|
Class 2 Device Recall LIFEPAK 20 |
|
Date Initiated by Firm |
January 08, 2008 |
Date Posted |
March 29, 2008 |
Recall Status1 |
Terminated 3 on September 04, 2013 |
Recall Number |
Z-1050-2008 |
Recall Event ID |
46351 |
510(K)Number |
K012274
|
Product Classification |
External Automatic Defibrillator - Product Code MKJ
|
Product |
Physio-Control LIFEPAK 20 defibrillator/monitor, AC powered external defibrillator with battery backup. Part Numbers: 3202487-073, 3202487-079, 3202488-036, Physio Control, Inc, Redmond, WA 98052 |
Code Information |
Serial Numbers (domestic units): 36133753, 36133760, 36139018, 36139023, 36139030, 36137448, 36138998, 36141356, 36141357, 36141358, 36141360, 36141361, 36141362, 36141363, 36141365, 36141366, 36141367, 36141370, 36141373, 36141378, 36141380, 36141381, 36141382, 36141383, 36141384, 36141393, 36141394, 36141395, 36141396, 36141399, 36141400, 36142717, 36142718, 36142730, 36142731, 36141386, 36141398, 36133739, 36133740, 36133742, 36133743, 36133744, 36133745, 36133746, 36133747, 36133748, 36133749, 36133750, 36133751, 36133752, 36133754, 36133755, 36133756, 36133757, 36133758, 36133759, 36137218, 36137219, 36137220, 36137222, 36137224, 36137225, 36137226, 36137227, 36137230, 36137231, 36137352, 36137353, 36137355, 36137356, 36137357, 36137359, 36137360, 36137361, 36137362, 36137363, 36137364, 36137365, 36137366, 36137367, 36137368, 36137369, 36139002, 36139010 Serial Numbers (international units): 36137354, 36137358, 36137370, 36137371, 36137372, 36138999, 36139000, 36139005, 36139006, 36139007, 36139009, 36139011, 36139012, 36139013, 36139014, 36139021, 36139024, 36139025, 36139031 |
Recalling Firm/ Manufacturer |
Physio Control, Inc. 11811 Willows Rd Ne Redmond WA 98052-2003
|
For Additional Information Contact |
Todd Bandy 425-867-4000 Ext. 4644
|
Manufacturer Reason for Recall |
Inoperable Device - Potential for delay in therapy or prevention of defibrillation therapy due to corrosion of the Printed Circuit Board Assemblies (PCBA).
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
The firm contacted Customers by phone and by Urgent-Medical Device Recall letters sent via return receipt requested to their consignees on 1/8/08. The letter advised consignees that the Printed Circuit Board Assemblies (PCBA) might not meet quality requirements. This manufacturing deviation could lead to eventual PCBA corrosion that could render the device inoperative and not ready for use in an emergency under battery power. The firm recommended keeping the unit in service until it could be replaced. Representatives will visit each consignee and replace units. |
Quantity in Commerce |
103 total devices - 84 domestic, 19 international |
Distribution |
Worldwide Distribution --- USA including states of OK, SC, TX, and WA, and country of Germany. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = MKJ and Original Applicant = MEDTRONIC PHYSIO-CONTROL CORP.
|
|
|
|