Date Initiated by Firm |
August 01, 2008 |
Date Posted |
February 03, 2009 |
Recall Status1 |
Terminated 3 on November 12, 2009 |
Recall Number |
Z-0805-2009 |
Recall Event ID |
48991 |
Product Classification |
Breathing Circuit - Product Code CAI
|
Product |
King Systems Universal Flex2 Breathing Circuit & Pack Kit, Latex Free, 50 per case, King Systems Corporation, Noblesville, IN; REF # F90757.
This breathing circuit is a device intended to administer medical gases and/or anesthetic gases to a patient during anesthesia for inhalation, or medical gases to a patient during respiratory care inhalation. |
Code Information |
Lots IKLB6, IKWU1 and ILBQ3. |
Recalling Firm/ Manufacturer |
King Systems Corp. 15011 Herriman Blvd Noblesville IN 46060
|
For Additional Information Contact |
317-776-6823
|
Manufacturer Reason for Recall |
The inner line may be severed or broken into two pieces, thus hindering the delivery of oxygen and anesthetics to the patient and possibly elevated patient carbon dioxide levels.
|
FDA Determined Cause 2 |
Pending |
Action |
Consignees were notified by an Urgent Product Recall letter on or about 7/29/2008 to cease sales, notify their customers to the hospital level, and to return product on hand. |
Quantity in Commerce |
Approx. 500,000 units of all products. |
Distribution |
Worldwide Distribution - including USA and countries of Australia, Colombia, Germany, Netherlands, and Peru. |