| | Class 3 Device Recall V92 |  |
| Date Initiated by Firm | July 03, 2025 |
| Date Posted | August 22, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2404-2025 |
| Recall Event ID |
97217 |
| 510(K)Number | K182857 K202216 |
| Product Classification |
System, imaging, pulsed doppler, ultrasonic - Product Code IYN
|
| Product | V9-2 Transducer Probe |
| Code Information |
Model No. 989605466201, 989606717071, 989605492481, 989605456291; UDI: (01)00884838088573(21)B3FMNT, (01)00884838102606(21)B3FLM3, (01)00884838102606(21)B411GD, (01)00884838102606(21)F053BY, (01)00884838102606(21)F0HT0R, 010088483810260621B3Y0MP, (01)00884838102606(21)F0KG9X, (01)00884838088573(21)B3FLWW, (01)00884838102606(21)F0HT11, (01)00884838102606(21)F04XCK, 010088483810260621F0FQZG, 010088483810260621F0FVH7, (01)00884838102606(21)F06PQM, (01)00884838102606(21)F06TGM, (01)00884838088573(21)B32TQN, (01)00884838102606(21)F0HJWD, (01)00884838102606(21)F0986F, (01)00884838102606(21)F07D1M, (01)00884838088573(21)B32TQZ, (01)00884838088573(21)B32TWD, (01)00884838088573(21)B32V4Y, (01)00884838102606(21)F09LH5, (01)00884838102606(21)F0JZJM, (01)00884838102606(21)F0576P, (01)00884838102606(21)F0CN2B, (01)00884838102606(21)F0F02Q, (01)00884838102606(21)F0JL42, (01)00884838088573(21)B3FLCN, (01)00884838088573(21)B3FMXB, (01)00884838088573(21)B2PYFR, (01)00884838088573(21)B2ZR1X, (01)00884838088573(21)B32V5Y, (01)00884838088573(21)B3FMPD, (01)00884838088573(21)B3FMPR, (01)00884838088573(21)B3FMPV, (01)00884838088573(21)B3FMQ5, (01)00884838088573(21)B3FMR6, (01)00884838102606(21)F0BHVR, (01)00884838102606(21)F0JWZG, (01)00884838102606(21)F0HJWH, (01)00884838088573(21)B3FMQ0, (01)00884838102606(21)F04XD2, (01)00884838102606(21)B3XYZG, (01)00884838102606(21)F0CHGT, (01)00884838088573(21)B32TWR, (01)00884838088573(21)B32V6B, (01)00884838102606(21)F073H5, (01)00884838088573(21)B3FMFT, (01)00884838102606(21)F0CN1C, (01)00884838102606(21)F0DP3N, (01)00884838102606(21)F0HX7Q, (01)00884838102606(21)F0J128, (01)00884838102606(21)F073H3, (01)00884838102606(21)F09BF3, (01)00884838102606(21)F0B49X, (01)00884838102606(21)F0CN22, (01)00884838102606(21)F08HC0, (01)00884838102606(21)F0BCZZ, (01)00884838102606(21)F0BD0Q, (01)00884838088573(21)B3FMPW, (01)00884838102606(21)F06THR, (01)00884838102606(21)B3FLFH, (01)00884838102606(21)B3Y06W, (01)00884838102606(21)F077M2, (01)00884838088573(21)B32V6Y, (01)00884838088573(21)B32V7T, (01)00884838088573(21)B32V9B, (01)00884838102606(21)F04XCZ, (01)00884838088573(21)B3FKXX, (01)00884838088573(21)B3LBCN, (01)00884838088573(21)B3FMT7, (01)00884838088573(21)B3FM94, (01)00884838088573(21)B3FMCT, (01)00884838102606(21)F0JTYL, (01)00884838102606(21)B3XZV4, (01)00884838102606(21)F09Z2F, (01)00884838102606(21)F077LB, (01)00884838102606(21)F04V3T, (01)00884838102606(21)B3Y0FQ, (01)00884838102606(21)F0F5TN, 010088483810260621F0FZ5Q, (01)00884838088573(21)B3FM8Y, (01)00884838088573(21)B3FL58, (01)00884838102606(21)F0BPYX, (01)00884838102606(21)F0785K, (01)00884838102606(21)F08K80, (01)00884838102606(21)F08MHQ, (01)00884838088573(21)B32VBM, (01)00884838088573(21)B3FMX2, (01)00884838102606(21)F08JLG, (01)00884838102606(21)F08XZ1, (01)00884838102606(21)F096BL, 010088483810260621B3Y0KV, (01)00884838088573(21)B3FL1R, (01)00884838088573(21)B3FL59, (01)00884838088573(21)B3FL7C, (01)00884838088573(21)B3FLC8, (01)00884838088573(21)B3FMW5, 010088483810260621B3FM16, (01)00884838102606(21)F077M8, (01)00884838102606(21)B3Y06F, (01)00884838102606(21)B3Y06P, (01)00884838102606(21)B3Y06R, (01)00884838102606(21)B3Y0N7, (01)00884838102606(21)F08W4H, (01)00884838102606(21)F08XZL, (01)00884838102606(21)F0BD08, (01)00884838102606(21)F0CQ5C, (01)00884838102606(21)F0CV7B, (01)00884838102606(21)F0DWF3, (01)00884838102606(21)F04P7C, (01)00884838102606(21)F09FYJ, (01)00884838102606(21)B3Y0L1, (01)00884838102606(21)B3XZH3, (01)00884838102606(21)F05HY6, (01)00884838102606(21)F0BGZZ, (01)00884838102606(21)F0BD00, (01)00884838088573(21)B32TP0, (01)00884838088573(21)B3FL7J, 010088483810260621B3Y0KY, (01)00884838102606(21)F06PR2, 010088483810260621F0G352, (01)00884838088573(21)B3FMWT, (01)00884838102606(21)F0J12C, (01)00884838102606(21)F0D436, (01)00884838102606(21)B3XZZJ, (01)00884838102606(21)B3XZM0, (01)00884838088573(21)B3FMNR, (01)00884838102606(21)F0B274, (01)00884838102606(21)F05BT9, (01)00884838102606(21)F05HY9, 010088483810260621B3Y0JC, (01)00884838088573(21)B3FMX0, (01)00884838088573(21)B32VBT, (01)00884838102606(21)B3Y04W, (01)00884838102606(21)F04XCT, (01)00884838102606(21)F07CCM, (01)00884838102606(21)F0CHGW, (01)00884838102606(21)B3FM1B, (01)00884838102606(21)F0BKH3, (01)00884838102606(21)F0JTFF, (01)00884838102606(21)F0JWZK, (01)00884838102606(21)F0JY22, 010088483810260621B3Y0N5, 010088483810260621F0FZV3, (01)00884838102606(21)F0BCZQ, (01)00884838102606(21)F0BHVY, (01)00884838102606(21)F0CN1F, (01)00884838102606(21)F0CQ52, (01)00884838102606(21)F0BT8G, (01)00884838102606(21)B3FM03, (01)00884838102606(21)B3FM0L, (01)00884838102606(21)B3FM3D, (01)00884838088573(21)B3FL03, (01)00884838102606(21)B3XZZM, (01)00884838102606(21)F08K7M, (01)00884838102606(21)F09QR2, (01)00884838088573(21)B3LBCQ, (01)00884838088573(21)B3L4T9, (01)00884838102606(21)F0JTYK, (01)00884838102606(21)F0D432, (01)00884838088573(21)B3FL28, (01)00884838102606(21)F08W46, (01)00884838088573(21)B3FL0G, (01)00884838102606(21)F07WT3, (01)00884838102606(21)F0F1LH, (01)00884838102606(21)F0FMX1, (01)00884838102606(21)F0F4CJ, (01)00884838102606(21)F0JDM2, (01)00884838088573(21)B32TTZ, (01)00884838102606(21)F0CQ4K, (01)00884838102606(21)F0JH10, (01)00884838102606(21)F0HT0L, 010088483810260621B3Y0M8, (01)00884838102606(21)F08MHC, (01)00884838102606(21)F053C1, (01)00884838102606(21)F0CHGZ, (01)00884838102606(21)F08F29, (01)00884838102606(21)F08R3Q, (01)00884838102606(21)F08R43, (01)00884838102606(21)F08W4M, (01)00884838102606(21)F0JH0V, (01)00884838102606(21)F0JH0Z, (01)00884838102606(21)F0JQTR, (01)00884838102606(21)F0JWZN, (01)00884838102606(21)F05LXL, (01)00884838102606(21)F06FBB, (01)00884838102606(21)F06THW, (01)00884838088573(21)B3FL05, 010088483810260621F0FQZT, (01)00884838088573(21)B32V1Y, (01)00884838102606(21)F0K1MF, (01)00884838088573(21)B2PYFC, (01)00884838088573(21)B3FL9H, (01)00884838088573(21)B32V8Y, (01)00884838102606(21)F09TK9, (01)00884838102606(21)B3FM00, (01)00884838102606(21)B3FM0M, (01)00884838102606(21)F04P80, (01)00884838102606(21)F05LY2, (01)00884838102606(21)F06BP6, (01)00884838088573(21)B32V1W, (01)00884838102606(21)B3XZ05, (01)00884838102606(21)B3Y05Z, (01)00884838102606(21)B3Y0G6, (01)00884838102606(21)B3Y0KQ, (01)00884838088573(21)B3FLN5, (01)00884838088573(21)B3FMNK, (01)00884838088573(21)B3FMNN, (01)00884838102606(21)F07WRY, (01)00884838102606(21)F0B9Y9, (01)00884838102606(21)B3XZQ8, (01)00884838102606(21)B3Y06X, (01)00884838102606(21)F0G34Z, (01)00884838102606(21)F09C06, (01)00884838088573(21)B32TR1, (01)00884838088573(21)B32TRF, (01)00884838088573(21)B3FMTQ, (01)00884838088573(21)B3FMMM, (01)00884838088573(21)B3FMV7, (01)00884838088573(21)B3FMVC, (01)00884838088573(21)B3FMVH, (01)00884838088573(21)B3FMVJ, (01)00884838088573(21)B3FMVK, (01)00884838088573(21)B3FMW8, (01)00884838088573(21)B3FMW9, (01)00884838088573(21)B3FMX1, (01)00884838102606(21)B3XZNX, (01)00884838102606(21)F04MZ5, (01)00884838102606(21)F0HXHH, (01)00884838102606(21)F0HXHR, (01)00884838102606(21)F0576B, (01)00884838102606(21)B3FLKX, (01)00884838102606(21)F0HNBJ, (01)00884838102606(21)F0JH14, (01)00884838102606(21)F0HXHD, (01)00884838102606(21)B415T7, (01)00884838102606(21)F0BCZV, (01)00884838102606(21)F0BHVK, (01)00884838102606(21)F0B27J, (01)00884838102606(21)B3XZ13, (01)00884838102606(21)B3Y06J, 010088483810260621F0FZV8, (01)00884838088573(21)B3FLG8, (01)00884838102606(21)B3Y0JQ, (01)00884838088573(21)B32V32, (01)00884838102606(21)F07857, (01)00884838102606(21)B3XYZ9, (01)00884838088573(21)B3FLWR, (01)00884838088573(21)B3LBCP, (01)00884838088573(21)B3LBDM, (01)00884838102606(21)F053C4, (01)00884838102606(21)F05BT7, (01)00884838088573(21)B3FMP9, (01)00884838088573(21)B3FMPH, (01)00884838102606(21)B3XZ1H, (01)00884838088573(21)B32V6X, (01)00884838102606(21)B3FLND, (01)00884838102606(21)B3XZWD, (01)00884838102606(21)F0DP3M, 010088483810260621B3Y0N1, (01)00884838088573(21)B3FM8L, (01)00884838102606(21)F0CQ5K, (01)00884838102606(21)B3Y02L, (01)00884838088573(21)B32TZL, (01)00884838088573(21)B32VD7, (01)00884838102606(21)F06FB6, (01)00884838088573(21)B2PYHH, (01)00884838102606(21)B3XYYF, (01)00884838102606(21)B3Y0J4, (01)00884838102606(21)F05HY0, (01)00884838088573(21)B32TYK, (01)00884838088573(21)B32TZZ, (01)00884838088573(21)B3FMLQ, (01)00884838088573(21)B3FMM1, (01)00884838102606(21)F08BBY, (01)00884838102606(21)F09C07, (01)00884838102606(21)F09LHQ, 010088483810260621B3FM2W, (01)00884838102606(21)F0HT2B, (01)00884838102606(21)B3XYTV, (01)00884838102606(21)B3XZ17, (01)00884838102606(21)F0CN2F, (01)00884838088573(21)B3FMPZ, (01)00884838088573(21)B3L519, (01)00884838102606(21)F0JQTF, 010088483810260621B3FM19, (01)00884838102606(21)B3XZ09, (01)00884838088573(21)B3FMK4, (01)00884838088573(21)B3FMN9, (01)00884838088573(21)B32TMB, (01)00884838088573(21)B32V8R, (01)00884838102606(21)F0G7NP, (01)00884838102606(21)F08MHD, (01)00884838102606(21)B3FLPD, (01)00884838102606(21)B3Y062, (01)00884838102606(21)F08XZF, (01)00884838088573(21)B3L4ZY, (01)00884838088573(21)B3L511, (01)00884838102606(21)F06FBH, (01)00884838102606(21)F06FBJ, (01)00884838102606(21)F06KTR, (01)00884838102606(21)F06N29, (01)00884838102606(21)B3XZYZ, 010088483810260621B3Y0M4, (01)00884838088573(21)B32V2M, (01)00884838088573(21)B3LB9X, (01)00884838102606(21)B3Y06K, (01)00884838102606(21)F0JH0T, (01)00884838102606(21)B3XZTL, (01)00884838088573(21)B32V81, (01)00884838102606(21)F0HNBF, (01)00884838102606(21)F073H1, (01)00884838102606(21)F05HY7, (01)00884838088573(21)B3FMB1, (01)00884838102606(21)F09L7Q, (01)00884838102606(21)F09LHB, (01)00884838102606(21)F04XCH, (01)00884838102606(21)B3FM3L, (01)00884838102606(21)B3Y0HH, (01)00884838102606(21)F08HC9, (01)00884838102606(21)F08JLD, (01)00884838102606(21)F0HNB3, (01)00884838102606(21)F0J5Z6, (01)00884838102606(21)B3Y04D, (01)00884838102606(21)B3XZTD, (01)00884838102606(21)B3FLL8, (01)00884838102606(21)F07855, (01)00884838102606(21)F07CCW, (01)00884838102606(21)B3FM47, (01)00884838102606(21)B3Y0FY, (01)00884838102606(21)F0CN2D, 010088483810260621B3FM4M, 010088483810260621B3FM5N, 010088483810260621B3FM63, 010088483810260621B3FM82, (01)00884838088573(21)B3FL01, (01)00884838088573(21)B3FL1Y, (01)00884838102606(21)F0B9YK, (01)00884838102606(21)F0BP4K, (01)00884838102606(21)F0BT7N, (01)00884838102606(21)F0CQ4W, (01)00884838102606(21)B3FM5Y, (01)00884838088573(21)B3FMT4, 010088483810260621B3FM7X, (01)00884838102606(21)F05HYK, (01)00884838102606(21)B3XYT2, (01)00884838102606(21)B3XYT8, (01)00884838102606(21)B3XYZH, (01)00884838102606(21)B3XZ0Y, (01)00884838102606(21)F08XZ4, (01)00884838088573(21)B3FL84, (01)00884838102606(21)F09L7X, (01)00884838088573(21)B2ZR2T, (01)00884838088573(21)B3FL1Q, (01)00884838102606(21)B415RB, (01)00884838088573(21)B3FL0K, (01)00884838102606(21)B411G8, (01)00884838102606(21)B41GPJ, (01)00884838088573(21)B2DHZK, (01)00884838088573(21)B2DHZN, (01)00884838088573(21)B32TC9, (01)00884838088573(21)B32V65, (01)00884838088573(21)B32V7V, (01)00884838088573(21)B3FMMC, (01)00884838102606(21)B3XZN1, (01)00884838102606(21)F07866, (01)00884838102606(21)F07CCR, (01)00884838102606(21)F07K7N, (01)00884838102606(21)F07NXK, (01)00884838102606(21)F07NXL, (01)00884838102606(21)F079WX, (01)00884838102606(21)F07CD7, (01)00884838102606(21)F07WTG, (01)00884838102606(21)B3XZ11, (01)00884838102606(21)F09866, (01)00884838088573(21)B3FL04, (01)00884838088573(21)B3FL1T, (01)00884838102606(21)B3XYTW, (01)00884838102606(21)F0F5RG, (01)00884838088573(21)B32V16, (01)00884838102606(21)F0F45C, (01)00884838102606(21)F09LH8, (01)00884838102606(21)F0JG5M, (01)00884838102606(21)F0CN1Y, (01)00884838102606(21)F0BCZH, (01)00884838102606(21)B3FM14, (01)00884838102606(21)B3Y03P, (01)00884838102606(21)B3Y05X, (01)00884838102606(21)B3Y0FP, (01)00884838102606(21)B3Y0FV, (01)00884838102606(21)B3FLW5, (01)00884838102606(21)F08F28, (01)00884838102606(21)F08HBY, (01)00884838102606(21)F09C04, (01)00884838102606(21)F09D2V, (01)00884838102606(21)F09LGQ, (01)00884838102606(21)F0B8CY, (01)00884838102606(21)F0BD06, (01)00884838102606(21)F0BD0F, (01)00884838102606(21)F0BH04, (01)00884838102606(21)F0BKH5, (01)00884838102606(21)F0D434, (01)00884838102606(21)F0HG5J, (01)00884838102606(21)F0HT24, (01)00884838102606(21)F0J4HT, (01)00884838102606(21)F0J5Z3, (01)00884838102606(21)F0785L, (01)00884838102606(21)F09LP7, (01)00884838088573(21)B3FLCW, (01)00884838088573(21)B3FM92, (01)00884838102606(21)B3XZKX, (01)00884838102606(21)B3XZL3, (01)00884838102606(21)F0G7NN, (01)00884838102606(21)B3Y01K, (01)00884838102606(21)B3Y02G, (01)00884838102606(21)F08W4X, (01)00884838102606(21)F0HFMN, (01)00884838088573(21)B3FMWD, (01)00884838088573(21)B3FMX7, 010088483810260621B3Y0H4, 010088483810260621B3Y0HZ, 010088483810260621B3Y0J6, (01)00884838102606(21)B3XZQ6, (01)00884838102606(21)F0BD0K, (01)00884838102606(21)F0BKH7, (01)00884838102606(21)B3XZQ3, (01)00884838102606(21)F0HG48, (01)00884838102606(21)F0HG56, (01)00884838088573(21)B32V38, (01)00884838088573(21)B32V0Y, (01)00884838102606(21)F07CD4, (01)00884838088573(21)B32V4H, (01)00884838102606(21)F08W55, (01)00884838102606(21)F05HYH, (01)00884838102606(21)F05HYW, (01)00884838102606(21)F0B9Y3, (01)00884838102606(21)F0BKGP, (01)00884838102606(21)F06N31, (01)00884838102606(21)F06PR6, (01)00884838102606(21)F0HT23, (01)00884838102606(21)F0J5Z2, (01)00884838102606(21)F077LZ, (01)00884838102606(21)B3XZW5, (01)00884838102606(21)F09BF6, (01)00884838102606(21)F09LH3, (01)00884838102606(21)B3FM0Q, (01)00884838102606(21)B3XZWN, (01)00884838102606(21)B3XZWP, (01)00884838102606(21)F08XZQ, (01)00884838102606(21)F09FXZ, (01)00884838102606(21)F05LXZ, (01)00884838102606(21)F06N2Y, 010088483810260621F0FP2F, 010088483810260621F0FQZR, 010088483810260621F0GCLL, (01)00884838102606(21)F0986Y, (01)00884838102606(21)F0BD0C, (01)00884838102606(21)F0HXHY, (01)00884838102606(21)F053C3, (01)00884838102606(21)F0HVXX, (01)00884838102606(21)F0JCQX, 010088483810260621F0FZV9, (01)00884838102606(21)F0KL6G, (01)00884838102606(21)F0JZJB, (01)00884838102606(21)B3Y0N4, (01)00884838102606(21)B3FLZZ, (01)00884838102606(21)B3FM0J, (01)00884838102606(21)F04XD5, (01)00884838102606(21)B3XYTQ, (01)00884838102606(21)B3XZF4, (01)00884838102606(21)F09FY7, (01)00884838102606(21)F0HNBD, 010088483810260621F0GCLK, (01)00884838088573(21)B32TPF, (01)00884838088573(21)B32TQW, (01)00884838088573(21)B32THK, (01)00884838088573(21)B32TP4, (01)00884838088573(21)B32TWP, (01)00884838088573(21)B32TX3, (01)00884838088573(21)B32TX8, (01)00884838088573(21)B32TXZ, (01)00884838102606(21)F09LGV, (01)00884838102606(21)F04MYZ, (01)00884838102606(21)F04V4D, (01)00884838102606(21)F0J38C, (01)00884838102606(21)F0J9JM, (01)00884838102606(21)F0JT57, (01)00884838102606(21)F06KTP, (01)00884838102606(21)F0B3FN, (01)00884838088573(21)B32VDD, (01)00884838102606(21)F077MB, (01)00884838088573(21)B32VBH, (01)00884838102606(21)B3Y066, (01)00884838088573(21)B3FLC5, (01)00884838088573(21)B3FMG2, (01)00884838088573(21)B32VD8, (01)00884838102606(21)F0CQ5P, (01)00884838102606(21)F0986K, (01)00884838102606(21)F0986L, (01)00884838102606(21)F09FXQ, (01)00884838102606(21)F06FB0, (01)00884838102606(21)F0HNBC, (01)00884838102606(21)F0HXJ1, (01)00884838102606(21)F096C4, (01)00884838088573(21)B3FKY7, (01)00884838088573(21)B3FMG3, (01)00884838102606(21)F06HZK, (01)00884838102606(21)F09FY6, (01)00884838102606(21)F0DP3H, (01)00884838088573(21)B3FMRL, (01)00884838102606(21)F06KTB, (01)00884838102606(21)F08K7T, (01)00884838088573(21)B32V98, (01)00884838102606(21)F07WTK, (01)00884838102606(21)F0HG5F, (01)00884838102606(21)B3FLJG, (01)00884838088573(21)B32TPH, (01)00884838102606(21)F08K8G, (01)00884838088573(21)B3L4Z2, (01)00884838102606(21)F09D34, (01)00884838088573(21)B32QCM, (01)00884838088573(21)B3FL9X, (01)00884838088573(21)B3FMWF, (01)00884838102606(21)F0JCQY, (01)00884838102606(21)B3XZG8, (01)00884838102606(21)F0K1M5, (01)00884838102606(21)B3XZ0B, (01)00884838088573(21)B3LBGV, (01)00884838102606(21)F0JQTB, (01)00884838102606(21)F0JTFG, (01)00884838102606(21)F0JWZF, (01)00884838088573(21)B32V5F, (01)00884838088573(21)B3FMV5, (01)00884838102606(21)B3XZPM, (01)00884838102606(21)F0HDK0, (01)00884838088573(21)B3FMN3, (01)00884838088573(21)B3FMN7, (01)00884838088573(21)B3FMN8, (01)00884838088573(21)B3FMNX, (01)00884838102606(21)F0JY1Y, (01)00884838102606(21)F0JZJ9, (01)00884838088573(21)B32V9W, (01)00884838088573(21)B3L6KJ, (01)00884838102606(21)F08R4H, (01)00884838102606(21)F095FR, (01)00884838102606(21)F095G0, (01)00884838102606(21)F09D32, (01)00884838088573(21)B3LBFL, (01)00884838088573(21)B3FMCH, (01)00884838102606(21)F09KJN, (01)00884838102606(21)F0CQ59, 010088483810260621F0FQZJ, 010088483810260621B3Y0HV, (01)00884838102606(21)F07NXN, (01)00884838102606(21)F08W48, (01)00884838088573(21)B3FMXC, (01)00884838102606(21)F0CN2K, (01)00884838102606(21)F06N28, (01)00884838102606(21)F08HBG, (01)00884838102606(21)F07862, (01)00884838102606(21)B3XZNF, (01)00884838088573(21)B3FKYG, (01)00884838102606(21)B3XZNM, (01)00884838102606(21)F08K8K, (01)00884838102606(21)F09865, (01)00884838088573(21)B3FL9W, (01)00884838088573(21)B3FL5B, (01)00884838088573(21)B3FM9Y, (01)00884838088573(21)B3FMMR, (01)00884838102606(21)F0CN2H, (01)00884838102606(21)F07RXT, 010088483810260621F0FP2M, (01)00884838102606(21)F09BF2, (01)00884838102606(21)F05HXQ, (01)00884838102606(21)F0B9YD, (01)00884838088573(21)B32V4L, (01)00884838102606(21)F0986Q, (01)00884838088573(21)B3FMPN, (01)00884838102606(21)F095F6, (01)00884838102606(21)F0CN2C, (01)00884838088573(21)B32TBP, (01)00884838102606(21)F0G7NK, (01)00884838102606(21)B3XZ1F, (01)00884838102606(21)F05HY3, (01)00884838102606(21)F09FY3, (01)00884838102606(21)F0HFN2, (01)00884838102606(21)F0HG4B, (01)00884838102606(21)F077M0, (01)00884838088573(21)B3LBCL, (01)00884838102606(21)F09FYH, (01)00884838088573(21)B3FL83, (01)00884838102606(21)F08JKR, (01)00884838102606(21)B3XZW9, (01)00884838088573(21)B32V2V, (01)00884838102606(21)F06PR3, (01)00884838102606(21)F07860, 010088483810260621B3FLQD, (01)00884838088573(21)B3FKY4, 010088483810260621B3FM61, (01)00884838088573(21)B3FMT9, (01)00884838102606(21)F08MH7, (01)00884838102606(21)F08MJ6, (01)00884838102606(21)F0HX7L, (01)00884838088573(21)B3FMRN, (01)00884838102606(21)F09FXV, (01)00884838102606(21)F06PQJ, (01)00884838102606(21)F09LHD, (01)00884838088573(21)B32V9P, (01)00884838088573(21)B32V9V, (01)00884838088573(21)B32VD0, (01)00884838088573(21)B3FMRK, (01)00884838102606(21)F0986P, (01)00884838102606(21)F0986X, (01)00884838088573(21)B3FMJM, (01)00884838102606(21)F0HT0T, (01)00884838102606(21)F0HNBG, (01)00884838102606(21)F0J38H, (01)00884838102606(21)F0CHGB, (01)00884838088573(21)B32TPM, (01)00884838088573(21)B32TZ3, (01)00884838102606(21)B415T9, (01)00884838102606(21)B41GQ4, (01)00884838102606(21)F04MYH, (01)00884838102606(21)F04MYL, (01)00884838102606(21)F04P6X, (01)00884838102606(21)F04XCJ, (01)00884838102606(21)F079Y6, (01)00884838102606(21)F0KL6K, (01)00884838102606(21)F0BD0D, (01)00884838102606(21)B3FLH7, (01)00884838102606(21)B3FLJ2, (01)00884838102606(21)B3FLJJ, (01)00884838102606(21)B3FLK2, (01)00884838102606(21)B3FM12, (01)00884838102606(21)B3XYVM, (01)00884838102606(21)B411GJ, (01)00884838102606(21)F08F2P, (01)00884838102606(21)F095F4, (01)00884838102606(21)F0986M, (01)00884838102606(21)F0986N, (01)00884838102606(21)F09D3B, (01)00884838102606(21)B3XZLQ, (01)00884838102606(21)B3XZNT, (01)00884838102606(21)B3Y002, (01)00884838102606(21)F09867, (01)00884838102606(21)F0BP4C, (01)00884838088573(21)B3L6KT, (01)00884838088573(21)B3FMV6, (01)00884838102606(21)F0HFN9, (01)00884838102606(21)F0CLFB, (01)00884838102606(21)F0CN1K, (01)00884838102606(21)F0CQ4J, (01)00884838102606(21)F0CQ56, (01)00884838102606(21)B3XZGP, (01)00884838102606(21)B3XZTM, (01)00884838102606(21)F0B49K, (01)00884838088573(21)B3L49P, (01)00884838102606(21)F08R3Y, (01)00884838102606(21)F08R4M, (01)00884838102606(21)F08W4Y, (01)00884838088573(21)B32TBJ, (01)00884838102606(21)F096BD, (01)00884838102606(21)F09LGR, (01)00884838102606(21)F09LH9, (01)00884838102606(21)F09TK2, (01)00884838102606(21)F0CHH4, (01)00884838102606(21)B3FLQ3, (01)00884838102606(21)B3Y0FZ, (01)00884838088573(21)B2PYFV, (01)00884838088573(21)B32TQV, (01)00884838088573(21)B32TTC, (01)00884838102606(21)B3Y0KT, (01)00884838088573(21)B32V17, (01)00884838088573(21)B3L4Z3, (01)00884838102606(21)B3XZZB, (01)00884838102606(21)B3XZMW, (01)00884838102606(21)B3FLPY, (01)00884838102606(21)B3FM70, (01)00884838102606(21)F0DR5G, (01)00884838102606(21)F07RXF, (01)00884838102606(21)F07RXG, (01)00884838102606(21)F07858, (01)00884838088573(21)B32TR8, (01)00884838088573(21)B32V8Z, (01)00884838102606(21)F0BKHG, (01)00884838102606(21)F0CT05, (01)00884838102606(21)F0JCQR, (01)00884838088573(21)B32TW1, (01)00884838102606(21)F0CQ48, (01)00884838088573(21)B32TZ4, (01)00884838102606(21)B3FLNB, (01)00884838102606(21)F0BHVC, (01)00884838102606(21)F0JDLY, 010088483810260621F0FZVF, (01)00884838102606(21)B3Y048, (01)00884838088573(21)B32TNZ, (01)00884838088573(21)B3FL91, (01)00884838088573(21)B3FMTH, (01)00884838102606(21)F06TGQ, (01)00884838102606(21)F04MYG, (01)00884838102606(21)B3XZN6, (01)00884838102606(21)B411GR, (01)00884838088573(21)B32TVQ, (01)00884838088573(21)B3FMPG, (01)00884838102606(21)F0HT0X, (01)00884838102606(21)F0JQTQ, (01)00884838102606(21)F0JQTT, (01)00884838102606(21)F0JQTW, (01)00884838102606(21)F0JT5B, (01)00884838102606(21)F0JTFB, (01)00884838102606(21)F0JTYM, (01)00884838088573(21)B3FMCC, (01)00884838102606(21)F07D16, (01)00884838102606(21)F07K79, (01)00884838088573(21)B32THV, (01)00884838102606(21)F073GY, (01)00884838088573(21)B3FMCL, (01)00884838088573(21)B3FLRD, (01)00884838088573(21)B3FLV6, (01)00884838102606(21)B3FLH8, (01)00884838102606(21)B3FLL5, (01)00884838102606(21)F08HBM, (01)00884838102606(21)B3XZKY, (01)00884838088573(21)B32V41, (01)00884838102606(21)F0KL6X, (01)00884838102606(21)F0KK51, (01)00884838102606(21)F0HT21, (01)00884838102606(21)F0J38G, (01)00884838102606(21)F0J4R5, (01)00884838088573(21)B3FMXG, (01)00884838102606(21)F0HVY5, (01)00884838088573(21)B3L49J, (01)00884838088573(21)B32TCL, (01)00884838088573(21)B32TKF, 010088483810260621B3FM38, 010088483810260621B3Y0MK, (01)00884838102606(21)F083YV, 010088483810260621F0FQZ7, 010088483810260621F0FQZQ, (01)00884838088573(21)B3FMW3, (01)00884838088573(21)B32TV8, (01)00884838088573(21)B3FLTB, (01)00884838102606(21)B3FLJC, (01)00884838102606(21)B3FLJH, (01)00884838102606(21)B3FLLN, (01)00884838102606(21)B3FLM4, (01)00884838088573(21)B3FMVF, (01)00884838088573(21)B3FMQB, (01)00884838088573(21)B3LBCK, (01)00884838102606(21)F0HT1N, (01)00884838088573(21)B3LBGW, (01)00884838102606(21)B3XZNV, (01)00884838102606(21)B3FL8F, (01)00884838102606(21)F08002, (01)00884838088573(21)B32V8V, (01)00884838102606(21)F09Z2Q, (01)00884838102606(21)F0BP4B, (01)00884838088573(21)B3L4T8, (01)00884838088573(21)B32VBK, (01)00884838088573(21)B32VCY, (01)00884838088573(21)B3FLC9, (01)00884838102606(21)F08R4B, (01)00884838102606(21)F09LJ0, (01)00884838102606(21)B3FLJ8, (01)00884838102606(21)B3FLJK, (01)00884838102606(21)B3FLJL, (01)00884838102606(21)B3XYTX, (01)00884838102606(21)B3XYVJ, (01)00884838102606(21)B3XYY0, (01)00884838102606(21)B411GM, (01)00884838102606(21)F04XD1, (01)00884838102606(21)F06PQG, (01)00884838102606(21)F06TH2, (01)00884838102606(21)F0BHVQ, (01)00884838102606(21)F06TGR, (01)00884838102606(21)B3XZP6, (01)00884838102606(21)B3XZQ9, (01)00884838102606(21)B3Y001, (01)00884838102606(21)B3Y02M, (01)00884838088573(21)B3FKXM, (01)00884838088573(21)B3FMM7, (01)00884838088573(21)B3FL72, (01)00884838102606(21)B3Y064, (01)00884838102606(21)F07WT2, (01)00884838102606(21)F07WTR, (01)00884838102606(21)F07WRV, (01)00884838102606(21)F08JKX, (01)00884838088573(21)B3FMV1, (01)00884838102606(21)F08W49, (01)00884838088573(21)B3FMVN, (01)00884838102606(21)F077M1, (01)00884838102606(21)F08JL7, (01)00884838088573(21)B2PYF9, (01)00884838088573(21)B3LB9T, (01)00884838088573(21)B3LBB2, (01)00884838088573(21)B3LBBF, (01)00884838088573(21)B3LBC7, (01)00884838088573(21)B3FL88, (01)00884838102606(21)F07CD6, (01)00884838102606(21)F07D12, (01)00884838088573(21)B3L4T3, (01)00884838102606(21)B3XYV9, (01)00884838088573(21)B3FKXV, (01)00884838088573(21)B3FMWR, (01)00884838102606(21)F0DR5K, (01)00884838102606(21)F0DZ2W, (01)00884838102606(21)F0DZ35, (01)00884838102606(21)F0F02W, (01)00884838102606(21)F0F5TC, (01)00884838102606(21)F083YD, (01)00884838102606(21)F083YW, (01)00884838088573(21)B32VDB, (01)00884838088573(21)B3FMWJ, (01)00884838102606(21)F04P7F, (01)00884838088573(21)B3FL87, (01)00884838088573(21)B3FKZQ, (01)00884838102606(21)F06N32, (01)00884838102606(21)F0BD03, (01)00884838102606(21)F0BD07, (01)00884838102606(21)F05768, (01)00884838102606(21)F0CLFG, (01)00884838102606(21)F0B8BZ, (01)00884838102606(21)B3XZDG, (01)00884838102606(21)B3XZH0, (01)00884838102606(21)B3XZH5, (01)00884838102606(21)F06HYF, (01)00884838102606(21)F06HYM, (01)00884838102606(21)F091RQ, (01)00884838102606(21)B3FM02, (01)00884838102606(21)B3FM04, (01)00884838102606(21)B41GQF, (01)00884838102606(21)B3FL8T, (01)00884838102606(21)F0784Z, (01)00884838102606(21)F083YF, (01)00884838102606(21)F08BBF, (01)00884838102606(21)F08BBP, (01)00884838102606(21)F08BC4, (01)00884838102606(21)F0BP4F, (01)00884838102606(21)F0CHGQ, (01)00884838102606(21)B3FLHF, (01)00884838088573(21)B3L6KN, (01)00884838090804(21)B31X22, (01)00884838102606(21)B3Y02H, (01)00884838102606(21)B3Y043, (01)00884838102606(21)B3Y049, (01)00884838102606(21)F04MYN, (01)00884838102606(21)F04P7Y, (01)00884838102606(21)F0JT56, (01)00884838088573(21)B3FMR4, (01)00884838088573(21)B3FMRJ, (01)00884838088573(21)B3FMV2, (01)00884838102606(21)F0B49D, (01)00884838102606(21)F0BCZX, (01)00884838088573(21)B3LBH5, (01)00884838088573(21)B3LBH6, (01)00884838102606(21)F08JL0, 010088483810260621B3FM5Z, (01)00884838102606(21)F04XD9, (01)00884838088573(21)B3FLCY, (01)00884838102606(21)F07GL8, 010088483810260621B3Y0HX, (01)00884838102606(21)F0JB5W, (01)00884838102606(21)F06THH, (01)00884838102606(21)F0BD0V, (01)00884838088573(21)B3FMBF, (01)00884838088573(21)B3FMHW, (01)00884838102606(21)F083YH, (01)00884838102606(21)F0CQ55, (01)00884838102606(21)B3FLPV, (01)00884838102606(21)F09FYM, (01)00884838102606(21)F0CQ5F, (01)00884838102606(21)B3XZ1J, (01)00884838102606(21)B3XZLH, (01)00884838102606(21)F04V3Y, (01)00884838102606(21)F04V45, (01)00884838102606(21)F04XCQ, (01)00884838102606(21)F053C6, (01)00884838102606(21)F0FMXB, (01)00884838102606(21)B3Y02C, (01)00884838102606(21)B3Y02D, (01)00884838102606(21)F0JH1F, (01)00884838102606(21)F0BT8H, (01)00884838102606(21)B3Y0KX, (01)00884838102606(21)F09LP0, (01)00884838102606(21)F0B4BJ, (01)00884838102606(21)F073GX, (01)00884838102606(21)B3FLJ4, (01)00884838102606(21)B3FLJM, (01)00884838102606(21)B3FLJQ, (01)00884838102606(21)B415R8, (01)00884838102606(21)B41GQ6, (01)00884838102606(21)F09LP5, (01)00884838102606(21)B3FM0Z, (01)00884838102606(21)B3XZKM, (01)00884838102606(21)B3XZMG, (01)00884838102606(21)B3XZP9, (01)00884838102606(21)B3XZPD, (01)00884838102606(21)F04V40, (01)00884838102606(21)F07NXM, (01)00884838102606(21)F0B3FB, (01)00884838102606(21)F0J4HW, (01)00884838102606(21)B3XYZP, (01)00884838102606(21)B3XZ0M, (01)00884838102606(21)B3XZDX, (01)00884838102606(21)F06HYT, (01)00884838102606(21)B3FM6B, (01)00884838102606(21)F09D31, (01)00884838102606(21)F0JQT9, (01)00884838102606(21)B3FLF2, (01)00884838102606(21)B3XZFC, (01)00884838102606(21)F04XD7, (01)00884838102606(21)F04XD8, (01)00884838088573(21)B3LB8M, (01)00884838088573(21)B3FL4T, (01)00884838088573(21)B3FKZT, (01)00884838088573(21)B32TCH, (01)00884838102606(21)B411GF, (01)00884838102606(21)F06PQF, (01)00884838102606(21)F079Y0, (01)00884838102606(21)F0KK56, (01)00884838088573(21)B3FL7W, (01)00884838102606(21)F0GV2B, (01)00884838102606(21)B3FLZH, (01)00884838102606(21)F05BT4, (01)00884838102606(21)F07K7F, (01)00884838088573(21)B3LB8P, (01)00884838088573(21)B3LB8R, (01)00884838102606(21)F06FBC, (01)00884838102606(21)F06KTV, (01)00884838102606(21)F077LH, (01)00884838102606(21)F077LW, (01)00884838102606(21)F07851, (01)00884838102606(21)F0785J, (01)00884838102606(21)F04P7M, (01)00884838088573(21)B32TVL, (01)00884838088573(21)B3FMRT, (01)00884838088573(21)B3L498, (01)00884838088573(21)B3L6KF, (01)00884838088573(21)B3LBCW, (01)00884838088573(21)B3LBD1, (01)00884838088573(21)B3LBH8, 010088483810260621F0FP2Q, (01)00884838102606(21)F04P72, (01)00884838088573(21)B3LB8Q, (01)00884838102606(21)F0LQH0, (01)00884838102606(21)B3FM1C, (01)00884838102606(21)F05BRJ, (01)00884838102606(21)F077LC, (01)00884838088573(21)B32V07, (01)00884838102606(21)F08K8F, (01)00884838088573(21)B32TWY, (01)00884838102606(21)F0CQ44, (01)00884838088573(21)B3FL8H, (01)00884838088573(21)B3L495, (01)00884838088573(21)B3L49Q, (01)00884838102606(21)B3XZP3, (01)00884838102606(21)F07CD1, (01)00884838088573(21)B32V69, (01)00884838102606(21)B415TB, (01)00884838102606(21)F0HG59, (01)00884838088573(21)B32V82, (01)00884838088573(21)B3FM9B, (01)00884838088573(21)B32V76, (01)00884838088573(21)B32V9Z, (01)00884838088573(21)B32TTP, (01)00884838088573(21)B3FMNM, (01)00884838088573(21)B32V4D, (01)00884838102606(21)F0HG4J, (01)00884838088573(21)B3FLVQ, (01)00884838088573(21)B3FMP2, (01)00884838088573(21)B3FMN2, (01)00884838102606(21)B3XZDK, (01)00884838088573(21)B3FL78, (01)00884838088573(21)B3FMD0, 010088483810260621B3Y0L2, (01)00884838088573(21)B32V0G, (01)00884838102606(21)B3FLQF, (01)00884838102606(21)B3XYYB, (01)00884838102606(21)F09C05, (01)00884838088573(21)B3FMQJ, (01)00884838102606(21)B3XZQD, (01)00884838102606(21)F0BD0P, (01)00884838088573(21)B3FMM5, (01)00884838102606(21)F0G350, (01)00884838088573(21)B3FLGV, 010088483810260621F0FQZ5, (01)00884838102606(21)B3FLJY, (01)00884838102606(21)B3FLK0, (01)00884838102606(21)F0JQT8, (01)00884838102606(21)B3FLJ5, (01)00884838102606(21)B3FLJF, (01)00884838102606(21)F07RXM, (01)00884838088573(21)B3LBGB, (01)00884838102606(21)B3FLL2, (01)00884838102606(21)F07D0V, (01)00884838102606(21)F07D0Y, (01)00884838102606(21)F08JL5, (01)00884838102606(21)B3FM06, (01)00884838102606(21)B3FM08, 010088483810260621B3FM6P, 010088483810260621B3FM7W, 010088483810260621B3Y0J8, 010088483810260621B3Y0KM, 010088483810260621B3Y0M7, 010088483810260621B3Y0MF, (01)00884838088573(21)B3FMV8, (01)00884838102606(21)B3XZRC, (01)00884838102606(21)B3XZTV, (01)00884838102606(21)F0JZJG, (01)00884838102606(21)F0CLF8, (01)00884838102606(21)F0CLFW, (01)00884838102606(21)F0CN26, (01)00884838102606(21)F0CN27, (01)00884838102606(21)F07WTM, (01)00884838102606(21)F0HFN7, (01)00884838102606(21)F0JQTJ, (01)00884838088573(21)B2PYF7, (01)00884838102606(21)F06KTQ, (01)00884838088573(21)B3FMB3, (01)00884838102606(21)B3FLJD, (01)00884838088573(21)B3FM90, (01)00884838102606(21)B3XZRG, (01)00884838102606(21)F07853, (01)00884838102606(21)F0DXHD, (01)00884838102606(21)F09LPC, (01)00884838102606(21)F0B8C3, (01)00884838102606(21)F0CHFV, (01)00884838102606(21)B3XZGX, (01)00884838088573(21)B32V85, (01)00884838088573(21)B32TPP, (01)00884838102606(21)F06TGP, (01)00884838088573(21)B3FMDT, (01)00884838088573(21)B3FKXN, (01)00884838102606(21)F07856, (01)00884838102606(21)F079Y2, (01)00884838102606(21)F0DR4V, (01)00884838102606(21)F06HYQ, (01)00884838102606(21)F0785G, (01)00884838102606(21)F0CHG6, (01)00884838102606(21)B3Y0FN, (01)00884838102606(21)F091RN, (01)00884838102606(21)F096BM, (01)00884838102606(21)F09869, (01)00884838102606(21)B3FLLM, (01)00884838102606(21)B3XZ10, (01)00884838102606(21)F06TH9, (01)00884838102606(21)F0DJL4, (01)00884838102606(21)F08XZR, (01)00884838102606(21)B3Y0G7, (01)00884838102606(21)F077MK, (01)00884838102606(21)F04XDH, (01)00884838102606(21)F04XDL, (01)00884838088573(21)B32V8M, (01)00884838088573(21)B3FKXT, (01)00884838102606(21)B41GQC, (01)00884838102606(21)F095FF, (01)00884838102606(21)F095FN, (01)00884838102606(21)F091RH, (01)00884838102606(21)B3XZ08, (01)00884838088573(21)B2PYGX, (01)00884838088573(21)B3FMRH, (01)00884838088573(21)B3FMRW, (01)00884838088573(21)B32TR0, (01)00884838088573(21)B2ZR2F, (01)00884838102606(21)F0K7MJ, (01)00884838088573(21)B3FLB9, (01)00884838102606(21)B3Y0FT, (01)00884838102606(21)F0BHVV, (01)00884838088573(21)B3FL57, (01)00884838102606(21)F0BD0G, (01)00884838102606(21)F0F4CV, (01)00884838088573(21)B3FKY2, (01)00884838088573(21)B3FL00, (01)00884838088573(21)B3FMPJ, (01)00884838102606(21)F0HG4R, (01)00884838102606(21)F0HX7R, (01)00884838102606(21)F09KJV, (01)00884838102606(21)F053BV, (01)00884838102606(21)F05BTF, (01)00884838088573(21)B3FL1P, (01)00884838102606(21)B3FLGY, (01)00884838102606(21)B3FLJB, (01)00884838088573(21)B32V63, (01)00884838088573(21)B32V6J, (01)00884838102606(21)F07RX9, (01)00884838102606(21)B3XZTZ, (01)00884838088573(21)B3L6L0, (01)00884838102606(21)F08K7V, (01)00884838088573(21)B3FMKG, (01)00884838088573(21)B3FMNC, (01)00884838088573(21)B3L491, (01)00884838102606(21)B3FLGZ, (01)00884838102606(21)B3FLK3, (01)00884838102606(21)B3XYTD, (01)00884838102606(21)B3FLF5, (01)00884838088573(21)B32TNK, (01)00884838102606(21)B3Y060, (01)00884838102606(21)B3Y068, (01)00884838102606(21)F0B4B0, (01)00884838088573(21)B32TNH, (01)00884838102606(21)F079XT, (01)00884838102606(21)F077LP, (01)00884838088573(21)B3FL1J, (01)00884838088573(21)B32V1X, (01)00884838088573(21)B3FKYD, (01)00884838088573(21)B3FMFP, (01)00884838088573(21)B3LBD4, (01)00884838102606(21)B3XYZJ, (01)00884838088573(21)B3FLVF, (01)00884838102606(21)F0DWDY, (01)00884838102606(21)B3XZN4, (01)00884838102606(21)B3XZWB, (01)00884838088573(21)B32TRL, (01)00884838102606(21)B415R3, (01)00884838102606(21)F08BBZ, (01)00884838102606(21)F08HC7, (01)00884838102606(21)F0CQ58, (01)00884838088573(21)B32TD3, (01)00884838088573(21)B2PYF6, (01)00884838088573(21)B3FMR1, (01)00884838088573(21)B3FMRP, (01)00884838102606(21)F08HBW, (01)00884838102606(21)F0JL45, (01)00884838102606(21)F0BKHJ, (01)00884838102606(21)B3XZF5, (01)00884838102606(21)B3XZ12, (01)00884838102606(21)B3XZGV, (01)00884838102606(21)F0J9JV, (01)00884838088573(21)B32VC3, (01)00884838102606(21)B3XYTK, (01)00884838102606(21)F0HFMH, (01)00884838102606(21)B415T4, (01)00884838102606(21)F0BD0N, (01)00884838102606(21)F0J9JT, (01)00884838102606(21)F0CHFY, (01)00884838088573(21)B3FM9C, (01)00884838102606(21)F09Z28, (01)00884838088573(21)B32TPL, (01)00884838102606(21)B3FM09, (01)00884838102606(21)B415T5, (01)00884838088573(21)B32V4C, (01)00884838102606(21)F0BCZT, (01)00884838088573(21)B3L51B, 010088483810260621F0FQZ9, (01)00884838102606(21)F0J9JN, (01)00884838088573(21)B3L4ZL, (01)00884838102606(21)F053BX, (01)00884838088573(21)B3FMT1, (01)00884838088573(21)B32VCT, (01)00884838088573(21)B3LB9Y, (01)00884838102606(21)B3XYZM, (01)00884838102606(21)B3XZ0W, (01)00884838102606(21)F0BCZL, (01)00884838102606(21)F0B8CW, (01)00884838102606(21)B3XZN2, (01)00884838088573(21)B32TRW, (01)00884838102606(21)B3FLJN, (01)00884838102606(21)B3FLL1, (01)00884838102606(21)F09TK4, (01)00884838102606(21)F0BPYH, (01)00884838088573(21)B3FLCJ, (01)00884838088573(21)B2ZR3Y, (01)00884838088573(21)B32TG8, (01)00884838088573(21)B32TT4, (01)00884838088573(21)B3L505, (01)00884838088573(21)B32TFZ, (01)00884838102606(21)B3Y0H8, (01)00884838088573(21)B32THR, (01)00884838088573(21)B32TDV, (01)00884838088573(21)B2ZR2P, (01)00884838088573(21)B32V45, (01)00884838088573(21)B32V59, (01)00884838088573(21)B32V61, (01)00884838088573(21)B32V74, (01)00884838102606(21)B3XZGG, (01)00884838102606(21)F06KRR, (01)00884838102606(21)F06TGX, (01)00884838070813(21)B34T4P, (01)00884838088573(21)B2ZR37, (01)00884838102606(21)F0G7NT, 010088483810260621B3FM64, (01)00884838102606(21)B3XYXQ, (01)00884838102606(21)F06KTK, (01)00884838088573(21)B32TM7, (01)00884838102606(21)B411G2, (01)00884838102606(21)F04P7R, (01)00884838088573(21)B32TQY, (01)00884838088573(21)B32TRH, (01)00884838088573(21)B32TR6, (01)00884838088573(21)B32TYF, (01)00884838088573(21)B32TZY, (01)00884838088573(21)B32V0L, (01)00884838088573(21)B32V0P, (01)00884838088573(21)B32V1D, (01)00884838088573(21)B32TT5, (01)00884838088573(21)B32TTM, (01)00884838088573(21)B32TTW, (01)00884838088573(21)B32TV5, (01)00884838088573(21)B32TVW, (01)00884838088573(21)B3FL98, (01)00884838088573(21)B3FL9D, (01)00884838102606(21)F0B8CT, (01)00884838102606(21)F07K7D, (01)00884838102606(21)B3FLZL, (01)00884838088573(21)B32V27, (01)00884838088573(21)B32V2J, (01)00884838102606(21)F0BKHB, (01)00884838102606(21)F06PQW, (01)00884838088573(21)B32V5D, (01)00884838102606(21)F0BT8M, (01)00884838088573(21)B32V4K, (01)00884838102606(21)B3FLJ1, (01)00884838102606(21)B3FLJ3, (01)00884838102606(21)B3FLJR, (01)00884838102606(21)B3FLJZ, (01)00884838102606(21)B3XYVH, (01)00884838102606(21)F095FX, (01)00884838102606(21)F0986B, (01)00884838102606(21)F09C01, (01)00884838102606(21)F09R4M, (01)00884838088573(21)B32V31, (01)00884838102606(21)F0D435, (01)00884838088573(21)B32TNY, (01)00884838088573(21)B32V6M, (01)00884838088573(21)B32V6T, (01)00884838088573(21)B3FMHR, (01)00884838088573(21)B32VC2, (01)00884838088573(21)B32V83, (01)00884838088573(21)B32VCB, (01)00884838088573(21)B32V7Y, (01)00884838102606(21)F0BH0C, (01)00884838102606(21)F0JCQZ, (01)00884838088573(21)B3FMM8, (01)00884838088573(21)B3FMVM, (01)00884838088573(21)B3FMWN, (01)00884838102606(21)F0JT59, (01)00884838102606(21)F079Y8, (01)00884838102606(21)F0DP3J, (01)00884838088573(21)B32V8X, (01)00884838088573(21)B32VBW, (01)00884838102606(21)F09R4N, (01)00884838102606(21)F0B13K, (01)00884838102606(21)B3FLZX, (01)00884838088573(21)B32V8W, (01)00884838088573(21)B32V91, (01)00884838102606(21)B3FLZN, (01)00884838088573(21)B32V8D, (01)00884838088573(21)B32TNT, (01)00884838088573(21)B32VCV, (01)00884838088573(21)B3FMWW, (01)00884838088573(21)B32VD9, (01)00884838102606(21)F0BH0D, (01)00884838088573(21)B3FMXD, (01)00884838102606(21)F09KJL, (01)00884838102606(21)F0JH16, (01)00884838088573(21)B3FMP8, (01)00884838102606(21)F06KV4, (01)00884838088573(21)B32QBW, (01)00884838102606(21)F096C8, (01)00884838088573(21)B3FMT2, (01)00884838102606(21)F0J389, (01)00884838088573(21)B3FMV3, (01)00884838088573(21)B32V90, (01)00884838088573(21)B3FMK3, (01)00884838102606(21)F0HDKG, (01)00884838088573(21)B3FMRX, (01)00884838102606(21)F09KJJ, (01)00884838088573(21)B3FMND, (01)00884838102606(21)B3Y06H, (01)00884838102606(21)B3Y0JD, (01)00884838088573(21)B3FMJW, (01)00884838088573(21)B3L4TK, (01)00884838102606(21)F053BH, (01)00884838088573(21)B3L4T2, (01)00884838088573(21)B3FMFQ, (01)00884838088573(21)B3LBGZ, (01)00884838102606(21)F095FJ, (01)00884838088573(21)B3L4RW, (01)00884838102606(21)F0BD0B, (01)00884838088573(21)B3L6LF, (01)00884838102606(21)F0CLFH, (01)00884838102606(21)F09FYG, (01)00884838088573(21)B3L49C, (01)00884838088573(21)B3L4T5, (01)00884838102606(21)B3XZLY, (01)00884838102606(21)B3XZN9, (01)00884838102606(21)B3XZP1, (01)00884838102606(21)F0FP1L, 010088483810260621F0FVHD, (01)00884838102606(21)B3XZDT, (01)00884838102606(21)B3XZLJ, (01)00884838088573(21)B3FL1G, (01)00884838102606(21)F0CQ5B, (01)00884838088573(21)B3FL0C, (01)00884838088573(21)B3FL0D, (01)00884838088573(21)B3FL1H, (01)00884838088573(21)B3FMNG, (01)00884838102606(21)F08W4K, (01)00884838102606(21)F09D3D, (01)00884838102606(21)F0HX7T, (01)00884838088573(21)B3FLB7, (01)00884838088573(21)B3FL9Q, (01)00884838088573(21)B3FLB1, (01)00884838088573(21)B3FLGT, (01)00884838102606(21)F0F02V, (01)00884838088573(21)B3FL2H, (01)00884838102606(21)B3FLY4, (01)00884838102606(21)B3FM1V, (01)00884838102606(21)B3FM20, (01)00884838102606(21)B3FM3P, (01)00884838088573(21)B3FLB6, (01)00884838088573(21)B3FLGC, (01)00884838102606(21)F0986Z, (01)00884838102606(21)B3Y0HP, (01)00884838088573(21)B32VC0, (01)00884838088573(21)B32VCF, (01)00884838088573(21)B3FMX8, (01)00884838102606(21)B3Y04M, (01)00884838102606(21)B3FLLD, (01)00884838102606(21)B3FLL3, (01)00884838102606(21)F05DQ1, (01)00884838102606(21)F05BRZ, (01)00884838102606(21)F09L7D, (01)00884838102606(21)B3FM07, (01)00884838102606(21)B3FLZP, (01)00884838102606(21)F04XCN, (01)00884838102606(21)B3XZV6, (01)00884838102606(21)B3Y018, (01)00884838088573(21)B32TKD, (01)00884838102606(21)B3Y06D, (01)00884838102606(21)B3Y06T, (01)00884838102606(21)B3Y0G1, (01)00884838102606(21)B3Y0GY, (01)00884838102606(21)B3Y0HF, (01)00884838102606(21)B3Y0HM, (01)00884838102606(21)B3Y0JX, (01)00884838102606(21)F08F2M, (01)00884838102606(21)F06N36, (01)00884838102606(21)B3FM1D, (01)00884838102606(21)B3FLJ0, (01)00884838102606(21)B3FLK4, (01)00884838102606(21)B3XZTQ, 010088483810260621B3Y0L0, (01)00884838102606(21)B3XZMD, (01)00884838102606(21)B3FLJV, (01)00884838102606(21)B3Y0H2, (01)00884838102606(21)F096C5, (01)00884838088573(21)B3FL26, (01)00884838102606(21)F077LL, (01)00884838102606(21)B3XZN5, (01)00884838102606(21)B41GQ8, (01)00884838102606(21)B411GH, (01)00884838102606(21)B3XZM8, (01)00884838102606(21)B3Y01B, (01)00884838102606(21)B3Y01H, (01)00884838102606(21)B3XZNB, (01)00884838102606(21)F09L7J, (01)00884838102606(21)B3XZTX, (01)00884838102606(21)F09D2Z, (01)00884838102606(21)F0BH07, (01)00884838102606(21)F0BT7G, (01)00884838102606(21)F0CT09, (01)00884838102606(21)F0CT0J, (01)00884838102606(21)B3XZF1, (01)00884838102606(21)F06TGT, (01)00884838102606(21)F06THV, (01)00884838102606(21)F06FB3, (01)00884838088573(21)B3FMK0, (01)00884838088573(21)B3FMML, (01)00884838102606(21)B3XZFD, (01)00884838102606(21)B41GPX, (01)00884838102606(21)F04P74, (01)00884838102606(21)F04V47, (01)00884838102606(21)B3XYXR, (01)00884838102606(21)F0986J, (01)00884838102606(21)F09LHP, (01)00884838102606(21)F04P7X, (01)00884838102606(21)F077MF, (01)00884838102606(21)F09D3C, (01)00884838102606(21)F0CQ4H, (01)00884838102606(21)F04XD0, (01)00884838102606(21)F09FXY, (01)00884838102606(21)F09Z2H, (01)00884838102606(21)F0B13R, (01)00884838102606(21)F0B49Z, (01)00884838102606(21)F077M3, (01)00884838102606(21)F07CCN, (01)00884838102606(21)F077LY, (01)00884838102606(21)B3XZGC, (01)00884838102606(21)F0FMWK, (01)00884838102606(21)F073GJ, (01)00884838102606(21)F0GV2F, (01)00884838102606(21)F0BT8J, (01)00884838102606(21)F07WT8, (01)00884838102606(21)F077L8, (01)00884838102606(21)F0JL43, (01)00884838102606(21)F07NX3, (01)00884838102606(21)F07WT6, (01)00884838102606(21)F079X1, (01)00884838088573(21)B32V77, (01)00884838102606(21)F073GM, (01)00884838102606(21)F07NWY, (01)00884838102606(21)F07WTP, (01)00884838102606(21)F07NXJ, (01)00884838102606(21)F08W4Q, (01)00884838102606(21)F07NWZ, (01)00884838102606(21)F08K8H, (01)00884838088573(21)B3FKZ0, (01)00884838102606(21)F0BKGY, (01)00884838102606(21)F053BL, (01)00884838102606(21)F08W4W, (01)00884838102606(21)F095FV, (01)00884838102606(21)F077LR, (01)00884838088573(21)B32TWB, (01)00884838102606(21)F06TH6, (01)00884838102606(21)F09868, (01)00884838102606(21)F0JT53, (01)00884838102606(21)F09L7P, (01)00884838102606(21)F0CQ4V, (01)00884838102606(21)F0HXHB, (01)00884838102606(21)F0HXHV, (01)00884838102606(21)F0HDKP, (01)00884838102606(21)F0JT50, (01)00884838102606(21)F0B273, (01)00884838102606(21)F0BKH1, (01)00884838102606(21)F0BKHM, (01)00884838088573(21)B3FM8V, (01)00884838102606(21)F09KJP, (01)00884838102606(21)F0B3FG, (01)00884838102606(21)F0CN24, (01)00884838102606(21)F09FYF, (01)00884838102606(21)F0JWZJ, (01)00884838102606(21)F09D2T, 010088483810260621F0FP26, 010088483810260621F0FQZ1, (01)00884838102606(21)F0BPZ2, (01)00884838088573(21)B32TTR, (01)00884838102606(21)F0B13J, (01)00884838102606(21)F0B3FR, (01)00884838102606(21)F0BCZG, (01)00884838102606(21)F0BH0B, (01)00884838102606(21)F0J4HV, (01)00884838102606(21)F0BT7J, (01)00884838102606(21)F0JT4Y, (01)00884838102606(21)F0CHG7, (01)00884838102606(21)F08BC1, (01)00884838102606(21)F0CQ43, (01)00884838088573(21)B3FL02, (01)00884838102606(21)F0BCZY, (01)00884838102606(21)F0BPYV, (01)00884838102606(21)F0B8C5, (01)00884838102606(21)F0CQ4D, (01)00884838102606(21)F0CHGX, (01)00884838102606(21)F0CN1D, (01)00884838102606(21)F0CQ4F, (01)00884838102606(21)F0CQ4Q, (01)00884838102606(21)F0CQ5M, (01)00884838102606(21)F09BF4, (01)00884838102606(21)F0CT0G, (01)00884838102606(21)F0H4D0, (01)00884838102606(21)F0CQ4R, (01)00884838102606(21)F0F5RL, (01)00884838102606(21)F08BBL, (01)00884838102606(21)F08JL9, (01)00884838102606(21)F0F030, 010088483810260621F0FP1W, (01)00884838102606(21)F0B49B, (01)00884838102606(21)F0F5R9, (01)00884838102606(21)F0F5RC, (01)00884838102606(21)F0F45J, (01)00884838102606(21)F0B9Y8, (01)00884838102606(21)F0B8CX, (01)00884838102606(21)F09KJH, (01)00884838102606(21)F0H8R2, (01)00884838102606(21)B3FLHC, (01)00884838102606(21)F0CQ4B, (01)00884838102606(21)F0986D, (01)00884838102606(21)F0HG58, (01)00884838102606(21)F0HG54, (01)00884838102606(21)F0GPHJ, (01)00884838102606(21)F0H6QM, (01)00884838102606(21)F0H6QJ, (01)00884838102606(21)F0HFMZ, (01)00884838102606(21)F0J38B, (01)00884838102606(21)F0J4R7, (01)00884838102606(21)F0B8D1, (01)00884838102606(21)F0JDLM, (01)00884838102606(21)F0JL3K, (01)00884838102606(21)F0JL3Q, (01)00884838102606(21)F0JCQV, (01)00884838102606(21)F0J38F, (01)00884838102606(21)F0J4R4, (01)00884838102606(21)F0J9JK, (01)00884838102606(21)F0JG5Q, (01)00884838102606(21)F08MHG, (01)00884838102606(21)F095F9, (01)00884838102606(21)F0CHHB, (01)00884838102606(21)F0CN29, (01)00884838102606(21)F095FH, (01)00884838102606(21)F08W4C, (01)00884838102606(21)F0CHGL, (01)00884838102606(21)F0BH02, (01)00884838102606(21)F08R42, (01)00884838102606(21)F08XZ2, (01)00884838102606(21)F0JTFH, (01)00884838102606(21)B3FLQC, (01)00884838102606(21)F0HN33, (01)00884838102606(21)F07864, (01)00884838102606(21)F0JTF9, (01)00884838102606(21)F0JQTC, (01)00884838102606(21)F0JT52, (01)00884838088573(21)B3FLB0, (01)00884838102606(21)F0KG9R, (01)00884838102606(21)F0JWZR, (01)00884838088573(21)B32TWX, (01)00884838088573(21)B3FLRG, (01)00884838102606(21)B3FLZT, (01)00884838102606(21)B3XZ0F, (01)00884838102606(21)F0HG5L, (01)00884838102606(21)F0KK57, (01)00884838102606(21)F0KL6L, (01)00884838088573(21)B3FKYC, (01)00884838102606(21)B3XYZC, (01)00884838102606(21)F08F2C, (01)00884838102606(21)F09TK5, (01)00884838102606(21)F0CHG0; Serial No. B3FMNT, B3FLM3, B411GD, F053BY, F0HT0R, B3Y0MP, F0KG9X, B3FLWW, F0HT11, F04XCK, F0FQZG, F0FVH7, F06PQM, F06TGM, B32TQN, F0HJWD, F0986F, F07D1M, B32TQZ, B32TWD, B32V4Y, F09LH5, F0JZJM, F0576P, F0CN2B, F0F02Q, F0JL42, B32QCN, B3FLCN, B3FMXB, B2PYFR, B2ZR1X, B32V5Y, B3FMPD, B3FMPR, B3FMPV, B3FMQ5, B3FMR6, F0BHVR, F0JWZG, F0HJWH, B3FMQ0, F04XD2, B3XYZG, F0CHGT, B32TWR, B32V6B, F073H5, B3FMFT, F0CN1C, F0DP3N, F0HX7Q, F0J128, F073H3, F09BF3, F0B49X, F0CN22, F0B8CV, F08HC0, F0BCZZ, F0BD0Q, B3FMPW, F06THR, B3FLFH, B3Y06W, F077M2, B32V6Y, B32V7T, B32V9B, F04XCZ, B3FKXX, B3LBCN, B3FMT7, B3FM94, B3FMCT, F0JTYL, B3XZV4, F09Z2F, F077LB, F04V3T, B3Y0FQ, F0F5TN, F0FZ5Q, B3FM8Y, B3FL58, F0BPYX, F0785K, F08K80, F08MHQ, B32VBM, B3FMX2, F08JLG, F08XZ1, F096BL, B3Y0KV, B3FL1R, B3FL59, B3FL7C, B3FLC8, B3FMW5, B3FM16, F077M8, B3Y06F, B3Y06P, B3Y06R, B3Y0N7, F08W4H, F08XZL, F0BD08, F0CQ5C, F0CV7B, F0DWF3, F04P7C, F09FYJ, B3Y0L1, B3XZH3, F05HY6, F0BGZZ, F0BD00, B32TP0, B3FL7J, B3Y0KY, F06PR2, F0G352, B3FMWT, F0J12C, F0D436, B3XZZJ, B3XZM0, B3FMNR, F0B274, F05BT9, F05HY9, B3Y0JC, B3FMX0, B32VBT, B3Y04W, F04XCT, F07CCM, F0CHGW, B3FM1B, F0BKH3, F0JTFF, F0JWZK, F0JY22, B3Y0N5, F0FZV3, F0BCZQ, F0BHVY, F0CN1F, F0CQ52, F0BT8G, B3FM03, B3FM0L, B3FM3D, B3FL03, B3XZZM, F08K7M, F09QR2, B3LBCQ, B3L4T9, F0JTYK, F0D432, B3FL28, F08W46, B32TFC, B3FL0G, F07WT3, F0F1LH, F0FMX1, F0F4CJ, F0JDM2, B32TTZ, F0CQ4K, F0JH10, F0HT0L, B3Y0M8, F08MHC, F053C1, F0CHGZ, F08F29, F08R3Q, F08R43, F08W4M, F0JH0V, F0JH0Z, F0JQTR, F0JWZN, F05LXL, F06FBB, F06THW, B3FL05, F0FQZT, B32V1Y, F0K1MF, B2PYFC, B3FL9H, B32V8Y, F09TK9, B3FM00, B3FM0M, F04P80, F05LY2, F06BP6, B32V1W, B3XZ05, B3Y05Z, B3Y0G6, B3Y0KQ, B3FLN5, B3FMNK, B3FMNN, F07WRY, F0B9Y9, B3XZQ8, B3Y06X, F0G34Z, F09C06, B32TR1, B32TRF, B3FMTQ, B3FMMM, B3FMV7, B3FMVC, B3FMVH, B3FMVJ, B3FMVK, B3FMW8, B3FMW9, B3FMX1, B3XZNX, F04MZ5, F0HXHH, F0HXHR, F0576B, B3FLKX, F0HNBJ, F0JH14, F0HXHD, B415T7, F0BCZV, F0BHVK, F0B27J, B3XZ13, B32QBH, B3Y06J, F0FZV8, B3FLG8, B3Y0JQ, B32V32, F07857, B3XYZ9, B3FLWR, B3LBCP, B3LBDM, F053C4, F05BT7, B3FMP9, B3FMPH, B3XZ1H, B32V6X, B3FLND, B3XZWD, F0DP3M, B3Y0N1, B3FM8L, F0CQ5K, B3Y02L, B32TZL, B32VD7, F06FB6, B2PYHH, B3XYYF, B3Y0J4, F05HY0, B32TYK, B32TZZ, B3FMLQ, B3FMM1, F08BBY, F09C07, F09LHQ, B3FM2W, F0HT2B, B3XYTV, B3XZ17, F0CN2F, B3FMPZ, B3L519, F0JQTF, B3FM19, B3XZ09, B3FMK4.B3FMN9, B32TMB, B32V8R, F0G7NP, F08MHD, B3FLPD, B3Y062, F08XZF, B3L4ZY, B3L511, F06FBH, F06FBJ, F06KTR, F06N29, F06KTT, B3XZYZ, B3Y0M4, B32V2M, B3LB9X, B3Y06K, F0JH0T, B3XZTL, B32V81, F0HNBF, F073H1, F05HY7, B3FMB1, F09L7Q, F09LHB, F04XCH, B3FM3L, B3Y0HH, F08HC9, F08JLD, F0HNB3, F0J5Z6, B3Y04D, B3XZTD, B3FLL8, F07855, F07CCW, B3FM47, B3Y0FY, F0CN2D, B3FM4M, B3FM5N, B3FM63, B3FM82, B3FM21, B3FL01, B3FL1Y, F0B9YK, F0BP4K, F0BT7N, F0CQ4W, B3FM5Y, B3FMT4, B3FM7X, F05HYK, B3XYT2, B3XYT8, B3XYZH, B3XZ0Y, F08XZ4, B3FL84, F09L7X, B2ZR2T, B3FL1Q, B415RB, B3FL0K, B411G8, B41GPJ, B2DHZK, B2DHZN, B32TC9, B32V65, B32V7V, B3FMMC, B3XZN1, F07866, F07CCR, F07K7N, F07NXK, F07NXL, F079WX, F07CD7, F07WTG, B3XZ11, F09866, B3FL04, B3FL1T, B3XYTW, F0F5RG, B32V16, F0F45C, F09LH8, F0JG5M, F0CN1Y, F0BCZH, B3FM14, B3Y03P, B3Y05X, B3Y0FP, B3Y0FV, B3FLW5, F08F28, F08HBY, F09C04, F09D2V, F09LGQ, F0B8CY, F0BD06, F0BD0F, F0BH04, F0BKH5, F0D434, F0HG5J, F0HT24, F0J4HT, F0J5Z3, F0785L, F09LP7, B3FLCW, B3FM92, B3XZKX, B3XZL3, F0G7NN, B3Y01K, B3Y02G, F08W4X, F0HFMN, B3FMWD, B3FMX7, B3Y0H4, B3Y0HZ, B3Y0J6, B3XZQ6, F0BD0K, F0BKH7, B3XZQ3, F0HG48, F0HG56, B32V38, B32V0Y, F07CD4, B32V4H, F08W55, F05HYH, F05HYW, F0B9Y3, F0BKGP, F06N31, F06PR6, F0HT23, F0J5Z2, F077LZ, B3XZW5, F09BF6, F09LH3, B3FM0Q, B3XZWN, B3XZWP, F08XZQ, F09FXZ, F05LXZ, F06N2Y, F0FP2F, F0FQZR, F0GCLL, F0986Y, B3FLDW, F0BD0C, F0HXHY, F0J123, F053C3, F0HVXX, F0JCQX, F0FZV9, F0KL6G, F0JZJB, B3Y0N4, B3FLZZ, B3FM0J, F04XD5, B3XYTQ, B3XZF4, F09FY7, F0HNBD, F0GCLK, B32TPF, B32TQW, B32THK, B32TP4, B32TWP, B32TX3, B32TX8, B32TXZ, F09LGV, F04MYZ, F04V4D, F0J38C, F0J9JM, F0JT57, F06KTP, F0B3FN, B32VDD, F077MB, B32VBH, B3Y066, B3FLC5, B3FMG2, B32VD8, F0CQ5P, F0986K, F0986L, F09FXQ, F06FB0, F0HNBC, F0HXJ1, F096C4, B3FKY7, B3FMG3, F06HZK, F09FY6, F0DP3H, B3FMRL, B32TGR, F06KTB, F08K7T, B32V98, F07WTK, F0HG5F, B3FLJG, B32TPH, F08K8G, B3L4Z2, F09D34, B32QCM, B3FL9X, B3FMWF, F0JCQY, B3XZG8, F0K1M5, B3XZ0B, B3LBGV, F0JQTB, F0JTFG, F0JWZF, B32V5F, B3FMV5, B3XZPM, F0HDK0, B3FMN3, B3FMN7, B3FMN8, B3FMNX, F0JY1Y, F0JZJ9, B32V9W, B3L6KJ, F08R4H, F095FR, F095G0, F09D32, B3LBFL, B3FMCH, F09KJN, F0CQ59, F0FQZJ, B3Y0HV, F07NXN, F08W48, B3FMXC, F0CN2K, F06N28, F08HBG, F07862, B3XZNF, B3FKYG, B3XZNM, F08K8K, F09865, B3FL9W, B3FL5B, B3FM9Y, B3FMMR, F0CN2H, F07RXT, F0FP2M, F09BF2, F05HXQ, F0B9YD, B32V4L, F0986Q, B3FMPN, F095F6, F0CN2C, B32TBP, F0G7NK, B3XZ1F, F05HY3, F09FY3, F0HFN2, F0HG4B, F077M0, B3LBCL, F09FYH, B3FL83, F08JKR, B3XZW9, B32V2V, F06PR3, F07860, B3FLQD, B3FKY4, B3FM61, B3FMT9, F08MH7, F08MJ6, F0HX7L, B3FMRN, F09FXV, F06PQJ, F09LHD, B32TG2, B32TDL, B32TDZ, B32V9P, B32V9V, B32VD0, B3FMRK, F0986P, F0986X, B3FMJM, F0HT0T, F0HNBG, F0J38H, F0CHGB, B32TPM, B32TZ3, B415T9, B41GQ4, F04MYH, F04MYL, F04P6X, F04XCJ, F079Y6, F0KL6K, F0BD0D, B3FLH7, B3FLJ2, B3FLJJ, B3FLK2, B3FM12, B3XYVM, B411GJ, F08F2P, F095F4, F0986M, F0986N, F09D3B, B3XZLQ, B3XZNT, B3Y002, F09867, F0BP4C, B3L6KT, B3FMV6, F0HFN9, B32TBN, F0CLFB, F0CN1K, F0CQ4J, F0CQ56, B3XZGP, B3XZTM, F0B49K, B3L49P, F08R3Y, F08R4M, F08W4Y, B32TBJ, F096BD, F09LGR, F09LH9, F09TK2, F0CHH4, B3FLQ3, B3Y0FZ, B2PYFV, B32TQV, B32TTC, B3Y0KT, B32V17, B3L4Z3, B3XZZB, B3XZMW, B3FLPY, B3FM70, F0DR5G, F07RXF, F07RXG, F07858, B32TR8, B32V8Z, F0BKHG, F0CT05, F0JCQR, F0CQ4X, B32TW1, F0CQ48, B32TZ4, B3FLNB, F0BHVC, F0JDLY, F0FZVF, B32QBX, B3Y048, B32QD0, B32TNZ, B3FL91, B3FMTH, F06TGQ, F04MYG, B3XZN6, B411GR, B32TVQ, B3FMPG, F0HT0X, F0JQTQ, F0JQTT, F0JQTW, F0JT5B, F0JTFB, F0JTYM, F0JT55, B3FMCC, F07D16, F07K79, F07NX0, F06TGV, B32THV, F073GY, B3FMCL, B3FLRD, B3FLV6, B3FLH8, B3FLL5, B32TFD, F08HBM, B3XZKY, B32V41, B32TRB, F0KL6X, F0KK51, F0HT21, F0J38G, F0J4R5, B3FMXG, F0HVY5, B3L49J, B32TCL, B32TKF, B3FM38, B3Y0MK, F083YV, F0FQZ7, F0FQZQ, B3FMW3, B32TV8, B3FLTB, B3FLJC, B3FLJH, B3FLLN, B3FLM4, B3FMVF, B3FMQB, B3LBCK, F0HT1N, B3LBGW, B3XZNV, B3FL8F, F08002, B32V8V, F09Z2Q, F0BP4B, B3L4T8, B32VBK, B32VCY, B3FLC9, F08R4B, F09LJ0, B3FLJ8, B3FLJK, B3FLJL, B3XYTX, B3XYVJ, B3XYY0, B411GM, F04XD1.F06PQG, F06TH2, F0BHVQ, F06TGR, B3XZP6, B3XZQ9, B3Y001, B3Y02M, B3FKXM, B3FMM7, B3FL72, B3Y064, F07WT2, F07WTR, F07WRV, F08JKX, B3FMV1, F08W49, B3FMVN, F077M1, F08JL7, B2PYF9, B3LB9T, B3LBB2, B3LBBF, B3LBC7, B3FL88, F07CD6, F07D12, B3L4T3, B3XYV9, B3FKXV, B3FMWR, F0DR5K, F0DZ2W, F0DZ35, F0F02W, F0F5TC, F083YD, F083YW, B32VDB, B3FMWJ, F04P7F, B3FL87, B3FKZQ, F06N32, F0BD03, F0BD07, F05768, F0CLFG, F0B8BZ, B3XZDG, B3XZH0, B3XZH5, F06HYF, F06HYM, F091RQ, B3FLFG, B3FM02, B3FM04, B41GQF, B3FL8T, F0784Z, F083YF, F08BBF, F08BBP, F08BC4, F0BP4F, F0CHGQ, B3FLHF, B3L6KN, B2PYGP, B3Y02H, B3Y043, B3Y049, F04MYN, F04P7Y, F0JT56, B3FMR4, B3FMRJ, B3FMV2, F0B49D, F0BCZX, B3LBH5, B3LBH6, F08JL0, B3FM5Z, F04XD9, B3FLCY, F07GL8, B3Y0HX, F0JB5W, B32VFM, B2ZR2N, B2ZR1Y, F06THH, F0BD0V, B3FMBF, B3FMHW, F083YH, F0CQ55, B3FLPV, F09FYM, F0CQ5F, B3XZ1J, B3XZLH, F04V3Y, F04V45, F04XCQ, F053C6, F0FMXB, B3Y02C, B3Y02D, F0JH1F, F0BT8H, B3Y0KX, F09LP0, F0B4BJ, F073GX, B3FLJ4, B3FLJM, B3FLJQ, B415R8, B41GQ6, F09LP5, B3FM0Z, B3XZKM, B3XZMG, B3XZP9, B3XZPD, F04V40, F07NXM, F0B3FB, F0J4HW, B3XYZP, B3XZ0M, B3XZDX, F06HYT, B3FM6B, F09D31, F0JQT9, B3FLF2, B3XZFC, F04XD7, F04XD8, B3LB8M, B3FL4T, B3FKZT, B32TCH, B411GF, F06PQF, F079Y0, F0KK56, B3FL7W, F0GV2B, B3FLZH, F05BT4, F07K7F, B3LB8P, B3LB8R, F06FBC, F06KTV, F077LH, F077LW, F07851, F0785J, F04P7M, B32TVL, B3FMRT, B3L498, B3L6KF, B3LBCW, B3LBD1, B3LBH8, F0FP2Q, F04P72, B3LB8Q, F0LQH0, B3FM1C, F05BRJ, F077LC, B32V07, F08K8F, B32TWY, F0CQ44, B3FL8H, B3L495, B3L49Q, B3XZP3, F07CD1, B32V69, B415TB, F0HG59, B32V82, B3FM9B, B32V76, B32V9Z, B32TTP, B3FMNM, B32V4D, F0HG4J, B3FLVQ, B3FMP2, B3FMN2, B3XZDK, B3FL78, B3FMD0, B3Y0L2, B32V0G, B3FLQF, B3XYYB, F09C05, B3FMQJ, B3XZQD, F0BD0P, B3FMM5, F0G350, B32TPD, B3FLGV, F0FQZ5, B3FLJY, B3FLK0, F0JQT8, B3FLJ5, B3FLJF, F07RXM, B3LBGB, B3FLL2, F07D0V, F07D0Y, F08JL5, B3FM06, B3FM08, B3FM6P, B3FM7W, B3Y0J8, B3Y0KM, B3Y0M7, B3Y0MF, B3FMV8, B3XZRC, B3XZTV, F0JZJG, F0CLF8, F0CLFW, F0CN26, F0CN27, F07WTM, F0HFN7, F0JQTJ, B2PYF7, F06KTQ, B3FMB3, B3FLJD, B3FM90, B3XZRG, F07853, F0DXHD, F09LPC, F0B8C3, F0CHFV, B3XZGX, B32V85, B32TPP, B32TH4, B2ZR38, F06TGP, B32TPB, B3FMDT, B3FKXN, F07856, F079Y2, F0DR4V, F06HYQ, F0785G, F0CHG6, B3Y0FN, F091RN, F096BM, F09869, B3FLLM, B3XZ10, F06TH9, F0DJL4, F08XZR, B3Y0G7, F077MK, F04XDH, F04XDL, B32V8M, B3FKXT, B41GQC, F095FF, F095FN, F091RH, B3XZ08, B2PYGX, B3FMRH, B3FMRW, B32TR0, B32QCP, B32TLD, B2ZR2F, F0K7MJ, B3FLB9, B3Y0FT, F0BHVV, B3FL57, F0BD0G, F0F4CV, B3FKY2, B3FL00, B3FMPJ, F0HG4R, F0HX7R, F09KJV, F053BV, F05BTF, B3FL1P, B3FLGY, B3FLJB, B32V63, B32V6J, F07RX9, B3XZTZ, B3L6L0, F08K7V, B3FMKG, B3FMNC, B3L491, B3FLGZ, B3FLK3, B3XYTD, B3FLF5, B32TNK, B3Y060, B3Y068, F0B4B0, B32TNH, F079XT, F077LP, B3FL1J, B32V1X, B3FKYD, B3FMFP, B3LBD4, B3XYZJ, B3FLVF, F0DWDY, B3XZN4, B3XZWB, B32TRL, B415R3, F08BBZ, F08HC7, F0CQ58, B32TD3, B32TFJ, B32TDD, B2PYF6, B3FMR1, B3FMRP, F08HBW, F0JL45, F0BKHJ, B3XZF5, B3XZ12, B3XZGV, F0J9JV, B32TDP, B32VC3, B3XYTK, F0HFMH, B415T4, F0BD0N, F0J9JT, F0CHFY, B2ZR2L, B32TDG, B2ZR27, B3FM9C, F09Z28, B32TPL, B3FM09, B415T5, B32V4C, F0BCZT, B3L51B, F0FQZ9, F0J9JN, B3L4ZL, F053BX, B3FMT1, B32VCT, B3LB9Y, B3XYZM, B3XZ0W, F0BCZL, F0B8CW, B3XZN2, B32TRW, B3FLJN, B3FLL1, F09TK4, F0BPYH, B3FLCJ, B2ZR3Y, B32TG8, B32TT4, B3L505, B32TFZ, B3Y0H8, B32THR, B32TDV, B2ZR2M, B2ZR2P, B32V45, B32V59, B32V61, B32V74, B3XZGG, F06KRR, F06TGX, B2PYGL, B2ZR37, F0G7NT, B3FM64, B3XYXQ, F06KTK, B32TM7, B411G2, F04P7R, B32TQY, B32TRH, B32TR6, B32TYF, B32TZY, B32V0L, B32V0P, B32V1D, B32TT5, B32TTM, B32TTW, B32TV5, B32TVW, B3FL98, B3FL9D, F0B8CT, F07K7D, B3FLZL, B32V27, B32V2J, F0BKHB, F06PQW, B32V5D, F0BT8M, B32V4K, B3FLJ1, B3FLJ3, B3FLJR, B3FLJZ, B3XYVH, F095FX, F0986B, F09C01, F09R4M, B32V31, F0D435, B32TNY, B32V6M, B32V6T, B3FMHR, B32VC2, B32V83, B32VCB, B32V7Y, F0BH0C, F0JCQZ, B3FMM8, B3FMVM, B3FMWN, F0JT59, F079Y8, F0DP3J, B32V8X, B32VBW, F09R4N, F0B13K, B3FLZX, B32V8W, B32V91, B3FLZN, B32V8D, B32TNT, B32VCV, B3FMWW, B32VD9, F0BH0D, B3FMXD, F09KJL.F0JH16, B3FMP8, F06KV4, B32QBW, F096C8, B3FMT2, F0J389, B3FMV3, B32V90, B3FMK3, F0HDKG, B3FMRX, F09KJJ, B3FMND, B3Y06H, B3Y0JD, B3FMJW, B3L4TK, F053BH, B3L4T2, B3FMFQ, B3LBGZ, F095FJ, B3L4RW, F0BD0B, B3L6LF, F0CLFH, F09FYG, B3L49C, B3L4T5, B3XZLY, B3XZN9, B3XZP1, F0FP1L, F0FVHD, B3XZDT, B3XZLJ, B3FL1G, F0CQ5B, B3FL0C, B3FL0D, B3FL1H, B3FMNG, F08W4K, F09D3D, F0HX7T, B3FLB7, B3FL9Q, B3FLB1, B3FLGT, F0F02V, B3FL2H, B3FLY4, B3FM1V, B3FM20, B3FM3P, B3FLB6, B3FLGC, F0986Z, B3Y0HP, B32VC0, B32VCF, B3FMX8, B3Y04M, B3FLLD, B3FLL3, F05DQ1, F05BRZ, F09L7D, B3FM07, B3FLZP, F04XCN, B3XZV6, B3Y018, B32TKD, B3Y06D, B3Y06T, B3Y0G1, B3Y0GY, B3Y0HF, B3Y0HM, B3Y0JX, F08F2M, F06N36, B3FM1D, B3FLJ0, B3FLK4, B3XZTQ, B3Y0L0, B3XZMD, B3FLJV, B3Y0H2, F096C5, B3FL26, F077LL, B3XZN5, B41GQ8, B411GH, B3XZM8, B3Y01B, B3Y01H, B3XZNB, F09L7J, B3XZTX, F09D2Z, F0BH07, F0BT7G, F0CT09, F0CT0J, B3XZF1, F06TGT, F06THV, F06FB3, B3FMK0, B3FMML, B3XZFD, B41GPX, F04P74, F04V47, B3XYXR, F0986J, F09LHP, F04P7X, F077MF, F09D3C, F0CQ4H, F04XD0, F09FXY, F09Z2H, F0B13R, F0B49Z, F077M3, F07CCN, F077LY, B3XZGC, F0FMWK, F073GJ, F0GV2F, F0BT8J, F07WT8, F077L8, F0JL43, F07NX3, F07WT6, F079X1, B32V77, F073GM, F07NWY, F07WTP, F07NXJ, F08W4Q, F07NWZ, F08K8H, B3FKZ0, F0BKGY, F053BL, F08W4W, F095FV, F077LR, B32TWB, F06TH6, F09868, F0JT53, F09L7P, F0CQ4V, B32TFV, F0HXHB, F0HXHV, F0HDKP, F0JT50, F0B273, F0BKH1, F0BKHM, B32TCZ, B3FM8V, F09KJP, B32TCQ, F0B3FG, F0CN24, F09FYF, F0JWZJ, F09D2T, F0FP26, F0FQZ1, F0BPZ2, B32TTR, F0B13J, F0B3FR, F0BCZG, F0BH0B, F0J4HV, F0BT7J, F0JT4Y, F0CHG7, B2ZR0Y, F08BC1, F0CQ43, B3FL02, F0BCZY, F0BPYV, F0B8C5, F0CQ4D, F0CHGX, F0CN1D, F0CQ4F, F0CQ4Q, F0CQ5M, F09BF4, B32QCW, F0CT0G, F0H4D0, F0CQ4R, F0F5RL, F08BBL, F08JL9, F0F030, F0FP1W, F0B49B, F0F5R9, F0F5RC, F0F45J, F0B9Y8, F0B8CX, B3FLDX, F09KJH, F0B9XY, F0H8R2, B3FLHC, F0CQ4B, F0986D, F0HG58, F0HG54, F0GPHJ, F0H6QM, F0H6QJ, F0HFMZ, F0J38B, F0J4R7, F0B8D1, F0JDLM, F0JL3K, F0JL3Q, F0JCQV, F0J38F, F0J4R4, F0J9JK, F0JG5Q, F08MHG, F095F9, F0CHHB, F0CN29, F095FH, F08W4C, F0CHGL, F0BH02, F08R42, F08XZ2, F0JTFH, B3FLQC, F0HN33, F07864, F0JTF9, F0JQTC, F0JT52, B3FLB0, F0KG9R, F0JWZR, B32TWX, B3FLRG, B3FLZT, B3XZ0F, F0HG5L, F0KK57, F0KL6L, B3FKYC, B3XYZC, F08F2C, F09TK5, F0CHG0.
|
| FEI Number |
2518586
|
Recalling Firm/ Manufacturer |
Philips Ultrasound, Inc 1 Echo Dr Reedsville PA 17084-8603
|
| For Additional Information Contact | Philips Customer Care Solutions Center 1-800-722-9377 |
Manufacturer Reason for Recall | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
FDA Determined Cause 2 | Device Design |
| Action | An URGENT Medical Device Correction Notice, dated 7/3/25, was mailed to consignees. This notice instructs consignees to locate all Philips Ultrasound transducers in their facility and enter their serial numbers into https://www.transducer.philips.com to obtain the manufacturing certificate for each of their devices. This certificate should be retained with its system documentation. Using the manufacturing certificate consignees are to calculate the useful life of their device. The expected useful life for Transesophageal transducers, X-Matrix transducers, and Lumify transducers is 5 years; The expected useful life for Endocavity transducers and Transcutaneous transducers is 7 years. If a transducer has exceeded its useful life, consignees are instructed to conduct transducer performance tests or contact Philips for guidance. Consignees are asked to complete and return the provided response form and circulate the recall notification to device users for their awareness. Consignees with any questions can contact Philips Customer Care Solutions Center at 1-800-722-9377. |
| Quantity in Commerce | 171,322 units |
| Distribution | US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = IYN
|
|
|
|