| | Class 2 Device Recall Varian brand Cseries Clinac, Trilogy and Novalis Tx, Software Versions 6.X and 7.X, |  |
| Date Initiated by Firm | June 10, 2010 |
| Date Posted | January 08, 2011 |
| Recall Status1 |
Terminated 3 on December 23, 2011 |
| Recall Number | Z-0886-2011 |
| Recall Event ID |
56347 |
| 510(K)Number | K070094 K904364 K913119 |
| Product Classification |
Accelerator, linear, medical - Product Code GKZ
|
| Product | Varian brand C-series Clinac, Trilogy and Novalis Tx, Software Versions 6.X and 7.X, Model Numbers: H14, H27, H29, HCX, Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA
The Trilogy Radiotherapy Delivery System is a radiation therapy accelerator intended to deliver megavoltage x-ray treatments for conventional radiotherapy (three dimensional conformal radiotherapy and intensity modulated radiotherapy) and stereotactic radiosurgery and radiotherapy. Stereotactic treatments are intended for therapy of lesions, e.g., arteriovenous malformations, primary tumors and metastases. Stereotactic treatments may be intracranial or extra cranial and consist of single-session or fractionated delivery. |
| Code Information |
CODE(S): H294145
H293620 ,H271973 ,H272328 ,H293393 ,H294467 ,H293944 ,H293764 ,H293725 H140117
H140734 ,H272492 ,H140070 ,H293653 ,H293998 ,H294049 ,H294199 ,H293726 H271877
H293666 ,H273065 ,H272283 ,H293823 ,H293999 ,H293697 ,H294367 ,H293728 H272434
H140754 ,H273130 ,H272370 ,H293529 ,H290201 ,H294103 ,H294440 ,H293698 H272828
H293724 ,H273170 ,H272327 ,H293787 ,H290555 ,H293760 ,H294474 ,H293774 H272889
H293817 ,H293381 ,H272329 ,H140772 ,H290553 ,H293682 ,H294496 ,H293776 H272922
H140768 ,H140709 ,H272299 ,H293883 ,H290549 ,H293009 ,H294597 ,H293773 H140203
H140775 ,H293411 ,H272425 ,H294113 ,H290552 ,H293942 ,H294641 ,H293829 H140204
H293843 ,H293412 ,H272421 ,H294129 ,H290551 ,H293765 ,H294650 ,H293830 H291054
H293847 ,H293463 ,H272407 ,H293908 ,H290559 ,H293758 ,H294699 ,H293854 H273072
H140786 ,H140729 ,H272857 ,H293909 ,H293396 ,H290557 ,H291075 ,H293839 H140208
H293850 ,H293499 ,H272859 ,H294134 ,H290547 ,H293770 ,H293368 ,H294073 H291098
H293968 ,H293500 ,H291022 ,H294186 ,H290541 ,H294136 ,H140756 ,H294118 H291114
H294022 ,H293668 ,H291011 ,H294092 ,H293365 ,H294143 ,H294034 ,H294104 H273027
H293970 ,H293841 ,H291015 ,H294273 ,H293435 ,H294239 ,H294527 ,H294200 H273107
H294017 ,H294164 ,H291082 ,H294204 ,H293387 ,H294185 ,H294107 ,H294203 H273102
H140796 ,H140882 ,H291021 ,H294210 ,H293482 ,H294248 ,H294535 ,H294256 H291160
H294065 ,H293567 ,H291016 ,H294317 ,H290554 ,H294290 ,H273048 ,H140856 H291191
H294062 ,H290278 ,H291085 ,H140836 ,H293434 ,H294390 ,H273140 ,H294471 H291164
H294175 ,H273187 ,H291018 ,H294364 ,H293555 ,H293766 ,H293371 ,H273150 H291165
H294232 ,H293595 ,H291154 ,H294403 ,H293506 ,H294153 ,H273208 ,H293898 H291163
H140819 ,H140763 ,H291166 ,H294383 ,H293569 ,H293961 ,H293305 ,H271763 H273103
H294409 ,H293929 ,H291190 ,H294414 ,H273031 ,H294250 ,H293681 ,H273117 H140229
H140843 ,H293958 ,H291172 ,H293906 ,H293853 ,H294422 ,H293842 ,H273182 H140234
H294448 ,H293450 ,H291231 ,H294347 ,H293455 ,HCX4138 ,H293924 ,H293374 H140231
H140844 ,H271159 ,H291179 ,H294585 ,H293498 ,H294011 ,H294462 ,H293601 H291212
H294584 ,H271157 ,H291178 ,H294586 ,H293633 ,H294445 ,H294525 ,H293602 H140236
H140865 ,H271569 ,H293307 ,H294553 ,H293477 ,H294480 ,H294532 ,H293645 H140235
H294667 ,H271624 ,H291195 ,H294551 ,H293489 ,H294400 ,H294608 ,H293661 H291261
H272188 ,H271662 ,H291194 ,H294674 ,H293626 ,HCX4400 ,H140096 ,H294098 H140232
H272946 ,H271663 ,H291180 ,H294683 ,H293705 ,H294490 ,H140097 ,H294123 H291262
H293494 ,H271941 ,H293509 ,H294684 ,H293676 ,H294488 ,H291115 ,H294296 H293422
H294593 ,H271926 ,H293598 ,H294675 ,H293710 ,H294449 ,H293424 ,H294297 H293442
H294594 ,H272157 ,H293549 ,H294352 ,H293907 ,H294568 ,H293444 ,H294304 H140713
H293796 ,H272216 ,H140728 ,H294750 ,H293586 ,H294634 ,H140751 ,H294543 H291259
H293940 ,H272159 ,H293521 ,H294751 ,H293745 ,H294611 ,H293662 ,H294461 H291260
H294526 ,H272162 ,H293656 ,H294758 ,H293838 ,H294015 ,H293663 ,H294592 H293512
H294502 ,H272311 ,H293654 ,H294766 ,H293978 ,H293495 ,H293650 ,H270794 H140732
H293324 ,H272309 ,H293655 ,H293336 ,H293738 ,H293759 ,H293727 ,H271582 H293561
H273186 ,H272310 ,H293784 ,H293337 ,H293753 ,H293761 ,H293612 ,H272277 H272372
H293407 ,H294201 ,H273164 ,H294146 ,H294140 ,H294602 ,H294428 ,H293363 H272473
H293474 ,H294109 ,H273162 ,H294114 ,H294133 ,H294636 ,H294460 ,H293362 H272458
H140711 ,H294228 ,H273191 ,H294368 ,H140124 ,H273185 ,H140857 ,H293380 H272494
H140704 ,H140822 ,H273200 ,H294356 ,H290533 ,H273036 ,H294432 ,H293410 H273011
H140717 ,H294241 ,H273203 ,H294346 ,H294329 ,H273168 ,H294587 ,H293440 H140119
H293578 ,H294230 ,H293321 ,H294264 ,H294433 ,H291199 ,H294647 ,H293408 H140114
H140154 ,H294262 ,H273202 ,H294233 ,H292832 ,H291268 ,H294183 ,H293452 H273077
H293533 ,H294242 ,H293379 ,H294357 ,H272834 ,H140715 ,H294371 ,H293430 H273080
H140707 ,H294280 ,H293388 ,H140842 ,H291174 ,H293502 ,H290282 ,H140719 H273106
H140742 ,H140850 ,H293378 ,H294084 ,H273172 ,H294074 ,H273064 ,H293409 H273110
H140731 ,H294057 ,H293376 ,H294231 ,H290556 ,H293353 ,H140129 ,H293454 H273121
H293576 ,H294529 ,H293392 ,H294274 ,H273066 ,H293947 ,H273089 ,H293312 H273126
H293625 ,H294419 ,H273210 ,H294415 ,H293333 ,H294042 ,H273039 ,H293449 H273131
H140740 ,H294519 ,H293464 ,H294160 ,H293304 ,H294043 ,H273042 ,H293451 H273104
H293644 ,H140863 ,H293389 ,H294477 ,H273178 ,H271802 ,H273092 ,H293527 H273128
H293675 ,H294581 ,H293597 ,H294310 ,H293524 ,H272195 ,H273134 ,H293523 H273147
H140759 ,H294627 ,H293640 ,H294536 ,H293665 ,H272985 ,H273127 ,H293566 H273160
H140748 ,H294605 ,H293686 ,H294601 ,H293844 ,H140125 ,H273122 ,H293557 H140140
H293624 ,H140872 ,H293619 ,H294591 ,H293897 ,H272932 ,H290548 ,H293558 H140142
H293651 ,H294646 ,H140747 ,H294545 ,H293755 ,H273113 ,H273111 ,H293623 H273177
H272905 ,H140875 ,H140758 ,H294511 ,H140774 ,H273083 ,H273116 ,H293641 H273167
H293779 ,H294678 ,H293530 ,H294369 ,H293946 ,H273115 ,H273119 ,H293634 H273175
H293646 ,H140871 ,H293735 ,H294547 ,H293793 ,H273145 ,H273144 ,H293719 H273171
H140764 ,H140887 ,H293763 ,H294548 ,H294121 ,H272964 ,H273132 ,H293720 H273165
H293861 ,H294068 ,H293736 ,H294473 ,H293876 ,H293370 ,H273133 ,H293721 H273204
H140766 ,H294067 ,H140760 ,H294589 ,H294124 ,H272973 ,H140138 ,H293730 H140144
H140779 ,H293534 ,H293845 ,H140866 ,H293677 ,H273183 ,H273124 ,H293729 H140148
H293860 ,H271370 ,H293732 ,H294645 ,H293310 ,H140724 ,H273156 ,H293762 H140146
H140788 ,H271378 ,H293465 ,H294671 ,H294351 ,H140712 ,H140137 ,H293788 H140145
H293918 ,H271531 ,H293955 ,H294685 ,H294387 ,H293419 ,H140141 ,H293797 H140151
H293922 ,H271593 ,H293982 ,H294724 ,H294212 ,H293443 ,H273123 ,H293792 H140150
H293949 ,H272137 ,H293575 ,H293643 ,H294374 ,H293491 ,H273125 ,H293799 H293351
H293954 ,H272111 ,H294036 ,H293857 ,H294465 ,H140762 ,H273163 ,H140767 H293373
H293963 ,H272497 ,H293917 ,H294198 ,H294618 ,H293781 ,H273194 ,H293813 H140139
H140794 ,H272875 ,H293791 ,H273166 ,H294423 ,H140769 ,H291257 ,H293800 H140153
H293992 ,H272909 ,H294085 ,H140862 ,H294060 ,H294100 ,H273197 ,H293809 H140708
H294023 ,H273101 ,H294167 ,H294292 ,H294542 ,H294161 ,H273138 ,H293832 H293391
H140798 ,H140136 ,H294119 ,H290561 ,H294515 ,H294223 ,H273193 ,H293798 H293334
H294173 ,H273151 ,H294005 ,H140152 ,H294664 ,H140789 ,H273189 ,H293807 H293369
H140815 ,H273174 ,H294117 ,H293585 ,H294238 ,H140841 ,H140238 ,H293816 H293815
H294342 ,H293808 ,H140839 ,H291083 ,H273082 ,H273180 ,H294215 ,H294349 H293814
H294334 ,H294656 ,H294335 ,H291127 ,H293306 ,H293320 ,H294229 ,H294588 H140773
H140828 ,H294659 ,H140739 ,H291128 ,H273181 ,H294144 ,H294315 ,H294705 H293827
H140834 ,H140873 ,H291064 ,H291141 ,H293375 ,H294205 ,H294370 ,H293618 H293828
H294318 ,H294610 ,H291227 ,H291168 ,H293532 ,H294401 ,H294569 ,H271265 H293881
H294309 ,H140889 ,H291044 ,H291204 ,H294080 ,H294402 ,H272033 ,H272219 H293895
H294293 ,H294600 ,H291132 ,H291205 ,H294435 ,H294495 ,H273084 ,H290386 H293896
H294307 ,H294693 ,H293540 ,H291223 ,H294434 ,H140886 ,H273108 ,H272405 H293920
H294338 ,H294723 ,H140755 ,H293447 ,H294746 ,H291130 ,H140147 ,H272362 H293911
H294363 ,H140892 ,H293959 ,H271860 ,H271881 ,H291137 ,H273179 ,H272333 H293943
H294283 ,H294635 ,H293678 ,H271976 ,H272023 ,H291055 ,H273201 ,H291037 H293887
H294396 ,H294643 ,H293997 ,H271681 ,H293349 ,H140135 ,H273198 ,H291238 H293812
H294308 ,H294666 ,H293616 ,H140847 ,H293545 ,H291198 ,H291254 ,H291237 H140785
H294450 ,H294668 ,H294097 ,H291250 ,H293801 ,H273159 ,H273184 ,H293395 H293884
H294447 ,H294676 ,H294247 ,H293469 ,H140840 ,H273153 ,H273195 ,H293448 H294002
H140849 ,H294700 ,H294313 ,H293468 ,H140874 ,H293421 ,H293479 ,H293394 H294027
H294319 ,H294701 ,H290208 ,H294344 ,H140700 ,H293456 ,H293572 ,H293423 H294046
H294451 ,H294706 ,H290274 ,HCX4344 ,H294044 ,H293364 ,H293559 ,H293604 H294152
H294312 ,H294713 ,H290400 ,H140090 ,H140878 ,H293704 ,H293570 ,H293657 H140804
H294407 ,H294715 ,H290399 ,H140101 ,H140803 ,H293553 ,H293390 ,H293767 H294071
H294260 ,H294726 ,H272359 ,H290517 ,H273112 ,H293631 ,H293599 ,H290372 H140814
H294066 ,H294728 ,H290501 ,H140121 ,H273176 ,H293628 ,H293446 ,H294037 H294154
H294463 ,H294731 ,H291211 ,H140062 ,H294108 ,H293679 ,H293382 ,H294082 H294195
H294503 ,H294732 ,H293596 ,H140143 ,H294190 ,H293794 ,H293383 ,H140790 H294220
H294087 ,H294736 ,H293741 ,H293384 ,H294191 ,H293386 ,H140730 ,H294219 H294176
H294213 ,H294737 ,H293866 ,H140705 ,H294520 ,H293471 ,H293712 ,H294681 H294165
H294580 ,H293603 ,H293964 ,H293328 ,H294606 ,H293629 ,H293497 ,H294682 H294258
H294583 ,H294093 ,H294331 ,H140134 ,H290256 ,H293785 ,H293849 ,H271400 H294271
H294492 ,H294491 ,H140877 ,H293637 ,H271890 ,H293786 ,H293852 ,H272397 H294270
H294582 ,H294279 ,H272418 ,H140745 ,H273173 ,H293870 ,H293951 ,H272427 H294276
H294508 ,H294533 ,H291009 ,H140750 ,H273155 ,H294038 ,H293941 ,H290511 H294277
H294590 ,H293550 ,H291103 ,H140741 ,H293372 ,H293904 ,H293972 ,H272911 H294278
H294632 ,H293938 ,H293531 ,H140749 ,H293977 ,H293905 ,H294193 ,H291033 H294289
H294339 ,H294489 ,H140752 ,H294070 ,H294055 ,H293510 ,H294192 ,H290540 H294216
H294094 ,H290262 ,H293574 ,H294072 ,H140743 ,H293869 ,H294208 ,H291213 H294217
H294637 ,H290263 ,H293996 ,H140805 ,H140771 ,H294047 ,H294211 ,H291232 H294281
H294537 ,H290352 ,H294081 ,H294069 ,H294214 ,H294059 ,H294389 ,H293577 H294272
H294649 ,H291046 ,H294523 ,H140859 ,H294779 ,H140818 ,H294014 ,H293593 H294306
H294619 ,H293630 ,H290423 ,H140861 ,H294565 ,H293878 ,H293990 ,H293492 H294288
H294663 ,H140823 ,H291007 ,H291059 ,H273142 ,H294147 ,H293991 ,H293649 H293667
H293511 ,H272908 ,H294102 ,H290249 ,H140001 ,H272014 ,H140046 ,H272401 H293742
H293339 ,H273004 ,H294076 ,H271393 ,H271760 ,H290362 ,H272238 ,H272394 H293652
H290560 ,H273049 ,H294077 ,H271407 ,H271762 ,H272009 ,H272237 ,H272801 H293818
H294639 ,H272469 ,H294078 ,H271413 ,H271785 ,H271975 ,H272236 ,H272456 H293834
H294640 ,H273063 ,H294079 ,H271385 ,H271780 ,H271964 ,H272259 ,H272470 H293778
H291233 ,H273105 ,H294218 ,H271439 ,H271797 ,H272052 ,H290392 ,H272500 H293914
H293879 ,H291152 ,H293867 ,H271452 ,H271835 ,H272000 ,H272271 ,H272803 H293928
H273149 ,H291151 ,H294257 ,H271438 ,H271838 ,H272043 ,H140058 ,H272454 H293965
H273152 ,H140224 ,H294354 ,H271425 ,H271831 ,H272035 ,H272301 ,H272466 H293981
H293323 ,H140225 ,H294263 ,H271502 ,H271783 ,H140038 ,H272306 ,H272479 H293993
H273114 ,H140226 ,H294323 ,H271515 ,H271858 ,H272045 ,H290397 ,H272491 H140797
H293315 ,H140227 ,H294321 ,H271519 ,H271856 ,H271951 ,H272307 ,H140088 H294305
H293584 ,H291271 ,H294320 ,H271526 ,H271862 ,H272049 ,H272291 ,H272480 H140864
H293739 ,H273206 ,H294322 ,H271507 ,H271833 ,H290376 ,H272321 ,H140092 H273143
H293803 ,H293325 ,H294268 ,H271523 ,H140011 ,H271942 ,H272325 ,H140201 H271994
H293687 ,H293326 ,H294539 ,H290287 ,H271879 ,H272062 ,H272293 ,H290505 H291008
H293795 ,H293207 ,H294541 ,H271497 ,H271825 ,H272069 ,H272326 ,H272816 H291182
H293865 ,H293356 ,H294171 ,H271559 ,H271870 ,H272001 ,H272302 ,H290503 H291108
H294110 ,H293327 ,H294540 ,H271587 ,H271893 ,H272086 ,H272346 ,H272476 H293311
H294111 ,H293416 ,H294538 ,H271588 ,H271896 ,H272078 ,H290405 ,H272811 H293331
H294040 ,H293417 ,H294621 ,H271420 ,H271900 ,H272089 ,H272345 ,H290509 H293556
H293301 ,H293377 ,H294620 ,H271594 ,H290341 ,H140043 ,H140073 ,H272837 H293507
H293314 ,H273055 ,H294716 ,H271607 ,H271903 ,H290371 ,H290402 ,H272822 H293747
H294345 ,H293441 ,H140893 ,H271626 ,H140018 ,H272081 ,H140071 ,H272824 H140795
H294412 ,H293583 ,H270788 ,H271669 ,H271921 ,H272129 ,H272336 ,H272488 H293740
H294411 ,H293674 ,H271041 ,H271664 ,H290326 ,H272134 ,H140067 ,H272854 H293980
H294379 ,H293671 ,H270862 ,H290304 ,H271906 ,H140045 ,H272354 ,H272860 H293872
H294438 ,H293672 ,H271066 ,H271691 ,H271918 ,H272143 ,H272365 ,H140098 H293875
H294439 ,H293670 ,H271105 ,H290313 ,H271895 ,H272172 ,H272341 ,H272841 H293960
H294388 ,H293716 ,H271169 ,H290308 ,H271935 ,H272187 ,H272355 ,H272862 H140829
H294644 ,H293715 ,H290241 ,H271678 ,H271966 ,H272178 ,H272377 ,H290510 H291272
H273056 ,H293717 ,H271167 ,H271618 ,H140026 ,H290379 ,H140076 ,H272873 H293496
H293752 ,H293718 ,H271242 ,H140020 ,H140027 ,H272202 ,H272388 ,H290508 H293600
H294255 ,H293669 ,H290222 ,H271711 ,H140031 ,H272174 ,H272410 ,H272866 H293638
H293751 ,H293804 ,H271362 ,H271693 ,H271993 ,H272209 ,H272422 ,H272868 H293848
H273148 ,H293820 ,H271194 ,H271712 ,H271961 ,H272220 ,H272428 ,H272872 H294141
H294657 ,H293782 ,H271308 ,H271720 ,H271974 ,H272204 ,H140078 ,H272847 H294265
H271954 ,H293826 ,H290224 ,H271729 ,H140025 ,H140047 ,H140077 ,H272885 H294284
H271872 ,H293713 ,H271334 ,H271714 ,H272029 ,H272231 ,H272426 ,H272914 H290550
H272826 ,H293714 ,H290243 ,H271724 ,H272003 ,H272223 ,H272444 ,H291002 H290600
H273029 ,H291076 ,H291048 ,H273135 ,H291219 ,H293303 ,H293318 ,H293432 H272893
H291028 ,H291077 ,H291088 ,H291140 ,H140233 ,H291236 ,H293357 ,H293425 H290520
H273030 ,H291052 ,H140211 ,H291045 ,H291202 ,H293332 ,H291229 ,H293300 H291006
H291032 ,H291079 ,H291099 ,H291109 ,H291169 ,H291266 ,H291249 ,H293438 H140109
H272971 ,H272978 ,H140215 ,H291117 ,H291200 ,H291247 ,H140149 ,H140706 H272899
H291027 ,H291073 ,H140133 ,H291111 ,H291218 ,H291256 ,H293313 ,H293366 H291001
H291025 ,H291067 ,H291120 ,H291116 ,H291170 ,H291270 ,H291252 ,H293404 H291004
H291026 ,H291078 ,H291136 ,H291157 ,H291196 ,H291253 ,H140703 ,H293348 H272926
H291038 ,H291063 ,H140219 ,H291161 ,H291221 ,H293316 ,H293345 ,H293486 H272938
H273016 ,H140216 ,H273120 ,H291113 ,H291222 ,H140240 ,H293429 ,H293397 H290519
H272982 ,H140213 ,H273109 ,H140220 ,H291207 ,H293308 ,H293431 ,H293490 H291003
H291023 ,H291090 ,H291149 ,H291092 ,H291209 ,H293322 ,H293385 ,H293503 H140111
H272990 ,H291068 ,H291134 ,H273137 ,H140228 ,H291224 ,H293426 ,H291176 H140202
H291029 ,H291074 ,H291096 ,H273136 ,H291133 ,H291243 ,H293342 ,H293439 H272960
H291020 ,H140217 ,H291139 ,H291162 ,H291181 ,H291225 ,H293360 ,H293508 H272981
H291042 ,H273026 ,H273118 ,H291183 ,H291171 ,H291267 ,H293418 ,H293485 H272842
H272825 ,H291072 ,H291142 ,H291150 ,H291185 ,H291245 ,H293428 ,H293518 H291013
H291035 ,H291080 ,H291094 ,H291131 ,H291216 ,H291258 ,H293414 ,H293406 H291005
H291039 ,H291097 ,H140218 ,H291053 ,H291203 ,H291255 ,H291226 ,H293354 H140205
H291043 ,H291093 ,H140221 ,H140223 ,H291210 ,H290558 ,H291248 ,H293459 H291010
H272987 ,H291100 ,H291101 ,H273158 ,H273199 ,H293343 ,H291263 ,H293476 H272846
H140207 ,H291105 ,H291146 ,H291159 ,H291208 ,H293358 ,H293436 ,H140720 H272848
H291041 ,H291066 ,H140222 ,H291135 ,H291234 ,H293329 ,H293403 ,H293466 H290504
H291049 ,H291069 ,H140209 ,H291186 ,H291235 ,H293340 ,H293460 ,H293483 H290512
H273052 ,H291081 ,H291119 ,H291192 ,H291215 ,H293335 ,H293413 ,H293405 H272956
H273040 ,H291061 ,H291123 ,H291184 ,H291220 ,H293350 ,H140239 ,H293488 H272963
H291050 ,H273075 ,H291129 ,H291187 ,H291206 ,H293309 ,H293402 ,H293420 H272993
H291056 ,H140214 ,H291121 ,H291144 ,H291241 ,H293346 ,H293472 ,H293433 H272999
H291065 ,H140206 ,H291122 ,H291143 ,H140237 ,H291230 ,H293398 ,H293484 H291024
H140212 ,H291106 ,H291112 ,H291167 ,H291175 ,H293352 ,H140714 ,H140722 H291012
H291057 ,H291102 ,H291145 ,H291158 ,H291217 ,H293302 ,H293453 ,H293401 H272986
H291058 ,H291084 ,H273141 ,H291177 ,H291264 ,H291240 ,H293445 ,H293541 H291019
H291070 ,H291086 ,H291153 ,H273157 ,H273190 ,H293361 ,H293437 ,H293548 H140115
H291071 ,H291087 ,H291118 ,H291148 ,H273209 ,H291269 ,H293427 ,H293344 H291017
H273053 ,H291089 ,H291155 ,H291188 ,H291242 ,H291251 ,H293457 ,H293480 H291014
H291047 ,H291107 ,H291125 ,H291197 ,H291173 ,H293317 ,H140716 ,H293481 H291030
H291051 ,H272996 ,H291126 ,H140230 ,H291246 ,H293330 ,H293461 ,H293478 H291034
H291062 ,H291104 ,H291138 ,H291189 ,H291214 ,H291239 ,H293462 ,H293536 H290530
H291031 ,H291040 ,H291147 ,H291193 ,H291273 ,H291265 ,H293355 ,H293581 H291036
H291060 ,H140210 ,H291156 ,H291124 ,H291228 ,H293399 ,H293415 ,H293319 H293501
H293591 ,H293673 ,H293695 ,H293873 ,H293919 ,H294032 ,H140806 ,H294184 H293517
H293562 ,H293690 ,H140733 ,H293891 ,H140791 ,H294045 ,H294105 ,H294166 H293519
H293504 ,H293587 ,H293771 ,H293890 ,H293926 ,H294030 ,H294112 ,H294178 H293473
H293590 ,H293689 ,H293756 ,H293880 ,H293932 ,H294006 ,H294053 ,H294209 H293458
H293592 ,H293706 ,H140736 ,H293855 ,H293925 ,H294033 ,H293986 ,H140813 H293367
H293538 ,H293647 ,H293611 ,H293868 ,H293948 ,H294041 ,H294125 ,H294194 H293568
H293608 ,H293700 ,H293783 ,H140780 ,H293956 ,H294051 ,H140807 ,H294234 H293564
H293539 ,H293696 ,H293790 ,H293877 ,H293930 ,H294026 ,H140808 ,H294227 H293579
H140701 ,H293694 ,H293750 ,H293885 ,H140793 ,H294048 ,H140810 ,H294207 H293475
H293610 ,H293554 ,H293806 ,H293864 ,H293962 ,H294039 ,H294130 ,H294116 H291095
H140746 ,H293702 ,H140765 ,H293882 ,H293966 ,H293974 ,H294095 ,H294101 H293359
H140718 ,H293692 ,H293708 ,H140784 ,H293899 ,H293994 ,H294106 ,H294197 H293544
H140735 ,H293693 ,H293658 ,H293701 ,H293967 ,H294004 ,H294122 ,H294181 H293552
H293522 ,H293699 ,H293769 ,H293862 ,H293936 ,H294031 ,H294142 ,H294196 H293565
H293543 ,H293703 ,H293821 ,H293913 ,H293953 ,H294008 ,H294132 ,H294188 H293589
H293594 ,H293746 ,H293819 ,H140770 ,H293971 ,H294063 ,H294157 ,H294237 H293347
H293505 ,H291201 ,H293824 ,H293888 ,H293957 ,H294003 ,H294151 ,H294131 H293487
H293642 ,H293733 ,H140776 ,H293902 ,H293934 ,H293984 ,H294099 ,H294202 H293514
H293648 ,H140757 ,H293789 ,H293900 ,H293950 ,H294024 ,H294054 ,H294225 H293516
H293588 ,H293709 ,H293825 ,H140783 ,H293969 ,H294028 ,H294135 ,H294029 H293467
H293606 ,H293734 ,H293754 ,H293886 ,H293995 ,H294025 ,H140809 ,H294148 H293513
H293537 ,H293685 ,H293802 ,H293910 ,H293975 ,H294061 ,H294137 ,H294162 H293515
H293520 ,H293535 ,H293573 ,H293912 ,H293987 ,H294052 ,H140811 ,H294163 H293546
H293551 ,H293659 ,H293805 ,H293889 ,H140782 ,H294050 ,H294156 ,H294240 H293614
H293621 ,H293707 ,H293836 ,H293874 ,H293952 ,H294056 ,H294159 ,H294243 H140725
H293622 ,H293743 ,H293822 ,H293893 ,H293988 ,H293983 ,H294168 ,H140821 H293525
H293660 ,H293723 ,H293846 ,H293903 ,H293985 ,H294086 ,H294058 ,H294244 H293617
H293639 ,H291244 ,H140778 ,H293915 ,H293921 ,H140799 ,H294179 ,H294252 H140721
H140738 ,H293744 ,H140777 ,H293935 ,H294009 ,H294088 ,H140801 ,H294149 H140726
H293636 ,H293748 ,H293856 ,H140787 ,H293979 ,H294075 ,H294169 ,H294126 H140727
H293607 ,H140753 ,H293831 ,H293772 ,H294010 ,H294064 ,H294172 ,H294127 H273196
H293571 ,H140761 ,H293851 ,H293931 ,H294021 ,H294000 ,H294155 ,H294128 H293615
H293683 ,H293737 ,H293840 ,H293894 ,H293989 ,H294089 ,H294177 ,H294245 H140723
H293563 ,H293341 ,H293859 ,H293923 ,H293976 ,H294091 ,H294187 ,H294254 H293609
H140710 ,H140737 ,H293835 ,H293901 ,H294016 ,H294120 ,H140800 ,H294269 H293560
H293684 ,H293749 ,H293863 ,H293892 ,H294012 ,H140802 ,H140817 ,H140820 H293470
H293493 ,H293688 ,H293664 ,H140781 ,H294018 ,H294007 ,H294182 ,H294267 H293635
H293613 ,H293780 ,H293858 ,H293916 ,H294013 ,H294083 ,H294170 ,H294226 H293627
H293691 ,H293775 ,H293837 ,H293933 ,H294001 ,H294035 ,H294189 ,H294266 H140744
H293605 ,H293777 ,H293871 ,H140792 ,H294020 ,H294090 ,H294180 ,H294285 H294261
H294360 ,H294397 ,H294521 ,H294572 ,H140883 ,H294282 ,H294394 ,H293833 H140825
H294158 ,H294421 ,H294443 ,H294577 ,H294698 ,H294294 ,H294365 ,H294486 H294275
H294326 ,H294427 ,H294509 ,H294458 ,H294660 ,H294328 ,H294393 ,H294475 H294150
H294366 ,H294425 ,H294406 ,H294507 ,H294679 ,H294324 ,H294381 ,H294430 H294235
H294358 ,H294429 ,H294530 ,H294571 ,H294696 ,H294325 ,H294380 ,H294479 H140824
H140838 ,H140848 ,H294534 ,H294574 ,H294711 ,H140831 ,H294395 ,H294494 H140827
H294246 ,H294436 ,H294517 ,H294236 ,H294720 ,H294336 ,H294399 ,H294418 H294287
H294333 ,H294437 ,H294470 ,H294598 ,H294719 ,H140832 ,H294362 ,H294478 H294286
H294355 ,H294424 ,H294510 ,H140855 ,H294712 ,H294332 ,H294372 ,H294482 H294295
H294361 ,H294426 ,H294513 ,H294603 ,H294677 ,H294348 ,H294398 ,H294504 H294224
H294373 ,H294469 ,H294516 ,H294604 ,H294686 ,H294350 ,H294420 ,H294485 H294299
H294340 ,H140851 ,H294524 ,H294616 ,H140897 ,H294353 ,H294408 ,H294481 H294298
H294392 ,H294444 ,H294441 ,H140816 ,H294738 ,H294341 ,H294385 ,H294499 H294301
H294376 ,H294468 ,H294544 ,H294599 ,H294735 ,H140833 ,H140846 ,H294483 H294316
H294174 ,H140852 ,H294528 ,H294595 ,H294729 ,H294337 ,H294413 ,H294514 H294302
H294377 ,H294359 ,H294518 ,H294612 ,H294742 ,H294343 ,H294442 ,H140854 H140812
H140830 ,H294472 ,H294206 ,H294629 ,H294727 ,H294330 ,H294404 ,H294484 H294303
H294378 ,H140826 ,H294546 ,H294625 ,H140891 ,H140835 ,H294410 ,H294505 H294314
H294384 ,H294416 ,H140845 ,H140867 ,H294743 ,H140837 ,H294417 ,H294501 H294476
H294615 ,H140881 ,H294493 ,H140868 ,H294747 ,H294221 ,H294386 ,H140853 H294554
H294576 ,H140895 ,H294531 ,H294555 ,H294753 ,H294327 ,H294382 ,H294497 H294556
H294630 ,H140896 ,H294487 ,H140869 ,H294689 ,H294498 ,H294617 ,H294690 H294552
H294613 ,H294730 ,H294300 ,H294624 ,H294761 ,H294500 ,H294642 ,H294763 H294506
H294653 ,H294771 ,H294431 ,H294652 ,H294783 ,H294391 ,H294567 ,H140702 H294549
H294658 ,H294773 ,H294311 ,H294654 ,H294794 ,H294573 ,H294694 ,H294253 H294575
H294680 ,H294782 ,H294446 ,H294665 ,H291091 ,H140858 ,H294626 ,HCX4019 H294550
H294695 ,H294291 ,H140876 ,H293542 |
| FEI Number |
2916710
|
Recalling Firm/ Manufacturer |
Varian Medical Systems Oncology Systems 911 Hansen Way Palo Alto CA 94304-1028
|
| For Additional Information Contact | 650-424-5731 |
Manufacturer Reason for Recall | The Auto goto or Auto setup functions in C-Series version 7 software ignore the couch angle and exceed motion zone (unless a tolerance is defined by the user); it has the potential for collision with the patient on the couch.
|
FDA Determined Cause 2 | Other |
| Action | Varian sent a Customer Technical Bulletin letter to all affected users on May 27, 2010, and was also posted to Varian's customer support site: http://www.MyVarian.com. The letter identified the product, the problem, and provided instructions on the recall.
Customers were instructed to return the proof of notification (via completed "Proof of Notification Form or appropriate Postal system certified receipt of delivery to Varian.
For questions regarding this recall call 650-483-3153. |
| Distribution | Worldwide Distribution - USA including CA, TX, IL, GA, NJ, and NV and the countries of Australia, Belgium, Brazil, Canada, China, Denmark, France, Germany, Hong Kong, India, Italy, Japan, Middle East, Netherlands, Russia, Spain, Switzerland, and the UK |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = GKZ 510(K)s with Product Code = GKZ
|
|
|
|