Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 031 Date of Entry 08/06/2013 
FR Recognition Number 3-115
Standard(Included in ASCA)
IEC  60601-2-34 Edition 3.0 2011-05
Medical electrical equipment - Part 2-34: Particular requirements for the basic safety, including essential performance, of invasive blood pressure monitoring equipment
Scope/Abstract
IEC 60601-2-34:2011 concerns the basic safety and essential performance of invasive blood pressure monitoring equipment. It amends and supplements IEC 60601-1 (third edition, 2005): Medical electrical equipment - Part 1: General requirements for basic safety and essential performance. The third edition of IEC 60601-2-34 cancels and replaces the second edition published in 2001 and constitutes a technical revision in order to align structurally with the 2005 edition of IEC 60601-1. IEC 60601-2-34 applies to basic safety and essential performance of invasive blood pressure monitoring equipment. It does not apply to catheter tubing, catheter needles, Luer locks, taps and tap tables connected. It does not apply to non-invasive blood pressure monitoring equipment.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Transition Period
FDA recognition of IEC 60601-2-34 Edition 3.0 2011-05 [Rec# 3-115] will be superseded by recognition of IEC 60601-2-34 Edition 4.0 2024-10 [Rec# 3-202]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 3-115] until December 19, 2027. After this transition period, declarations of conformity to [Rec# 3-115] will not be accepted.
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
§870.1110 Computer, Blood-Pressure Class 2 DSK
§870.2060 Amplifier And Signal Conditioner, Transducer Signal Class 2 DRQ
§870.2850 Transducer, Blood-Pressure, Extravascular Class 2 DRS
§870.2870 Transducer, Pressure, Catheter Tip Class 2 DXO
Relevant FDA Guidance and/or Supportive Publications*
AAMI CR500:2019 Basic Introduction to the IEC 60601 Series.

Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contact
 Gavin D'Souza
  FDA/OC/CDRH/OPEQ/OHTII
  301-796-2693
  Gavin.DSouza@fda.hhs.gov
Standards Development Organization
IEC International Electrotechnical Commission http://www.iec.ch/
FDA Specialty Task Group (STG)
Cardiovascular
*These are provided as examples and others may be applicable.
-
-