|
Class 2 Device Recall UniCel DxH 800 Coulter Cellular Analysis System |
|
Date Initiated by Firm |
April 14, 2011 |
Date Posted |
May 20, 2011 |
Recall Status1 |
Terminated 3 on May 23, 2012 |
Recall Number |
Z-2227-2011 |
Recall Event ID |
58584 |
510(K)Number |
K081930
|
Product Classification |
Differential Cell Counter - Product Code GKZ
|
Product |
Unicel¿ DxH 800 Coulter¿ Cellular Analysis System.
A quantitative, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. |
Code Information |
Part # 629029 |
Recalling Firm/ Manufacturer |
Beckman Coulter Inc. 250 S Kraemer Blvd Brea CA 92821-6232
|
For Additional Information Contact |
Clair K. O'Donovan, Ph.D. 714-961-4483
|
Manufacturer Reason for Recall |
The recall was initiated because Beckman Coulter has confirmed the auto prune will not remove patient results if the patient record includes demographics where the results are less than 30 days old.
|
FDA Determined Cause 2 |
Software design |
Action |
The recall communication was initiated on 4/12/11 with Beckman Coulter forwarding an Urgent Product Correction (PC) letter with attached PC Response Form (via US Postal Service for US customers) to all customers who purchased the UniCel¿ DxH" 800 Coulter¿ Cellular Analysis System PN 629029. The letter provides customers with an explanation of the problem identified and an action to be taken. For Issue 1 & 2, customers were instructed to perform manual database maintenance procedures to locate and delete unnecessary patient result and demographic information that extends beyond the life (in days) of the control lots. Ensure an alternative method of analysis is available during database maintenance.
For Issue 3, customers were instructed to not attempt to transmit the files again after receiving an incomplete report. Instead, submit the missing file(s) by one of the following methods:
- Submit your data at the electronic IQAP website (eIQAP) at
(http://www.beckmancoulter.com/qap/index.jsp)
- Fax printouts of the missing files to 786-639-3999
- Email an electronic copy (.CSV) of the control file to IQAP.Reports@beckman.com.
- Include the IQAP number on any communication. A corrected report will be returned to the customers' laboratories.
For issue 4, customers were instructed to submit their control data to IQAP by one of the following methods:
-Submit their data at the electronic IQAP website (eIQAP) at
(http://www.beckmancoulter.com/qap/index.jsp)
-Fax printouts of the missing files to 786-639-3999
-Email an electronic copy (.CSV) of the control file to IQAP.Reports@beckman.com.
-Include your IQAP number on any communication.
Customers were instructed to complete and return the enclosed Response Form within ten days. Customers were instructed to contact Beckman Coulter Customer Service at 800-526-7694 in the United States or Canada, or contact their local Beckman Coulter representative. |
Quantity in Commerce |
673 units |
Distribution |
Worldwide Distribution -- US, Canada, Australia, Belgium, Colombia, Croatia, Czech Republic, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Republic of Korea, Macao, Malaysia, Myanmar, Netherlands, Norway, Philippines, Puerto Rico, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, and United Kingdom. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = GKZ and Original Applicant = BECKMAN COULTER, INC.
|
|
|
|