| Class 2 Device Recall CELLDYN Sapphire Hematology Analyzer |  |
Date Initiated by Firm | April 05, 2010 |
Date Posted | September 14, 2010 |
Recall Status1 |
Terminated 3 on May 17, 2011 |
Recall Number | Z-2421-2010 |
Recall Event ID |
56422 |
510(K)Number | K051215 |
Product Classification |
Counter, Differential Cell - Product Code GKZ
|
Product | CELL-DYN Sapphire Probe Detector PCB Assembly (also called the Aspiration Bottom Sensor) used on the Cell-Dyn Sapphire Hematology Analyzer, Manufactured by Abbott Laboratories, Santa Clara, CA |
Code Information |
List 08H00-01, 08H00-03, PCB Assembly part number 8960167101 |
Recalling Firm/ Manufacturer |
Abbott Laboratories 5440 Patrick Henry Dr Santa Clara CA 95054-1113
|
For Additional Information Contact | 408-982-4800 |
Manufacturer Reason for Recall | Under certain conditions in closed mode of operations, the Aspiration Probe Detector PCB assembly may fail to operate properly and provide inaccurate results. |
FDA Determined Cause 2 | Other |
Action | Abbott Diagnostics Division of Santa Clara, CA initiated a field correction of the CELL-DYN Sapphire Probe Detector Assembly used on the CELL-DYN Sapphire Hemotology Analyzer, on April 5, 2010. The correction is being addressed through field service representatives and a mandatory Technical Service Bulletin (TSB). The TSB requires inspection, testing, and potential replacement of the affected assembly. The firm will not be issuing a seperate notification. For questions regarding this correction please contact (408) 982-4800, or fax (408) 982-4863. |
Quantity in Commerce | 926 units |
Distribution | Product was distributed to 926 retailers throughout the US and to argentina, Australia, Brazil, Canada, Chile, China, Columbia, Germany, Hong King, Japan, Mexico, New Zealand, Panama, Paraguay, Peru, Singapore, South Korea, Taiwan and Venezuela. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = GKZ
|
|
|
|