 Class 2 Device Recall IsantiGliadin IgG Enzyme Immunoassay Test Kitsee related information
Date Initiated by FirmMarch 21, 2008
Date PostedSeptember 16, 2008
Recall Status1 Terminated 3 on February 11, 2009
Recall NumberZ-1684-2008
Recall Event ID 47430
510(K)NumberK012795 
Product Classification Gliadin Antibodies - Product Code MST
ProductIs-anti-Gliadin IgG Enzyme Immunoassay Test Kit packaged in cardboard boxes with individual components secured within a foam insert. Catalog Number # 720-800. - Laboratory In-Vitro Diagnostic Assay intended for the semi-quantitative detection of IgG antibodies to Gliadin in human serum by indirect enzyme immunoassay as an aid in the diagnosis of celiac disease.
Code Information Catalog Number # 720-800 -- Lot # S0207, Expiration Date: Nov. 30, 2008.
FEI Number 1044713
Recalling Firm/
Manufacturer
Diamedix Corporation
2140 N. Miami Avenue
Miami FL 33127
For Additional Information Contact
305-324-2300
Manufacturer Reason
for Recall
Standards and controls in the Is-anti-Gliadin IgG Test Kits are exhibiting low activity. A drop in the O.D.s the Positive Control could potentially recover low out of range (which would invalidate the test), or false positive test results could be obtained.
FDA Determined
Cause 2
Environmental control
ActionA Customer Notification Letter was sent to each customer via Certified Mail for domestic customers and via Federal Express for international customer requesting them to contact Diamedix Technical Services to arrange for the return and replacement of any kits of the lot in question. If you have any questions please contact Glenn Gerstenfeld at 1-800-327-4565 or (305) 324-2300.
Quantity in Commerce174 Test Kits
DistributionClass II Recall - Worldwide Distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = MST
-
-