Date Initiated by Firm |
November 27, 2007 |
Date Posted |
January 27, 2009 |
Recall Status1 |
Terminated 3 on August 01, 2013 |
Recall Number |
Z-0930-2009 |
Recall Event ID |
50169 |
510(K)Number |
K023589
|
Product Classification |
Powered Wheelchair - Product Code ITI
|
Product |
Invacare Storm TDX SP Power Wheelchair with Stability Lock, Model numbers: TDXSP, TDXSP-CG, and TDXSP-MCG
To provide mobility to a person restricted to a sitting position. |
Code Information |
Serial Numbers: 07EE006037 07GE001214 07GE005740 07HE003498 06JE003608 07CE002600 07DE002131 07GE004997 07GE004998 07GE006523 07AE005120 07CE002390 06IE000118 07FE004477 07EE006245 07BE001102 07BE001317 07BE001318 07CE002595 07EE001795 07FE007661 07FE007659 07FE001517 07GE000030 07EE006036 07GE003396 07HE001436 07HE002698 07EE001586 06LE003477 07BE003359 07HE006445 07HE001075 06LE003480 07FE006897 07DE002127 07DE005069 07GE005156 07GE006527 07GE006536 07FE007531 07GE005875 06HE010677 07DE003566 07AE001341 07DE002490 07EE004963 07FE006900 07DE002284 07EE006259 06HE010682 07HE005502 07FE003815 07AE006319 07BE001333 07BE003357 07DE002500 07EE006262 07FE007283 07GE005158 07HE002851 07HE004228 07HE006761 06JE001557 07EE001484 07BE006828 07DE001930 07FE007282 07GE001382 07GE005171 06LE006034 06HE008565 07EE006269 07FE002703 07BE003353 07DE001641 07EE001928 07GE004167 07EE000542 07GE003500 07FE004471 07GE003231 07GE004181 07EE006272 07GE004169 07EE001939 07IE000138 07IE000705 07HE001768 07HE006179 07HE006185 07EE004942 07GE001142 07GE006524 07HE002308 07HE002828 07HE003051 07EE004629 07EE002818 07BE001123 07HE004998 07GE006742 07CE008701 07BE003358 06JE003901 07AE001010 07BE006722 07GE000060 07GE004183 07CE002612 07CE007711 07FE000069 06JE006574 06IE008343 06JE006264 07GE005530 07GE006874 07GE006876 07HE005361 07HE005604 07BE005561 07DE005076 07EE001318 07AE003548 07EE006996 06HE010436 06JE009097 07BE001104 07BE005147 06JE005455 07AE002036 06LE002631 07BE001120 07CE005290 06IE004481 07EE004971 07BE005138 06HE008573 07CE006299 07CE005278 07AE003012 07HE001772 07FE006661 07FE004822 07EE005196 06KE006599 07CE006357 07FE000071 07DE003169 07GE001975 07GE005160 07EE004621 07HE002829 07EE005171 07FE003581 06LE004873 07BE005760 07BE006222 07DE004798 07EE005174 07AE003013 07AE004640 07DE004804 07GE001370 07IE000113 07IE001335 07GE007317 07HE004214 07HE004755 06KE003402 07FE000862 07EE001092 07EE001096 07FE005137 07GE000043 07DE003993 07GE006521 07HE005492 07HE006451 07FE006907 07GE003515 07FE000884 07GE002441 06LE002012 07DE002120 07EE004631 07GE001131 07HE005654 07BE001323 07HE002619 06JE008362 06JE009104 06IE004485 07EE006021 07GE000062 07GE000804 07HE004743 07HE006180 07HE006181 07CE008189 07DE005062 07AE002034 07DE002290 07FE006901 07FE006918 07FE005903 06HE009622 07FE006908 07GE004925 07GE001374 07FE002714 07BE005356 07BE005148 07FE006275 07GE000793 07FE005889 07EE006011 07HE007050 07GE006526 07AE003737 07CE004835 07BE001117 07IE000105 07GE006944 07HE001767 07AE003552 07DE002487 07FE001077 07GE002442 07HE001213 07HE004477 06LE007225 06LE001325 07EE007012 07FE000023 07EE007013 07DE002497 07DE002496 06JE007421 07EE005407 07GE002243 07FE006383 07GE001977 07GE001981 07HE007035 07GE004171 07GE006948 07GE007323 07GE000066 07GE004906 07GE004905 07EE002509 07FE002202 07GE003777 07HE005265 06IE008338 07GE004995 07CE005784 06IE004749 07EE006611 07FE002713 07BE004937 06KE004473 07FE003584 07HE001452 07AE005911 07FE006064 07GE005902 07HE004469 07HE005260 07GE005874 07HE003344 07HE001455 07CE004833 07FE006063 07CE001573 07IE000960 07IE001328 07FE006065 07EE003205 07DE003991 07EE007019 07EE006604 07FE006913 07GE005166 07HE001073 07HE003716 07FE001072 06LE001322 07AE006937 07CE007185 07CE001560 07EE006997 07DE002280 06KE008637 07FE005117 07GE002232 07FE000876 07GE004918 07GE005891 07GE006759 07EE001330 06IE004748 07FE006066 07FE007537 07HE001450 07AE001605 07CE006302 07AE002038 07BE001328 06LE006735 07CE001570 07CE002605 07CE007697 07DE005255 07CE002608 07CE003326 07DE002129 07DE005413 07FE002881 07FE002441 07GE001973 07IE000405 07IE001340 07GE006752 07HE002608 07HE003063 07HE004226 07EE006255 06JE009103 07DE004594 07CE008940
07DE003985 07DE004598 07EE003203 07FE002207 07EE005190 07GE001970 07EE003823 07DE005406 07DE002281 06LE004874 07CE002386 06KE008638 07CE001576 07FE002706 07DE005264 07GE001979 07GE001978 07FE005138 07HE001052 06KE007465 07AE002032 07EE004607 07DE003821 07CE008936 07EE004615 07CE005541 07EE003204 07DE004813 07EE002819 07FE005892 07FE005129 07GE000799 07GE002427 07GE004932 07EE001095 07EE002511 07GE000802 07GE003237 06HE010685 07GE005443 07GE004940 07GE005430 07GE005650 07GE005907 07HE002316 07HE002320 07HE003064 07HE004990 07HE006206 07EE004471 07EE002976 07DE003402 07EE002975 07FE000015 07HE003057 07HE005642 06KE003130 07CE000068 07GE001130 07HE004738 07CE007692 07DE003562 07EE000530 07CE002192 07DE003819 07DE006051 07FE004807 07HE005493 07AE003550 07FE004467 06HE010425 07CE001565 07CE006021 07DE004797 07GE002422 07GE000049 07GE000055 07HE001217 07HE006763 06LE007423 07BE001105 07GE006746 07BE005149 07BE001321 07HE001943 07HE003725 07DE000586 06KE003132 06KE007712 07AE005372 07AE005371 07DE001391 07FE003821 07GE002437 07CE004023 07GE003767 07EE005177 07DE003836 07CE005535 07AE004635 07EE001927 07FE005118 07AE002251 06IE004761 07AE007125 07BE006827 07CE005791 07DE003834 07EE001331 07GE005646 07FE005128 07HE002824 07HE002839 07HE004476 07BE004734 07AE001609 07GE002424 07FE002194 07FE003822 07FE004480 07GE004188 07GE006753 07EE001767 07DE004809 07FE002204 07EE002977 07CE008948 07GE006763 07EE004968 07BE003598 07CE007705 07DE006454 07FE004474 07GE005445 07FE004473 07IE000103 06KE003981 07CE005530 07AE007129 07EE006256 07GE006736 07CE002197 07DE004605 07EE002506 07EE001482 07GE005652 07GE002244 07HE002620 07HE003346 06LE003475 06LE007426 06LE005909 06LE007421 07AE002696 07AE005608 07BE001592 07DE003563 07CE003801 07EE004945 07GE003496 07EE005194 07EE003983 07FE002887 07GE000071 07HE001068 07HE001069 07BE001827 07AE005910 07EE001478 07EE004946 07FE002190 07AE006939 07DE005414 07EE002825 07FE003826 07EE005403 07GE002433 07FE007280 07HE007036 07IE000401 07HE000002 07HE001058 07HE001218 07HE006172 07HE006174 07HE006192 07HE001453 07AE003545 07HE004973 07HE006752 07DE002509 07EE001102 07CE006313 07CE007184 07EE001490 07GE000032 06LE001533 07GE005655 07HE002314 07HE002606 07DE002274 07BE003350 06KE006353 07EE001766 07FE006769 07EE001337 07GE005163 07CE007696 07CE002596 07FE007525 07DE004808 07FE003585 07GE003233 07GE004179 07IE000950 07GE006747 07HE004977 07BE005767 07DE005420 07AE007348 07EE007014 07EE005179 07FE006651 07EE004943 07GE003398 07FE004464 07AE003734 07AE007630 07HE006755 07AE003727 07AE005366 07DE003818 07DE003982 07DE001395 07HE002315 07HE002602 06LE005916 07GE002248 07GE006934 07EE006999 07EE007009 07FE000873 07FE007548 07DE003817 07GE005659 07DE004796 07EE005183 07HE001762 07HE002595 07HE004462 06LE003725 07BE006822 07CE005536 07CE005783 07FE002695 06LE001329 07DE005265 07DE005056 07BE002024 07BE002032 07HE000003 07HE001212 06JE005451 07GE002426 07EE006603 07HE006175 07FE002208 07HE006447 07DE002286 07GE001128 07DE003577 07IE000954 07HE003741 06HE010687 07BD006824 07DE004811 07DE004814 07CE002207 07GE006540 07HE001771 07HE001788 07HE006460 07GE005168 07GE003774 07FE005905 07GE003770 07GE003759 07FE003828 07GE003024 07BE006826 07DE003833 07DE002493 07EE006605 07HE002312 07EE006258 07GE005886 07AE003006 07DE001637 07CE000069 07FE000014 07FE006070 07GE004177 07HE004752 07HE005640 06LE001717 07DE005054 07DE000588 07FE002427 06KE006075 07IE000379 07HE001776 07HE006747 07HE006760 06JE008691 06LE002632 07FE003811 07DE003560 07EE004606 06LE003283 07GE005642 07GE003753 07EE004475 07GE003504 07FE006909 07IE000383 07GE004924 07GE004931 07GE005898 07HE006866 07HE006867 07HE006868 07HE006869 07HE006870 07HE006871 07HE006872 07HE006873 07HE006874 07HE006875 07CE006024 07DE003164 07EE001485 07GE003013 07DE006263 07EE002826 07GE006735 07HE002292 07HE002307 07GE005152 07GE005150 07BE005355 06LE001332 07CE007190 07FE006272 07DE005064 07FE004469 07FE000027 07GE004937 07GE003023 07HE003058 07HE003728 07HE006467 07DE006046 07EE006246 07HE000012 07HE001761 07FE002439 07EE004598 07BE002031 07CE008699 07GE001976 07EE006010 07FE001076 07GE003232 07GE005442 07GE003760 07HE000018 07HE001774 07HE001944 07HE003062 07HE004225 07HE005499 07HE006750 07EE005406 07BE004732 07AE003018 07EE004653 07FE004816 07HE004227 07CE006020 07FE007530 07AE007346 07FE005125 07DE004595 07BE005560 07DE003400 07CE005276 07IE000112 07HE002966 07HE004231 06LE003280 07AE003730 07EE001479 07EE006015 07FE000032 07DE003832 07EE007018 07FE004813 07IE000375 07HE001444 07HE002306 07HE005652 07HE005833 07HE005834 07HE006203 07GE003755 07GE004180 07GE004184 07FE007552 07DE003394 07FE005353 06KE008635 07GE005441 07DE002279 07BE005757 07AE005364 06JE003391 07CE002394 07BE006486 07CE005549 07CE004842 06KE003980 07EE007000 07HE004218 07HE001934 07DE003405 07DE006252 07HE006468 07AE001017 07AE001016 07AE001512 07AE001511 07CE006349 07CE006347 07CE006350 07CE002382 07CE002594 07CE006348 07GE005732 07FE007663 07GE002553 07GE003399 07GE005730 07GE005731 07GE005733 07GE005734 07GE005735 07GE005736 07GE005737 07GE005738 07GE005739 07AE001019 07AE001018 07AE006938 07DE006473 07AE002033 07CE002400 07FE006283 07DE001632 06LE005161 06LE006743 06LE006739 07FE002428 07GE000037 06LE004002 06LE007425 06IE003837 06IE002119 06JE007420 07BE005359 07CE006025 07AE002484 07DE001929 07CE008831 07FE006660 07DE006260 07EE005408 07GE003764 07GE003769 07GE004186 07FE003601 07GE003517 07GE006952 07HE001447 07HE005503 07HE006758 06IE004750 07FE003588 07GE004929 07IE000128 07IE001318 06KE007258 07FE001518 06KE007711 07FE004810 07DE005407 07FE007267 07GE003002 07AE004645 07CE008692 07BE006829 07GE005438 07EE004972 07HE000005 07HE004983 06LE001328 07BE003600 07FE005897 07HE001458 07AE002037 07AE003023 06LE007037 07FE005893 07BE003594 07BE002263 07BE006821 07DE003558 07EE004472 07DE003163 07GE003494 07CE007698 07CE008186 07CE008945 07EE002978 07DE003168 07GE005446 07GE005439 07FE000872 07FE005894 07FE002879 07GE003503 07GE004921 07GE004173 07IE000110 07GE006933 07HE001773 07HE001937 07HE002611 06JE007419 06JE008034 07EE006264 07HE001933 07EE001924 06LE001722 07HE001196 07HE002287 07CE008932 07DE003177 07CE006032 07DE002498 07FE002442 07HE005656 07DE004816 07AE003005 07FE007024 07EE005193 07FE005143 06JE008689 07AE003547 07HE003348 07CE008183 07DE003398 07HE003343 07HE005256 07HE001195 06JE006328 07CE001563 07CE005788 07EE006017 07GE005879 07EE006602 07FE002210 07HE002853 07FE005895 07HE002836 07DE004592 07HE006454 07GE006937 07FE003602 07EE000010 07EE000011 07EE004966 07DE022294 07HE003735 07BE005362 07DE002130 06IE003836 07CE006569 06KE008634 06LE005917 07IE000681 07HE002293 07AE004636 07BE004934 07DE003576 07FE006261 07BE004932 07FE002874 07FE005901 06JE006259 06IE008339 07FE005354 07GE001367 07GE002421 07CE003614 07FE006247 07FE007539 07GE002246 07GE001980 07GE003507 07GE003756 07HE001216 07HE005003 07AE001342 07EE006023 07AE006100 07BE005557 07BE006485 07BE006786 07GE004920 07FE000881 07HE000011 07FE006253 07DE005271 06JE009095 07BE003596 07GE002428 07GE005639 06JE005910 06JE007418 07FE006279 07HE005646 07HE005648 07GE003223 07DE003401 07HE000008 07HE001782 07HE002297 07HE002826 07GE000789 07BE003599 07DE002128 07EE004637 07HE007039 07GE006960 07CE005279 07AE007633 07CE002611 07EE006257 07DE004604 07DE005060 07EE000007 07FE001075 07GE003499 07DE002302 07EE001937 07GE000072 07IE000136 07HE004747 07HE006169 07BE001595 07CE006033 07EE000540 07GE003780 07GE007335 07HE001221 07HE003736 07HE006465 07FE006252 07FE002882 07FE001525 07GE004175 07FE006657 07IE000685 07GE006956 07HE003732 07FE007526 07IE001324 07IE001326 07DE004591 07FE002191 07FE001074 07EE002984 07CE006355 07FE002189 07FE003817 07GE005167 07FE006652 07HE004212 06LE005750 07BE004075 07FE005124 07BE001103 07EE002982 07DE005262 07EE001933 07GE000045 07HE005262 07AE005909 07AE005908 07AE005119 07AE005118 07FE002455 07EE005205 07AE002692 07AE002693 06HE009619 07BE006231 07GE007312 07HE002609 07HE003743 06IE009397 07HE003720 07HE007032 07GE006530 07HE001441 07HE002607 06JE005450 06LE000185 06LE004661 07AE003010 07CE003605 07GE005429 07CE006308 07CE005288 07DE002287 07BE006229 07EE006012 06IE004483 07GE005161 07EE006013 07GE004936 07IE000394 07GE005449 07GE005878 07GE005895 07GE005906 07HE002831 07HE002835 07HE004473 07EE001930 07CE008686 07FE002770 07GE002430 07GE005882 07GE006761 07HE003722 07CE004840 07AE005367 07DE002491 07DE001932 07GE000058 07HE004230 07GE001120 07EE003210 07DE005058 07CE001561 07EE001316 07EE004474 07GE000067 07GE001137 07GE004935 07HE006198 07HE001446 07EE002812 07IE000699 07GE007327 07FE001521 07EE001476 07HE000035 07HE001459 07FE002870 07GE001136 07EE003217 07HE002617 07HE002852 07HE003729 07HE006176 07DE003561 07DE005254 07FE004809 07GE001971 07GE003498 07GE006939 07GE007314 07HE000004 07HE006441 07BE001326 06LE004003 07AE003546 07DE003162 07HE002849 06LE001718 07GE005435 07BE005360 07FE001530 07GE006738 07FE000007 07FE003808 07EE006995 07CE003324 07EE006268 07FE000877 07IE000495 06HE010689 07EE003208 07BE006479 07BE005766 07DE001651 07FE001079 07GE002432 07GE002431 07FE005907 07FE005908 07GE003513 07HE003751 07HE005495 07IE000391 07EE004470 07FE005890 07BE006488 07FE007543 07GE003505 07GE000070 07GE001982 07IE000386 07GE005894 07GE005897 07DE002495 07FE007547 07FE007553 06IE004470 07BE001111 07FE000879 06JE008283 07AE006936 07EE007038 07CE008688 07CE008180 07DE001924 07GE001362 07GE003225 07FE007535 07DE006467 06KE006641 07EE004633 07GE003240 07EE002987 07GE006758 07HE006168 07FE000002 07CE002399 07DE006058 07AE001184 07HE006188 07AE002472 07AE002249 07BE005353 07DE002275 07AE002250 07DE003574 07AE007627 07EE004617 07GE000797 07IE000377 07GE004907 07HE003050 07DE002121 06LE000186 07BE004079 07CE005795 07FE006068 07FE005358 07EE002983 07FE006903 06HE009620 07EE004599 07GE005644 07BE004733 07DE004597 07GE004916 07GE006957 07HE001454 07HE001766 07BE004927 07DE005061 07EE004970 07FE000030 07FE005140 07EE005412 07EE004975 07IE000122 07IE000376 07GE006954 07HE000001 07HE000032 07HE001061 07HE001063 07HE001457 07HE002614 07HE003352 07HE004220 07HE005267 07HE005649 07HE006191 07CE009275 07BE006224 07FE002694 07CE002087 07CE003327 07GE005654 07BE006221 07BE003597 07EE002811 07EE003200 07FE002435 07FE002445 07FE004824 07FE006902 07GE006935 07HE004740 07HE004987 07AE003019 07CE009280 07FE005900 07DE003813 07EE004959 07DE004596 07GE000065 07IE000404 07HE004981 07FE001520 06KE002881 07BE005354 07DE003393 07FE002193 07EE006609 07FE002873 07FE006139 06KE003406 07GE006739 07HE002299 07FE006067 07FE001524 07FE006262 07GE005440 07GE001380 07GE001139 07GE003516 07GE004923 07IE000104 07IE000111 07IE000115 07IE000129 07IE000130 07IE000385 07IE000397 07IE000686 07IE000697 07GE005437 07HE001197 07HE001938 07HE003066 07HE003342 07HE003350 07HE003353 07HE003727 07HE004213 07HE004988 07HE006189 07HE006754 07CE001559 07GE001365 07FE006905 07IE000964 07HE004739 07BE001596 07EE004967 07GE004941 07GE002434 07FE004476 06KE000131 06LE002628 07FE004806 07FE004804 07FE004802 07FE004803 06LE001503 06KE006594 06KE006593 07AE005359
07AE005360 07AE005361 07AE005362 07AE005358 07AE005349 07AE005351 07AE005353 07AE005355 07AE005357 07FE002692 07AE005350 07AE005352 07AE005354 07AE005356 07AE005348 07EE003223 07EE002008 07FE003842 07FE003841 07FE007660 07GE003855 07GE002112 07EE006277 07EE004982 07EE003222 07FE003605 07FE000068 07IE000781 07GE003602 07GE003603 07GE004265 07GE005289 07GE005529 07GE006012 07HE000119 07GE003853 06KE007673 07CE004897 07DE006071 06LE006732 06JE008282 06KE007669 06JE002731 06JE004162 06KE003106 07AE006317 07BE006475 07BE005755 07BE006720 07BE006476 07CE005822 07CE005823 07CE005824 07CE005825 07CE005826 07CE005827 07CE005828 07CE005829 07CE009408 07DE002115 07DE002272 07CE002381 07CE005831 07CE005832 07CE005833 07CE005834 07CE005830 07CE006354 07CE006353 07CE006352 07CE006351 07CE005836 07CE005837 07CE005838 07CE005839 07CE005835 07EE007043 07EE004644 07FE006137 07FE006136 07FE006135 07FE006131 07FE006130 07FE006132 07FE005981 07EE007050 07EE007051 07EE007052 07EE007048 07EE007049 07FE006133 07FE006379 07EE007053 07EE007054 07EE007047 07EE007045 07EE007046 07EE007044 07GE003109 07GE003108 07CE005819 07CE005820 07CE005821 07DE003555 07EE007056 07EE007055 07FE006134 07CE005555 07CE005556 07EE002809 07EE004646 07EE004647 07EE004648 07FE005455 07DE003556 07FE004487 07HE007142 07IE000489 07FE004488 07FE004489 07FE004490 07FE007017 07FE007018 07FE007019 07FE007020 07FE007021 07FE007022 07FE007405 07FE007406 07FE007407 07GE003110 07GE003386 07GE003387 07GE003388 07GE003389 07GE003390 07GE003391 07GE003392 07GE003393 07GE003394 07GE003395 07GE006626 07EE007041 07EE007042 07HE001141 07HE002967 07HE002968 07HE002969 07HE002970 07HE002971 07HE002973 07HE002974 07HE002975 07HE002976 07HE003157 07HE004308 07HE005122 07HE005358 07HE005359 07HE005360 07HE000023 07FE002877 07BE001604 07DE002283 07BE006484 07GE003222 07HE005263 07AE007344 06IE004474 07EE006024 07CE003609 07CE009407 07GE005884 07GE006744 07CE008928 07IE000959 07AE007126 07BE001109 07GE000038 07FE004466 07EE001333 07GE003227 07EE001774 07BE004936 07FE006650 07GE007319 07HE001067 07AE006098 06LE002635 06LE007224 07CE006307 07BE004931 07CE009282 07DE003399 07BE002266 07EE007021 07GE006949 07DE000591 07HE003715 07BE004929 07BE001322 06KE007306 07AE007122 07DE003816 07EE006251 07EE001922 07DE005068 07DE005067 07FE003816 07BE001122 07HE005643 06KE008639 07FE004812 07AE003020 07AE003015 07CE002198 07FE002888 07DE000589 07CE007701 07FE003813 07HE003052 07FE002429 07DE002276 07CE005782 07DE006048 07DE006253 07GE003224 07GE003228 06LE004875 07AE005369 07CE007186 07CE007187 07BE001597 07CE005801 07DE003176 07EE001932 07FE002206 07FE002704 07GE005447 07EE002820 07FE006270 07GE000041 07GE005647 07HE002288 07HE002600 07HE004978 06IE009765 07AE005117 07AE005116 07EE005411 07EE005397 06JE009094 07BE006723 07EE002504 07FE000003 07FE006645 07CE009279 07EE001778 07FE007540 07HE001451 07HE002834 07BE004073 07GE002444 07HE004215 07CE005560 07CE005322 07CE005323 07CE005558 07CE005321 07CE005559 07CE005325 07CE005324 07EE004651 07EE007039 07EE004652 07EE001585 07DE006459 07FE000871 07HE001208 07HE006469 07CE002194 07EE004596 07HE003738 07DE002494 07GE003854 07HE001053 07GE005904 07IE000691 07IE000700 07DE005415 07GE005656 07HE004237 07FE002431 07GE000059 07GE001134 07EE004632 07GE000057 07IE000396 07GE006534 07GE006945 07HE004974 07CE007183 07GE000047 07GE000040 07AE002248 07FE002863 07GE004163 07CE005798 07FE000033 07GE004996 07FE003589 07EE001487 07IE000116 07BE005756 06IE004472 07GE000031 07CE006026 07EE004957 07GE002436 07AE003008 07EE003216 07FE002878 07IE000123 07GE006955 07HE001062 07HE006753 06IE008341 07BE001101 06JE006268 06JE006263 06IE008345 06JE006265 06JE006255 07HE004310 06LE003474 07AE007342 07CE008931 07DE001640 07CE008937 07GE002435 07EE004616 07CE003620 07DE001663 07HE007053 07GE006953 07GE004913 06HE010430 07HE002843 07EE005393 07EE000001 07EE005396 07CE005800 07GE003497 07HE003730 07BE002274 06KE006912 06IE004477 07FE007538 07GE001135 07CE002402 07FE006281 07FE006282 07CE002401 07FE007290 07HE002837 07DE001389 07HE004975 06JE005909 07AE001340 07GE007313 07HE004459 07DE004806 07DE005422 07GE003242 07GE003229 07FE005135 07FE003823 06HE010420 07CE004017 07EE001104 07DE004607 07FE004481 07BE002273 07EE000539 07GE002438 07GE000805 07GE005660 07HE003340 07EE003209 07EE003215 07EE007016 07DE006466 07CE008698 07CE008702 07CE008737 07BE001319 07BE001320 07BE004071 07HE006877 07CE007712 07HE003499 07HE003500 07AE001015 07BE006218 07CE003603 07HE001437 07HE003156 07BE005754 07BE005753 07AE005363 06LE005907 06LE005908 07AE001185 07DE006370 07BE006474 07DE005090 07DE003557 07EE001343 07EE001584 07DE001416 07CE002593 07DE006371 07DE004821 07DE003855 07FE006395 07GE006875 07HE001142 07HE002020 07GE005531 06LE006033 07AE003543 07AE003544 07AE004634 07CE001557 07EE001151 07EE003224 06JE005865 07FE000070 07AE005607 07CE002085 07CE006315 07CE006316 07CE005548 07EE000545 07FE002715 07GE003779 07HE000021 07HE003742 07FE000878 07AE005368 06HE010438 07FE006914 07GE005881 07GE005887 07HE001764 07FE002872 07FE004463 06KE008636 07DE003822 07FE005896 07GE000050 07DE003406 07CE008181 07FE005898 07FE001528 07FE006267 07HE004482 07BE005558 07BE005765 07DE001394 07FE005357 07IE001336 07FE001073 07AE007621 07CE002075 06LE006741 06LE006742 06LE007040 07DE003567 07EE006614 07DE001655 07EE000008 07EE001777 07FE000011 07GE000052 07FE004827 07GE003763 06IE004484 07HE004972 06LE007420 07EE005423 07CE005277 07EE001336 07BE002262 07CE004834 07AE006940 07AE007130 07CE007194 07DE003829 07EE003824 07GE002235 07HE006751 07HE004236 06KE004475 07CE004025 06H008567 07CE005286 07CE002200 07CE005298 07DE003181 07FE006663 07EE005170 06KE005554 07BE003351 07DE002273 07FE003580 07EE001921 07FE006248 07EE000002 07GE003491 07AE002478 06LE005167 06JE003392 06LE001536 07AE003549 06LE004877 07EE000009 07EE006606 07DE001928 07EE001779 07EE004964 07GE003012 07GE001384 07CE005297 07HE006765 07GE006548 07GE006947 07HE000028 07HE001440 07HE001783 07HE001939 07HE001941 07HE001947 07HE002612 07HE004466 07HE004479 07HE004737 07HE006457 07HE006473 07EE005169 07CE001564 07EE001477 07EE001480 07BE004930 07EE005186 07GE003506 07IE000696 07GE007316 07HE002603 07HE004980 07FE003827 06HE010688 07BE002270 06LE001534 07DE002296 07DE003831 07EE007015 07EE004636 07EE006019 07FE000010 06LE003279 07CE005274 07DE001636 07DE005405 07AE003732 07FE007532 06LE007039 07AE004639 07AE004644 07BE002029 07CE005283 07CE005284 07DE003404 07DE004803 07DE004812 07EE000534 07EE000533 07EE006261 07FE002707 07BE005365 07BE005559 07CE005287 07CE005545 07DE001653 07DE005258 07DE005257 07EE003211 07DE002505 07DE005070 07EE004955 07EE006006 07GE000044 07FE000024 07FE000017 07DE006464 07DE005259 07EE005176 07FE000016 07FE000026 07FE000031 07FE000875 07FE000012 07GE000063 07GE003236 07FE007542 07FE007545 07IE000119 07GE006531 07GE007330 07HE001945 07HE003349 07HE004748 07HE005000 06HE011045 07EE003198 07CE002383 07EE004630 07GE002229 07BE006481 07CE003611 07DE005073 07FE002436 07FE002434 07BE001327 07IE001339 07HE000019 07HE004216 06KE006636 06LE001716 07AE001013 07CE009405 07DE005057 07FE004814 06HE010674 07FE004465 06KE007305 07HE007034 07DE006251 07FE005352 07DE003397 07FE006906 07HE001940 07GE000036 07GE005451 07EE004973 07FE003818 06HE011046 07CE005285 07GE001386 07GE005653 06LE002636 07DE001393 07DE001654 07EE001773 07FE005359 07DE004807 07HE001072 07BE006219 07CE009276 07FE002430 07HE003048 07IE000704 07IE000692 06LE004872 07BE004926 07EE004951 07FE005122 07FE005119 07CE004022 07EE006267 07FE006071 07GE006877 07CE007181 07DE002118 07EE001796 07EE006278 07EE007040 07EE006279 07FE007023 07GE004266 07EE000579 07EE003226 07CE007180 07CE007179 07GE001464 07HE003850 07HE003851 07HE005123 07GE005888 07HE003740 07IE000117 07HE002596 07GE003397 06JE006327 06LE002629 06LE003997 07CE007695 07CE008184 07DE005404 07GE004934 07GE000800 07DE001938 07HE000007 07HE004456 07HE006443 07GE005873 07HE004741 07IE000680 07IE000688 07HE004474 07HE007033 07HE007038 07GE005883 07GE006958 07HE001059 07IE000106 07IE000694 07HE007040 07HE000022 07HE001204 07HE001448 07HE002618 07HE005274 07HE005644 07HE006449 07HE006459 07HE004470 07IE000684 07HE003746 07HE005269 07IE001321 07IE000118 07HE004756 07HE005658 07IE000382 07IE000952 07GE005901 07GE006519 07GE006520 07GE006541 07GE007331 07HE002604 07HE003065 07HE003347 07HE006167 07HE006448 07IE000114 07IE000955 07IE000957 07IE001735 07HE002290 07HE002598 07IE001329 07GE005885 07HE004471 07HE004472 07HE003733 07IE000125 07HE004746 07HE006749 07HE002841 07HE004222 07HE005254 07GE007324 07HE003339 07HE006450 07IE000693 07GE005900 07HE000015 07HE002286 07HE005268 07HE006746 07HE003714 07IE000108 07GE006938 07HE002317 07HE007051 07HE003749 07HE003053 07HE004463 07HE004475 07HE004745 07HE006193 07HE001946 07IE001317 07HE006184 07IE000109 07HE001071 07HE002854 07HE004993 07HE006452 07HE003723 07HE004982 07HE004224 07HE004458 07HE006199 07IE000389 07IE000381 07HE001215 07IE000132 07HE006764 07HE007030 07GE005645 07FE006646 07FE006921 07HE002832 07HE002856 07HE001781 07HE003713 07HE006170 07FE006257 07FE007549 07GE001143 07IE000384 07HE002311 06LE006733 07EE001315 07EE001088 07CE005534 07DE003824 07FE002200 07GE003239 07AE005609 07FE006260 07GE003018 07GE003010 07EE005422 07CE002614 07GE006533 07EE005410 07EE002985 07GE001373 07FE006265 07EE005172 07FE000865 07FE007268 07CE004839 06KE008632 07CE001566 07DE002488 07EE007001 07GE005153 07EE002970 07FE007529 07AE001014 07CE005294 07CE005543 07CE008696 07FE002709 07DE005063 07FE006271 07FE007550 07GE004170 07DE005423 07AE007123 07GE005428 07AE003731 07BE006477 07CE001567 07FE000001 07EE003201 07GE001363 07EE005395 07CE007188 07CE008690 07GE001375 07GE004922 07GE003761 07EE006247 07EE002501 07GE001118 06LE005751 07BE001832 07EE006613 07EE004961 07GE001372 07FE002772 07EE000003 07EE002972 07EE006248 07FE002868 07DE006457 07EE002814 07GE001364 07FE000020 07GE001377 07GE001369 07FE006659 07CE008938 07EE004950 07CE002384 07CE005533 07DE002124 07DE005053 07DE003980 07EE002500 07FE000864 07FE006062 07GE002231 07FE002866 07FE005120 07FE005123 07EE003206 07GE002247 07GE001127 07EE004948 07DE003564 07EE005173 07AE001606 07GE005148 07FE004462 07GE005149 07GE000035 07GE001969 07FE006893 07DE005261 07DE005272 07FE005130 07DE002506 07CE002076 07EE001322 07DE000732 07EE006994 07GE005427 07GE003235 07AE003007 07FE003579 07DE006456 07EE006002 07GE001124 07GE004933 07FE006653 07EE002503 07FE000009 07FE007273 07GE003015 07DE004593 07FE005351 07GE001125 07EE004962 07FE005368 07FE004479 07GE006745 07EE000538 07FE002884 07FE006912 07FE002699 06LE005915 07GE003492 06LE004878 07DE003990 07FE003594 07BE001115 07FE004472 07GE001385 07GE003011 07HE001198 07HE004754 07FE000004 07DE003565 07GE000795 07DE004801 07DE005421 07FE002864 06LE007424 07EE000527 06KE004877 07EE005405 07DE003989 07FE003593 07GE000048 07FE006266 07FE007284 07GE004168 06JE008358 07GE003005 07FE002880 07EE004954 07EE001492 07FE003597 07GE003490 07CE008946 07DE006052 07FE005899 07FE005134 07BE005139 07CE004020 07CE002392 07EE006607 07FE000867 07CE004830 07DE001392 07AE005365 07GE004164 07CE002606 07DE003828 07FE006658 07DE005409 07EE001770 07GE005432 07EE002971 07FE007270 07FE007266 07CE007189 07DE003827 07DE003578 07DE002132 07DE005074 07FE002447 07BE002269 07EE001100 07GE001371 07GE005164 07FE007289 07DE005049 07IE000387 07GE004166 07EE006270 07HE001778 07HE006187 06HE008569 07GE000064 07HE004467 07CE005280 07CE006018 07FE007536 07HE001449 06LE003476 06LE003727 07GE002238 07FE003583 07CE005275 07EE006252 07EE001769 07GE001121 07HE003712 07HE003721 07FE003812 06JE003900 07HE002610 07DE002119 07GE004915 07IE000120 07GE005880 07DE002125 06LE005919 07GE003493 07FE007271 07GE000056 07GE002250 07HE001214 07AE007343 06KE003982 07HE001066 07HE002857 06KE001938 07BE003352 07AE001012 06HE008566 07IE000949 07GE002420 07CE006029 07FE002876 07FE004820 07HE004735 07GE000791 06LE006737 06LE006738 07BE006228 07CE006311 07HE001932 07FE006258 06LE006032 06LE006031 07HE004542 06KE003629 06LE001324 06LE001323 07AE007124 07CE003317 06LE007422 07AE002485 07DE005419 07CE004829 07AE007634 06KE002880 07CE006019 07EE002508 07BE006227 07DE002499 07GE006748 06JE005911 07AE007128 07BE003356 07DE001646 07HE002599 07GE002245 07EE004605 07GE005151 07FE006919 07HE003744 06LE002010 07CE007702 07FE002705 07GE006946 07GE006959 07GE007333 07HE000020 07HE007049 07FE007533 07HE001770 07GE007428 07FE002711 07EE005184 07EE003202 06LE003481 07AE003021 07HE004543 07CE002610 07EE006244 06LE005164 06IE005861 07AE007345 07CE003613 07HE006466 07FE006069 07HE001442 07BE001825 07EE002815 07GE005893 07HE001076 07EE006273 06KE003401 07IE000390 07IE000403 07GE000794 07CE008188 07DE006055 07GE002237 07FE004826 07FE005370 07HE001201 07BE003361 06LE000182 07CE007693 07DE003165 07BE001107 07CE006023 07CE006022 07DE005052 07DE003395 07EE002810 07EE004949 07FE006896 07FE002700 07BE005364 07EE004638 07EE004958 07FE002199 07EE006016 07FE006915 07GE002440 07FE005367 07HE003731 07GE005899 07GE006756 07DE006050 07BE005357 07DE005416 07FE000021 07HE001064 07HE001070 07HE005494 07IE000392 07BE004729 07DE000590 07CE008944 07DE006056 07HE001779 07GE002425 07DE003837 07EE004941 07FE006264 07GE007315 07GE006544 06LE004876 06KE007468 06LE004881 07CE004838 07GE005657 07GE001141 07HE003060 07HE003750 07HE004989 07HE005270 07GE007321 07IE000678 07AE002031 06LE007036 07BE004728 06LE001726 07CE008187 07GE003773 07GE004999 06LE007227 06LE007226 07DE006369 07DE000584 07DE000585 07DE006072 07EE005420 07EE005421 07FE004805 07HE005124 07FE000018 07EE005188 07FE005131 06IE007956 07GE006532 07FE000863 07GE001359 07BE002268 07CE002387 07DE005260 07EE005178 07GE002241 07HE004483 07CE003319 07CE006572 07FE005350 07GE003226 06HE010680 07EE001489 06KE003630 06KE004880 06KE006640 06KE003407 07CE003608 07DE002295 07DE003984 07DE002492 07FE002702 07AE004643 07DE003820 07CE002393 07EE006018 07EE005192 07FE005362 07HE001205 07HE004461 06LE006740 07EE004969 07EE004623 07HE002615 07AE002253 07BE005363 07GE005905 07EE004480 06KE004876 07EE002813 07EE004602 06KE004476 06LE001719 06LE006035 06KE003984 06JE009102 07FE002769 07DE005055 07EE001494 07EE001491 07EE001936 07EE001781 07FE000869 07GE000054 07GE000796 07EE004624 07CE007704 07FE005144 07HE002847 07CE008689 07GE004930 07DE003992 07DE006254 07HE007044 07HE002305 07HE006474 07GE003241 07FE006647 07FE007534 07DE002282 07BE005761 07HE006177 07CE004018 07CE002607 07GE006749 07DE004815 07EE001099 07HE000014 07HE006446 07HE000031 07HE004744 07DE003173 07HE000030 07CE002609 07GE001133 06LE005165 07CE008694 07BE004939 07DE005263 07DE003839 07GE004189 07AE002474 07DE005412 07GE005170 07FE005369 07GE004928 07FE002446 07HE005275 07FE001531 06LE001720 07IE000965 07FE004825 07HE001935 07HE002601 07EE001103 07BE001118 07EE001768 07IE000703 06KE007713 07EE006275 07HE006205 07HE003724 06KE002287 07GE001376 07IE000134 06JE009099 07AE006103 07HE002302 07EE005424 07CE008939 07EE005182 07CE002597 07DE006471 07HE001074 07CE005787 07EE000531 07FE003590 07HE000027 07HE002855 07HE006766 07GE006950 07HE004753 07AE006101 07GE000033 07HE004238 07HE005660 07BE001602 07DE002507 07GE004939 07FE006904 07IE000689 07GE001967 07HE001199 07AE002694 07CE001571 07CE001568 06LE004879 07CE005547 07EE001329 07DE003986 07FE004815 07GE005890 07HE003059 07IE000121 07DE005048 07EE004953 07FE005355 07GE001358 07FE001527 07EE005189 07FE000874 07CE006037 07EE001938 07HE001065 07GE005162 07GE005876 07GE006543 07GE006545 07HE001443 07EE005400 07EE000526 07EE000525 06LE002014 06KE001940 07CE006309 07EE006266 07GE002308 07BE001100 07GE006525 07CE003320 07EE006608 07CE006304 07EE003968 07EE005416 07DE001659 07CE008704 07EE001782 07HE003726 06LE003724 07CE006573 06LE002637 07GE001132 07GE003014 07DE006265 07FE006284 07HE007043 07HE007052 07IE000679 07IE000956 07IE001490 07GE005663 07AE003009 07CE002598 07CE005531 07FE006649 07FE006648 07GE001122 06JE005914 06KE000177 07AE005131 07CE005546 07GE001379 07GE000806 07DE003835 07EE002988 07EE004479 07FE007291 07HE001203 07HE001222 07HE003747 07HE006196 07HE006201 07HE006202 07HE006762 07GE006741 07HE002310 07HE005264 07IE001320 07CE003805 07DE005065 07FE006268 07HE001769 07HE007042 06KE000175 07BE001829 07FE007272 07GE003004 07GE004172 07HE000006 07HE002623 06KE005841 07AE002247 07AE006318 07BE004731 07DE003171 07BE004730 07AE003729 07GE000034 07DE002285 07DE001643 07EE001326 07HE005657 07CE006570 07EE006615 07HE001936 07IE000948 07GE006528 07AE006102 07BE005143 07CE002388 07CE008691 06JE007828 07DE001931 07FE007551 07GE000061 06JE007827 07FE004470 07FE006276 06JE004204 07GE007322 07GE007328 06JE004203 07FE002697 07CE005538 07DE006049 07FE005121 06LE001535 07FE004811 06LE001727 07GE000042 07HE000010 06IE008344 06LE001333 07HE002309 07FE006269 07DE001639 07FE006249 07IE000961 07FE002867 07AE006942 07CE003325 07GE003017 07GE001972 07EE001101 07FE003599 07DE006264 07HE004992 07BE004727 07CE003321 07DE003814 07CE006301 07EE004600 07AE005126 07FE005891 07GE002230 07GE001366 07GE003006 06LE003282 07AE007350 07BE005144 07BE005358 07DE006261 07FE000028 07FE007279 07FE002432 07HE003745 06LE002634 06KE001939 07BE001112 06HE010691 07DE000741 07IE000124 06IE002116 07BE001830 06HE010435 07BE002271 07DE006461 07HE003061 06KE003983 07EE004974 06IE004479 07FE002433 07CE003616 07FE006895 07HE001055 07DE003396 07EE001483 07GE005658 07GE003001 07GE005155 07EE001089 07CE006017 07BE002030 07HE006458 07FE007288 07GE005909 06LE004869 06LE004868 06LE004871 06LE004870 06LE004662 07BE006478 07GE005434 07GE001126 07HE005497 07GE005651 07CE006298 07CE005529 07CE005532 06LE005162 07EE002512 07FE004482 07HE001763 07HE004976 07HE005641 07FE006061 07CE003329 07DE006047 06LE001505 06LE001506 07EE003225 07DE002116 07EE006040 07HE002700 07HE005603 07CE009281 06JE009098 06LE007038 07DE002489 07HE001056 07GE005877 07FE006920 07GE005000 07AE003726 06KE007253 06LE001691 06KE006598 07EE001587 07FE006396 07FE006767 07HE001439 07HE001438 07GE006627 07IE000494 06JE009096 07EE001923 07HE000024 07HE006183 07IE000374 07IE000378 07IE000380 07GE005436 07CE006300 07DE001387 07DE002123 07FE002198 07GE002228 07FE006259 07GE000788 07FE002192 07FE001522 07FE004808 07DE004600 07HE003067 06LE001723 07GE007325 07HE004217 07HE006444 07IE000107 06LE003281 07HE006748 07HE005272 07IE000137 07GE005165 06JE008688 07AE001009 06HE010678 07FE005356 07FE004817 06KE003404 06KE006905 06KE006916 07BE006226 06IE004482 07EE005409 07EE006610 07HE005271 07BE001828 06JE006329 07CE004019 07FE002197 07EE001765 06LE001724 07CE007193 07FE003586 07EE000535 07EE001929 07FE000034 07GE007334 07HE007041 07GE005641 07GE006737 06JE002581 06KE002879 06LE002011 06LE005910 07AE005121 07AE005122 07BE004924 07CE008930 07DE002486 07GE005640 07DE005403 07BE001108 07GE003003 07GE004908 07CE008942 06IE009764 07DE003823 07EE004619 07DE005072 07EE000006 07GE003510 07FE000882 07GE004912 07CE004841 07CE006038 07DE005075 07FE003600 07FE006072 07FE006916 07HE002616 07HE003737 07HE006166 07FE000866 07DE003826 07GE002249 07FE000870 07BE005368 07GE005903 06HE011336 06LE001721 07CE005792 07AE002040 07GE004926 07GE004927 07HE007055 07FE007285 07FE000008 07GE001968 07CE006310 07GE001360 06LE004001 07DE005425 06HE010672 07EE002979 07HE006173 07GE006750 07DE005050 07GE001361 07HE000009 07HE004229 07IE000951 07IE000682 07IE000687 07GE001129 07GE002423 07GE004174 07HE004232 07DE001638 07BE004077 07GE004178 07FE002871 06JE009100 07FE006251 07AE002246 07FE006655 07HE006303 07HE006304 07HE006305 07HE006306 07HE006307 07HE006308 07HE006309 07HE006310 07HE006311 07HE006312 07HE006313 07HE006314 07HE006316 07HE006317 07HE006318 07HE006319 07HE006320 07HE006321 07HE006322 07HE006323 07HE006324 06LE003998 07FE004819 07GE006760 06JE007097 07FE004821 07EE006617 07HE003069 07HE004742 07HE004984 07HE005498 07FE001078 07HE002825 07HE005491 06IE008348 07GE006751 07FE002865 07GE000039 07HE005647 07BE006785 07CE001562 07FE002440 07DE004603 07CE002080 07BE002272 06IE006809 07HE000033 07HE000034 07HE006190 07HE005002 07HE006470 07HE006756 07AE005912 07AE002473 07FE005139 06KE003131 07CE009277 07BE003595 07AE005610 07EE004960 07FE006286 07HE004991 07HE002622 07CE007191 07HE006462 06LE000184 07FE004478 07FE006273 07GE000046 07FE005127 07GE003771 07FE005142 07HE002321 07HE003055 07FE005116 07DE005256 07CE009278 07FE002712 07EE001094 07DE003994 07HE001445 07HE004464 07HE006200 07EE000005 07EE004597 07EE004947 07GE002233 07GE000792 07EE006254 07GE005154 07DE006465 07GE000053 07FE007281 07EE005187 07HE001054 07HE002294 07HE002304 07HE003719 07EE001090 07DE003830 07HE007031 07BE005759 07EE004944 07FE007269 06IE004762 07EE006260 07DE006469 07FE002437 07GE003022 07EE004639 07GE003778 07HE002298 07HE004221 07HE005253 07HE006171 07HE006204 07GE005889 07CE003804 07HE001463 07HE006461 07CE006571 07CE003323 07DE000593 07DE000592 07EE002505 07HE003718 07IE001325 07DE005253 07GE003754 07EE006001 07FE005126 07GE003509 07FE004475 07HE001775 07HE001787 07HE004481 07HE005273 06LE003479 07DE006462 07HE002301 07FE006911 07HE004468 07GE006529 07BE004072 07FE003825 07BE004925 07CE007694 07CE001558 07DE003166 07EE001319 07FE002693 07AE004641 07AE003014 07BE001598 07CE007703 06LE005753 07DE002502 07BE006487 07CE005289 07CE006027 07EE001332 07FE000029 07FE001529 07FE004468 07GE000068 07FE001523 07GE005159 07FE000013 07GE000798 07GE006941 07HE002319 07HE002844 07HE003345 07HE004457 07HE004979 07HE005258 07HE006182 06JE003899 07FE003820 07FE005902 07FE007275 07EE004604 07DE005410 07DE004805 07EE005175 07EE000536 07EE002817 07GE003762 07DE002511 07DE002508 07EE005195 07HE003049 07HE003056 07HE005255 07HE005490 07HE006197 07HE006442 07FE002771 07DE005051 07EE001481 07GE001140 07EE005402 07FE006278 07GE006943 07HE005259 07HE006186 07GE006522 07FE003582 07BE004923 06HE010426 07EE007008 07FE004828 07EE004976 07IE000683 07DE002122 07EE004603 07EE006250 07FE006898 07BE005562 07EE002507 07EE004627 07EE006007 07EE002823 07GE003016 07GE003007 07EE000537 07CE008947 07HE003054 07EE001317 07BE004935 07EE001780 07FE004818 07EE002824 07HE004460 07DE002501 07HE002318 07BE005136 07DE001388 07GE005147 07CE003803 07GE006961 07DE002300 07EE005198 07AE007626 07AE003736 07CE005293 07DE006460 07FE002885 07GE006743 07EE006998 07GE003511 07GE003781 07DE000595 07CE006035 07GE004165 07CE007182 07DE001925 07GE002234 07CE007192 07CE007195 07CE005542 07CE006034 07DE001935 07GE005892 07FE006656 07GE004176 07FE006285 07HE002840 07HE005496 07DE001647 07GE000069 07GE007329 06HE010671 07DE004799 06IE009398 07DE006463 06KE005555 07DE001667 07EE002981 07FE002869 07EE006022 07CE004832 07CE008929 07AE001613 07DE005411 07FE002698 07BE002026 07DE003170 07EE001325 07EE004620 07DE006054 06JE009101 07AE005129 07CE004024 06IE009396 06IE009761 06HE008571 06IE007958 07DE001934 07FE004829 07DE005066 07HE004736 07IE000127 07EE004952 07FE003810 07EE000528 07GE005643 07HE002289 06LE000183 07BE002264 06HE010417 07AE003733 06KE002882 07AE005613 07AE005612 07AE005614 06HE010429 06KE008633 07DE006455 07BE003355 07DE003172 07FE006662 07CE003315 07DE005071 07HE002842 07BE006223 06LE005749 07CE002599 07GE001123 06HE010432 07HE001207 07DE000587 07AE001607 07AE004637 06KE007467 07BE004076 07GE002236 07FE007286 07GE005648 07FE000868 06LE003996 07AE007620 07CE004831 07FE000006 07GE003000 07GE001119 07AE003011 07DE004810 07EE007020 07EE001324 07FE000880 07GE004909 07GE003514 07GE006878 07GE006936 07HE001942 07HE003717 07HE003734 07HE006195 07HE006456 07BE006721 07EE001091 06LE001330 07BE005366 07DE003988 06JE007829 07FE003814 07HE007046 07HE007029 07EE003199 07HE002303 07HE007047 07GE005661 07DE003559 07EE001775 07CE008941 07CE008687 07FE000005 07BE002025 07CE008182 07DE001386 07FE003809 07BE002027 07CE002195 07DE005059 07DE003575 07EE003213 07FE005132 07FE007274 07HE000029 07HE002295 07HE005261 07DE006256 07BE005146 06KE006076 07EE001335 06IE008359 07GE003495, and 07EE006271. |
Recalling Firm/ Manufacturer |
Invacare Corporation 1200 Taylor St Elyria OH 44035-4190
|
For Additional Information Contact |
Carroll L. Martin 440-329-6356
|
Manufacturer Reason for Recall |
Invacare has become aware that on some chairs, the stability lock feature may not be engaging properly or consistently. If the stability lock feature does not engage properly, the wheelchair may veer to one side or it may rock forward onto its front riggings.
|
FDA Determined Cause 2 |
Pending |
Action |
On 11/27/2007 the firm sent Urgent Field Correction Needed recall notification letters and response forms to their customers informing them of the problem. They instructed their customers to immediately inspect and adjust the Stability Lock feature, distribute the enclosed update to the Owner's Manual to all end-users, and return the enclosed response form notifying them that the inspection and/or adjustments have been completed. |
Quantity in Commerce |
4171 Wheelchairs |
Distribution |
Worldwide Distribution --- including USA and countries of Canada, Australia, Germany, New Zealand and the United Kingdom. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = ITI and Original Applicant = INVACARE CORP.
|