Class 2 Device Recall APEX Hip System Bipolar Head |
|
Date Initiated by Firm |
July 10, 2009 |
Date Posted |
September 10, 2009 |
Recall Status1 |
Terminated 3 on June 28, 2012 |
Recall Number |
Z-1970-2009 |
Recall Event ID |
52662 |
510(K)Number |
K082468
|
Product Classification |
Hip Joint Femoral (hemi-hip) Metal/Polymer Cemented or Uncemented Prosthesis - Product Code KWY
|
Product |
Apex Hip System Bipolar Head, 22.225mm ID, 42 mm OD Product Code: H3-82242.
Intended for use in combination with an Apex Hip System femoral for uncemented primary or revision hemiarthroplasty of the hip. |
Code Information |
Lot Number: 5250 |
Recalling Firm/ Manufacturer |
Omni Life Science 175 Paramont Drive Raynham MA 02767
|
Manufacturer Reason for Recall |
The APEX Hip System Bipolar Head implant may not properly capture the femoral head assembly of the bipolar head onto the femoral head.
|
FDA Determined Cause 2 |
Device Design |
Action |
Omni Life Science notified distributors and implanting surgeons by an Urgent, APEX Hip System Bipolar Head Voluntary Recall letter dated June 29, 2009. The method of notification is both e-mail and FedEx overnight. The letter requires a signature acknowledging receiving the letter and the letter is to be faxed back to Omni Life Science. The Distributors/Sale Representatives are instructed to fill in the "Quantity Returned" on the recall letter and fax back to Omni. They are to ship all existing Apex Hip System Bipolar Head inventory by 07/10/2009 to the directed address. |
Quantity in Commerce |
18 units |
Distribution |
Nationwide Distribution -- Including states of AZ, CA, CO, FL, IL, MI, NJ, OK, UT, VA, and WI. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = KWY and Original Applicant = OMNI LIFE SCIENCE, INC.
|
|
|
|