Class 2 Device Recall Cook MultiUse Holmium Laser Fibers

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Cook MultiUse Holmium Laser Fiberssee related information
Date Initiated by FirmMay 09, 2017
Date PostedJune 05, 2017
Recall Status1 Terminated 3 on January 26, 2018
Recall NumberZ-2539-2017
Recall Event ID 77304
510(K)NumberK124030 
Product Classification Powered laser surgical instrument - Product Code GEX
ProductCook Multi-Use Holmium Laser Fibers, Supplied sterilized by ethylene oxide gas in peel-open packages
Code Information HLF-M273-H30  HLF-M365-H30  HLF-M550-H30  HLF-M940-H30  HLF-M273-HSMA  HLF-M365-HSMA  HLF-M550-HSMA  HLF-M940-HSMA  
FEI Number 1820334
Recalling Firm/
Manufacturer
Cook Inc.
750 N Daniels Way
Bloomington IN 47404-9120
For Additional Information ContactCook Medical Customer Relations Departme
812-339-2235
Manufacturer Reason
for Recall
Correction to the product Instructions for Use because the current reprocessing instructions do not provide sufficient detailed information for the cleaning, disinfection, and sterilization of these products.
FDA Determined
Cause 2
Device Design
ActionCook is initiating a voluntary recall of multiple products in the form of a correction to the product Instructions for Use. Cook Inc. has identified the current reprocessing instructions do not provide sufficient detailed information for the cleaning, disinfection, and sterilization of these products. Potential adverse events that may occur if the products are not adequately reprocessed include urological infections and systemic infections from a urological origin as well as events resulting from chemical residual exposure. Cook distributed notices via courier on May 9, 2017. Customers were instructed to do the following: Distribution of the Cook Multi-Use Holmium Laser Fibers will not occur until the reprocessing instructions in the Instructions for Use have been corrected. Single use fibers are available. Contact your local sales representative for further information. You can continue to use your inventory of Cook Multi-Use Holmium Laser Fibers by following the attached Suggested Fiber Reprocessing Instructions. Action to be taken: 1. Examine your inventory immediately to identify and quarantine those affected products. 2. Implement the updated Suggested Fiber Reprocessing Instructions. 3. Return the required Acknowledgement and Receipt Form within 30 days. 4. Even if you do not have affected products, you must still complete the Acknowledgement and Receipt Form and send it via fax (812.339.7316) or email (fieldactionsna@cookmedical.com). 5. Immediately report any adverse events to Cook Medical Customer Relations at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30 am and 5:00 pm (Eastern Time), or by email at customerrelationsna@cookmedical.com. Adverse events or quality problems experienced with the use of this product may also be reported to the FDA: Online at: http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail) Call the FDA at: 1-800-FDA-1088 Should you have any medical questions or concer
Quantity in Commerce4369
DistributionDomestic: AL AR AZ CA CO FL GA HI ID IL IN KY LA MD MI MO MS NC NM NY OH OK OR PA SC TN TX UT VA WA WI Foreign: Austria, Australia, Argentina, Azerbaijan, Belgium, Bulgaria, Bahrain, Canada, Chile, China, Cyprus, Czech Republic, Denmark, Djibouti, Egypt, France, Hong Kong, Honduras, Croatia, Greece, Germany, Hungary, India, Ireland, Italy, Latvia, Malaysia, Morocco, Netherlands, Poland, Portugal, Puerto Rico, Saudi Arabia, Serbia, Spain, Slovenia, South Africa, Sri Lanka, Sweden, Singapore, Taiwan, Thailand, Tunisia, United Kingdom, and Uruguay
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = GEX
-
-