| | Class 2 Device Recall HA FlexTrak |  |
| Date Initiated by Firm | May 03, 2024 |
| Date Posted | June 04, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1948-2024 |
| Recall Event ID |
94603 |
| 510(K)Number | K213583 |
| Product Classification |
System, nuclear magnetic resonance imaging - Product Code LNH
|
| Product | HA FlexTrak-Patient transport functionality to transport the patient from the preparation area to the MR system,
Product Number: 989710006411 |
| Code Information |
UDI: N/A
Serial Numbers:
3
5
6
9
11
12
13
14
15
16
17
18
19
20
22
23
25
26
29
30
31
32
33
34
35
37
38
39
41
42
43
44
45
49
50
51
52
53
55
56
58
61
63
64
66
67
68
70
71
72
73
75
77
79
80
81
82
83
84
86
87
88
89
91
93
94
95
97
98
100
101
102
103
104
105
107
108
109
110
111
112
113
114
115
116
117
118
119
120
122
123
124
125
126
129
130
132
133
135
136
137
138
139
140
141
142
143
146
148
149
150
151
152
153
154
155
156
158
159
160
161
162
163
164
165
167
169
170
171
172
173
174
175
177
180
181
183
184
185
186
188
189
190
191
192
193
195
196
197
198
200
201
202
203
204
205
206
207
208
209
211
212
213
215
216
218
219
220
221
222
223
224
225
226
228
229
232
233
234
235
236
237
238
240
241
242
243
244
246
248
250
251
253
254
255
257
258
259
260
262
263
264
265
267
269
270
272
273
274
275
278
279
281
282
284
286
287
288
289
290
291
292
293
294
295
296
297
298
301
302
303
304
305
307
308
309
310
311
312
313
314
315
317
318
319
320
322
323
324
325
326
328
331
332
334
335
336
337
338
339
340
341
342
343
344
345
346
347
350
351
352
355
356
358
359
360
361
362
363
364
366
367
368
369
370
371
372
373
374
375
379
380
381
382
384
386
387
388
390
391
392
394
395
396
397
399
400
401
402
403
404
405
406
408
409
411
412
413
414
415
416
417
419
420
421
422
423
424
426
427
428
429
431
432
434
435
436
437
438
439
440
441
442
444
445
446
448
449
450
451
452
453
454
456
457
460
461
462
463
464
465
466
467
468
469
472
473
474
475
476
477
478
479
480
481
482
483
485
487
489
491
492
493
494
496
497
498
499
500
501
502
503
504
506
507
508
509
510
511
512
515
516
517
518
520
522
523
524
525
528
529
531
532
534
535
536
537
538
541
542
543
544
545
546
547
548
549
550
552
553
554
555
556
558
559
560
561
562
564
565
566
567
568
569
571
574
575
577
579
582
583
584
585
587
588
591
592
593
594
596
597
598
599
600
601
603
605
607
608
609
610
611
613
614
618
619
620
621
623
625
627
628
629
630
631
633
634
635
639
640
641
642
643
645
646
647
648
650
651
652
653
656
658
659
660
661
662
663
664
665
666
667
668
670
671
672
673
674
675
676
677
679
680
681
682
683
684
686
687
688
689
692
694
696
697
698
700
702
705
706
708
709
710
711
712
716
717
718
719
723
724
725
727
728
729
730
731
732
733
734
735
737
738
739
740
742
743
744
745
746
747
748
749
750
752
753
754
755
756
757
758
759
760
762
763
764
765
766
768
769
770
771
772
773
774
775
777
778
779
780
781
782
783
784
787
788
789
790
791
792
793
794
795
796
797
798
799
800
801
802
803
804
807
808
809
810
811
813
814
815
816
817
819
820
821
822
823
825
826
827
828
829
830
831
832
834
835
836
837
838
839
840
841
842
843
844
845
847
848
849
850
852
853
854
856
857
858
859
860
862
863
865
867
868
869
870
871
872
873
874
875
876
877
880
881
882
883
884
885
886
887
888
889
890
891
892
893
894
895
896
897
898
899
901
903
904
905
906
908
909
910
911
912
914
915
916
918
919
921
922
924
925
926
927
928
929
930
931
932
933
934
935
936
937
938
939
940
941
942
943
944
945
947
948
949
950
953
954
955
956
957
958
961
962
963
968
969
970
971
973
974
975
977
979
980
981
982
984
985
986
987
988
989
990
991
992
993
994
995
996
997
1000
1001
1002
1003
1004
1005
1006
1007
1008
1010
1012
1013
1015
1016
1017
1018
1019
1020
1021
1022
1023
1024
1025
1026
1027
1028
1031
1032
1033
1034
1035
1036
1037
1038
1039
1040
1043
1044
1045
1046
1047
1048
1049
1050
1051
1053
1054
1055
1056
1057
1059
1060
1061
1062
1063
1064
1065
1066
1068
1069
1071
1073
1074
1076
1077
1078
1079
1080
1083
1084
1085
1086
1088
1090
1091
1092
1094
1095
1096
1097
1098
1099
1100
1101
1103
1104
1105
1106
1107
1108
1109
1110
1113
1114
1117
1118
1119
1120
1121
1123
1124
1125
1126
1130
1131
1132
1133
1135
1136
1138
1139
1140
1141
1142
1143
1145
1146
1147
1148
1149
1150
1151
1153
1154
1155
1156
1157
1159
1160
1161
1162
1163
1164
1165
1166
1167
1168
1169
1170
1172
1174
1175
1177
1178
1179
1180
1181
1182
1183
1184
1185
1186
1188
1189
1190
1191
1194
1197
1198
1199
1200
1201
1203
1204
1205
1206
1207
1208
1209
1210
1211
1212
1214
1215
1216
1217
1218
1219
1220
1221
1222
1223
1224
1226
1227
1228
1229
1231
1233
1234
1235
1236
1237
1238
1239
1240
1241
1242
1243
1244
1247
1248
1249
1250
1252
1253
1254
1256
1258
1260
1261
1262
1263
1264
1265
1268
1270
1272
1273
1274
1275
1276
1277
1278
1280
1281
1283
1285
1286
1287
1288
1290
1291
1293
1295
1296
1297
1298
1299
1300
1303
1304
1305
1306
1307
1308
1309
1311
1313
1314
1315
1316
1317
1318
1319
1320
1321
1322
1323
1324
1325
1326
1327
1328
1329
1330
1331
1332
1334
1335
1336
1337
1338
1339
1341
1342
1343
1347
1348
1349
1351
1352
1353
1354
1355
1356
1357
1358
1359
1360
1361
1362
1363
1364
1365
1366
1367
1368
1369
1370
1371
1372
1373
1374
1375
1377
1378
1379
1380
1381
1382
1383
1385
1386
1387
1388
1390
1393
1394
1396
1397
1398
1399
1400
1402
1403
1404
1405
1406
1407
1410
1411
1412
1413
1414
1415
1417
1418
1419
1421
1422
1423
1424
1425
1426
1427
1428
1429
1430
1431
1432
1433
1436
1437
1439
1440
1441
1442
1443
1444
1445
1446
1447
1448
1449
1450
1451
1452
1454
1455
1457
1458
1460
1461
1462
1463
1464
1465
1466
1467
1468
1469
1470
1471
1472
1473
1474
1475
1476
1477
1478
1479
1481
1482
1483
1484
1485
1486
1487
1488
1489
1490
1491
1492
1493
1494
1495
1496
1497
1499
1500
1501
1502
1503
1504
1507
1508
1509
1510
1511
1512
1513
1514
1515
1516
1517
1518
1519
1520
1521
1522
1523
1524
1525
1526
1527
1529
1530
1531
1532
1533
1534
1535
1536
1537
1570
1571
1572
1573
1574
1575
1577
1578
1579
1580
1581
1582
1583
1586
1589
1590
1592
1594
1595
1596
1597
1598
1600
1601
1602
1606
1607
1608
1609
1610
1611
1613
1614
1615
1616
1617
1619
1620
1621
1622
1623
1624
1625
1626
1627
1628
1629
1630
1632
1633
1651
1652
1656
1657
1658
1659
1660
1662
1664
1665
1666
1699
1703
1704
1705
1706
1708
1712
1713
1714
1715
1717
1718
1720
1721
1722
1723
1726
1727
1772
2060
101683
101684
101685
101686
101688
101689
101690
101691
101692
101693
101694
101696
101701
101707
101711
101719
101725
101728
101729
101730
101731
101733
101734
101738
101739
101740
101741
101742
101743
101744
101745
101763
101767
101768
101769
101770
101771
101773
101775
101777
101778
|
| FEI Number |
3016618143
|
Recalling Firm/ Manufacturer |
Philips North America Llc
222 Jacobs St
Cambridge
MA
02141-2289
|
| For Additional Information Contact | Customer Care Solution Center 800-722-9377 |
Manufacturer Reason for Recall | Oil may leak from the HA FlexTrack Trolley onto the floor when pressing the hydraulic pedal, resulting in a potential slipping/falling hazard |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Philips NA issued URGENT Medical Device Correction letter to US consigneess dated May 3, 2024 via FedEx. Philips Markets Organizations are responsible for distributing the letters outside of the US.
Letter states reason for recall, health risk and action to take:
A. If you see hydraulic fluid (oil) leakage:
" First remove fluid from the floor using a paper towel. Then remove any remaining oil film
from the floor using a cleaning agent (e.g. soap and water) and paper towel.
" Contact your local Philips service representative.
B. You may continue to use your HA FlexTrak Trolley in accordance with the intended use.
C. Circulate this notice to all users of this device so that they are aware of the issue and
associated hazard/harm until this issue has been resolved.
D. Please post this letter near the associated MR system(s) for ease of reference.
E. Complete and return the attached response form to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the URGENT Medical Device Correction, understanding of the issue, and required actions to be taken.
A Philips representative will contact you to schedule a time for a Field Service Engineer (FSE) to visit your site to inspect the HA FlexTrack Trolley and correct if required (reference FCO78100580).
If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377, 8AM-8PM EST, Monday-Friday). |
| Quantity in Commerce | 1348 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain,
Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark,
Estonia, Fiji, Finland, France, French Guiana, Germany, Guadeloupe, Honduras, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Jordan, Lithuania, Martinique, Mexico, Morocco,
Netherlands, Norway, Oman, Peru, Poland, Russian Fed., Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = LNH
|
|
|
|