| | Class 2 Device Recall Terumo APS 1 |  |
| Date Initiated by Firm | October 12, 2004 |
| Date Posted | December 30, 2006 |
| Recall Status1 |
Terminated 3 on February 01, 2007 |
| Recall Number | Z-0330-2007 |
| Recall Event ID |
36396 |
| 510(K)Number | K022947 |
| Product Classification |
Perfusion System - Product Code DTQ
|
| Product | Terumo Advanced Perfusion System 1 Level Detection; Catalog number 802111. |
| Code Information |
Serial numbers 00023, 00025 through 00027, 00055, 00061 through 00081, 00083 through 00099, 00101 through 00109, 00111 through 00183, 00185 through 00260, 00262 through 00269. 00271 through 00293, 00295 through 00307 and 00309 through 00357. |
Recalling Firm/ Manufacturer |
Terumo Cardiovascular Systems Corp 6200 Jackson Rd Ann Arbor MI 48103-9586
|
| For Additional Information Contact | 800-521-2818 |
Manufacturer Reason for Recall | Device module(s) may fail to properly initialize during start up, thus resulting in that single module being inoperable and a red 'X' being displayed on the monitor screen. |
FDA Determined Cause 2 | Other |
| Action | U.S. consignees were sent an addendum to the operators manual on October 12, 2004 instructing them to cycle power to the unit to clear the error. International distributors were requested to provide this information to their customers on 12/15/04, but Terumo has not yet received confirmation that this was completed. |
| Quantity in Commerce | 4,740 for all affected products |
| Distribution | Worldwide Distribution-USA and countries Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = DTQ
|
|
|
|