| | Class 1 Device Recall Thermo Scientific Sensititre Plate, GN4F, |  |
| Date Initiated by Firm | October 20, 2022 |
| Date Posted | December 08, 2022 |
| Recall Status1 |
Terminated 3 on September 30, 2024 |
| Recall Number | Z-0327-2023 |
| Recall Event ID |
91104 |
| 510(K)Number | K060621 K073424 K860753 K880641 K936305 K983244 |
| Product Classification |
Gram negative identification panel - Product Code JWY
|
| Product | Thermo SCIENTIFIC Sensititre Plate, GN4F, Gram Negative IVD AST |
| Code Information |
Lot Numbers: B2232B, B2197, B2012, B1355, B1181, B1173, B1125A |
| FEI Number |
1924669
|
Recalling Firm/ Manufacturer |
Remel Inc
12076 Santa Fe Trail Dr
Lenexa
KS
66215-3519
|
| For Additional Information Contact | Tom Campbell 913-895-4077 |
Manufacturer Reason for Recall | An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam. |
FDA Determined Cause 2 | Process control |
| Action | The firm issued an "Urgent Medical Device Correction" notice to its consignees on 10/25/2022 by US mail. The notice explained the issue, risk to health, and notified the consignees of an update to the IFU, specifically, "to add limitations for Carbapenems (Imipenem, Doripenem, Meropenem, Ertapenem), Cefepime, Piperacillin/Tazobactam and Aztreonam added for users to "NOT REPORT" any result for each member of the Protea tribe (including Proteus spp., Providencia spp., Morganella Morganii) to address false susceptible results."
Please complete and return the Required Customer Reply form via email to MBD.vigilance@thermofisher.com. Please keep this notification on file.
The notice needs to be passed on to all who need to be aware within your organization or to any organization where the potentially affected products have been transferred.
This correction notification is being made and performed by the US agent, Remel Inc., on behalf of the Foreign Manufacturer of the affected product, Trek Diagnostic Systems Ltd., located in East Grinstead, UK. Trek Diagnostic Systems and Remel Inc., both part of Thermo Fisher Scientific, are investigating the root cause and will implement corrective actions to prevent future recurrence.
If you have any questions, please contact our Technical Services Department at 800-255-6730 (US) or email microbiology.ts.us@thermofisher.com.
|
| Quantity in Commerce | 1917 units |
| Distribution | US Nationwide Distribution |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|