| | Class 1 Device Recall Intersurgical |  |
| Date Initiated by Firm | November 03, 2026 |
| Date Posted | November 24, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0493-2027 |
| Recall Event ID |
100283 |
| Product Classification |
Circuit, breathing (w connector, adaptor, y piece) - Product Code CAI
|
| Product | Intersurgical Smoothbore Breathing Systems, Model Numbers:
1) REF: 5018000, 22mm Smoothbore tube with soft end connectors, 22F-22F, e 1.8m;
2) REF: 5804000, 22mm Smoothbore bilevel breathing system with swivel CO leak port, e 1.8m;
3) REF: 5805000, 22mm Smoothbore bilevel breathing system with monitoring line and swivel CO leak port, e1.6m;
4) REF: 5805030, 22mm Smoothbore bilevel breathing system with swivel CO
leak port and Clear-Guard Midi filter e1.8m; breathing circuit |
| Code Information |
REF: 5018000 UDI/DI Single unit: 5030267068944; Case: 05030267028597, Lot Codes: 32511811;
REF: 5804000 UDI/DI Single unit: 5030267070480; Case: 05030267028559, Lot Codes: 32508034, 32509327, 32510909, 32516895, 32518659, 32519175, 32525543;
REF: 5805000 UDI/DI Single unit: 5030267070534; Case: 05030267153558, Lot Codes: 32516315, 32517877;
REF: 5805030 UDI/DI Single unit: 5030267156030; Case: 05030267156047, Lot Codes: 32505469, 32506841, 32513929, 32514276, 32515579, 32518484, 32519342, 32519468, 32519979, 32522143, 32522415, 32523651, 32525498, 32525628, 32526171, 32526270, 32527299, 32528409. |
| FEI Number |
1832562
|
Recalling Firm/ Manufacturer |
Pulmodyne Inc.
2055 Executive Dr
Indianapolis
IN
46241-4311
|
| For Additional Information Contact | Customer Support 800-828-8633 |
Manufacturer Reason for Recall | Soft connectors may disconnect from the tube resulting in a loss of ventilation or compromised ventilation. |
FDA Determined Cause 2 | Process control |
| Action | Intersurgical Inc. issued an "URGENT MEDICAL DEVICE RECALL" notice to its consignees on 11/03/2026 via email. The notice explained the issue, potential risk, and requested the following:
"1. Immediately discontinue use and quarantine any stock with the product/lot numbers described above in this recall notification.
2. Please share this notice with all users of these products within your facility. If you have distributed any of these products, please immediately contact your customers, advise them of the recall situation and have them return their outstanding recalled stock to you.
3. Contact Intersurgical Customer Support at 800-828-9633 for a Return Goods Authorization (RGA) Number and return label. Hours of operation are Monday through Thursday, 8:00 a.m. to 4:00 p.m. EST and Friday 8:00 a.m. to 3:00 p.m. EST.
a. Once you have received the RGA Number, please enter it in the space provided on the Customer Product Recall Acknowledgement Form:
i. Return the completed form via email to Intersurgical Inc. at recalls@intersurgicalinc.com.
ii. Please ship all product back to Intersurgical using the return label supplied by Customer Support.
iii. The subject line should contain your Company name and RGA number.
1. This will allow us to document your receipt of this letter and the amount of product you have on-hand for return.
b. Intersurgical will credit or replace your product free of charge when the product is received.
4. Please note: If you do not have the recalled product in stock, please check the appropriate box on the Customer Product Recall Acknowledgement Form.
a. Return the completed form via email to Intersurgical Inc. at recalls@intersurgicalinc.com.
b. The subject line should contain your Company name and Intersurgical Recall.
5. Please continue to report to Intersurgical any adverse events involving these products."
For questions, contact Customer Support at 1-800-828-9633 or email at recalls@intersurgicalinc.com.
|
| Quantity in Commerce | 28368 units |
| Distribution | US Distribution: FL, OH, MN, TX, VA, NY, NJ, CA, CO, IN. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|