Date Initiated by Firm December 18, 2009
Date Posted June 21, 2010
Recall Status1 Terminated 3 on December 08, 2011
Recall Number Z-1810-2010
Recall Event ID 55603
510(K)Number K031987  
Product Classification automated external defibrillators (non-wearable) - Product Code MKJ
Product Cardiac Science Powerheart 9390A automated external defibrillator. This fully automatic model does not require the user to press a shock button in order for the device to deliver therapy (when appropriate).

Intended for use in either in-hospital or out-of-hospital settings for emergency treatment of victims exhibiting symptoms of sudden cardiac arrest who are unresponsive and not breathing.
Code Information Serial Numbers: 4259985, 4260065, 4260137, 4260148, 4260174, 4260179, 4260181, 4261110, 4261111, 4261112, 4261113, 4261114, 4261115, 4261116, 4261117, 4261118, 4261119, 4261120, 4261121, 4261122, 4261123, 4261125, 4261126, 4261127, 4261128, 4261129, 4261130, 4261131, 4261132, 4261133, 4261134, 4261136, 4261138, 4261139, 4261140, 4261141, 4261142, 4261143, 4261144, 4261145, 4261147, 4261149, 4261150, 4261151, 4261153, 4261154, 4261155, 4261156, 4261157, 4261158, 4261160, 4261161, 4261162, 4261163, 4261164, 4261166, 4261168, 4261169, 4261170, 4261172, 4261174, 4261175, 4261177, 4261178, 4261179, 4261180, 4261181, 4261182, 4261183, 4261184, 4261185, 4261186, 4261188, 4261189, 4261190, 4261216, 4261217, 4261284, 4261285, 4261292, 4261299, 4261300, 4261301, 4261305, 4261306, 4261308, 4261312, 4261313, 4261317, 4261319, 4261320, 4261321, 4261325, 4261326, 4261329, 4261330, 4261332, 4261333, 4261342, 4261343, 4261348, 4261351, 4261352, 4261355, 4261356, 4261358, 4261360, 4261361, 4261365, 4261370, 4261381, 4261382, 4261383, 4261384, 4261385, 4261386, 4261387, 4261388, 4261394, 4261395, 4261396, 4261397, 4261398, 4261399, 4261400, 4261401, 4261402, 4261403, 4261404, 4261405, 4261406, 4261407, 4261408, 4261409, 4261410, 4261411, 4261412, 4261413, 4261414, 4261415, 4261416, 4261417, 4261420, 4261422, 4261423, 4261425, 4261449, 4261450, 4261451, 4261463, 4261465, 4261473, 4261475, 4261476, 4261477, 4261478, 4261479, 4261480, 4261481, 4261482, 4261483, 4261484, 4261485, 4261486, 4261487, 4261488, 4261489, 4261490, 4261492, 4261495, 4261496, 4261498, 4261499, 4261500, 4261501, 4261502, 4261503, 4261505, 4261506, 4261507, 4261508, 4261512, 4261513, 4261514, 4261515, 4261516, 4261517, 4261518, 4261519, 4261520, 4261522, 4261523, 4261524, 4261526, 4261527, 4261528, 4261529, 4261530, 4261531, 4261532, 4261534, 4261535, 4261536, 4261537, 4261538, 4261539, 4261540, 4261541, 4261542, 4261543, 4261544, 4261545, 4261546, 4261547, 4261548, 4261549, 4261551, 4261552, 4261554, 4261555, 4261557, 4261558, 4261559, 4261560, 4261561, 4261562, 4261563, 4261564, 4261565, 4261566, 4261567, 4261568, 4261570, 4261571, 4261572, 4261573, 4261574, 4261575, 4261576, 4261577, 4261579, 4261580, 4261586, 4261587, 4261592, 4261597, 4261598, 4261605, 4261608, 4261610, 4261611, 4261613, 4261614, 4261615, 4261617, 4261618, 4261621, 4261623, 4261624, 4261626, 4261628, 4261630, 4261631, 4261632, 4261633, 4261634, 4261635, 4261636, 4261637, 4261638, 4261639, 4261640, 4261641, 4261642, 4261643, 4261644, 4261645, 4261646, 4261647, 4261648, 4261649, 4261650, 4261651, 4261652, 4261653, 4261654, 4261656, 4261657, 4261660, 4261662, 4261663, 4261664, 4261665, 4261666, 4261670, 4261672, 4261676, 4261678, 4261679, 4261680, 4261683, 4261688, 4261689, 4261690, 4261691, 4261692, 4261695, 4261696, 4261697, 4261700, 4261709, 4261711, 4261712, 4261713, 4261714, 4261716, 4261718, 4261721, 4261726, 4261727, 4261729, 4261733, 4261734, 4261737, 4261739, 4261744, 4261745, 4261746, 4261747, 4261748, 4261750, AND 4261751.
Recalling Firm/
Manufacturer
Cardiac Science Corporation
3303 Monte Villa Pkwy
Bothell WA 98021-8969
For Additional Information Contact
425-402-2000
Manufacturer Reason
for Recall
The K302 relay switch of the G3 series AED product line may experience an early failure. If the K302 contacts are open at the time of rescue, there is a possible delay in therapy or failure to deliver therapy.
FDA Determined
Cause 2
Other
Action Customer notification letters were mailed to US end users and distributors (final consignees) between December 18 and 23, 2009 by USPS certified mail, return receipt requested. Notification letters were sent by FedEx to Canadian distributors (final consignees) on December 23, 2009. Secondary consignee (distributor) letters were mailed on December 23, 2009. Letters advise user to check the status indicator on front of AED to determine if it is rescue ready. Users are advised to keep the AED in service and to perform the Daily and Monthly Scheduled Maintenance as outlined in the Operator and Service Manual. Questions about this communication should be addressed to 1.877.901.1788 toll free from inside the US, or to email .
Quantity in Commerce 332 units
Distribution Domestic: 323 units to distributors and end users. Canada: 9 units to 2 distributors.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = MKJ and Original Applicant = CARDIAC SCIENCE
-
-