| Class 2 Device Recall Panorama HFO | |
Date Initiated by Firm | February 05, 2010 |
Date Posted | March 01, 2011 |
Recall Status1 |
Terminated 3 on August 10, 2011 |
Recall Number | Z-1458-2011 |
Recall Event ID |
57561 |
510(K)Number | K041602 |
Product Classification |
System, nuclear magnetic resonance imaging - Product Code LNH
|
Product | The Panorama HFO system has the Traceable Item Identification (12NC):
9896 030 17741-SW R 2.5.1 (shipped as R 2.5.1, upgraded to R 2.5.3 via FCO).
9896 030 19641-SW R 2.5.3
9896 030 19861-SW R 2.6.1
9896 030 19862-SW R 2.6.1
9896 030 19863-SW R 2.6.3 |
Code Information |
Site Numbers: 533860, 533900, 535230, 535664, 538574, 539374, 540482, 541065, 541851, 542236, 544039, 544148, 544696, 545034, 545732, 545981, 546206, 547843, 547897, 548040, 548064, 549389, 549909, 550324, 550439, 551025, 551583, 552453, 552526, 552609, 553116, 554715, 556146, 556161, 556895, 557599, 41443708, 41444456, 41444458, 41444745, 41445015, 41445018, 41445530, 41445594, 41445595, 41446119, 41455905, 41455908, 41455913, 41455915, 41455923, 41455938, 41455940, 41455954, 42047935, 42057444, 42293607, 42392891, 42608348, 42850814, 42892497, 42917294, 43708886, 44062048, 44247559, 44605983, 45130117, 45245963, 45246494, 45246504, 45445404, 47157157, 47230785, 47455618, 47480028, 47848863, 49060365, 49060404, 49416204, 49424702, 49776644, and 49861074. |
Recalling Firm/ Manufacturer |
Philips Medical Systems North America Co. Phillips 22100 Bothell Everett Hwy Bothell WA 98021-8431
|
For Additional Information Contact | Philips Call Center 800-722-9377 |
Manufacturer Reason for Recall | In the ExamCard MobiFlex the Patient Position is "Prone" which must be "Supine". The result is mislabeling of left and right side of the patient. If this is not detected by the operator there is a risk of wrong treatment or the need of additional X-RAY during treatment. |
FDA Determined Cause 2 | Other |
Action | Philips began sendng out the URGENT-Device Correction Notice letter dated February 5, 2010, ro all affected consignees on February 10, 2010.
The consignees were advised to change the patient position from "Prone" immediately to "Supine" and store the corrected ExamCard in the hospital database if they use the ExamCard, MobiFlex dual Injection.
Consignees were instructed to call Philips Healthcare Call Center at 800-722-9377 #5 (then follow prompts) and also reference "FCO 78100318" with any communication. |
Quantity in Commerce | 82 units |
Distribution | Nationwide Distribution including the following states: AL, CA, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, NE, NH, NJ, NY, OH, OR, SC, TN, TX, VA, VT, WI, and WV.AL, CA, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, NE, NH, NJ, NY, OH, OR, SC, TN, TX, VA, VT, WI, and WV.
There was no foreign consignees. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = LNH
|
|
|
|