Class 2 Device Recall Philips XRay System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Philips XRay Systemsee related information
Date Initiated by FirmFebruary 19, 2019
Create DateApril 13, 2019
Recall Status1 Terminated 3 on October 14, 2020
Recall NumberZ-1140-2019
Recall Event ID 82459
510(K)NumberK031333 K033737 K041949 K050151 K984545 
Product Classification System, x-ray, angiographic - Product Code IZI
ProductFoot Switches used with Philips MultiDiagnost Eleva with Flat Detector; Product Codes: 708037, 708038 Product The Allura Xper FD Series are intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures.
Code Information Foot Switches: 452270000141 Footswitch CV 3p 4m; 452270000151 Footswitch MD 3p 6m; 452270000142 Footswitch CV 3p 4m; 452270000381 Footswitch CV 3p 8m; 452270000382 Footswitch CV 3p 8m; 459800076001 Footswitch CV 4p+2 4m; 459800076021 Footswitch CV 4p+2 8m; 459800415571 Wireless Footswitch 4p+2; 459800415581 Wireless footswitch set 4p+2; 459800415531 Wireless Footswitch 3P; 459800415561 Wireless footswitch set 3P.
FEI Number 3002807880
Recalling Firm/
Manufacturer
Philips Medical Systems Nederlands
Veenpluis 4-6
Best Netherlands
For Additional Information ContactTechnical Support Line
800-7229377
Manufacturer Reason
for Recall
Additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.
FDA Determined
Cause 2
Component design/selection
ActionThe firm previously notified customers on September 24, 2014 as part of Recall Z-1280-2016. On February 21, 2019, the firm sent an updated recall letter to newly identified affected consignees. Customers were alerted that bent pedal of footswitch might cause interruption of live Flouroscopy imaging or exposure. Action to be taken by Customer/User: 1. Before starting a procedure, the user should check the footswitch for possible bent pedals. If bent pedals are found the procedure should not continue and Local Service should be contacted. 2. Do not use the footswitch on an anti-fatigue mat or in the pedestal. 3. Please ensure that all staff with access to the affected systems are informed of the content of this Medical Device Correction. 4. Include this Medical Device Correction together with the documentation of the system until the system has been corrected by Philips. The firm stated that the problem will be resolved by an update of the mechanical hardware of the footswitch which will prevent the pedals of the footswitch from becoming bent. The mechanical hardware was available as of February 2019. Customers will be notified by their local Philips representative to schedule a date for installation of the mechanical hardware update. Technical Support Line: 1-800-722-9377
Quantity in Commerce7209 total
DistributionUS Nationwide Distribution in the states of CA, FL, MI, OR, PA, TX and WA.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.