| | Class 2 Device Recall NovoThor |  |
| Date Initiated by Firm | September 13, 2022 |
| Date Posted | December 01, 2023 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0453-2024 |
| Recall Event ID |
93344 |
| Product Classification |
Lamp, infrared, therapeutic heating - Product Code ILY
|
| Product | NovoTHOR Gen 2.0 whole body red light therapy bed, model S1176 |
| Code Information |
Serial numbers: 2003, 2004, 2005, 2006, 2007, 2009, 2010, 2011, 2017, 2018, 2015, 2020, 2019, 2001, 2021, 2028, 2033, 2036, 2030, 2029. |
| FEI Number |
3003899624
|
Recalling Firm/ Manufacturer |
Thor Photomedicine Ltd
Watermeadow House
Chesham
United Kingdom
|
Manufacturer Reason for Recall | Failure of ball stud components causing the gas strut on one side to detach from the canopy, making the canopy difficult to lift. |
FDA Determined Cause 2 | Device Design |
| Action | The "IMPORTANT MEDICAL DEVICE CORRECTION" notification issued October 05, 2023 by email with follow-up phone call. The notice advises users (1) how to recognize the issue. (2) THOR is replacing all ball studs on all affected devices.
Should you continue to have users/patients use the NovoTHOR, then be aware it may potentially cause harm
should the failure occur whilst they are lifting or lowering the canopy.
" Ensure you warn all users of the potential risk
" Ensure someone is present to lift and lower the canopy
" Ensure the canopy is lifted and lowered gently and slowly
" Ensure someone is present in the room or immediately outside the room during use.
A service representative will contact customers within 6 weeks to schedule correction. Customers are requested to respond to the email notice with their acknowledgement. Contact phone number for Thor is (877) 355 3151. |
| Quantity in Commerce | 19 |
| Distribution | Worldwide Distribution: US (Nationwide): Arizona, California, Colorado, Connecticut, Florida, Hawaii, Idaho, Illinois, Iowa, Maryland, Michigan, Minnesota, Missouri, Nevada, New Hampshire, North Carolina, Ohio, Oregon, Texas, Utah, Virginia, West Virginia.
OUS (foreign) to countries of: Australia, Canada, Netherlands, New Zealand, Singapore, United Kingdom.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|