|
Class 2 Device Recall Command Module |
|
Date Initiated by Firm |
July 22, 2008 |
Date Posted |
September 19, 2008 |
Recall Status1 |
Terminated 3 on July 25, 2011 |
Recall Number |
Z-2343-2008 |
Recall Event ID |
48759 |
510(K)Number |
K972502
|
Product Classification |
Arrhythmia Detector and alarm - Product Code DSI
|
Product |
Ultraview SL Command Module, Model 91496 The product is intended for use with the Patient Care Monitoring System (PCMS) to acquire, monitor, and process various clinical parameters from an adult or neonate/infant populations in any type of clinical environment other than home use. Physiological parameters that may be monitored include cardiac activity, respiration, invasive and noninvasive pressure, temperature, oxygen saturation (SpO2) and cardiac output. Acquired data may then be communicated to an information network for display, recording, editing and analysis.
|
Code Information |
Serial Numbers 1496-000016, 1496-000018, 1496-000059, 1496-000062, 1496-000065, 1496-000066, 1496-000069, 1496-000092, 1496-000137, 1496-000138, 1496-000140, 1496-000143, 1496-000145, 1496-000148, 1496-000149, 1496-000151, 1496-000152, 1496-000153, 1496-000154, 1496-000155, 1496-000156, 1496-000157, 1496-000158, 1496-000159, 1496-000160, 1496-000161, 1496-000162, 1496-000163, 1496-000166, 1496-000167, 1496-000168, 1496-000169, 1496-000170, 1496-000171, 1496-000172, 1496-000173, 1496-000204, 1496-000205, 1496-000206, 1496-000207, 1496-000208, 1496-000210, 1496-000211, 1496-000212, 1496-000213, 1496-000214, 1496-000215, 1496-000216, 1496-000217, 1496-000218, 1496-000219, 1496-000220, 1496-000221, 1496-000222, 1496-000223, 1496-000224, 1496-000225, 1496-000226, 1496-000227, 1496-000228, 1496-000229, 1496-000230, 1496-000231, 1496-000232, 1496-000233, 1496-000234, 1496-000235, 1496-000236, 1496-000237, 1496-000238, 1496-000239, 1496-000240, 1496-000243, 1496-000244, 1496-000245, 1496-000246, 1496-000247, 1496-000248, 1496-000249, 1496-000250, 1496-000251, 1496-000253, 1496-000254, 1496-000255, 1496-000256, 1496-000257, 1496-000258, 1496-000265, 1496-000266, 1496-000267, 1496-000268, 1496-000269, 1496-000274, 1496-000276, 1496-000277, 1496-000278, 1496-000279, 1496-000289, 1496-000438, 1496-000460, 1496-000461, 1496-000465, 1496-000474, 1496-000476, 1496-000499, 1496-000500, 1496-000501, 1496-000502, 1496-000503, 1496-000504, 1496-000546, 1496-000584, 1496-000585, 1496-000586, 1496-000588, 1496-000589, 1496-000591, 1496-000607, 1496-000624, 1496-000625, 1496-000626, 1496-000627, 1496-000628, 1496-000629, 1496-000630, 1496-000631, 1496-000632, 1496-000633, 1496-000634, 1496-000635, 1496-000697, 1496-000698, 1496-000699, 1496-000700, 1496-000702, 1496-000703, 1496-000721, 1496-000722, 1496-000724, 1496-000725, 1496-000726, 1496-000727, 1496-000728, 1496-000736, 1496-000747, 1496-000748, 1496-000749, 1496-000755, 1496-000759, 1496-000761, 1496-000762, 1496-000763, 1496-000764, 1496-000765, 1496-000766, 1496-000767, 1496-000768, 1496-000769, 1496-000770, 1496-000771, 1496-000782, 1496-000783, 1496-000784, 1496-000785, 1496-000786, 1496-000787, 1496-000788, 1496-000789, 1496-000790, 1496-000791, 1496-000792, 1496-000793, 1496-000794, 1496-000806, 1496-000807, 1496-000813, 1496-000816, 1496-000836, 1496-000840, 1496-000841, 1496-000842, 1496-000858, 1496-000859, 1496-000886, 1496-000890, 1496-000891, 1496-000892, 1496-000893, 1496-000913, 1496-000919, 1496-000920, 1496-000921, 1496-000922, 1496-000923, 1496-000924, 1496-000925, 1496-000926, 1496-000927, 1496-000928, 1496-000929, 1496-000930, 1496-000931, 1496-000932, 1496-000933, 1496-000934, 1496-000935, 1496-000936, 1496-000937, 1496-000938, 1496-000939, 1496-000940, 1496-000941, 1496-000942, 1496-000943, 1496-000944, 1496-000946, 1496-000948, 1496-000949, 1496-000950, 1496-000951, 1496-000952, 1496-000953, 1496-000954, 1496-000955, 1496-000956, 1496-000957, 1496-000958, 1496-000959, 1496-000961, 1496-000962, 1496-000963, 1496-000970, 1496-000971, 1496-000972, 1496-000981, 1496-000990, 1496-000994, 1496-000996, 1496-000997, 1496-001098, 1496-001102, 1496-001103, 1496-001104, 1496-001105, 1496-001106, 1496-001107, 1496-001108, 1496-001109, 1496-001110, 1496-001112, 1496-001113, 1496-001114, 1496-001115, 1496-001116, 1496-001117, 1496-001118, 1496-001122, 1496-001123, 1496-001124, 1496-001125, 1496-001127, 1496-001128, 1496-001129, 1496-001149, 1496-001150, 1496-001151, 1496-001154, 1496-001155, 1496-001160, 1496-001161, 1496-001162, 1496-001163, 1496-001164, 1496-001165, 1496-001166, 1496-001167, 1496-001168, 1496-001169, 1496-001170, 1496-001171, 1496-001172, 1496-001173, 1496-001174, 1496-001175, 1496-001176, 1496-001177, 1496-001178, 1496-001179, 1496-001180, 1496-001181, 1496-001182, 1496-001183, 1496-001184, 1496-001186, 1496-001187, 1496-001188, 1496-001190, 1496-001191, 1496-001192, 1496-001193, 1496-001194, 1496-001196, 1496-001197, 1496-001198, 1496-001199, 1496-001200, 1496-001201, 1496-001210, 1496-001213, 1496-001225, 1496-001228, 1496-001245, 1496-001246, 1496-001247, 1496-001248, 1496-001249, 1496-001250, 1496-001251, 1496-001252, 1496-001255, 1496-001276, 1496-001283, 1496-001284, 1496-001285, 1496-001286, 1496-001287, 1496-001288, 1496-001289, 1496-001290, 1496-001291, 1496-001292, 1496-001293, 1496-001294, 1496-001295, 1496-001296, 1496-001297, 1496-001298, 1496-001310, 1496-001311, 1496-001312, 1496-001313, 1496-001314, 1496-001315, 1496-001316, 1496-001317, 1496-001318, 1496-001319, 1496-001320, 1496-001321, 1496-001322, 1496-001323, 1496-001324, 1496-001325, 1496-001326, 1496-001334, 1496-001344, 1496-001345, 1496-001346, 1496-001347, 1496-001348, 1496-001349, 1496-001350, 1496-001351, 1496-001352, 1496-001356, 1496-001357, 1496-001358, 1496-001360, 1496-001361, 1496-001363, 1496-001364, 1496-001365, 1496-001366, 1496-001367, 1496-001369, 1496-001370, 1496-001371, 1496-001372, 1496-001373, 1496-001374, 1496-001375, 1496-001376, 1496-001377, 1496-001378, 1496-001379, 1496-001380, 1496-001382, 1496-001383, 1496-001384, 1496-001385, 1496-001386, 1496-001387, 1496-001388, 1496-001390, 1496-001431, 1496-001436, 1496-001438, 1496-001444, 1496-001445, 1496-001446, 1496-001447, 1496-001448, 1496-001449, 1496-001450, 1496-001451, 1496-001453, 1496-001454, 1496-001455, 1496-001456, 1496-001457, 1496-001458, 1496-001459, 1496-001460, 1496-001462, 1496-001463, 1496-001468, 1496-001469, 1496-001470, 1496-001471, 1496-001472, 1496-001473, 1496-001474, 1496-001475, 1496-001476, 1496-001477, 1496-001478, 1496-001479, 1496-001480, 1496-001481, 1496-001482, 1496-001483, 1496-001484, 1496-001485, 1496-001486, 1496-001487, 1496-001488, 1496-001489, 1496-001490, 1496-001491, 1496-001492, 1496-001493, 1496-001494, 1496-001495, 1496-001496, 1496-001497, 1496-001498, 1496-001499, 1496-001500, 1496-001501, 1496-001502, 1496-001503, 1496-001505, 1496-001506, 1496-001507, 1496-001508, 1496-001509, 1496-001510, 1496-001511, 1496-001512, 1496-001513, 1496-001514, 1496-001515, 1496-001516, 1496-001517, 1496-001518, 1496-001519, 1496-001520, 1496-001533, 1496-001534, 1496-001535, 1496-001536, 1496-001537, 1496-001538, 1496-001539, 1496-001540, 1496-001542, 1496-001543, 1496-001544, 1496-001545, 1496-001546, 1496-001547, 1496-001548, 1496-001549, 1496-001550, 1496-001552, 1496-001553, 1496-001554, 1496-001555, 1496-001556, 1496-001557, 1496-001558, 1496-001559, 1496-001560, 1496-001570, 1496-001571, 1496-001572, 1496-001573, 1496-001575, 1496-001576, 1496-001577, 1496-001578, 1496-001579, 1496-001580, 1496-001581, 1496-001598, 1496-001599, 1496-001602, 1496-001603, 1496-001604, 1496-001605, 1496-001606, 1496-001612, 1496-001631, 1496-001632, 1496-001633, 1496-001634, 1496-001635, 1496-001636, 1496-001637, 1496-001638, 1496-001639, 1496-001640, 1496-001641, 1496-001642, 1496-001643, 1496-001644, 1496-001645, 1496-001646, 1496-001647, 1496-001648, 1496-001650, 1496-001651, 1496-001659, 1496-001660, 1496-001661, 1496-001662, 1496-001663, 1496-001668, 1496-001671, 1496-001690, 1496-001697, 1496-001699, 1496-001700, 1496-001701, 1496-001703, 1496-001705, 1496-001708, 1496-001709, 1496-001715, 1496-001716, 1496-001717, 1496-001718, 1496-001719, 1496-001720, 1496-001721, 1496-001722, 1496-001723, 1496-001724, 1496-001725, 1496-001726, 1496-001728, 1496-001729, 1496-001730, 1496-001731, 1496-001732, 1496-001733, 1496-001737, 1496-001740, 1496-001743, 1496-001745, 1496-001746, 1496-001747, 1496-001748, 1496-001749, 1496-001750, 1496-001751, 1496-001752, 1496-001753, 1496-001757, 1496-001759, 1496-001760, 1496-001761, 1496-001762, 1496-001763, 1496-001764, 1496-001765, 1496-001766, 1496-001767, 1496-001768, 1496-001769, 1496-001770, 1496-001771, 1496-001772, 1496-001773, 1496-001774, 1496-001775, 1496-001776, 1496-001777, 1496-001778, 1496-001779, 1496-001780, 1496-001781, 1496-001782, 1496-001783, 1496-001784, 1496-001785, 1496-001786, 1496-001787, 1496-001788, 1496-001789, 1496-001791, 1496-001792, 1496-001793, 1496-001795.1496-001796, 1496-001799, 1496-001800, 1496-001801, 1496-001803, 1496-001804, 1496-001814, 1496-001815, 1496-001816, 1496-001817, 1496-001818, 1496-001819, 1496-001820, 1496-001821, 1496-001822, 1496-001823, 1496-001824, 1496-001825, 1496-001826, 1496-001827, 1496-001828, 1496-001829, 1496-001830, 1496-001831, 1496-001832, 1496-001833, 1496-001834, 1496-001843, 1496-001844, 1496-001850, 1496-001851, 1496-001892, 1496-001893, 1496-001900, 1496-001901, 1496-001902, 1496-001903, 1496-001904, 1496-001908, 1496-001909, 1496-001910, 1496-001911, 1496-001915, 1496-001916, 1496-001917, 1496-001918, 1496-001919, 1496-001920, 1496-001921, 1496-001922, 1496-001924, 1496-001930, 1496-001931, 1496-001932, 1496-001933, 1496-001934, 1496-001935, 1496-001937, 1496-001938, 1496-001939, 1496-001940, 1496-001941, 1496-001964, 1496-001977, 1496-001978, 1496-001979, 1496-001980, 1496-001981, 1496-001982, 1496-002044, 1496-002045, 1496-002046, 1496-002047, 1496-002048, 1496-002049, 1496-002050, 1496-002051, 1496-002052, 1496-002053, 1496-002054, 1496-002055, 1496-002056, 1496-002057, 1496-002151, 1496-002153, 1496-002154, 1496-002155, 1496-002182, 1496-002183, 1496-002184, 1496-002186, 1496-002199, 1496-002238, 1496-002245, 1496-002268, 1496-002269, 1496-002270, 1496-002271, 1496-002272, 1496-002273, 1496-002275, 1496-002277, 1496-002278, 1496-002279, 1496-002280, 1496-002281, 1496-002300, 1496-002301, 1496-002302, 1496-002303, 1496-002304, 1496-002305, 1496-002325, 1496-002326, 1496-002351, 1496-002471, 1496-002472, 1496-002558, 1496-002589, 1496-002599, 1496-002606, 1496-002609, 1496-002610, 1496-002611, 1496-002612, 1496-002613, 1496-002614, 1496-002615, 1496-002616, 1496-002617, 1496-002618, 1496-002619, 1496-002620, 1496-002621, 1496-002622, 1496-002623, 1496-002624, 1496-002625, 1496-002626, 1496-002627, 1496-002628, 1496-002629, 1496-002630, 1496-002631, 1496-002632, 1496-002633, 1496-002634, 1496-002635, 1496-002636, 1496-002637, 1496-002638, 1496-002639, 1496-002640, 1496-002641, 1496-002642, 1496-002643, 1496-002644, 1496-002645, 1496-002646, 1496-002647, 1496-002648, 1496-002649, 1496-002650, 1496-002652, 1496-002653, 1496-002654, 1496-002655, 1496-002657, 1496-002658, 1496-002659, 1496-002660, 1496-002661, 1496-002662, 1496-002665, 1496-002666, 1496-002667, 1496-002668, 1496-002671, 1496-002672, 1496-002673, 1496-002674, 1496-002699, 1496-002700, 1496-002701, 1496-002702, 1496-002704, 1496-002742, 1496-002744, 1496-002745, 1496-002746, 1496-002747, 1496-002748, 1496-002749, 1496-002750, 1496-002751, 1496-002755, 1496-002756, 1496-002757, 1496-002758, 1496-002759, 1496-002760, 1496-002761, 1496-002762, 1496-002763, 1496-002764, 1496-002765, 1496-002766, 1496-002767, 1496-002768, 1496-002769, 1496-002771, 1496-002776, 1496-002777, 1496-002778, 1496-002784, 1496-002808, 1496-002809, 1496-002810, 1496-002811, 1496-002812, 1496-002813, 1496-002842, 1496-002874, 1496-002875, 1496-002906, 1496-002907, 1496-002908, 1496-002909, 1496-002910, 1496-002911, 1496-002912, 1496-002913, 1496-002914, 1496-002915, 1496-002916, 1496-002943, 1496-002945, 1496-002946, 1496-002947, 1496-002948, 1496-002949, 1496-002950, 1496-002951, 1496-002952, 1496-002953, 1496-002954, 1496-002955, 1496-002956, 1496-002957, 1496-002965, 1496-002966, 1496-002967, 1496-002988, 1496-003008, 1496-003009, 1496-003010, 1496-003011, 1496-003012, 1496-003013, 1496-003014, 1496-003015, 1496-003016, 1496-003017, 1496-003054, 1496-003079, 1496-003080, 1496-003081, 1496-003082, 1496-003083, 1496-003084, 1496-003085, 1496-003086, 1496-003087, 1496-003088, 1496-003103, 1496-003104, 1496-003105, 1496-003106, 1496-003107, 1496-003108, 1496-003109, 1496-003110, 1496-003111, 1496-003112, 1496-003113, 1496-003114, 1496-003115, 1496-003116, 1496-003117, 1496-003119, 1496-003121, 1496-003125, 1496-003126, 1496-003127, 1496-003128, 1496-003129, 1496-003130, 1496-003131, 1496-003132, 1496-003133, 1496-003134, 1496-003142, 1496-003143, 1496-003144, 1496-003145, 1496-003146, 1496-003147, 1496-003148, 1496-003149, 1496-003153, 1496-003154, 1496-003173, 1496-003174, 1496-003185, 1496-003203, 1496-003204, 1496-003205, 1496-003206, 1496-003207, 1496-003208, 1496-003209, 1496-003212, 1496-003213, 1496-003214, 1496-003216, 1496-003217, 1496-003218, 1496-003219, 1496-003220, 1496-003221, 1496-003263, 1496-003264, 1496-003265, 1496-003266, 1496-003267, 1496-003268, 1496-003269, 1496-003270, 1496-003271, 1496-003272, 1496-003273, 1496-003274, 1496-003275, 1496-003276, 1496-003277, 1496-003278, 1496-003279, 1496-003280, 1496-003282, 1496-003283, 1496-003284, 1496-003285, 1496-003298, 1496-003299, 1496-003300, 1496-003301, 1496-003302, 1496-003303, 1496-003304, 1496-003305, 1496-003321, 1496-003322, 1496-003341, 1496-003346, 1496-003347, 1496-003415, 1496-003417, 1496-003418, 1496-003422, 1496-003423, 1496-003424, 1496-003425, 1496-003426, 1496-003427, 1496-003428, 1496-003429, 1496-003430, 1496-003431, 1496-003432, 1496-003433, 1496-003449, 1496-003450, 1496-003451, 1496-003452, 1496-003453, 1496-003454, 1496-003455, 1496-003456, 1496-003457, 1496-003458, 1496-003459, 1496-003460, 1496-003461, 1496-003462, 1496-003463, 1496-003467, 1496-003533, 1496-003534, 1496-003535, 1496-003536, 1496-003537, 1496-003539, 1496-003540, 1496-003541, 1496-003554, 1496-003586, 1496-003604, 1496-003610, 1496-003612, 1496-003613, 1496-003615, 1496-003616, 1496-003617, 1496-003618, 1496-003619, 1496-003647, 1496-003668, 1496-003669, 1496-003670, 1496-003671, 1496-003672, 1496-003707, 1496-003708, 1496-003709, 1496-003710, 1496-003711, 1496-003712, 1496-003713, 1496-003714, 1496-003715, 1496-003716, 1496-003717, 1496-003718, 1496-003719, 1496-003720, 1496-003721, 1496-003722, 1496-003723, 1496-003724, 1496-003725, 1496-003726, 1496-003727, 1496-003730, 1496-003749, 1496-003769, 1496-003772, 1496-003773, 1496-003774, 1496-003777, 1496-003778, 1496-003779, 1496-003780, 1496-003781, 1496-003782, 1496-003783, 1496-003784, 1496-003789, 1496-003790, 1496-003791, 1496-003792, 1496-003826, 1496-003853, 1496-003854, 1496-003855, 1496-003856, 1496-003857, 1496-003858, 1496-003859, 1496-003860, 1496-003861, 1496-003862, 1496-003863, 1496-004025, 1496-004026, 1496-004038, 1496-004039, 1496-004040, 1496-004041, 1496-004042, 1496-004043, 1496-004044, 1496-004045, 1496-004046, 1496-004047, 1496-004048, 1496-004049, 1496-004050, 1496-004051, 1496-004052, 1496-004053, 1496-004054, 1496-004055, 1496-004073, 1496-004119, 1496-004120, 1496-004170, 1496-004171, 1496-004176, 1496-004183, 1496-004184, 1496-004189, 1496-004190, 1496-004195, 1496-004196, 1496-004197, 1496-004198, 1496-004199, 1496-004200, 1496-004201, 1496-004202, 1496-004203, 1496-004204, 1496-004205, 1496-004214, 1496-004215, 1496-004216, 1496-004217, 1496-004218, 1496-004246, 1496-004247, 1496-004248, 1496-004249, 1496-004250, 1496-004251, 1496-004304, 1496-004407, 1496-004409, 1496-004410, 1496-004411, 1496-004412, 1496-004419, 1496-004420, 1496-004421, 1496-004422, 1496-004423, 1496-004424, 1496-004425, 1496-004468, 1496-004538, 1496-004565, 1496-004566, 1496-004567, 1496-004568, 1496-004597, 1496-004654, 1496-004656, 1496-004716, 1496-004717, 1496-004719, 1496-004720, 1496-004721, 1496-004745, 1496-004755, 1496-004764, 1496-004765, 1496-004766, 1496-004767, 1496-004768, 1496-004769, 1496-004770, 1496-004771, 1496-004772, 1496-004773, 1496-004774, 1496-004775, 1496-004776, 1496-004777, 1496-004778, 1496-004786, 1496-004787, 1496-004788, 1496-004805, 1496-004806, 1496-004813, 1496-004814, 1496-004815, 1496-004816, 1496-004817, 1496-004818, 1496-004824, 1496-004844, 1496-004845, 1496-004846, 1496-004847, 1496-004848, 1496-004849, 1496-004850, 1496-004851, 1496-004852, 1496-004853, 1496-004854, 1496-004855, 1496-004856, 1496-004857, 1496-004858, 1496-004859, 1496-004860, 1496-004861, 1496-004862, 1496-004863, 1496-004864, 1496-004865, 1496-004866, 1496-004867, 1496-004875, 1496-004880, 1496-004893, 1496-004912, 1496-004913, 1496-004914, 1496-004915, 1496-004932, 1496-004933, 1496-004934, 1496-004939, 1496-004940, 1496-004948, 1496-004949, 1496-004982, 1496-004983, 1496-004985, 1496-004987, 1496-004988, 1496-004989, 1496-005014, 1496-005015, 1496-005018, 1496-005019, 1496-005021, 1496-005022, 1496-005023, 1496-005024, 1496-005025, 1496-005026, 1496-005027, 1496-005031, 1496-005032, 1496-005033, 1496-005034, 1496-005035, 1496-005038, 1496-005039, 1496-005040, 1496-005041, 1496-005042, 1496-005043, 1496-005044, 1496-005045, 1496-005046, 1496-005047, 1496-005048, 1496-005049, 1496-005050, 1496-005051, 1496-005052, 1496-005053, 1496-005054, 1496-005055, 1496-005056, 1496-005057, 1496-005058, 1496-005059, 1496-005060, 1496-005061, 1496-005062, 1496-005063, 1496-005064, 1496-005065, 1496-005066, 1496-005079, 1496-005080, 1496-005081, 1496-005082, 1496-005083, 1496-005094, 1496-005095, 1496-005096, 1496-005110, 1496-005134, 1496-005135, 1496-005136, 1496-005137, 1496-005138, 1496-005139, 1496-005140, 1496-005141, 1496-005142, 1496-005143, 1496-005144, 1496-005145, 1496-005146, 1496-005147, 1496-005148, 1496-005155, 1496-005158, 1496-005159, 1496-005162, 1496-005163, 1496-005164, 1496-005165, 1496-005166, 1496-005167, 1496-005168, 1496-005169, 1496-005171, 1496-005172, 1496-005173, 1496-005174, 1496-005185, 1496-005190, 1496-005191, 1496-005192, 1496-005193, 1496-005210, 1496-005212, 1496-005213, 1496-005214, 1496-005215, 1496-005216, 1496-005217, 1496-005218, 1496-005219, 1496-005220, 1496-005221, 1496-005222, 1496-005223, 1496-005249, 1496-005250, 1496-005253, 1496-005277, 1496-005290, 1496-005354, 1496-005417, 1496-005418, 1496-005419, 1496-005420, 1496-005421, 1496-005430, 1496-005459, 1496-005461, 1496-005467, 1496-005489, 1496-005490, 1496-005495, 1496-005503, 1496-005504, 1496-005508, 1496-005509, 1496-005510, 1496-005511, 1496-005516, 1496-005517, 1496-005518, 1496-005519, 1496-005520, 1496-005522, 1496-005523, 1496-005524, 1496-005525, 1496-005526, 1496-005527, 1496-005528, 1496-005529, 1496-005530, 1496-005531, 1496-005532, 1496-005533, 1496-005534, 1496-005535, 1496-005536, 1496-005537, 1496-005538, 1496-005539, 1496-005540, 1496-005541, 1496-005542, 1496-005543, 1496-005544, 1496-005545, 1496-005546, 1496-005547, 1496-005548, 1496-005549, 1496-005550, 1496-005551, 1496-005552, 1496-005553, 1496-005554, 1496-005559, 1496-005560, 1496-005561, 1496-005562, 1496-005563, 1496-005564, 1496-005565, 1496-005566, 1496-005567, 1496-005568, 1496-005569, 1496-005570, 1496-005571, 1496-005572, 1496-005573, 1496-005575, 1496-005576, 1496-005578, 1496-005579, 1496-005580, 1496-005581, 1496-005582, 1496-005584, 1496-005586, 1496-005587, 1496-005588, 1496-005589, 1496-005592, 1496-005593, 1496-005594, 1496-005595, 1496-005596, 1496-005597, 1496-005598, 1496-005599, 1496-005600, 1496-005605, 1496-005606, 1496-005608, 1496-005609, 1496-005610, 1496-005631, 1496-005632, 1496-005633, 1496-005634, 1496-005635, 1496-005636, 1496-005637, 1496-005638, 1496-005639, 1496-005640, 1496-005641, 1496-005642, 1496-005643, 1496-005644, 1496-005645, 1496-005646, 1496-005647, 1496-005648, 1496-005649, 1496-005659, 1496-005660, 1496-005661, 1496-005662, 1496-005663, 1496-005664, 1496-005665, 1496-005666, 1496-005667, 1496-005670, 1496-005684, 1496-005694, 1496-005695, 1496-005696, 1496-005697, 1496-005698, 1496-005699, 1496-005700, 1496-005701, 1496-005763, 1496-005764, 1496-005765, 1496-005766, 1496-005767, 1496-005768, 1496-005769, 1496-005770, 1496-005771, 1496-005772, 1496-005773, 1496-005774, 1496-005775, 1496-005776, 1496-005777, 1496-005795, 1496-005813, 1496-005814, 1496-005815, 1496-005816, 1496-005817, 1496-005818, 1496-005819, 1496-005820, 1496-005821, 1496-005822, 1496-005823, 1496-005824, 1496-005825, 1496-005826, 1496-005827, 1496-005828, 1496-005829, 1496-005830, 1496-005831, 1496-005832, 1496-005833, 1496-005834, 1496-005835, 1496-005836, 1496-005838, 1496-005839, 1496-005840, 1496-005862, 1496-005871, 1496-005872, 1496-005873, 1496-005874, 1496-005875, 1496-005876, 1496-005877, 1496-005878, 1496-005882, 1496-005883, 1496-005884, 1496-005885, 1496-005886, 1496-005887, 1496-005888, 1496-005935, 1496-005936, 1496-005940, 1496-005941, 1496-005942, 1496-005943, 1496-005944, 1496-005945, 1496-005946, 1496-005947, 1496-005948, 1496-005949, 1496-005950, 1496-005951, 1496-005981, 1496-006011, 1496-006013, 1496-006014, 1496-006015, 1496-006042, 1496-006043, 1496-006044, 1496-006045, 1496-006046, 1496-006047, 1496-006060, 1496-006070, 1496-006073, 1496-006080, 1496-006091, 1496-006092, 1496-006093, 1496-006094, 1496-006095, 1496-006096, 1496-006114, 1496-006115, 1496-006116, 1496-006137, 1496-006138, 1496-006145, 1496-006146, 1496-006147, 1496-006148, 1496-006149, 1496-006355, 1496-006356, 1496-006357, 1496-006372, 1496-006375, 1496-006383, 1496-006393, 1496-006394, 1496-006395, 1496-006396, 1496-006399, 1496-006411, 1496-006433, 1496-006451, 1496-006460, 1496-006485, 1496-006486, 1496-006487, 1496-006488, 1496-006489, 1496-006490, 1496-006491, 1496-006499, 1496-006500, 1496-006501, 1496-006502, 1496-006518, 1496-006519, 1496-006520, 1496-006521, 1496-006522, 1496-006524, 1496-006531, 1496-006532, 1496-006533, 1496-006534, 1496-006535, 1496-006536, 1496-006537, 1496-006538, 1496-006539, 1496-006540, 1496-006541, 1496-006542, 1496-006545, 1496-006561, 1496-006562, 1496-006563, 1496-006564, 1496-006565, 1496-006566, 1496-006567, 1496-006568, 1496-006569, 1496-006574, 1496-006598, 1496-006617, 1496-006618, 1496-006619, 1496-006620, 1496-006626, 1496-006627, 1496-006682, 1496-006683, 1496-006684, 1496-006685, 1496-006687, 1496-006688, 1496-006689, 1496-006690, 1496-006691, 1496-006692, 1496-006693, 1496-006694, 1496-006695, 1496-006696, 1496-006697, 1496-006699, 1496-006700, 1496-006701, 1496-006702, 1496-006703, 1496-006704, 1496-006705, 1496-006813, 1496-006814, 1496-006815, 1496-006816, 1496-006817, 1496-006818, 1496-006819, 1496-006820, 1496-006821, 1496-006822, 1496-006823, 1496-006824, 1496-006825, 1496-006826, 1496-006827, 1496-006834, 1496-006857, 1496-006858, 1496-006859, 1496-006860, 1496-006861, 1496-006862, 1496-006863, 1496-006872, 1496-006873, 1496-006874, 1496-006875, 1496-006884, 1496-006894, 1496-006901, 1496-006902, 1496-006903, 1496-006904, 1496-006905, 1496-006917, 1496-006918, 1496-006919, 1496-006920, 1496-006921, 1496-006922, 1496-006923, 1496-006926, 1496-006927, 1496-006928, 1496-006938, 1496-006939, 1496-006959, 1496-006960, 1496-006961, 1496-006962, 1496-006964, 1496-006965, 1496-006966, 1496-006999, 1496-007000, 1496-007001, 1496-007002, 1496-007003, 1496-007004, 1496-007005, 1496-007006, 1496-007007, 1496-007009, 1496-007010, 1496-007011, 1496-007012, 1496-007013, 1496-007014, 1496-007015, 1496-007016, 1496-007017, 1496-007018, 1496-007019, 1496-007020, 1496-007021, 1496-007022, 1496-007023, 1496-007024, 1496-007025, 1496-007026, 1496-007027, 1496-007028, 1496-007029, 1496-007030, 1496-007042, 1496-007053, 1496-007054, 1496-007055, 1496-007056, 1496-007057, 1496-007058, 1496-007059, 1496-007060, 1496-007088, 1496-007098, 1496-007106, 1496-007115, 1496-007116, 1496-007117, 1496-007118, 1496-007119, 1496-007120, 1496-007121, 1496-007122, 1496-007132, 1496-007137, 1496-007141, 1496-007144, 1496-007146, 1496-007197, 1496-007198, 1496-007199, 1496-007200, 1496-007201, 1496-007202, 1496-007203, 1496-007204, 1496-007205, 1496-007206, 1496-007207, 1496-007208, 1496-007209, 1496-007210, 1496-007213, 1496-007214, 1496-007215, 1496-007216, 1496-007217, 1496-007218, 1496-007219, 1496-007220, 1496-007221, 1496-007222, 1496-007223, 1496-007224, 1496-007225, 1496-007226, 1496-007231, 1496-007232, 1496-007233, 1496-007234, 1496-007235, 1496-007236, 1496-007237, 1496-007239, 1496-007240, 1496-007241, 1496-007242, 1496-007243, 1496-007261, 1496-007262, 1496-007263, 1496-007264, 1496-007265, 1496-007266, 1496-007267, 1496-007268, 1496-007269, 1496-007270, 1496-007271, 1496-007272, 1496-007273, 1496-007274, 1496-007275, 1496-007276, 1496-007277, 1496-007278, 1496-007281, 1496-007282, 1496-007283, 1496-007284, 1496-007285, 1496-007286, 1496-007287, 1496-007288, 1496-007289, 1496-007290, 1496-007291, 1496-007292, 1496-007293, 1496-007294, 1496-007300, 1496-007301, 1496-007302, 1496-007303, 1496-007304, 1496-007305, 1496-007306, 1496-007307, 1496-007314, 1496-007315, 1496-007316, 1496-007317, 1496-007341, 1496-007342, 1496-007343, 1496-007344, 1496-007345, 1496-007346, 1496-007347, 1496-007348, 1496-007349, 1496-007352, 1496-007353, 1496-007354, 1496-007355, 1496-007356, 1496-007357, 1496-007359, 1496-007360, 1496-007361, 1496-007362, 1496-007363, 1496-007364, 1496-007365, 1496-007366, 1496-007367, 1496-007368, 1496-007369, 1496-007370, 1496-007371, 1496-007372, 1496-007380, 1496-007381, 1496-007382, 1496-007388, 1496-007389, 1496-007390, 1496-007391, 1496-007392, 1496-007393, 1496-007394, 1496-007395, 1496-007397, 1496-007398, 1496-007399, 1496-007400, 1496-007401, 1496-007402, 1496-007403, 1496-007404, 1496-007405, 1496-007406, 1496-007407, 1496-007408, 1496-007409, 1496-007410, 1496-007418, 1496-007419, 1496-007420, 1496-007452, 1496-007471, 1496-007472, 1496-007473, 1496-007474, 1496-007477, 1496-007478, 1496-007479, 1496-007481, 1496-007482, 1496-007483, 1496-007484, 1496-007485, 1496-007486, 1496-007487, 1496-007488, 1496-007489, 1496-007490, 1496-007491, 1496-007492, 1496-007493, 1496-007498, 1496-007499, 1496-007500, 1496-007501, 1496-007502, 1496-007503, 1496-007504, 1496-007505, 1496-007506, 1496-007513, 1496-007514, 1496-007515, 1496-007516, 1496-007517, 1496-007518, 1496-007519, 1496-007520, 1496-007521, 1496-007522, 1496-007523, 1496-007524, 1496-007525, 1496-007526, 1496-007527, 1496-007528, 1496-007529, 1496-007530, 1496-007531, 1496-007532, 1496-007533, 1496-007534, 1496-007535, 1496-007546, 1496-007547, 1496-007548, 1496-007549, 1496-007550, 1496-007551, 1496-007552, 1496-007553, 1496-007554, 1496-007555, 1496-007556, 1496-007557, 1496-007558, 1496-007559, 1496-007560, 1496-007561, 1496-007563, 1496-007564, 1496-007565, 1496-007566, 1496-007567, 1496-007568, 1496-007569, 1496-007570, 1496-007571, 1496-007572, 1496-007573, 1496-007574, 1496-007575, 1496-007576, 1496-007577, 1496-007578, 1496-007579, 1496-007580, 1496-007581, 1496-007582, 1496-007583, 1496-007584, 1496-007585, 1496-007602, 1496-007604, 1496-007605, 1496-007606, 1496-007607, 1496-007608, 1496-007609, 1496-007610, 1496-007611, 1496-007615, 1496-007616, 1496-007617, 1496-007618, 1496-007619, 1496-007620, 1496-007621, 1496-007622, 1496-007623, 1496-007624, 1496-007625, 1496-007626, 1496-007627, 1496-007634, 1496-007636, 1496-007637, 1496-007638, 1496-007646, 1496-007649, 1496-007682, 1496-007696, 1496-007697, 1496-007699, 1496-007701, 1496-007702, 1496-007703, 1496-007705, 1496-007707, 1496-007712, 1496-007724, 1496-007725, 1496-007726, 1496-007727, 1496-007758, 1496-007759, 1496-007760, 1496-007761, 1496-007762, 1496-007763, 1496-007764, 1496-007765, 1496-007773, 1496-007774, 1496-007775, 1496-007776, 1496-007778, 1496-007797, 1496-007800, 1496-007801, 1496-007819, 1496-007820, 1496-007821, 1496-007822, 1496-007823, 1496-007848, 1496-007852, 1496-007853, 1496-007854, 1496-007855, 1496-007856, 1496-007858, 1496-007863, 1496-007864, 1496-007865, 1496-007866, 1496-007867, 1496-007868, 1496-007889, 1496-007891, 1496-007892, 1496-007893, 1496-007894, 1496-007911, 1496-007912, 1496-007913, 1496-007914, 1496-007915, 1496-007916, 1496-007919, 1496-007920, 1496-007925, 1496-007942, 1496-007943, 1496-007944, 1496-007945, 1496-007946, 1496-007947, 1496-007948, 1496-007949, 1496-007950, 1496-007951, 1496-007953, 1496-007954, 1496-007980, 1496-007981, 1496-007989, 1496-007990, 1496-007991, 1496-007992, 1496-007993, 1496-007994, 1496-007995, 1496-007996, 1496-007997, 1496-007998, 1496-007999, 1496-008000, 1496-008001, 1496-008002, 1496-008003, 1496-008042, 1496-008043, 1496-008045, 1496-008062, 1496-008063, 1496-008064, 1496-008065, 1496-008066, 1496-008067, 1496-008068, 1496-008069, 1496-008070, 1496-008071, 1496-008072, 1496-008073, 1496-008074, 1496-008075, 1496-008076, 1496-008077, 1496-008078, 1496-008079, 1496-008080, 1496-008081, 1496-008082, 1496-008083, 1496-008084, 1496-008085, 1496-008086, 1496-008087, 1496-008088, 1496-008089, 1496-008090, 1496-008091, 1496-008092, 1496-008093, 1496-008094, 1496-008095, 1496-008096, 1496-008097, 1496-008098, 1496-008099, 1496-008100, 1496-008101, 1496-008102, 1496-008103, 1496-008104, 1496-008105, 1496-008106, 1496-008107, 1496-008108, 1496-008109, 1496-008110, 1496-008111, 1496-008112, 1496-008113, 1496-008114, 1496-008115, 1496-008116, 1496-008117, 1496-008120, 1496-008121, 1496-008122, 1496-008123, 1496-008124, 1496-008126, 1496-008127, 1496-008128, 1496-008129, 1496-008130, 1496-008131, 1496-008132, 1496-008133, 1496-008134, 1496-008135, 1496-008137, 1496-008138, 1496-008139, 1496-008140, 1496-008141, 1496-008142, 1496-008143, 1496-008144, 1496-008145, 1496-008146, 1496-008147, 1496-008148, 1496-008149, 1496-008150, 1496-008151, 1496-008152, 1496-008153, 1496-008154, 1496-008155, 1496-008156, 1496-008157, 1496-008158, 1496-008159, 1496-008160, 1496-008161, 1496-008162, 1496-008169, 1496-008170, 1496-008171, 1496-008172, 1496-008173, 1496-008174, 1496-008175, 1496-008176, 1496-008177, 1496-008178, 1496-008179, 1496-008180, 1496-008181, 1496-008182, 1496-008183, 1496-008184, 1496-008185, 1496-008186, 1496-008187, 1496-008188, 1496-008189, 1496-008190, 1496-008191, 1496-008192, 1496-008193, 1496-008194, 1496-008195, 1496-008196, 1496-008197, 1496-008198, 1496-008199, 1496-008200, 1496-008201, 1496-008202, 1496-008203, 1496-008204, 1496-008205, 1496-008206, 1496-008207, 1496-008208, 1496-008209, 1496-008210, 1496-008211, 1496-008212, 1496-008213, 1496-008214, 1496-008215, 1496-008216, 1496-008224, 1496-008225, 1496-008226, 1496-008227, 1496-008228, 1496-008229, 1496-008230, 1496-008231, 1496-008232, 1496-008233, 1496-008234, 1496-008235, 1496-008237, 1496-008238, 1496-008239, 1496-008240, 1496-008248, 1496-008251, 1496-008252, 1496-008253, 1496-008254, 1496-008255, 1496-008256, 1496-008257, 1496-008258, 1496-008259, 1496-008260, 1496-008261, 1496-008262, 1496-008263, 1496-008264, 1496-008265, 1496-008266, 1496-008267, 1496-008268, 1496-008269, 1496-008272, 1496-008330, 1496-008331, 1496-008332, 1496-008333, 1496-008334, 1496-008335, 1496-008336, 1496-008337, 1496-008338, 1496-008339, 1496-008340, 1496-008341, 1496-008342, 1496-008343, 1496-008344, 1496-008345, 1496-008399, 1496-008427, 1496-008428, 1496-008429, 1496-008431, 1496-008473, 1496-008475, 1496-008476, 1496-008477, 1496-008478, 1496-008479, 1496-008480, 1496-008481, 1496-008482, 1496-008484, 1496-008485, 1496-008486, 1496-008487, 1496-008488, 1496-008489, 1496-008490, 1496-008492, 1496-008493, 1496-008494, 1496-008506, 1496-008507, 1496-008508, 1496-008509, 1496-008510, 1496-008511, 1496-008512, 1496-008513, 1496-008514, 1496-008515, 1496-008516, 1496-008517, 1496-008518, 1496-008519, 1496-008520, 1496-008521, 1496-008522, 1496-008523, 1496-008524, 1496-008525, 1496-008526, 1496-008528, 1496-008565, 1496-008566, 1496-008567, 1496-008568, 1496-008569, 1496-008570, 1496-008571, 1496-008572, 1496-008573, 1496-008574, 1496-008575, 1496-008578, 1496-008579, 1496-008631, 1496-008632, 1496-008633, 1496-008634, 1496-008666, 1496-008697, 1496-008710, 1496-008714, 1496-008716, 1496-008717, 1496-008718, 1496-008719, 1496-008720, 1496-008721, 1496-008722, 1496-008723, 1496-008724, 1496-008725, 1496-008731, 1496-008732, 1496-008733, 1496-008734, 1496-008735, 1496-008736, 1496-008737, 1496-008738, 1496-008739, 1496-008740, 1496-008741, 1496-008742, 1496-008743, 1496-008744, 1496-008745, 1496-008746, 1496-008747, 1496-008748, 1496-008749, 1496-008750, 1496-008751, 1496-008752, 1496-008753, 1496-008754, 1496-008765, 1496-008766, 1496-008788, 1496-008812, 1496-008813, 1496-008814, 1496-008815, 1496-008816, 1496-008817, 1496-008818, 1496-008819, 1496-008820, 1496-008821, 1496-008822, 1496-008823, 1496-008824, 1496-008825, 1496-008826, 1496-008867, 1496-008868, 1496-008869, 1496-008870, 1496-008871, 1496-008872, 1496-008873, 1496-008874, 1496-008877, 1496-008878, 1496-008879, 1496-008880, 1496-008881, 1496-008882, 1496-008883, 1496-008884, 1496-008885, 1496-008886, 1496-008887, 1496-008888, 1496-008889, 1496-008890, 1496-008891, 1496-008892, 1496-008893, 1496-008894, 1496-008895, 1496-008896, 1496-008897, 1496-008898, 1496-008899, 1496-008900, 1496-008901, 1496-008902, 1496-008903, 1496-008904, 1496-008905, 1496-008906, 1496-008907, 1496-008908, 1496-008909, 1496-008910, 1496-008911, 1496-008912, 1496-008925, 1496-008926, 1496-008927, 1496-008928, 1496-008929, 1496-008944, 1496-008947, 1496-008948, 1496-008949, 1496-008950, 1496-008951, 1496-008952, 1496-008953, 1496-008954, 1496-008955, 1496-008956, 1496-008957, 1496-008958, 1496-008959, 1496-008960, 1496-008964, 1496-008965, 1496-008966, 1496-008967, 1496-008968, 1496-008969, 1496-008970, 1496-008971, 1496-008972, 1496-008973, 1496-008974, 1496-009139, 1496-009140, 1496-009185, 1496-009279, 1496-009280, 1496-009281, 1496-009282, 1496-009283, 1496-009284, 1496-009285, 1496-009286, 1496-009287, 1496-009288, 1496-009289, 1496-009290, 1496-009362, 1496-009363, 1496-009364, 1496-009365, 1496-009366, 1496-009367, 1496-009368, 1496-009369, 1496-009370, 1496-009371, 1496-009372, 1496-009373, 1496-009408, 1496-009409, 1496-009410, 1496-009411, 1496-009412, 1496-009413, 1496-009414, 1496-009415, 1496-009416, 1496-009417, 1496-009418, 1496-009419, 1496-009420, 1496-009421, 1496-009422, 1496-009423, 1496-009424, 1496-009425, 1496-009426, 1496-009427, 1496-009428, 1496-009429, 1496-009430, 1496-009431, 1496-009432, 1496-009433, 1496-009462, 1496-009463, 1496-009481, 1496-009482, 1496-009483, 1496-009484, 1496-009485, 1496-009486, 1496-009487, 1496-009488, 1496-009489, 1496-009491, 1496-009493, 1496-009494, 1496-009495, 1496-009496, 1496-009497, 1496-009544, 1496-009550, 1496-009551, 1496-009552, 1496-009553, 1496-009554, 1496-009555, 1496-009556, 1496-009571, 1496-009572, 1496-009573, 1496-009574, 1496-009575, 1496-009636, 1496-009642, 1496-009644, 1496-009645, 1496-009646, 1496-009647, 1496-009648, 1496-009649, 1496-009650, 1496-009651, 1496-009701, 1496-009702, 1496-009703, 1496-009704, 1496-009705, 1496-009706, 1496-009707, 1496-009708, 1496-009709, 1496-009711, 1496-009712, 1496-009713, 1496-009714, 1496-009715, 1496-009716, 1496-009717, 1496-009718, 1496-009719, 1496-009720, 1496-009721, 1496-009722, 1496-009723, 1496-009724, 1496-009725, 1496-009726, 1496-009727, 1496-009728, 1496-009729, 1496-009731, 1496-009733, 1496-009734, 1496-009735, 1496-009736, 1496-009737, 1496-009738, 1496-009739, 1496-009740, 1496-009741, 1496-009777, 1496-009790, 1496-009791, 1496-009798, 1496-009799, 1496-009800, 1496-009801, 1496-009802, 1496-009803, 1496-009804, 1496-009805, 1496-009807, 1496-009810, 1496-009811, 1496-009821, 1496-009825, 1496-009826, 1496-009827, 1496-009828, 1496-009829, 1496-009830, 1496-009896, 1496-009923, 1496-009924, 1496-009925, 1496-009926, 1496-009927, 1496-009932, 1496-009949, 1496-009951, 1496-009952, 1496-009954, 1496-009956, 1496-009957, 1496-009958, 1496-009959, 1496-009960, 1496-009961, 1496-009962, 1496-009964, 1496-009965, 1496-009967, 1496-009974, 1496-009975, 1496-009976, 1496-009986, 1496-009990, 1496-009994, 1496-009995, 1496-009996, 1496-009997, 1496-009998, 1496-010000, 1496-010001, 1496-010002, 1496-010004, 1496-010005, 1496-010006, 1496-010007, 1496-010046, 1496-010047, 1496-010048, 1496-010051, 1496-010052, 1496-010139, 1496-010140, 1496-010142, 1496-010143, 1496-010144, 1496-010146, 1496-010147, 1496-010162, 1496-010163, 1496-010164, 1496-010165, 1496-010166, 1496-010167, 1496-010168, 1496-010169, 1496-010170, 1496-010171, 1496-010172, 1496-010173, 1496-010174, 1496-010175, 1496-010176, 1496-010177, 1496-010178, 1496-010179, 1496-010180, 1496-010181, 1496-010182, 1496-010183, 1496-010184, 1496-010185, 1496-010186, 1496-010187, 1496-010205, 1496-010206, 1496-010207, 1496-010208, 1496-010209, 1496-010210, 1496-010211, 1496-010214, 1496-010215, 1496-010216, 1496-010217, 1496-010218, 1496-010219, 1496-010220, 1496-010221, 1496-010222, 1496-010223, 1496-010224, 1496-010225, 1496-010226, 1496-010227, 1496-010228, 1496-010229, 1496-010230, 1496-010231, 1496-010232, 1496-010233, 1496-010234, 1496-010235, 1496-010236, 1496-010237, 1496-010238, 1496-010239, 1496-010240, 1496-010241, 1496-010242, 1496-010243, 1496-010244, 1496-010245, 1496-010246, 1496-010247, 1496-010248, 1496-010249, 1496-010253, 1496-010254, 1496-010255, 1496-010256, 1496-010262, 1496-010263, 1496-010264, 1496-010265, 1496-010266, 1496-010267, 1496-010268, 1496-010269, 1496-010270, 1496-010308, 1496-010309, 1496-010310, 1496-010311, 1496-010312, 1496-010314, 1496-010315, 1496-010336, 1496-010402, 1496-010405, 1496-010406, 1496-010408, 1496-010409, 1496-010410, 1496-010411, 1496-010412, 1496-010413, 1496-010414, 1496-010415, 1496-010416, 1496-010417, 1496-010418, 1496-010419, 1496-010420, 1496-010421, 1496-010422, 1496-010431, 1496-010432, 1496-010433, 1496-010434, 1496-010435, 1496-010436, 1496-010437, 1496-010452, 1496-010453, 1496-010455, 1496-010456, 1496-010457, 1496-010458, 1496-010459, 1496-010460, 1496-010461, 1496-010462, 1496-010463, 1496-010464, 1496-010465, 1496-010481, 1496-010497, 1496-010498, 1496-010499, 1496-010500, 1496-010501, 1496-010502, 1496-010503, 1496-010504, 1496-010505, 1496-010506, 1496-010507, 1496-010508, 1496-010509, 1496-010510, 1496-010511, 1496-010512, 1496-010514, 1496-010515, 1496-010520, 1496-010521, 1496-010522, 1496-010599, 1496-010600, 1496-010601, 1496-010626, 1496-010666, 1496-010668, 1496-010669, 1496-010670, 1496-010671, 1496-010672, 1496-010673, 1496-010675, 1496-010680, 1496-010681, 1496-010682, 1496-010683, 1496-010684, 1496-010685, 1496-010686, 1496-010687, 1496-010688, 1496-010693, 1496-010700, 1496-010722, 1496-010723, 1496-010724, 1496-010727, 1496-010730, 1496-010731, 1496-010732, 1496-010733, 1496-010734, 1496-010736, 1496-010737, 1496-010739, 1496-010741, 1496-010743, 1496-010745, 1496-010746, 1496-010748, 1496-010749, 1496-010750, 1496-010751, 1496-010752, 1496-010762, 1496-010780, 1496-010781, 1496-010782, 1496-010783, 1496-010784, 1496-010785, 1496-010786, 1496-010788, 1496-010789, 1496-010790, 1496-010791, 1496-010792, 1496-010793, 1496-010800, 1496-010801, 1496-010802, 1496-010803, 1496-010812, 1496-010814, 1496-010815, 1496-010816, 1496-010817, 1496-010820, 1496-010836, 1496-010837, 1496-010838, 1496-010840, 1496-010841, 1496-010842, 1496-010843, 1496-010844, 1496-010845, 1496-010870, 1496-010871, 1496-010872, 1496-010873, 1496-010874, 1496-010875, 1496-010876, 1496-010877, 1496-010879, 1496-010994, 1496-011003, 1496-011013, 1496-011014, 1496-011028, 1496-011029, 1496-011030, 1496-011031, 1496-011032, 1496-011033, 1496-011034, 1496-011035, 1496-011036, 1496-011037, 1496-011038, 1496-011039, 1496-011040, 1496-011041, 1496-011042, 1496-011043, 1496-011044, 1496-011045, 1496-011048, 1496-011049, 1496-011050, 1496-011079, 1496-011080, 1496-011081, 1496-011086, 1496-011087, 1496-011089, 1496-011090, 1496-011091, 1496-011092, 1496-011093, 1496-011121, 1496-011122, 1496-011123, 1496-011129, 1496-011130, 1496-011131, 1496-011132, 1496-011133, 1496-011134, 1496-011135, 1496-011136, 1496-011137, 1496-011138, 1496-011139, 1496-011140, 1496-011141, 1496-011164, 1496-011165, 1496-011167, 1496-011168, 1496-011169, 1496-011170, 1496-011171, 1496-011172, 1496-011173, 1496-011174, 1496-011175, 1496-011176, 1496-011178, 1496-011179, 1496-011180, 1496-011181, 1496-011182, 1496-011183, 1496-011184, 1496-011185, 1496-011186, 1496-011187, 1496-011216, 1496-011217, 1496-011218, 1496-011261, 1496-011262, 1496-011263, 1496-011264, 1496-011265, 1496-011266, 1496-011267, 1496-011268, 1496-011269, 1496-011270, 1496-011271, 1496-011272, 1496-011273, 1496-011274, 1496-011275, 1496-011276, 1496-011277, 1496-011278, 1496-011279, 1496-011280, 1496-011281, 1496-011282, 1496-011283, 1496-011284, 1496-011285, 1496-011286, 1496-011316, 1496-011317, 1496-011319, 1496-011321, 1496-011323, 1496-011324, 1496-011325, 1496-011326, 1496-011327, 1496-011328, 1496-011329, 1496-011330, 1496-011331, 1496-011332, 1496-011333, 1496-011334, 1496-011335, 1496-011344, 1496-011345, 1496-011346, 1496-011347, 1496-011348, 1496-011349, 1496-011350, 1496-011351, 1496-011352, 1496-011353, 1496-011369, 1496-011370, 1496-011371, 1496-011372, 1496-011373, 1496-011374, 1496-011375, 1496-011381, 1496-011386, 1496-011387, 1496-011388, 1496-011389, 1496-011390, 1496-011391, 1496-011392, 1496-011393, 1496-011394, 1496-011395, 1496-011396, 1496-011397, 1496-011398, 1496-011399, 1496-011400, 1496-011401, 1496-011402, 1496-011404, 1496-011405, 1496-011406, 1496-011408, 1496-011409, 1496-011410, 1496-011411, 1496-011412, 1496-011413, 1496-011414, 1496-011415, 1496-011416, 1496-011417, 1496-011418, 1496-011419, 1496-011420, 1496-011421, 1496-011422, 1496-011423, 1496-011424, 1496-011425, 1496-011426, 1496-011427, 1496-011428, 1496-011429, 1496-011430, 1496-011431, 1496-011432, 1496-011433, 1496-011434, 1496-011435, 1496-011436, 1496-011437, 1496-011438, 1496-011439, 1496-011441, 1496-011442, 1496-011443, 1496-011444, 1496-011445, 1496-011446, 1496-011447, 1496-011448, 1496-011449, 1496-011450, 1496-011451, 1496-011453, 1496-011454, 1496-011456, 1496-011457, 1496-011458, 1496-011460, 1496-011461, 1496-011462, 1496-011463, 1496-011465, 1496-011466, 1496-011467, 1496-011468, 1496-011480, 1496-011481, 1496-011482, 1496-011483, 1496-011484, 1496-011485, 1496-011486, 1496-011487, 1496-011488, 1496-011489, 1496-011490, 1496-011491, 1496-011492, 1496-011493, 1496-011494, 1496-011495, 1496-011496, 1496-011497, 1496-011498, 1496-011499, 1496-011500, 1496-011501, 1496-011502, 1496-011503, 1496-011504, 1496-011505, 1496-011506, 1496-011517, 1496-011518, 1496-011519, 1496-011520, 1496-011521, 1496-011522, 1496-011523, 1496-011524, 1496-011525, 1496-011526, 1496-011527, 1496-011528, 1496-011529, 1496-011530, 1496-011531, 1496-011532, 1496-011533, 1496-011534, 1496-011535, 1496-011536, 1496-011537, 1496-011538, 1496-011539, 1496-011540, 1496-011541, 1496-011542, 1496-011543, 1496-011544, 1496-011545, 1496-011546, 1496-011547, 1496-011548, 1496-011549, 1496-011550, 1496-011551, 1496-011564, 1496-011565, 1496-011566, 1496-011579, 1496-011580, 1496-011581, 1496-011582, 1496-011583, 1496-011584, 1496-011585, 1496-011586, 1496-011587, 1496-011588, 1496-011589, 1496-011600, 1496-011613, 1496-011617, 1496-011618, 1496-011619, 1496-011620, 1496-011621, 1496-011622, 1496-011623, 1496-011624, 1496-011625, 1496-011635, 1496-011642, 1496-011643, 1496-011644, 1496-011645, 1496-011646, 1496-011647, 1496-011648, 1496-011649, 1496-011650, 1496-011651, 1496-011652, 1496-011653, 1496-011654, 1496-011655, 1496-011656, 1496-011657, 1496-011658, 1496-011659, 1496-011660, 1496-011661, 1496-011662, 1496-011663, 1496-011664, 1496-011665, 1496-011666, 1496-011667, 1496-011668, 1496-011669, 1496-011671, 1496-011672, 1496-011673, 1496-011674, 1496-011675, 1496-011676, 1496-011677, 1496-011678, 1496-011679, 1496-011680, 1496-011681, 1496-011682, 1496-011683, 1496-011689, 1496-011714, 1496-011715, 1496-011716, 1496-011717, 1496-011718, 1496-011719, 1496-011720, 1496-011721, 1496-011722, 1496-011723, 1496-011727, 1496-011728, 1496-011729, 1496-011730, 1496-011731, 1496-011733, 1496-011734, 1496-011735, 1496-011736, 1496-011737, 1496-011742, 1496-011744, 1496-011745, 1496-011746, 1496-011747, 1496-011748, 1496-011749, 1496-011750, 1496-011751, 1496-011752, 1496-011753, 1496-011754, 1496-011755, 1496-011756, 1496-011757, 1496-011758, 1496-011759, 1496-011760, 1496-011767, 1496-011768, 1496-011769, 1496-011770, 1496-011771, 1496-011772, 1496-011773, 1496-011774, 1496-011775, 1496-011776, 1496-011777, 1496-011778, 1496-011779, 1496-011781, 1496-011782, 1496-011783, 1496-011784, 1496-011785, 1496-011786, 1496-011787, 1496-011792, 1496-011793, 1496-011794, 1496-011795, 1496-011796, 1496-011797, 1496-011798, 1496-011799, 1496-011800, 1496-011801, 1496-011802, 1496-011803, 1496-011804, 1496-011805, 1496-011806, 1496-011807, 1496-011808, 1496-011819, 1496-011820, 1496-011821, 1496-011822, 1496-011823, 1496-011843, 1496-011844, 1496-011845, 1496-011846, 1496-011847, 1496-011848, 1496-011849, 1496-011850, 1496-011852, 1496-011853, 1496-011854, 1496-011855, 1496-011858, 1496-011861, 1496-011862, 1496-011863, 1496-011864, 1496-011865, 1496-011866, 1496-011867, 1496-011868, 1496-011869, 1496-011870, 1496-011871, 1496-011872, 1496-011873, 1496-011874, 1496-011875, 1496-011876, 1496-011877, 1496-011878, 1496-011879, 1496-011929, 1496-011931, 1496-011940, 1496-011941, 1496-011942, 1496-011944, 1496-011945, 1496-011946, 1496-011947, 1496-011948, 1496-011949, 1496-011950, 1496-011951, 1496-011952, 1496-011953, 1496-011954, 1496-011955, 1496-011956, 1496-011958, 1496-011967, 1496-011968, 1496-011969, 1496-011970, 1496-011971, 1496-011972, 1496-011973, 1496-011974, 1496-011975, 1496-011976, 1496-011977, 1496-011978, 1496-011979, 1496-011980, 1496-011981, 1496-011982, 1496-011983, 1496-011984, 1496-011985, 1496-011986, 1496-011987, 1496-011988, 1496-011989, 1496-011990, 1496-011991, 1496-011992, 1496-011993, 1496-011994, 1496-011995, 1496-011996, 1496-011997, 1496-011999, 1496-012000, 1496-012001, 1496-012003, 1496-012004, 1496-012005, 1496-012006, 1496-012007, 1496-012008, 1496-012009, 1496-012010, 1496-012024, 1496-012025, 1496-012043, 1496-012044, 1496-012045, 1496-012046, 1496-012047, 1496-012048, 1496-012049, 1496-012050, 1496-012051, 1496-012052, 1496-012053, 1496-012055, 1496-012056, 1496-012057, 1496-012058, 1496-012059, 1496-012060, 1496-012061, 1496-012062, 1496-012063, 1496-012064, 1496-012065, 1496-012066, 1496-012067, 1496-012068, 1496-012069, 1496-012070, 1496-012071, 1496-012072, 1496-012073, 1496-012074, 1496-012075, 1496-012076, 1496-012077, 1496-012078, 1496-012079, 1496-012080, 1496-012081, 1496-012082, 1496-012083, 1496-012084, 1496-012085, 1496-012086, 1496-012087, 1496-012088, 1496-012089, 1496-012090, 1496-012091, 1496-012092, 1496-012093, 1496-012094, 1496-012095, 1496-012096, 1496-012097, 1496-012098, 1496-012099, 1496-012101, 1496-012102, 1496-012103, 1496-012105, 1496-012106, 1496-012107, 1496-012108, 1496-012109, 1496-012110, 1496-012111, 1496-012112, 1496-012113, 1496-012114, 1496-012115, 1496-012116, 1496-012117, 1496-012119, 1496-012120, 1496-012121, 1496-012122, 1496-012123, 1496-012124, 1496-012125, 1496-012126, 1496-012127, 1496-012128, 1496-012130, 1496-012131, 1496-012132, 1496-012133, 1496-012134, 1496-012135, 1496-012137, 1496-012145, 1496-012159, 1496-012160, 1496-012162, 1496-012163, 1496-012164, 1496-012165, 1496-012166, 1496-012167, 1496-012168, 1496-012169, 1496-012170, 1496-012171, 1496-012173, 1496-012186, 1496-012191, 1496-012199, 1496-012200, 1496-012201, 1496-012202, 1496-012203, 1496-012204, 1496-012205, 1496-012206, 1496-012207, 1496-012208, 1496-012209, 1496-012210, 1496-012212, 1496-012213, 1496-012214, 1496-012215, 1496-012216, 1496-012217, 1496-012218, 1496-012219, 1496-012220, 1496-012221, 1496-012222, 1496-012223, 1496-012224, 1496-012225, 1496-012226, 1496-012227, 1496-012228, 1496-012229, 1496-012230, 1496-012231, 1496-012232, 1496-012233, 1496-012303, 1496-012342, 1496-012343, 1496-012345, 1496-012346, 1496-012347, 1496-012350, 1496-012366, 1496-012378, 1496-012379, 1496-012396, 1496-012397, 1496-012398, 1496-012399, 1496-012400, 1496-012401, 1496-012402, 1496-012403, 1496-012404, 1496-012405, 1496-012406, 1496-012408, 1496-012422, 1496-012427, 1496-012430, 1496-012433, 1496-012436, 1496-012438, 1496-012439, 1496-012440, 1496-012441, 1496-012442, 1496-012443, 1496-012444, 1496-012445, 1496-012446, 1496-012447, 1496-012448, 1496-012449, 1496-012450, 1496-012451, 1496-012452, 1496-012453, 1496-012454, 1496-012455, 1496-012456, 1496-012457, 1496-012463, 1496-012474, 1496-012475, 1496-012476, 1496-012477, 1496-012478, 1496-012479, 1496-012480, 1496-012481, 1496-012482, 1496-012483, 1496-012484, 1496-012485, 1496-012486, 1496-012488, 1496-012489, 1496-012490, 1496-012491, 1496-012492, 1496-012493, 1496-012494, 1496-012509, 1496-012511, 1496-012522, 1496-012660, 1496-012661, 1496-012662, 1496-012663, 1496-012664, 1496-012665, 1496-012666, 1496-012667, 1496-012668, 1496-012669, 1496-012670, 1496-012671, 1496-012672, 1496-012674, 1496-012675, 1496-012676, 1496-012677, 1496-012678, 1496-012679, 1496-012680, 1496-012681, 1496-012682, 1496-012683, 1496-012684, 1496-012685, 1496-012686, 1496-012702, 1496-012705, 1496-012706, 1496-012711, 1496-012718, 1496-012719, 1496-012720, 1496-012721, 1496-012722, 1496-012723, 1496-012724, 1496-012725, 1496-012726, 1496-012727, 1496-012728, 1496-012734, 1496-012735, 1496-012736, 1496-012737, 1496-012738, 1496-012739, 1496-012740, 1496-012742, 1496-012743, 1496-012744, 1496-012745, 1496-012746, 1496-012747, 1496-012748, 1496-012749, 1496-012750, 1496-012751, 1496-012765, 1496-012767, 1496-012768, 1496-012769, 1496-012770, 1496-012771, 1496-012772, 1496-012773, 1496-012774, 1496-012775, 1496-012776, 1496-012777, 1496-012778, 1496-012779, 1496-012781, 1496-012782, 1496-012783, 1496-012785, 1496-012786, 1496-012787, 1496-012788, 1496-012791, 1496-012792, 1496-012887, 1496-012888, 1496-012889, 1496-012890, 1496-012891, 1496-012892, 1496-012893, 1496-012895, 1496-012940, 1496-012941, 1496-012942, 1496-012943, 1496-012944, 1496-012945, 1496-012946, 1496-012947, 1496-012948, 1496-012949, 1496-012950, 1496-012951, 1496-012952, 1496-012953, 1496-012954, 1496-012955, 1496-012976, 1496-012977, 1496-012978, 1496-012979, 1496-012980, 1496-013000, 1496-013001, 1496-013002, 1496-013003, 1496-013004, 1496-013005, 1496-013006, 1496-013018, 1496-013024, 1496-013025, 1496-013026, 1496-013027, 1496-013028, 1496-013029, 1496-013030, 1496-013031, 1496-013033, 1496-013034, 1496-013035, 1496-013036, 1496-013037, 1496-013038, 1496-013040, 1496-013041, 1496-013043, 1496-013044, 1496-013046, 1496-013047, 1496-013049, 1496-013054, 1496-013076, 1496-013077, 1496-013078, 1496-013079, 1496-013080, 1496-013081, 1496-013082, 1496-013083, 1496-013085, 1496-013086, 1496-013087, 1496-013088, 1496-013089, 1496-013090, 1496-013093, 1496-013094, 1496-013095, 1496-013107, 1496-013145, 1496-013146, 1496-013147, 1496-013148, 1496-013149, 1496-013150, 1496-013151, 1496-013152, 1496-013153, 1496-013154, 1496-013156, 1496-013157, 1496-013167, 1496-013168, 1496-013169, 1496-013170, 1496-013171, 1496-013172, 1496-013173, 1496-013213, 1496-013214, 1496-013215, 1496-013227, 1496-013229, 1496-013230, 1496-013231, 1496-013232, 1496-013233, 1496-013234, 1496-013235, 1496-013236, 1496-013237, 1496-013238, 1496-013239, 1496-013240, 1496-013242, 1496-013243, 1496-013244, 1496-013245, 1496-013246, 1496-013247, 1496-013248, 1496-013249, 1496-013250, 1496-013251, 1496-013252, 1496-013253, 1496-013254, 1496-013255, 1496-013256, 1496-013257, 1496-013258, 1496-013259, 1496-013260, 1496-013261, 1496-013262, 1496-013263, 1496-013264, 1496-013265, 1496-013266, 1496-013267, 1496-013269, 1496-013270, 1496-013271, 1496-013272, 1496-013273, 1496-013275, 1496-013276, 1496-013277, 1496-013278, 1496-013295, 1496-013297, 1496-013298, 1496-013299, 1496-013300, 1496-013301, 1496-013302, 1496-013305, 1496-013306, 1496-013324, 1496-013346, 1496-013347, 1496-013348, 1496-013349, 1496-013350, 1496-013351, 1496-013352, 1496-013353, 1496-013354, 1496-013355, 1496-013356, 1496-013357, 1496-013358, 1496-013359, 1496-013360, 1496-013361, 1496-013362, 1496-013363, 1496-013364, 1496-013365, 1496-013366, 1496-013367, 1496-013368, 1496-013370, 1496-013371, 1496-013372, 1496-013373, 1496-013374, 1496-013375, 1496-013376, 1496-013377, 1496-013378, 1496-013379, 1496-013406, 1496-013407, 1496-013408, 1496-013409, 1496-013410, 1496-013411, 1496-013412, 1496-013413, 1496-013414, 1496-013415, 1496-013416, 1496-013417, 1496-013418, 1496-013419, 1496-013436, 1496-013442, 1496-013445, 1496-013452, 1496-013453, 1496-013454, 1496-013455, 1496-013456, 1496-013457, 1496-013458, 1496-013459, 1496-013460, 1496-013461, 1496-013464, 1496-013465, 1496-013466, 1496-013467, 1496-013468, 1496-013469, 1496-013470, 1496-013471, 1496-013472, 1496-013473, 1496-013474, 1496-013475, 1496-013476, 1496-013482, 1496-013483, 1496-013484, 1496-013485, 1496-013486, 1496-013487, 1496-013488, 1496-013489, 1496-013490, 1496-013491, 1496-013493, 1496-013495, 1496-013497, 1496-013498, 1496-013499, 1496-013500, 1496-013501, 1496-013521, 1496-013522, 1496-013523, 1496-013524, 1496-013525, 1496-013526, 1496-013527, 1496-013528, 1496-013534, 1496-013535, 1496-013536, 1496-013537, 1496-013544, 1496-013545, 1496-013546, 1496-013547, 1496-013548, 1496-013549, 1496-013645, 1496-013646, 1496-013648, 1496-013649, 1496-013684, 1496-013685, 1496-013686, 1496-013687, 1496-013688, 1496-013689, 1496-013690, 1496-013691, 1496-013692, 1496-013693, 1496-013694, 1496-013695, 1496-013696, 1496-013697, 1496-013698, 1496-013699, 1496-013719, 1496-013736, 1496-013737, 1496-013738, 1496-013739, 1496-013740, 1496-013741, 1496-013742, 1496-013743, 1496-013744, 1496-013745, 1496-013746, 1496-013747, 1496-013748, 1496-013749, 1496-013750, 1496-013751, 1496-013753, 1496-013762, 1496-013763, 1496-013764, 1496-013765, 1496-013766, 1496-013843, 1496-013844, 1496-013845, 1496-013853, 1496-013854, 1496-013855, 1496-013856, 1496-013857, 1496-013858, 1496-013859, 1496-013860, 1496-013861, 1496-013862, 1496-013863, 1496-013864, 1496-013865, 1496-013866, 1496-013867, 1496-013868, 1496-013869, 1496-013870, 1496-013871, 1496-013872, 1496-013873, 1496-013874, 1496-013875, 1496-013876, 1496-013877, 1496-013878, 1496-013879, 1496-013880, 1496-013881, 1496-013882, 1496-013883, 1496-013884, 1496-013885, 1496-013886, 1496-013887, 1496-013888, 1496-013889, 1496-013890, 1496-013891, 1496-013892, 1496-013893, 1496-013894, 1496-013895, 1496-013896, 1496-013897, 1496-013898, 1496-013899, 1496-013900, 1496-013901, 1496-013902, 1496-013903, 1496-013904, 1496-013906, 1496-013907, 1496-013908, 1496-013909, 1496-013910, 1496-013911, 1496-013912, 1496-013917, 1496-013918, 1496-013919, 1496-013920, 1496-014022, 1496-014066, 1496-014067, 1496-014068, 1496-014069, 1496-014070, 1496-014071, 1496-014080, 1496-014081, 1496-014082, 1496-014085, 1496-014092, 1496-014093, 1496-014094, 1496-014095, 1496-014096, 1496-014097, 1496-014098, 1496-014099, 1496-014100, 1496-014101, 1496-014102, 1496-014103, 1496-014104, 1496-014105, 1496-014106, 1496-014107, 1496-014108, 1496-014109, 1496-014110, 1496-014111, 1496-014112, 1496-014113, 1496-014114, 1496-014115, 1496-014116, 1496-014117, 1496-014118, 1496-014119, 1496-014120, 1496-014121, 1496-014122, 1496-014123, 1496-014124, 1496-014125, 1496-014126, 1496-014127, 1496-014137, 1496-014138, 1496-014139, 1496-014140, 1496-014141, 1496-014142, 1496-014197, 1496-014201, 1496-014211, 1496-014212, 1496-014213, 1496-014214, 1496-014215, 1496-014216, 1496-014226, 1496-014227, 1496-014228, 1496-014229, 1496-014230, 1496-014231, 1496-014232, 1496-014233, 1496-014234, 1496-014235, 1496-014236, 1496-014237, 1496-014238, 1496-014239, 1496-014240, 1496-014241, 1496-014242, 1496-014243, 1496-014244, 1496-014245, 1496-014246, 1496-014247, 1496-014248, 1496-014249, 1496-014250, 1496-014251, 1496-014252, 1496-014307, 1496-014308, 1496-014309, 1496-014310, 1496-014311, 1496-014312, 1496-014313, 1496-014314, 1496-014315, 1496-014316, 1496-014317, 1496-014318, 1496-014319, 1496-014320, 1496-014321, 1496-014322, 1496-014323, 1496-014324, 1496-014325, 1496-014326, 1496-014327, 1496-014328, 1496-014329, 1496-014330, 1496-014331, 1496-014332, 1496-014333, 1496-014334, 1496-014364, 1496-014365, 1496-014366, 1496-014367, 1496-014368, 1496-014369, 1496-014397, 1496-014399, 1496-014400, 1496-014401, 1496-014402, 1496-014403, 1496-014404, 1496-014407, 1496-014408, 1496-014409, 1496-014414, 1496-014421, 1496-014442, 1496-014443, 1496-014444, 1496-014452, 1496-014453, 1496-014454, 1496-014455, 1496-014465, 1496-014472, 1496-014473, 1496-014474, 1496-014475, 1496-014476, 1496-014477, 1496-014499, 1496-014500, 1496-014501, 1496-014502, 1496-014507, 1496-014508, 1496-014509, 1496-014510, 1496-014511, 1496-014512, 1496-014513, 1496-014556, 1496-014590, 1496-014591, 1496-014592, 1496-014593, 1496-014594, 1496-014595, 1496-014596, 1496-014613, 1496-014622, 1496-014641, 1496-014642, 1496-014643, 1496-014644, 1496-014645, 1496-014646, 1496-014647, 1496-014648, 1496-014649, 1496-014650, 1496-014651, 1496-014652, 1496-014653, 1496-014654, 1496-014655, 1496-014656, 1496-014657, 1496-014658, 1496-014659, 1496-014660, 1496-014661, 1496-014662, 1496-014663, 1496-014664, 1496-014665, 1496-014666, 1496-014667, 1496-014668, 1496-014669, 1496-014670, 1496-014671, 1496-014672, 1496-014673, 1496-014674, 1496-014675, 1496-014676, 1496-014677, 1496-014710, 1496-014711, 1496-014713, 1496-014714, 1496-014715, 1496-014716, 1496-014717, 1496-014718, 1496-014719, 1496-014720, 1496-014721, 1496-014722, 1496-014777, 1496-014807, 1496-014830, 1496-014834, 1496-014836, 1496-014837, 1496-014838, 1496-014839, 1496-014840, 1496-014841, 1496-014842, 1496-014843, 1496-014844, 1496-014846, 1496-014874, 1496-014875, 1496-014876, 1496-014877, 1496-014910, 1496-014914, 1496-014915, 1496-014916, 1496-014917, 1496-014918, 1496-014919, 1496-014920, 1496-014921, 1496-014922, 1496-014923, 1496-014924, 1496-014925, 1496-014926, 1496-014927, 1496-014928, 1496-014929, 1496-014930, 1496-014932, 1496-014933, 1496-014934, 1496-014935, 1496-014938, 1496-014939, 1496-014940, 1496-014941, 1496-014942, 1496-014943, 1496-014944, 1496-014945, 1496-014946, 1496-014947, 1496-014948, 1496-014949, 1496-014989, 1496-014990, 1496-014991, 1496-014993, 1496-014994, 1496-014995, 1496-014996, 1496-014997, 1496-015017, 1496-015039, 1496-015040, 1496-015043, 1496-015044, 1496-015045, 1496-015047, 1496-015049, 1496-015050, 1496-015051, 1496-015052, 1496-015053, 1496-015054, 1496-015055, 1496-015056, 1496-015057, 1496-015058, 1496-015060, 1496-015061, 1496-015062, 1496-015063, 1496-015064, 1496-015065, 1496-015066, 1496-015067, 1496-015068, 1496-015069, 1496-015070, 1496-015071, 1496-015072, 1496-015073, 1496-015074, 1496-015075, 1496-015085, 1496-015128, 1496-015129, 1496-015130, 1496-015142, 1496-015143, 1496-015144, 1496-015145, 1496-015146, 1496-015147, 1496-015148, 1496-015149, 1496-015150, 1496-015151, 1496-015152, 1496-015153, 1496-015154, 1496-015155, 1496-015156, 1496-015157, 1496-015158, 1496-015159, 1496-015160, 1496-015161, 1496-015162, 1496-015163, 1496-015164, 1496-015165, 1496-015166, 1496-015167, 1496-015168, 1496-015169, 1496-015170, 1496-015171, 1496-015172, 1496-015173, 1496-015174, 1496-015175, 1496-015176, 1496-015177, 1496-015178, 1496-015179, 1496-015180, 1496-015182, 1496-015187, 1496-015188, 1496-015193, 1496-015198, 1496-015215, 1496-015216, 1496-015313, 1496-015314, 1496-015315, 1496-015316, 1496-015319, 1496-015320, 1496-015321, 1496-015322, 1496-015323, 1496-015325, 1496-015326, 1496-015327, 1496-015328, 1496-015329, 1496-015330, 1496-015331, 1496-015333, 1496-015334, 1496-015335, 1496-015336, 1496-015337, 1496-015338, 1496-015339, 1496-015340, 1496-015341, 1496-015342, 1496-015343, 1496-015344, 1496-015345, 1496-015346, 1496-015347, 1496-015348, 1496-015349, 1496-015350, 1496-015351, 1496-015352, 1496-015353, 1496-015354, 1496-015355, 1496-015358, 1496-015359, 1496-015360, 1496-015361, 1496-015362, 1496-015363, 1496-015364, 1496-015365, 1496-015366, 1496-015367, 1496-015368, 1496-015369, 1496-015370, 1496-015442, 1496-015443, 1496-015496, 1496-015498, 1496-015499, 1496-015500, 1496-015501, 1496-015502, 1496-015535, 1496-015536, 1496-015537, 1496-015538, 1496-015562, 1496-015574, 1496-015579, 1496-015580, 1496-015581, 1496-015582, 1496-015584, 1496-015585, 1496-015586, 1496-015587, 1496-015588, 1496-015589, 1496-015591, 1496-015593, 1496-015597, 1496-015611, 1496-015612, 1496-015613, 1496-015614, 1496-015615, 1496-015616, 1496-015617, 1496-015618, 1496-015619, 1496-015620, 1496-015621, 1496-015622, 1496-015623, 1496-015624, 1496-015626, 1496-015627, 1496-015636, 1496-015653, 1496-015655, 1496-015658, 1496-015662, 1496-015667, 1496-015694, 1496-015699, 1496-015701, 1496-015703, 1496-015704, 1496-015705.1496-015706, 1496-015707, 1496-015708, 1496-015709, 1496-015710, 1496-015711, 1496-015712, 1496-015713, 1496-015714, 1496-015715, 1496-015717, 1496-015718, 1496-015719, 1496-015720, 1496-015721, 1496-015722, 1496-015723, 1496-015724, 1496-015725, 1496-015726, 1496-015727, 1496-015728, 1496-015729, 1496-015730, 1496-015731, 1496-015732, 1496-015733, 1496-015734, 1496-015735, 1496-015736, 1496-015737, 1496-015738, 1496-015739, 1496-015752, 1496-015753, 1496-015754, 1496-015766, 1496-015767, 1496-015768, 1496-015769, 1496-015775, 1496-015776, 1496-015777, 1496-015778, 1496-015779, 1496-015780, 1496-015781, 1496-015782, 1496-015784, 1496-015785, 1496-015786, 1496-015787, 1496-015788, 1496-015789, 1496-015790, 1496-015791, 1496-015792, 1496-015793, 1496-015794, 1496-015795, 1496-015796, 1496-015797, 1496-015798, 1496-015799, 1496-015800, 1496-015801, 1496-015803, 1496-015804, 1496-015805, 1496-015806, 1496-015807, 1496-015808, 1496-015809, 1496-015811, 1496-015813, 1496-015814, 1496-015815, 1496-015828, 1496-015831, 1496-015836, 1496-015839, 1496-015840, 1496-015841, 1496-015842, 1496-015843, 1496-015844, 1496-015845, 1496-015846, 1496-015847, 1496-015848, 1496-015849, 1496-015850, 1496-015851, 1496-015852, 1496-015853, 1496-015857, 1496-015859, 1496-015870, 1496-015874, 1496-015879, 1496-015880, 1496-015881, 1496-015882, 1496-015883, 1496-015884, 1496-015885, 1496-015889, 1496-015890, 1496-015891, 1496-015892, 1496-015893, 1496-015894, 1496-015895, 1496-015896, 1496-015897, 1496-015898, 1496-015899, 1496-015900, 1496-015901, 1496-015902, 1496-015903, 1496-015905, 1496-015906, 1496-015907, 1496-015918, 1496-015919, 1496-015920, 1496-015921, 1496-015922, 1496-015923, 1496-015924, 1496-015926, 1496-015927, 1496-015928, 1496-015929, 1496-015930, 1496-015931, 1496-015936, 1496-015937, 1496-015938, 1496-015939, 1496-015940, 1496-015945, 1496-015947, 1496-015957, 1496-016040, 1496-016041, 1496-016042, 1496-016043, 1496-016045, 1496-016046, 1496-016047, 1496-016050, 1496-016060, 1496-016061, 1496-016065, 1496-016067, 1496-016069, 1496-016070, 1496-016072, 1496-016079, 1496-016081, 1496-016082, 1496-016083, 1496-016084, 1496-016085, 1496-016087, 1496-016088, 1496-016089, 1496-016090, 1496-016091, 1496-016092, 1496-016093, 1496-016094, 1496-016095, 1496-016096, 1496-016097, 1496-016098, 1496-016099, 1496-016100, 1496-016101, 1496-016102, 1496-016103, 1496-016104, 1496-016105, 1496-016106, 1496-016107, 1496-016108, 1496-016109, 1496-016111, 1496-016114, 1496-016115, 1496-016117, 1496-016118, 1496-016120, 1496-016121, 1496-016122, 1496-016123, 1496-016124, 1496-016125, 1496-016126, 1496-016127, 1496-016128, 1496-016129, 1496-016130, 1496-016131, 1496-016132, 1496-016133, 1496-016134, 1496-016139, 1496-016141, 1496-016144, 1496-016153, 1496-016162, 1496-016165, 1496-016169, 1496-016197, 1496-016203, 1496-016204, 1496-016205, 1496-016206, 1496-016207, 1496-016208, 1496-016209, 1496-016210, 1496-016211, 1496-016212, 1496-016237, 1496-016238, 1496-016239, 1496-016240, 1496-016241, 1496-016242, 1496-016243, 1496-016244, 1496-016245, 1496-016246, 1496-016247, 1496-016248, 1496-016265, 1496-016296, 1496-016297, 1496-016298, 1496-016299, 1496-016300, 1496-016301, 1496-016308, 1496-016309, 1496-016310, 1496-016311, 1496-016312, 1496-016313, 1496-016314, 1496-016321, 1496-016322, 1496-016324, 1496-016348, 1496-016349, 1496-016352, 1496-016358, 1496-016359, 1496-016360, 1496-016361, 1496-016386, 1496-016387, 1496-016403, 1496-016404, 1496-016405, 1496-016407, 1496-016408, 1496-016409, 1496-016410, 1496-016411, 1496-016412, 1496-016413, 1496-016414, 1496-016415, 1496-016416, 1496-016417, 1496-016419, 1496-016420, 1496-016421, 1496-016422, 1496-016423, 1496-016424, 1496-016425, 1496-016426, 1496-016427, 1496-016428, 1496-016429, 1496-016430, 1496-016431, 1496-016432, 1496-016433, 1496-016434, 1496-016435, 1496-016472, 1496-016473, 1496-016474, 1496-016475, 1496-016476, 1496-016477, 1496-016497, 1496-016500, 1496-016501, 1496-016502, 1496-016503, 1496-016504, 1496-016505, 1496-016506, 1496-016507, 1496-016508, 1496-016510, 1496-016511, 1496-016512, 1496-016513, 1496-016514, 1496-016515, 1496-016516, 1496-016517, 1496-016518, 1496-016528, 1496-016531, 1496-016533, 1496-016534, 1496-016535, 1496-016536, 1496-016537, 1496-016539, 1496-016540, 1496-016541, 1496-016542, 1496-016543, 1496-016544, 1496-016581, 1496-016582, 1496-016583, 1496-016593, 1496-016594, 1496-016613, 1496-016614, 1496-016615, 1496-016616, 1496-016617, 1496-016618, 1496-016619, 1496-016620, 1496-016622, 1496-016624, 1496-016625, 1496-016626, 1496-016631, 1496-016639, 1496-016640, 1496-016641, 1496-016642, 1496-016643, 1496-016644, 1496-016645, 1496-016646, 1496-016647, 1496-016648, 1496-016649, 1496-016650, 1496-016651, 1496-016652, 1496-016653, 1496-016654, 1496-016655, 1496-016657, 1496-016663, 1496-016664, 1496-016665, 1496-016666, 1496-016667, 1496-016668, 1496-016669, 1496-016671, 1496-016672, 1496-016673, 1496-016693, 1496-016694, 1496-016695, 1496-016696, 1496-016698, 1496-016699, 1496-016700, 1496-016701, 1496-016703, 1496-016704, 1496-016705, 1496-016707, 1496-016708, 1496-016709, 1496-016711, 1496-016712, 1496-016713, 1496-016714, 1496-016717, 1496-016718, 1496-016719, 1496-016720, 1496-016721, 1496-016722, 1496-016723, 1496-016734, 1496-016735, 1496-016781, 1496-016786, 1496-016787, 1496-016788, 1496-016789, 1496-016790, 1496-016791, 1496-016792, 1496-016797, 1496-016819, 1496-016820, 1496-016821, 1496-016822, 1496-016823, 1496-016824, 1496-016825, 1496-016826, 1496-016827, 1496-016828, 1496-016829, 1496-016830, 1496-016831, 1496-016832, 1496-016833, 1496-016834, 1496-016835, 1496-016836, 1496-016837, 1496-016838, 1496-016839, 1496-016840, 1496-016841, 1496-016844, 1496-016847, 1496-016848, 1496-016849, 1496-016850, 1496-016851, 1496-016852, 1496-016855, 1496-016879, 1496-016880, 1496-016881, 1496-016883, 1496-016885, 1496-016889, 1496-016903, 1496-016904, 1496-016905, 1496-016906, 1496-016907, 1496-016908, 1496-016909, 1496-016910, 1496-016912, 1496-016913, 1496-016914, 1496-016915, 1496-016916, 1496-016917, 1496-016918, 1496-016919, 1496-016920, 1496-016921, 1496-016922, 1496-016923, 1496-016924, 1496-016925, 1496-016926, 1496-016927, 1496-016928, 1496-016929, 1496-017006, 1496-017008, 1496-017009, 1496-017010, 1496-017011, 1496-017012, 1496-017013, 1496-017024, 1496-017025, 1496-017026, 1496-017027, 1496-017028, 1496-017030, 1496-017038, 1496-017044, 1496-017045, 1496-017065, 1496-017066, 1496-017067, 1496-017068, 1496-017069, 1496-017070, 1496-017071, 1496-017072, 1496-017073, 1496-017074, 1496-017075, 1496-017076, 1496-017077, 1496-017078, 1496-017079, 1496-017080, 1496-017107, 1496-017202, 1496-017203, 1496-017232, 1496-017306, 1496-017308, 1496-017309, 1496-017310, 1496-017312, 1496-017313, 1496-017314, 1496-017315, 1496-017316, 1496-017317, 1496-017318, 1496-017319, 1496-017320, 1496-017321, 1496-017322, 1496-017323, 1496-017324, 1496-017325, 1496-017326, 1496-017327, 1496-017328, 1496-017335, 1496-017356, 1496-017358, 1496-017359, 1496-017360, 1496-017361, 1496-017362, 1496-017393, 1496-017451, 1496-017452, 1496-017454, 1496-017455, 1496-017456, 1496-017457, 1496-017458, 1496-017459, 1496-017460, 1496-017461, 1496-017462, 1496-017463, 1496-017495, 1496-017496, 1496-017497, 1496-017498, 1496-017499, 1496-017500, 1496-017502, 1496-017503, 1496-017504, 1496-017505, 1496-017512, 1496-017513, 1496-017514, 1496-017515, 1496-017516, 1496-017517, 1496-017518, 1496-017519, 1496-017520, 1496-017521, 1496-017522, 1496-017523, 1496-017524, 1496-017525, 1496-017565, 1496-017566, 1496-017567, 1496-017568, 1496-017569, 1496-017570, 1496-017571, 1496-017572, 1496-017573, 1496-017574, 1496-017575, 1496-017576, 1496-017577, 1496-017578, 1496-017579, 1496-017580, 1496-017581, 1496-017582, 1496-017583, 1496-017584, 1496-017585, 1496-017586, 1496-017587, 1496-017588, 1496-017589, 1496-017590, 1496-017591, 1496-017592, 1496-017593, 1496-017594, 1496-017595, 1496-017600, 1496-017601, 1496-017602, 1496-017603, 1496-017604, 1496-017605, 1496-017606, 1496-017607, 1496-017608, 1496-017609, 1496-017610, 1496-017611, 1496-017612, 1496-017613, 1496-017614, 1496-017615, 1496-017616, 1496-017617, 1496-017618, 1496-017620, 1496-017621, 1496-017622, 1496-017623, 1496-017624, 1496-017625, 1496-017626, 1496-017638, 1496-017640, 1496-017649, 1496-017657, 1496-017658, 1496-017669, 1496-017670, 1496-017671, 1496-017672, 1496-017673, 1496-017675, 1496-017676, 1496-017685, 1496-017698, 1496-017716, 1496-017717, 1496-017718, 1496-017719, 1496-017720, 1496-017721, 1496-017722, 1496-017723, 1496-017724, 1496-017725, 1496-017726, 1496-017727, 1496-017728, 1496-017729, 1496-017730, 1496-017731, 1496-017732, 1496-017733, 1496-017734, 1496-017735, 1496-017736, 1496-017737, 1496-017738, 1496-017739, 1496-017740, 1496-017742, 1496-017743, 1496-017744, 1496-017745, 1496-017747, 1496-017748, 1496-017749, 1496-017750, 1496-017751, 1496-017752, 1496-017753, 1496-017754, 1496-017755, 1496-017756, 1496-017757, 1496-017758, 1496-017759, 1496-017761, 1496-017762, 1496-017763, 1496-017764, 1496-017766, 1496-017767, 1496-017768, 1496-017769, 1496-017770, 1496-017771, 1496-017772, 1496-017773, 1496-017774, 1496-017776, 1496-017777, 1496-017780, 1496-017781, 1496-017782, 1496-017788, 1496-017789, 1496-017790, 1496-017791, 1496-017797, 1496-017798, 1496-017799, 1496-017800, 1496-017801, 1496-017803, 1496-017804, 1496-017805, 1496-017806, 1496-017807, 1496-017808, 1496-017809, 1496-017822, 1496-017824, 1496-017828, 1496-017829, 1496-017830, 1496-017831, 1496-017832, 1496-017833, 1496-017857, 1496-017858, 1496-017859, 1496-017860, 1496-017861, 1496-017862, 1496-017863, 1496-017864, 1496-017865, 1496-017866, 1496-017867, 1496-017868, 1496-017869, 1496-017870, 1496-017871, 1496-017873, 1496-017874, 1496-017875, 1496-017876, 1496-017881, 1496-017931, 1496-017940, 1496-017941, 1496-017947, 1496-017948, 1496-017949, 1496-018008, 1496-018017, 1496-018018, 1496-018019, 1496-018020, 1496-018021, 1496-018022, 1496-018023, 1496-018024, 1496-018026, 1496-018027, 1496-018028, 1496-018029, 1496-018030, 1496-018031, 1496-018032, 1496-018033, 1496-018034, 1496-018035, 1496-018036, 1496-018037, 1496-018039, 1496-018040, 1496-018041, 1496-018042, 1496-018043, 1496-018044, 1496-018054, 1496-018055, 1496-018056, 1496-018057, 1496-018059, 1496-018060, 1496-018061, 1496-018062, 1496-018063, 1496-018064, 1496-018065, 1496-018066, 1496-018072, 1496-018073, 1496-018074, 1496-018075, 1496-018076, 1496-018077, 1496-018078, 1496-018079, 1496-018080, 1496-018081, 1496-018082, 1496-018083, 1496-018084, 1496-018085, 1496-018086, 1496-018087, 1496-018088, 1496-018089, 1496-018090, 1496-018091, 1496-018092, 1496-018093, 1496-018094, 1496-018095, 1496-018096, 1496-018097, 1496-018098, 1496-018099, 1496-018100, 1496-018101, 1496-018102, 1496-018103, 1496-018104, 1496-018105, 1496-018106, 1496-018107, 1496-018108, 1496-018109, 1496-018110, 1496-018111, 1496-018112, 1496-018113, 1496-018114, 1496-018115, 1496-018116, 1496-018118, 1496-018119, 1496-018120, 1496-018121, 1496-018122, 1496-018123, 1496-018124, 1496-018125, 1496-018126, 1496-018127, 1496-018128, 1496-018129, 1496-018131, 1496-018132, 1496-018133, 1496-018135, 1496-018136, 1496-018137, 1496-018138, 1496-018339, 1496-018340, 1496-018341, 1496-018342, 1496-018389, 1496-018408, 1496-018409, 1496-018410, 1496-018411, 1496-018412, 1496-018413, 1496-018414, 1496-018415, 1496-018416, 1496-018417, 1496-018418, 1496-018419, 1496-018420, 1496-018421, 1496-018422, 1496-018423, 1496-018424, 1496-018425, 1496-018426, 1496-018429, 1496-018430, 1496-018431, 1496-018432, 1496-018433, 1496-018434, 1496-018435, 1496-018436, 1496-018437, 1496-018438, 1496-018439, 1496-018440, 1496-018441, 1496-018442, 1496-018443, 1496-018444, 1496-018469, 1496-018470, 1496-018471, 1496-018472, 1496-018473, 1496-018474, 1496-018480, 1496-018481, 1496-018482, 1496-018483, 1496-018484, 1496-018485, 1496-018486, 1496-018487, 1496-018489, 1496-018490, 1496-018491, 1496-018492, 1496-018518, 1496-018519, 1496-018520, 1496-018521, 1496-018522, 1496-018523, 1496-018524, 1496-018525, 1496-018526, 1496-018527, 1496-018528, 1496-018529, 1496-018530, 1496-018531, 1496-018532, 1496-018533, 1496-018534, 1496-018536, 1496-018537, 1496-018538, 1496-018539, 1496-018540, 1496-018541, 1496-018542, 1496-018543, 1496-018544, 1496-018545, 1496-018553, 1496-018555, 1496-018563, 1496-018564, 1496-018565, 1496-018566, 1496-018567, 1496-018568, 1496-018569, 1496-018570, 1496-018571, 1496-018572, 1496-018573, 1496-018574, 1496-018575, 1496-018576, 1496-018577, 1496-018578, 1496-018579, 1496-018580, 1496-018583, 1496-018592, 1496-018596, 1496-018597, 1496-018598, 1496-018599, 1496-018600, 1496-018601, 1496-018606, 1496-018607, 1496-018608, 1496-018609, 1496-018611, 1496-018612, 1496-018613, 1496-018614, 1496-018615, 1496-018616, 1496-018617, 1496-018618, 1496-018619, 1496-018620, 1496-018621, 1496-018622, 1496-018623, 1496-018624, 1496-018625, 1496-018626, 1496-018627, 1496-018628, 1496-018629, 1496-018630, 1496-018631, 1496-018632, 1496-018633, 1496-018634, 1496-018635, 1496-018636, 1496-018637, 1496-018638, 1496-018639, 1496-018640, 1496-018641, 1496-018642, 1496-018643, 1496-018644, 1496-018645, 1496-018646, 1496-018647, 1496-018648, 1496-018649, 1496-018650, 1496-018651, 1496-018652, 1496-018677, 1496-018678, 1496-018679, 1496-018680, 1496-018681, 1496-018682, 1496-018685, 1496-018686, 1496-018687, 1496-018688, 1496-018689, 1496-018690, 1496-018691, 1496-018692, 1496-018693, 1496-018694, 1496-018695, 1496-018696, 1496-018697, 1496-018698, 1496-018699, 1496-018700, 1496-018701, 1496-018702, 1496-018703, 1496-018704, 1496-018705, 1496-018706, 1496-018707, 1496-018708, 1496-018709, 1496-018710, 1496-018711, 1496-018712, 1496-018713, 1496-018714, 1496-018732, 1496-018733, 1496-018734, 1496-018735, 1496-018736, 1496-018737, 1496-018738, 1496-018739, 1496-018740, 1496-018741, 1496-018742, 1496-018743, 1496-018744, 1496-018745, 1496-018746, 1496-018747, 1496-018748, 1496-018749, 1496-018750, 1496-018751, 1496-018752, 1496-018753, 1496-018758, 1496-018759, 1496-018760, 1496-018761, 1496-018762, 1496-018763, 1496-018764, 1496-018765, 1496-018766, 1496-018767, 1496-018768, 1496-018769, 1496-018782, 1496-018783, 1496-018785, 1496-018786, 1496-018787, 1496-018790, 1496-018791, 1496-018792, 1496-018793, 1496-018794, 1496-018796, 1496-018797, 1496-018798, 1496-018799, 1496-018805, 1496-018809, 1496-018976, 1496-018977, 1496-018978, 1496-018979, 1496-018980, 1496-018981, 1496-019008, 1496-019009, 1496-019010, 1496-019011, 1496-019012, 1496-019013, 1496-019014, 1496-019015, 1496-019016, 1496-019017, 1496-019018, 1496-019019, 1496-019020, 1496-019021, 1496-019022, 1496-019023, 1496-019024, 1496-019025, 1496-019026, 1496-019027, 1496-019028, 1496-019029, 1496-019031, 1496-019035, 1496-019036, 1496-019037, 1496-019038, 1496-019040, 1496-019041, 1496-019042, 1496-019043, 1496-019044, 1496-019046, 1496-019048, 1496-019049, 1496-019050, 1496-019051, 1496-019052, 1496-019053, 1496-019054, 1496-019055, 1496-019056, 1496-019057, 1496-019058, 1496-019059, 1496-019060, 1496-019061, 1496-019062, 1496-019063, 1496-019064, 1496-019065, 1496-019066, 1496-019067, 1496-019068, 1496-019069, 1496-019070, 1496-019071, 1496-019072, 1496-019073, 1496-019074, 1496-019075, 1496-019076, 1496-019083, 1496-019084, 1496-019085, 1496-019086, 1496-019087, 1496-019088, 1496-019089, 1496-019090, 1496-019091, 1496-019092, 1496-019093, 1496-019094, 1496-019095, 1496-019096, 1496-019097, 1496-019098, 1496-019099, 1496-019100, 1496-019101, 1496-019102, 1496-019103, 1496-019104, 1496-019105, 1496-019106, 1496-019107, 1496-019108, 1496-019109, 1496-019110, 1496-019111, 1496-019112, 1496-019113, 1496-019114, 1496-019115, 1496-019116, 1496-019117, 1496-019118, 1496-019119, 1496-019120, 1496-019123, 1496-019124, 1496-019125, 1496-019126, 1496-019127, 1496-019128, 1496-019129, 1496-019130, 1496-019131, 1496-019132, 1496-019133, 1496-019134, 1496-019135, 1496-019136, 1496-019137, 1496-019138, 1496-019139, 1496-019142, 1496-019143, 1496-019144, 1496-019145, 1496-019146, 1496-019147, 1496-019148, 1496-019149, 1496-019150, 1496-019172, 1496-019173, 1496-019325, 1496-019342, 1496-019343, 1496-019344, 1496-019345, 1496-019346, 1496-019347, 1496-019348, 1496-019349, 1496-019350, 1496-019351, 1496-019352, 1496-019353, 1496-019354, 1496-019355, 1496-019356, 1496-019357, 1496-019358, 1496-019359, 1496-019360, 1496-019361, 1496-019362, 1496-019363, 1496-019364, 1496-019365, 1496-019366, 1496-019367, 1496-019368, 1496-019369, 1496-019370, 1496-019371, 1496-019372, 1496-019373, 1496-019374, 1496-019375, 1496-019376, 1496-019377, 1496-019378, 1496-019379, 1496-019380, 1496-019381, 1496-019382, 1496-019383, 1496-019384, 1496-019385, 1496-019388, 1496-019389, 1496-019390, 1496-019391, 1496-019392, 1496-019393, 1496-019394, 1496-019395, 1496-019396, 1496-019397, 1496-019399, 1496-019400, 1496-019401, 1496-019402, 1496-019403, 1496-019404, 1496-019405, 1496-019406, 1496-019407, 1496-019408, 1496-019409, 1496-019410, 1496-019411, 1496-019412, 1496-019413, 1496-019414, 1496-019415, 1496-019416, 1496-019417, 1496-019418, 1496-019422, 1496-019423, 1496-019424, 1496-019425, 1496-019426, 1496-019427, 1496-019428, 1496-019429, 1496-019430, 1496-019431, 1496-019432, 1496-019433, 1496-019434, 1496-019435, 1496-019436, 1496-019437, 1496-019438, 1496-019439, 1496-019440, 1496-019441, 1496-019442, 1496-019443, 1496-019444, 1496-019445, 1496-019446, 1496-019447, 1496-019448, 1496-019449, 1496-019453, 1496-019454, 1496-019455, 1496-019456, 1496-019457, 1496-019458, 1496-019459, 1496-019462, 1496-019463, 1496-019464, 1496-019465, 1496-019466, 1496-019467, 1496-019468, 1496-019469, 1496-019470, 1496-019471, 1496-019472, 1496-019473, 1496-019474, 1496-019475, 1496-019476, 1496-019477, 1496-019478, 1496-019479, 1496-019480, 1496-019481, 1496-019482, 1496-019483, 1496-019484, 1496-019485, 1496-019514, 1496-019515, 1496-019516, 1496-019517, 1496-019518, 1496-019519, 1496-019520, 1496-019521, 1496-019522, 1496-019523, 1496-019598, 1496-019599, 1496-019600, 1496-019601, 1496-019602, 1496-019603, 1496-019604, 1496-019605, 1496-019606, 1496-019607, 1496-019617, 1496-019618, 1496-019619, 1496-019620, 1496-019621, 1496-019622, 1496-019623, 1496-019627, 1496-019628, 1496-019666, 1496-019667, 1496-019668, 1496-019669, 1496-019670, 1496-019671, 1496-019672, 1496-019673, 1496-019674, 1496-019675, 1496-019676, 1496-019677, 1496-019678, 1496-019679, 1496-019680, 1496-019681, 1496-019682, 1496-019683, 1496-019684, 1496-019685, 1496-019686, 1496-019687, 1496-019688, 1496-019689, 1496-019690, 1496-019691, 1496-019694, 1496-019695, 1496-019697, 1496-019698, 1496-019699, 1496-019700, 1496-019701, 1496-019702, 1496-019703, 1496-019704, 1496-019705, 1496-019706, 1496-019707, 1496-019708, 1496-019709, 1496-019710, 1496-019711, 1496-019712, 1496-019713, 1496-019714, 1496-019715, 1496-019716, 1496-019717, 1496-019718, 1496-019719, 1496-019720, 1496-019721, 1496-019722, 1496-019723, 1496-019724, 1496-019725, 1496-019726, 1496-019727, 1496-019728, 1496-019734, 1496-019735, 1496-019736, 1496-019758, 1496-019759, 1496-019760, 1496-019761, 1496-019762, 1496-019763, 1496-019789, 1496-019790, 1496-019791, 1496-019792, 1496-019793, 1496-019794, 1496-019795, 1496-019797, 1496-019798, 1496-019799, 1496-019800, 1496-019801, 1496-019802, 1496-019803, 1496-019804, 1496-019805, 1496-019806, 1496-019807, 1496-019808, 1496-019809, 1496-019810, 1496-019813, 1496-019814, 1496-019815, 1496-019816, 1496-019817, 1496-019818, 1496-019819, 1496-019820, 1496-019821, 1496-019822, 1496-019823, 1496-019824, 1496-019825, 1496-019826, 1496-019827, 1496-019829, 1496-019831, 1496-019832, 1496-019833, 1496-019834, 1496-019835, 1496-019836, 1496-019837, 1496-019838, 1496-019839, 1496-019840, 1496-019841, 1496-019842, 1496-019843, 1496-019844, 1496-019845, 1496-019846, 1496-019847, 1496-019848, 1496-019849, 1496-019850, 1496-019851, 1496-019852, 1496-019853, 1496-019854, 1496-019855, 1496-019856, 1496-019857, 1496-019858, 1496-019859, 1496-019860, 1496-019861, 1496-019862, 1496-019863, 1496-019864, 1496-019865, 1496-019866, 1496-019867, 1496-019868, 1496-019869, 1496-019874, 1496-019875, 1496-019876, 1496-019877, 1496-019878, 1496-019879, 1496-019880, 1496-019881, 1496-019882, 1496-019883, 1496-019884, 1496-019904, 1496-019952, 1496-019954, 1496-019955, 1496-019956, 1496-019957, 1496-019958, 1496-019959, 1496-019960, 1496-019961, 1496-019962, 1496-019963, 1496-019964, 1496-019965, 1496-019966, 1496-019967, 1496-019968, 1496-019969, 1496-019971, 1496-019972, 1496-019973, 1496-020093, 1496-020094, 1496-020095, 1496-020096, 1496-020097, 1496-020098, 1496-020099, 1496-020100, 1496-020101, 1496-020103, 1496-020104, 1496-020105, 1496-020106, 1496-020107, 1496-020108, 1496-020109, 1496-020110, 1496-020111, 1496-020112, 1496-020113, 1496-020114, 1496-020115, 1496-020116, 1496-020117, 1496-020118, 1496-020119, 1496-020120, 1496-020121, 1496-020122, 1496-020123, 1496-020124, 1496-020125, 1496-020126, 1496-020127, 1496-020128, 1496-020130, 1496-020139, 1496-020140, 1496-020141, 1496-020142, 1496-020143, 1496-020144, 1496-020145, 1496-020191, 1496-020192, 1496-020193, 1496-020194, 1496-020195, 1496-020196, 1496-020197, 1496-020198, 1496-020200, 1496-020201, 1496-020202, 1496-020208, 1496-020209, 1496-020273, 1496-020274, 1496-020275, 1496-020302, 1496-020320, 1496-020321, 1496-020322, 1496-020323, 1496-020324, 1496-020325, 1496-020326, 1496-020327, 1496-020328, 1496-020332, 1496-020335, 1496-020336, 1496-020337, 1496-020338, 1496-020339, 1496-020340, 1496-020341, 1496-020342, 1496-020343, 1496-020344, 1496-020345, 1496-020346, 1496-020347, 1496-020351, 1496-020369, 1496-020370, 1496-020373, 1496-020374, 1496-020396, 1496-020398, 1496-020399, 1496-020400, 1496-020401, 1496-020402, 1496-020403, 1496-020404, 1496-020405, 1496-020406, 1496-020407, 1496-020408, 1496-020409, 1496-020410, 1496-020411, 1496-020412, 1496-020432, 1496-020437, 1496-020646, 1496-020649, 1496-020788, 1496-020789, 1496-020790, 1496-020791, 1496-020792, 1496-020793, 1496-020794, 1496-020795, 1496-020796, 1496-020797, 1496-020798, 1496-020799, 1496-020805, 1496-020806, 1496-020808, 1496-020810, 1496-020811, 1496-020812, 1496-020814, 1496-020815, 1496-020817, 1496-020832, 1496-020842, 1496-020848, 1496-020850, 1496-020865, 1496-020873, 1496-020874, 1496-020875, 1496-020876, 1496-020877, 1496-020878, 1496-020879, 1496-020880, 1496-020881, 1496-020882, 1496-020883, 1496-020884, 1496-020885, 1496-020886, 1496-020887, 1496-020888, 1496-020889, 1496-020890, 1496-020899, 1496-020900, 1496-020901, 1496-020902, 1496-020903, 1496-020904, 1496-020905, 1496-020906, 1496-020907, 1496-020908, 1496-020910, 1496-020911, 1496-020912, 1496-020913, 1496-020914, 1496-020915, 1496-020916, 1496-020917, 1496-020918, 1496-020919, 1496-020920, 1496-020921, 1496-020922, 1496-020923, 1496-020924, 1496-020925, 1496-020926, 1496-020927, 1496-020928, 1496-020929, 1496-020930, 1496-020931, 1496-020933, 1496-020934, 1496-020935, 1496-020936, 1496-020939, 1496-020941, 1496-020942, 1496-020943, 1496-020944, 1496-020945, 1496-020946, 1496-020947, 1496-020948, 1496-020949, 1496-020950, 1496-021001, 1496-021002, 1496-021003, 1496-021004, 1496-021005, 1496-021006, 1496-021007, 1496-021008, 1496-021009, 1496-021010, 1496-021011, 1496-021012, 1496-021013, 1496-021014, 1496-021015, 1496-021016, 1496-021017, 1496-021018, 1496-021019, 1496-021020, 1496-021021, 1496-021022, 1496-021023, 1496-021024, 1496-021025, 1496-021026, 1496-021027, 1496-021028, 1496-021029, 1496-021030, 1496-021031, 1496-021032, 1496-021033, 1496-021034, 1496-021035, 1496-021036, 1496-021037, 1496-021038, 1496-021039, 1496-021040, 1496-021041, 1496-021042, 1496-021043, 1496-021044, 1496-021045, 1496-021046, 1496-021047, 1496-021048, 1496-021049, 1496-021050, 1496-021101, 1496-021102, 1496-021103, 1496-021104, 1496-021105, 1496-021106, 1496-021107, 1496-021108, 1496-021109, 1496-021110, 1496-021111, 1496-021112, 1496-021113, 1496-021114, 1496-021115, 1496-021116, 1496-021117, 1496-021118, 1496-021119, 1496-021120, 1496-021121, 1496-021122, 1496-021149, 1496-021150, 1496-021151, 1496-021152, 1496-021153, 1496-021155, 1496-021156, 1496-021157, 1496-021158, 1496-021159, 1496-021160, 1496-021161, 1496-021162, 1496-021163, 1496-021164, 1496-021165, 1496-021166, 1496-021167, 1496-021168, 1496-021169, 1496-021170, 1496-021171, 1496-021172, 1496-021306, 1496-021307, 1496-021308, 1496-021365, 1496-021366, 1496-021367, 1496-021368, 1496-021369, 1496-021370, 1496-021371, 1496-021372, 1496-021373, 1496-021374, 1496-021375, 1496-021376, 1496-021377, 1496-021378, 1496-021379, 1496-021380, 1496-021381, 1496-021382, 1496-021392, 1496-021393, 1496-021394, 1496-021397, 1496-021398, 1496-021399, 1496-021400, 1496-021401, 1496-021402, 1496-021403, 1496-021404, 1496-021405, 1496-021406, 1496-021407, 1496-021408, 1496-021409, 1496-021410, 1496-021411, 1496-021412, 1496-021413, 1496-021414, 1496-021415, 1496-021438, 1496-021439, 1496-021440, 1496-021441, 1496-021442, 1496-021444, 1496-021445, 1496-021446, 1496-021447, 1496-021448, 1496-021449, 1496-021450, 1496-021452, 1496-021453, 1496-021515, 1496-021519, 1496-021572, 1496-021573, 1496-021574, 1496-021575, 1496-021576, 1496-021577, 1496-021578, 1496-021579, 1496-021580, 1496-021581, 1496-021593, 1496-021608, 1496-021610, 1496-021614, 1496-021615, 1496-021616, 1496-021617, 1496-021618, 1496-021619, 1496-021620, 1496-021621, 1496-021622, 1496-021623, 1496-021624, 1496-021625, 1496-021626, 1496-021627, 1496-021628, 1496-021629, 1496-021630, 1496-021631, 1496-021632, 1496-021633, 1496-021634, 1496-021635, 1496-021636, 1496-021637, 1496-021638, 1496-021639, 1496-021640, 1496-021641, 1496-021643, 1496-021644, 1496-021645, 1496-021646, 1496-021648, 1496-021649, 1496-021650, 1496-021651, 1496-021652, 1496-021653, 1496-021654, 1496-021655, 1496-021656, 1496-021657, 1496-021658, 1496-021659, 1496-021660, 1496-021661, 1496-021662, 1496-021663, 1496-021664, 1496-021665, 1496-021666, 1496-021667, 1496-021668, 1496-021669, 1496-021670, 1496-021672, 1496-021673, 1496-021674, 1496-021676, 1496-021727, 1496-021728, 1496-021729, 1496-021730, 1496-021731, 1496-021732, 1496-021733, 1496-021735, 1496-021736, 1496-021737, 1496-021738, 1496-021739, 1496-021740, 1496-021741, 1496-021742, 1496-021744, 1496-021745, 1496-021746, 1496-021747, 1496-021752, 1496-021753, 1496-021807, 1496-021808, 1496-021809, 1496-021810, 1496-021811, 1496-021812, 1496-021813, 1496-021814, 1496-021815, 1496-021816, 1496-021817, 1496-021818, 1496-021819, 1496-021820, 1496-021821, 1496-021822, 1496-021823, 1496-021824, 1496-021825, 1496-021826, 1496-021827, 1496-021828, 1496-021829, 1496-021830, 1496-021831, 1496-021832, 1496-021833, 1496-021834, 1496-021835, 1496-021836, 1496-021837, 1496-021838, 1496-021839, 1496-021840, 1496-021841, 1496-021842, 1496-021843, 1496-021852, 1496-021853, 1496-021854, 1496-021855, 1496-021856, 1496-021857, 1496-021858, 1496-021859, 1496-021860, 1496-021861, 1496-021862, 1496-021863, 1496-021864, 1496-021865, 1496-021866, 1496-021867, 1496-021894, 1496-021895, 1496-021896, 1496-021897, 1496-021898, 1496-021899, 1496-021901, 1496-021902, 1496-021904, 1496-021905, 1496-021906, 1496-021907, 1496-021908, 1496-021909, 1496-021910, 1496-021911, 1496-021912, 1496-021913, 1496-021914, 1496-021915, 1496-021916, 1496-021917, 1496-021918, 1496-021919, 1496-021920, 1496-021921, 1496-021926, 1496-021927, 1496-021928, 1496-021929, 1496-021930, 1496-021931, 1496-021932, 1496-021933, 1496-021934, 1496-021935, 1496-021936, 1496-021937, 1496-021939, 1496-021940, 1496-021941, 1496-021942, 1496-021943, 1496-021944, 1496-021945, 1496-021946, 1496-021947, 1496-021948, 1496-021949, 1496-021950, 1496-021951, 1496-021952, 1496-021953, 1496-021954, 1496-021955, 1496-021956, 1496-021957, 1496-021959, 1496-021960, 1496-021961, 1496-021962, 1496-021963, 1496-021964, 1496-021965, 1496-021966, 1496-021967, 1496-021968, 1496-021969, 1496-021970, 1496-021971, 1496-021985, 1496-021986, 1496-021987, 1496-021988, 1496-021989, 1496-021990, 1496-021991, 1496-021992, 1496-021993, 1496-021994, 1496-021995, 1496-021996, 1496-022001, 1496-022002, 1496-022003, 1496-022015, 1496-022017, 1496-022018, 1496-022019, 1496-022020, 1496-022021, 1496-022022, 1496-022023, 1496-022024, 1496-022025, 1496-022026, 1496-022081, 1496-022082, 1496-022083, 1496-022084, 1496-022085, 1496-022086, 1496-022116, 1496-022117, 1496-022118, 1496-022119, 1496-022120, 1496-022121, 1496-022122, 1496-022124, 1496-022125, 1496-022126, 1496-022127, 1496-022128, 1496-022130, 1496-022131, 1496-022134, 1496-022168, 1496-022169, 1496-022178, 1496-022179, 1496-022180, 1496-022181, 1496-022182, 1496-022183, 1496-022184, 1496-022185, 1496-022186, 1496-022187, 1496-022188, 1496-022189, 1496-022190, 1496-022191, 1496-022192, 1496-022193, 1496-022194, 1496-022207, 1496-022208, 1496-022209, 1496-022210, 1496-022211, 1496-022212, 1496-022213, 1496-022214, 1496-022215, 1496-022216, 1496-022217, 1496-022218, 1496-022222, 1496-022223, 1496-022224, 1496-022225, 1496-022226, 1496-022227, 1496-022228, 1496-022229, 1496-022230, 1496-022231, 1496-022232, 1496-022233, 1496-022234, 1496-022235, 1496-022236, 1496-022237, 1496-022238, 1496-022239, 1496-022240, 1496-022241, 1496-022242, 1496-022243, 1496-022244, 1496-022245, 1496-022246, 1496-022247, 1496-022248, 1496-022249, 1496-022260, 1496-022269, 1496-022270, 1496-022271, 1496-022272, 1496-022325, 1496-022326, 1496-022335, 1496-022336, 1496-022337, 1496-022338, 1496-022339, 1496-022348, 1496-022350, 1496-022351, 1496-022352, 1496-022353, 1496-022354, 1496-022355, 1496-022356, 1496-022357, 1496-022381, 1496-022382, 1496-022383, 1496-022384, 1496-022385, 1496-022386, 1496-022387, 1496-022470, 1496-022471, 1496-022472, 1496-022473, 1496-022474, 1496-022478, 1496-022479, 1496-022480, 1496-022487, 1496-022488, 1496-022489, 1496-022490, 1496-022491, 1496-022492, 1496-022493, 1496-022494, 1496-022495, 1496-022496, 1496-022497, 1496-022498, 1496-022499, 1496-022500, 1496-022501, 1496-022502, 1496-022503, 1496-022504, 1496-022505, 1496-022506, 1496-022507, 1496-022508, 1496-022509, 1496-022510, 1496-022511, 1496-022512, 1496-022513, 1496-022514, 1496-022515, 1496-022516, 1496-022517, 1496-022518, 1496-022519, 1496-022520, 1496-022521, 1496-022522, 1496-022523, 1496-022524, 1496-022525, 1496-022526, 1496-022527, 1496-022528, 1496-022532, 1496-022533, 1496-022534, 1496-022535, 1496-022536, 1496-022537, 1496-022538, 1496-022539, 1496-022540, 1496-022541, 1496-022542, 1496-022543, 1496-022544, 1496-022545, 1496-022546, 1496-022547, 1496-022548, 1496-022549, 1496-022550, 1496-022551, 1496-022552, 1496-022553, 1496-022554, 1496-022555, 1496-022556, 1496-022557, 1496-022558, 1496-022559, 1496-022560, 1496-022561, 1496-022563, 1496-022564, 1496-022565, 1496-022566, 1496-022567, 1496-022568, 1496-022569, 1496-022570, 1496-022571, 1496-022622, 1496-022623, 1496-022624, 1496-022628, 1496-022629, 1496-022630, 1496-022649, 1496-022650, 1496-022651, 1496-022652, 1496-022653, 1496-022663, 1496-022664, 1496-022675, 1496-022676, 1496-022677, 1496-022678, 1496-022679, 1496-022680, 1496-022681, 1496-022682, 1496-022683, 1496-022684, 1496-022685, 1496-022686, 1496-022720, 1496-022721, 1496-022757, 1496-022758, 1496-022759, 1496-022760, 1496-022761, 1496-023048, 1496-023049, 1496-023051, 1496-023052, 1496-023053, 1496-023054, 1496-023056, 1496-023057, 1496-023059, 1496-023150, 1496-023163, 1496-023164, 1496-023165, 1496-023166, 1496-023183, 1496-023184, 1496-023185, 1496-023186, 1496-023187, 1496-023188, 1496-023189, 1496-023190, 1496-023191, 1496-023208, 1496-023231, 1496-023292, 1496-023300, 1496-023301, 1496-023302, 1496-023303, 1496-023304, 1496-023305, 1496-023306, 1496-023307, 1496-023308, 1496-023309, 1496-023310, 1496-023314, 1496-023315, 1496-023316, 1496-023317, 1496-023318, 1496-023319, 1496-023320, 1496-023321, 1496-023322, 1496-023323, 1496-023324, 1496-023325, 1496-023326, 1496-023327, 1496-023328, 1496-023329, 1496-023330, 1496-023331, 1496-023332, 1496-023333, 1496-023334, 1496-023335, 1496-023336, 1496-023337, 1496-023338, 1496-023339, 1496-023340, 1496-023341, 1496-023342, 1496-023343, 1496-023344, 1496-023345, 1496-023346, 1496-023347, 1496-023348, 1496-023349, 1496-023350, 1496-023351, 1496-023352, 1496-023353, 1496-023354, 1496-023355, 1496-023356, 1496-023357, 1496-023358, 1496-023359, 1496-023360, 1496-023363, 1496-023364, 1496-023365, 1496-023366, 1496-023367, 1496-023368, 1496-023369, 1496-023370, 1496-023371, 1496-023372, 1496-023373, 1496-023374, 1496-023375, 1496-023376, 1496-023377, 1496-023378, 1496-023382, 1496-023383, 1496-023384, 1496-023385, 1496-023386, 1496-023387, 1496-023393, 1496-023394, 1496-023395, 1496-023396, 1496-023399, 1496-023400, 1496-023401, 1496-023402, 1496-023403, 1496-023404, 1496-023405, 1496-023406, 1496-023407, 1496-023408, 1496-023409, 1496-023410, 1496-023417, 1496-023418, 1496-023419, 1496-023420, 1496-023422, 1496-023423, 1496-023424, 1496-023425, 1496-023426, 1496-023427, 1496-023428, 1496-023429, 1496-023430, 1496-023431, 1496-023432, 1496-023433, 1496-023434, 1496-023435, 1496-023436, 1496-023437, 1496-023438, 1496-023439, 1496-023460, 1496-023463, 1496-023464, 1496-023465, 1496-023466, 1496-023467, 1496-023468, 1496-023469, 1496-023470, 1496-023471, 1496-023472, 1496-023473, 1496-023474, 1496-023475, 1496-023476, 1496-023477, 1496-023478, 1496-023479, 1496-023480, 1496-023481, 1496-023482, 1496-023483, 1496-023484, 1496-023508, 1496-023509, 1496-023510, 1496-023511, 1496-023512, 1496-023513, 1496-023514, 1496-023515, 1496-023516, 1496-023517, 1496-023518, 1496-023519, 1496-023520, 1496-023572, 1496-023573, 1496-023574, 1496-023575, 1496-023576, 1496-023577, 1496-023578, 1496-023579, 1496-023580, 1496-023581, 1496-023582, 1496-023583, 1496-023584, 1496-023585, 1496-023586, 1496-023587, 1496-023588, 1496-023589, 1496-023590, 1496-023591, 1496-023592, 1496-023600, 1496-023601, 1496-023602, 1496-023603, 1496-023605, 1496-023606, 1496-023607, 1496-023608, 1496-023609, 1496-023613, 1496-023614, 1496-023615, 1496-023616, 1496-023617, 1496-023618, 1496-023619, 1496-023620, 1496-023621, 1496-023622, 1496-023623, 1496-023624, 1496-023625, 1496-023626, 1496-023627, 1496-023628, 1496-023629, 1496-023630, 1496-023631, 1496-023632, 1496-023636, 1496-023637, 1496-023643, 1496-023656, 1496-023657, 1496-023658, 1496-023659, 1496-023660, 1496-023661, 1496-023664, 1496-023665, 1496-023666, 1496-023667, 1496-023669, 1496-023670, 1496-023671, 1496-023672, 1496-023673, 1496-023674, 1496-023675, 1496-023676, 1496-023677, 1496-023678, 1496-023679, 1496-023680, 1496-023731, 1496-023732, 1496-023733, 1496-023734, 1496-023735, 1496-023736, 1496-023737, 1496-023738, 1496-023744, 1496-023757, 1496-023758, 1496-023759, 1496-023760, 1496-023761, 1496-023762, 1496-023763, 1496-023764, 1496-023765, 1496-023766, 1496-023767, 1496-023768, 1496-023769, 1496-023770, 1496-023771, 1496-023772, 1496-023773, 1496-023774, 1496-023775, 1496-023776, 1496-023777, 1496-023778, 1496-023779, 1496-023780, 1496-023781, 1496-023782, 1496-023783, 1496-023784, 1496-023785, 1496-023786, 1496-023787, 1496-023788, 1496-023789, 1496-023790, 1496-023791, 1496-023792, 1496-023793, 1496-023794, 1496-023796, 1496-023797, 1496-023798, 1496-023799, 1496-023800, 1496-023830, 1496-023831, 1496-023832, 1496-023833, 1496-023834, 1496-023835, 1496-023836, 1496-023837, 1496-023838, 1496-023839, 1496-023840, 1496-023841, 1496-023842, 1496-023843, 1496-023845, 1496-023846, 1496-023847, 1496-023848, 1496-023849, 1496-023850, 1496-023851, 1496-023852, 1496-023853, 1496-023854, 1496-023855, 1496-023939, 1496-023940, 1496-023941, 1496-023942, 1496-023943, 1496-023944, 1496-023945, 1496-023946, 1496-023947, 1496-023948, 1496-023953, 1496-023954, 1496-023955, 1496-023956, 1496-023957, 1496-023958, 1496-023959, 1496-023960, 1496-023961, 1496-023962, 1496-023963, 1496-023964, 1496-023965, 1496-023966, 1496-023967, 1496-023968, 1496-023969, 1496-023970, 1496-023971, 1496-023972, 1496-023973, 1496-023974, 1496-023975, 1496-023976, 1496-023978, 1496-023979, 1496-023980, 1496-023981, 1496-023982, 1496-023983, 1496-023984, 1496-023985, 1496-023986, 1496-023987, 1496-024028, 1496-024096, 1496-024097, 1496-024098, 1496-024099, 1496-024100, 1496-024101, 1496-024102, 1496-024104, 1496-024105, 1496-024106, 1496-024107, 1496-024108, 1496-024109, 1496-024110, 1496-024111, 1496-024112, 1496-024113, 1496-024114, 1496-024115, 1496-024116, 1496-024117, 1496-024118, 1496-024119, 1496-024120, 1496-024121, 1496-024122, 1496-024167, 1496-024168, 1496-024169, 1496-024170, 1496-024171, 1496-024173, 1496-024175, 1496-024176, 1496-024177, 1496-024178, 1496-024179, 1496-024180, 1496-024181, 1496-024182, 1496-024183, 1496-024184, 1496-024185, 1496-024186, 1496-024187, 1496-024188, 1496-024190, 1496-024191, 1496-024192, 1496-024193, 1496-024194, 1496-024195, 1496-024196, 1496-024197, 1496-024198, 1496-024199, 1496-024200, 1496-024201, 1496-024202, 1496-024203, 1496-024204, 1496-024205, 1496-024206, 1496-024207, 1496-024208, 1496-024209, 1496-024210, 1496-024211, 1496-024212, 1496-024213, 1496-024214, 1496-024215, 1496-024216, 1496-024217, 1496-024218, 1496-024219, 1496-024220, 1496-024222, 1496-024223, 1496-024224, 1496-024225, 1496-024226, 1496-024227, 1496-024228, 1496-024229, 1496-024230, 1496-024231, 1496-024232, 1496-024233, 1496-024234, 1496-024235, 1496-024236, 1496-024237, 1496-024238, 1496-024239, 1496-024240, 1496-024241, 1496-024245, 1496-024246, 1496-024247, 1496-024250, 1496-024251, 1496-024252, 1496-024253, 1496-024254, 1496-024255, 1496-024303, and 1496-024304. |
Recalling Firm/ Manufacturer |
Spacelabs Healthcare, Incorporated 5150 220th Ave Se Issaquah WA 98029-6834
|
For Additional Information Contact |
425-657-7200 Ext. 5970
|
Manufacturer Reason for Recall |
Potential for module to fail to obtain a NIBP reading on hypertension patients.
|
FDA Determined Cause 2 |
Process control |
Action |
On July 22, 2008 an "Urgent MEDICAL DEVICE CORRECTION" letter was mailed to all the firm's customers. The letter states the problem at hand, when the problem would likely to occur and instructions of a way to mitigate the risk, until the modules can be recalibrated . The firm's Customer Connectivity Center will confirm all affected customers have been notified. Spacelabs Technical Support will be recalibrating all modules personally and ensuring all corrections have been applied to all affected customers of record. Contact Spacelabs Healthcare at 1-800-522-7025 for assistance. |
Quantity in Commerce |
22,079 |
Distribution |
The product is distributed nationwide and internationally to medical facilities. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = DSI and Original Applicant = SPACELABS MEDICAL, INC.
|
|
|
|