| Class 2 Device Recall Zimmer VERSYS |  |
Date Initiated by Firm | August 02, 2006 |
Date Posted | August 31, 2006 |
Recall Status1 |
Terminated 3 on May 04, 2007 |
Recall Number | Z-1467-06 |
Recall Event ID |
35943 |
510(K)Number | K961378 |
Product Classification |
hip prosthesis - Product Code LWJ
|
Product | Zimmer VERSYS Hip System Femoral Stem, press-fit, enhanced taper, 12/14 neck taper, collarless, size 14, 135 mm stem length, tivanium TI-6AL-4V alloy, sterile, w/calcicoat ceramic coating; Cat. no. 65-7861-14-04 (65786101404). |
Code Information |
Lot 60460460. |
Recalling Firm/ Manufacturer |
Zimmer Inc. 345 E Main St Warsaw IN 46580-2746
|
For Additional Information Contact | Connie Morgan 800-846-4637 |
Manufacturer Reason for Recall | Lack of assurance of proper metal fatigue strength due to a metal grain structure anomaly. |
FDA Determined Cause 2 | Other |
Action | Consignees were notified via letter dated 8/2/06 to cease use of and to return the products. Zimmer distributors were to examine stock and return product on hand. |
Quantity in Commerce | 181 of all devices |
Distribution | Worldwide, including USA, Australia, China, Denmark, Japan, Jordan, Latin Ameria, Russia, Singapore and Switzerland. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = LWJ
|
|
|
|