| | Class 2 Device Recall Philips EPIQ series Diagnostic Ultrasound Systems |  |
| Date Initiated by Firm | December 21, 2023 |
| Date Posted | November 17, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0473-2027 |
| Recall Event ID |
100131 |
| 510(K)Number | K212704 |
| Product Classification |
System, imaging, pulsed doppler, ultrasonic - Product Code IYN
|
| Product | EPIQ Ultrasound Systems, REF: 795200, 795201, 795204, 795205, 795206, 795231, 795232, 795234, 795236, 795237, 795239, 795240, 795241, 795242, 795244, 795245, 795247, 795248, 795251, 795253, 795257, 795258, 795263. |
| Code Information |
UDI-DI: 00884838106703, 00884838106741, 00884838107441, 00884838107502, 00884838107519, 00884838107526, 00884838107533, 00884838107540, 00884838108271, 00884838108288, 00884838108301, 00884838108318, 00884838108332, 00884838108349, 00884838108370, 00884838108394, 00884838109995, 00884838110090, 00884838110755, 0088483811100, 00884838111059, 00884838111066, 00884838111097, 00884838112162, 00884838112179.
See main system label - Systems manufactured between 2023-09-18 to 2023-11-08.
|
| FEI Number |
3019216
|
Recalling Firm/ Manufacturer |
Philips Ultrasound, LLC
22100 Bothell Everett Hwy
Bothell
WA
98021-8431
|
| For Additional Information Contact | Philips Customer Care Solutions Center 800-722-9377 |
Manufacturer Reason for Recall | Diagnostic ultrasound systems have slide controls that adjust brightness range of gain compensation over a unique 2D image region, which may become unresponsive after multiple system start-ups/reboots. Rebooting systems may not recover TGC Slide Controls. The technical solution to correct affected systems was released on 15-Mar-2024. |
FDA Determined Cause 2 | Software change control |
| Action | On 12/21/2023, Important Product Notices were mailed to customers who were informed of the following:
As a workaround, the following methods are available for 2D image gain compensation:
1) Use the 2D gain rotary knob to increase or decrease the entire 2D image display brightness by rotation of the knob.
2) Use the iSCAN feature to automatically sample the entire 2D image and adjust in gain compensation, where manipulation of the individual TGC Slide Control is not required.
3) Enable the AutoSCAN feature which uses the real time frame by frame analysis and adjusts the 2D gain compensation. Enabling the AutoSCAN feature does not require manipulation of the individual TGC Slide Control. The AutoSCAN feature, located in the 2D tab in the system touchscreen is enabled when the button is highlighted in amber.
Firm contacted customers and scheduled times for a Field Service Engineers(FSE) to visit sites and install the solution to resolve this issue (reference FCO79500567).
If you need any further information or support, contact firm Customer Care Solutions Center (1-800-722-9377).
|
| Quantity in Commerce | 808 |
| Distribution | US Nationwide distribution including in the states and territory of NY, CT, FL, IL, AL, CA, VA, OH, MN, AZ, IN, KY, MI, WI, MA, MO, WV, NC, GA, DC, TX, TN, PA, MD, KS, MT, HI, NH, SC, AR, ME, NE, WA, NJ, RI, CO, OK, MS, UT, PR and the countries of India, France, China, Mexico, Australia, New Zealand, Japan, Thailand, Spain, South Korea, Switzerland, Taiwan, Poland, Indonesia, Iran, Singapore, Austria, Italy, Germany, Greece, Portugal, Canada, United Kingdom, Egypt, Iceland, South Africa, Romania, Czech Republic, Costa Rica, Brazil, Lebanon, Ireland, Malaysia, Sweden, Netherlands, Slovakia, Bangladesh.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|