| Date Initiated by Firm | March 21, 2008 |
| Date Posted | September 16, 2008 |
| Recall Status1 |
Terminated 3 on February 11, 2009 |
| Recall Number | Z-1684-2008 |
| Recall Event ID |
47430 |
| 510(K)Number | K012795 |
| Product Classification |
Gliadin Antibodies - Product Code MST
|
| Product | Is-anti-Gliadin IgG Enzyme Immunoassay Test Kit packaged in cardboard boxes with individual components secured within a foam insert. Catalog Number # 720-800. - Laboratory In-Vitro Diagnostic Assay intended for the semi-quantitative detection of IgG antibodies to Gliadin in human serum by indirect enzyme immunoassay as an aid in the diagnosis of celiac disease. |
| Code Information |
Catalog Number # 720-800 -- Lot # S0207, Expiration Date: Nov. 30, 2008. |
| FEI Number |
1044713
|
Recalling Firm/ Manufacturer |
Diamedix Corporation
2140 N. Miami Avenue
Miami
FL
33127
|
| For Additional Information Contact | 305-324-2300 |
Manufacturer Reason for Recall | Standards and controls in the Is-anti-Gliadin IgG Test Kits are exhibiting low activity. A drop in the O.D.s the Positive Control could potentially recover low out of range (which would invalidate the test), or false positive test results could be obtained. |
FDA Determined Cause 2 | Environmental control |
| Action | A Customer Notification Letter was sent to each customer via Certified Mail for domestic customers and via Federal Express for international customer requesting them to contact Diamedix Technical Services to arrange for the return and replacement of any kits of the lot in question. If you have any questions please contact Glenn Gerstenfeld at 1-800-327-4565 or (305) 324-2300. |
| Quantity in Commerce | 174 Test Kits |
| Distribution | Class II Recall - Worldwide Distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|