Model Number STANDARD
Device Problems Crack (1135); Device Contamination with Body Fluid (2317)
Patient Problems Abdominal Pain (1685); Exposure to Body Fluids (1745)
Event Date 02/28/2015
Event Type  malfunction  
Event Description
Male patient was in the emergency department for abdominal pain.Physician noted patient's gown had blood soaked spot (quarter-sized).Patient assessed and noted to have no open wounds.It was determined that the mattress cover was cracked and mattress soaked with blood.Previous patient had been a vaginal hemmorrhage.
=
manufacturer response for emergency room patient bed, pioneer (7 beds) (per site reporter).
=
mfr inquired about cleaning methods/materials and suggested bleach was pooled on cover causing cracking and breaking of cover.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STANDARD MODEL BED
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
STRYKER MEDICAL
3800 east centre avenue
portage MI 49002
MDR Report Key4619942
MDR Text Key5594610
Report Number4619942
Device Sequence Number2
Product Code FNL
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Type of Report Initial
Report Date 03/12/2015
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Date FDA Received03/13/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Invalid Data
Device Model NumberSTANDARD
Device Catalogue Number07855-034-303
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/13/2015
Device Age2 YR
Event Location Hospital
Date Report to Manufacturer03/20/2015
Patient Sequence Number1
Patient Age17 YR
Patient Weight73
-
-