| Date Posted | November 03, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0349-2027 |
| Recall Event ID |
100098 |
| PMA Number | P970051S225 |
| Product Classification |
Implant, cochlear - Product Code MCM
|
| Product | Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode
REF P1979576
Model Number: CI1000
UDI-DI codes: 9321502069847
Cochlear Nucleus CI1000 Series implants are prescription only, single use devices
intended for long term implantation under the skin in the mastoid region on either side of the head.
The cochlear implant system works by converting sound in the environment into electrical code. It transmits this code to the auditory nerve and on to the brain where it is interpreted as sound. The Cochlear Implant System consists of internal (implanted) and external components. The implanted component consists of a receiver/stimulator and an electrode array. The external components consist of a sound processor and their associated transmitting coils and cables and must be worn for the implant to work. |
| Code Information |
Model: CI1000
Product/Catalog: P1979576
UDI-DI code: 9321502069847
Serial Numbers:
1020210051972
1020210070461
1020210145046
1020210048286 |
| FEI Number |
3009092818
|
Recalling Firm/ Manufacturer |
Cochlear Limited
1 University Ave
Macquarie University
Australia
|
| For Additional Information Contact | Laura Blair 720-301-1034 |
Manufacturer Reason for Recall | Due to a manufacturing deviation of incomplete electrode welds of cochlear implants and resulting in one or more open circuits |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | On 09/30/2026, an "URGENT Field Safety Corrective Action Notification" was sent to U.S. customers informing them that a potential condition has been identified that could result in one or more open circuits affecting individual electrode pathways with the implant.
Customers were instucted to not implant and return affected products to Cochlear Limited.
For questions contact CNA_Quality@cochlear.com. |
| Quantity in Commerce | 4 device in the U.S. |
| Distribution | U.S.: NJ, NY, and PA.
O.U.S.: Not provided |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| PMA Database | PMAs with Product Code = MCM
|