| Part B: Supplementary Information Sheet (SIS) |
|
FR Recognition List Number
|
055
|
Date of Entry 12/21/2020
|
|
FR Recognition Number
|
11-370
|
| Standard | |
ISO 14879-1 Second edition 2020-07 Implants for surgery - Total knee-joint prostheses - Part 1: Determination of endurance properties of knee tibial trays |
|
Scope/AbstractThis document specifies a test method for determining the endurance properties, under specified laboratory conditions, of tibial trays used in knee-joint prostheses to support and secure the plastic articulating surface. It applies to tibial trays which cover both the medial and lateral plateaux of the tibia.
The test method does not apply to tibial components manufactured solely from plastic materials.
This document does not cover methods of examining and reporting the final condition of the test specimen; these can be the subject of agreement between the test laboratory and the parties submitting the specimen for test.
NOTE Correlation of test results with in vivo performance has not been established. |
|
| Extent of Recognition
|
Rationale for Recognition
| This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies. |
|
Public Law, CFR Citation(s) and Procode(s)*
| Regulation Number |
Device Name |
Device Class |
Product Code |
| §888.3480 |
Metal Cemented Constrained Femorotibial Knee Prosthesis
|
Class 3
|
KRN
|
| §888.3510 |
Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer
|
Class 2
|
KRO
|
| §888.3520 |
Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer
|
Class 2
|
HSX
|
| §888.3530 |
Prosthesis, Knee, Femorotibial, Semi-Constrained, Cemented, Metal/Polymer
|
Class 2
|
HRY
|
| §888.3550 |
Prosthesis, Knee, Patello/Femorotibial, Constrained, Cemented, Polymer/Metal/Metal
|
Class 3
|
KRP
|
| §888.3550 |
Prosthesis, Knee, Patello/Femorotibial, Constrained, Cemented, Polymer/Metal/Polymer
|
Class 2
|
KRQ
|
| §888.3560 |
Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
|
Class 2
|
JWH
|
| §888.3560 |
Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uhmwpe, Pegged, Cemented, Polymer/Metal/Polymer
|
Class 2
|
MBV
|
| §888.3565 |
Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer
|
Class 2
|
MBH
|
| N/A |
Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Polymer/Metal/Polymer |
Class 3
|
LXY
|
| N/A |
Prosthesis, Knee, Patello/Femorotibial, Unconstrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer |
Class 3
|
MBD
|
|
Relevant FDA Guidance and/or Supportive Publications*
Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA, January 16, 2003
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018. |
|
| FDA Technical Contact
|
| Standards Development Organization
|
| FDA Specialty Task Group (STG)
|
|
|
| *These are provided as examples and others may be applicable. |
|
|