• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Medline

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Medlinesee related information
Date Initiated by FirmOctober 06, 2026
Date PostedNovember 25, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0555-2027
Recall Event ID 100175
Product Classification Circuit, breathing (w connector, adaptor, y piece) - Product Code CAI
ProductMedline Convenience Kits Containing ICU Medical IV Tubing Set Drip Chambers 10 YEARS AND UP ADULT CIRCUIT DYNJAP9865A 8 TO 11 YEARS OLD PEDS CIRC DYNJAP9864A CARDIAC ANES VCH DYNJ39265I
Code Information DYNJAP9865A UDI-DI 10888277872943 Lots 21KBB065 21LBJ200 22BBR826 22CBM921 22EBT661 22GBY067 22IBT343 23ABA861 23ABS746 23BBR456 23CBW149 23DBG905 23FBL821 23GBT937 23HBE475 23IBS302 23JBN435 23KBG660 23LBV397 24CBI254 24EBD599 24GBD178 24GBR709 24HBN932 24IBR211 24JBL083 24KBL996 25ABO991 25CBF170 25DBJ600 25EBV555 25GBD712 25HBO762 25KBB166 25KBR530 26ABB971 26ABT605 26BBT623 26EBT213 DYNJAP9864A UDI-DI 10888277872936 Lots 21KBY459 21LBM857 22CBL756 22FBH318 22GBY066 22IBT422 22JBP539 23CBI117 23DBH976 23FBF503 23GBU138 23HBU899 23KBG670 24ABU871 24EBC651 24FBH092 24GBM291 24IBR042 24KBR439 25ABP731 25CBJ267 25DBK958 25EBN627 25HBC222 25IBA257 25JBH497 25LBO431 26BBJ224 26DBR422 DYNJ39265I UDI-DI 10193489448030 Lots 22LMH247 23CMA141 23DMH907 23GMD355 23HMB609 23HMG671 23IMA276 23JMG800 24AMI697 24BMD517 24CMA820 24FMJ962 24HMB832 24HMI760 24JMB132 25AMI743 25CMF045 25DME519 25FMI127 25GMA301 25IMF425 26DMF827 26FMB584 26FMJ392 26GMK920 26HMJ978
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries, LP
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactJeremy Napshin
800-633-5463
Manufacturer Reason
for Recall
Kits contain recalled ICU Medical IV Tubing Set Drip Chambers. Drip chambers may exhibit discoloration that is not expected to affect device safety, performance, or intended use. In some cases, user may elect to replace discolored product before use, which could result in delay to therapy.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionFirm began notifying customers on 10/6/26 via email and first class mail. Medline Industries LP instructed customers that they may continue to use the ICU Medical IV Tubing Set Drip Chambers and Medline kits containing the affected component as intended and in accordance with facility protocols, per the instructions provided in ICU Medical s notification.
Quantity in Commerce116,705
DistributionUS Nationwide distribution in the countries of CA, FL, GA, IA, IN, KS, KY, NV, NY, and TX.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-