| Class 2 Device Recall GEMINI TF PET/CT Scanners |  |
Date Initiated by Firm | November 24, 2008 |
Date Posted | November 17, 2009 |
Recall Status1 |
Terminated 3 on September 24, 2010 |
Recall Number | Z-0287-2010 |
Recall Event ID |
50135 |
510(K)Number | K033357 K052640 |
Product Classification |
PET/CT Scanner accessory - Product Code JAK
|
Product | Respiratory Gating System, Catalog Number: NPTA540, Model #4535 674 26701, for the following affected CT/PET scanning device systems : GEMINI TF CT/PET Systems 16 Slice, GEMINI TF CT/PET System 16 Slice Mobile; GEMINI TF CT/PET System 64 Slice Mobile; and GEMINI TF CT/PET System 64 Slice. (Model Numbers: 16 slice: 4535 679 83931; 64 slice Mobile: 4535 674 47551; and 64 slice: 4535 679 94741. |
Code Information |
The affected Serial Numbers are: Gemini TF 16 Slice: 7005, 7052, 7087, 7112, 7121, 7138, 7115, 7135, 7006, 7008, 7015, 7021, 7020, 7026, 7035, 7039, 7041, 7049, 7053, 7058, 7060, 7069, 7075, 7080, 7083, 7084, 7098, 7117, 7118, 7142; Gemini TF 16 Slice Mobile: 7059, 7104; Gemini TF 64 Slice Mobile: 7092; Gemini TF 64 Slice: 7012, 7013, 7032, 7071, 7082, 7091, 7126, 7132, 7088, 7103, 7136, 7027, 7109, 7105, 7028, 7167, 7078, and 7110. |
Recalling Firm/ Manufacturer |
Philips Medical Systems (Cleveland) Inc 595 Miner Road Cleveland OH 44143-2131
|
For Additional Information Contact | 440-483-7000 |
Manufacturer Reason for Recall | A leak between the Tube Interface and the Outlet Tube of the transducer for the Respiratory Gating System component may result in a failure of the device to produce respiratory correlated images. |
FDA Determined Cause 2 | Process design |
Action | Philips Medical Systems Nuclear Medicine division initiated this recall by sending a 'Customer Advisory Notification to their customer accounts on 11/24/2008. This notification informs the customer of the potential for the device equipped with the Pulmonary Gating Option to develop an air leak which could result in the device being unable to produce the desired respiratory correlated images. The notification provides field test instructions and recommends that the customer test the bellows on each day of use in order to assess that the unit is functioning properly. If the bellows device fails the test the recalling firm recommends that the unit not be used and that the recalling firm be contacted via telephone or E-mail. The firm intends to conduct a field correction of the affected units in which their Field Service Engineers will install upgrade kits on the affected devices tentatively scheduled to begin in September, 2009. |
Quantity in Commerce | 51 units |
Distribution | Worldwide distribution: USA, Puerto Rico, Canada, Belgium, China, Denmark, Egypt, France, Germany, Hong Kong, India, Japan, Netherlands, Pakistan, Philippines, Poland, Saudi Arabia, Singapore, South Korea, Spain, and Sweden. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = JAK
|
|
|
|