 Class 2 Device Recall Medline convenience kitssee related information
Date Initiated by FirmSeptember 24, 2026
Date PostedNovember 10, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0422-2027
Recall Event ID 100112
Product Classification Orthopedic tray - Product Code OJH
ProductMedline convenience kits containing BD Urinary Catheter Trays that include CliniClean Pericare Wipes ENT-MICRO DYNJ912347 LAMINECTOMY DYNJ911420A NON MB MJR NEURO CRANI SM PK DYNJT6104 SPINE DYNJ911856
Code Information DYNJ912347 UDI-DI 10198459699757 Lot 26EMA878 DYNJ911420A UDI-DI 10198459745416 Lot 26HMM122 DYNJT6104 UDI-DI 10198459539374 Lot 26BBE722 DYNJ911856 UDI-DI 10198459559068 Lot 26DMI925
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries, LP
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactKarin Johnson
800-633-5463
Manufacturer Reason
for Recall
Convenience kits contain additional product codes and lots of recalled CliniClean Pericare Wipes identified by manufacturer. Wipes may be contaminated with a lubricant used in the manufacturing process. Use of affected wipes may cause localized skin effects such as irritation, erythema, or mild inflammation. Failure to perform pericare may increase the risk of localized infection or catheter-associated urinary tract infection.
FDA Determined
Cause 2
Process change control
ActionThis is an update to a previously issued letter due to the manufacturer/supplier identifying additional lots of affected product. On September 24, 2026, customers were notified of the additional products and lots affected. Medline Industries LP instructed customers to request stickers to over-label the affected convenience kits with a sticker indicating to remove the affected component and replace it with product from supply. No kits are to be returned.
Quantity in Commerce7,043 kits
DistributionUS Nationwide distribution in the states of AL, AR, CA, FL, GA, IL, IN, LA, MA, MN, MO, MS, MT, NC, NE, NV, NY, OH, OK, RI, SC, TN, TX, WI.
Total Product Life CycleTPLC Device Report