| | Class 1 Device Recall Impella RP with SmartAssist |  |
| Date Initiated by Firm | December 12, 2024 |
| Date Posted | January 15, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0829-2025 |
| Recall Event ID |
95780 |
| PMA Number | P170011 |
| Product Classification |
Temporary non-roller type right heart support blood pump - Product Code PYX
|
| Product | Impella RP with SmartAssist; Product Number: 0046-0035; |
| Code Information |
Product Number: 0046-0035; UDI-DI: 00813502011869; Serial Numbers: 492884, 492882, 492883, 522056, 522054, 522055, 522106, 522107, 522110, 522111, 522105, 522109, 520014, 520017, 520018, 520019, 520020, 520021, 520022, 520023, 520747, 520748, 520743, 520744, 520746, 520739, 520740, 520664, 520742, 520660, 520661, 520663, 520666, 520668, 518497, 520659, 520665, 518500, 518501, 518496, 518494, 518499, 518502, 518503, 518963, 518960, 518961, 518962, 518959, 518958, 518955, 518956, 507392, 518954, 507386, 507389, 507390, 507391, 507385, 507384, 507393, 492683, 492682, 492681, 492680, 489923, 489921, 489919, 489918, 489916, 489917, 489915, 492677, 492679, 492674, 509155, 509156, 509157, 509151, 509154, 509149, 491454, 509148, 509150, 491456, 491457, 491458, 491455, 491460, 491461, 491462, 491463, 486724, 486717, 486718, 486721, 476036, 476038, 476034, 476035, 476033, 476032, 476031, 489894, 489892, 489893, 489889, 491433, 491437, 489886, 489887, 489888, 491431, 491432, 491436, 484301, 491435, 484300, 483987, 483986, 483984, 483985, 483979, 483981, 483988, 484302, 484294, 484295, 484296, 484298, 486727, 486725, 486726, 486729, 486728, 486732, 486733, 486734, 469495, 469494, 469492, 469493, 469497, 469496, 469488, 469489, 469490, 469491, 476257, 484772, 472427, 472429, 476254, 484773, 484774, 484775, 472430, 476253, 476255, 476250, 476256, 484769, 484771, 467632, 472081, 476259, 472073, 472074, 476251, 476258, 484767, 472075, 472077, 472076, 472078, 472079, 472432, 484776, 471719, 471724, 476303, 471722, 476307, 471717, 471725, 476304, 476306, 470544A, 475366, 475367, 474979, 475361, 475362, 481841, 479952, 474975, 474977, 474978, 474980, 474972, 476301, 479950, 462945, 479947, 481839, 462942, 462944, 476300, 479948, 481843, 474973, 474974, 481844, 481845, 475363, 462946, 481838, 481840, 479944, 475305, 479949, 472352, 475309, 479945, 479946, 475302, 475306, 475303, 478251, 478252, 478253, 478255, 478258, 475308, 475310, 478260, 475307, 475311, 472350, 472355, 472348, 472351, 472354, 472347, 472349, 472356, 473130, 473126, 473127, 472145, 472146, 473128, 472150, 472152, 464043, 465462, 472353, 465592, 470129, 470135, 473125, 473132, 466807A, 469668, 468237, 469850, 472148, 472149, 472151, 469851, 469857, 472143, 472144, 473129, 473133, 473134, 470546, 470579, 470585, 470542, 470545, 470578, 470580, 472437, 472438, 473122, 470581, 472439, 472441, 473124, 472442, 472445, 473117, 473119, 473116, 473118, 473120, 469520, 473106, 473121, 469518, 470459, 470461, 473107, 470458, 470457, 470547, 473112, 473115, 470540, 470541, 470553, 473110, 473111, 473113, 470548, 470549, 470552, 473109, 472440, 472443, 469519, 470537, 473108, 469515, 470538, 470539, 469513, 469514, 469516, 469517, 469521, 470555, 470377, 470376, 470374, 470372, 460028, 460026, 460022, 460023, 460021, 460025, 470530, 470536, 470531, 470534, 470528, 470529, 470378, 470380, 470381, 460024, 460020, 469854, 469673, 469848, 469849, 469853, 470130, 470131, 459388A, 470132, 470133, 470128, 470137, 469856, 469852, 469556, 469956, 469957, 468233, 469953, 469955, 468231, 468234, 468235, 468239, 468238, 469557, 469949, 469952, 469671, 469950, 469550, 469670, 464554, 464600, 469548, 469549, 469551, 469552, 469554, 469555, 464553, 469669, 469672, 469553, 464048, 464042, 464044, 464045, 464049, 464601, 464602, 464607, 464603, 464606, 464556, 462956, 459565, 462778, 462779, 454319, 454322, 461876, 461877, 461881, 464548, 464549, 464550, 464551, 464552, 464557, 462955, 462957, 462771, 462774, 462775, 462776, 462777, 462950, 462951, 464555, 462770, 462772, 454323, 454326, 465599, 465600, 465598, 465601, 465593, 465594, 465595, 465596, 465597, 463653, 464207, 463655, 464202, 464208, 463654, 464203, 464204, 464205, 464206, 463656, 463657, 463658, 463660, 463662, 463633, 463634, 468019, 463635, 463640, 463641, 468016, 468017, 468018, 468023, 468020, 468022, 463185, 463188, 463189, 463187, 464034, 464035, 464036, 464037, 464038, 463191, 463193, 464031, 464032, 464039, 464040, 463192, 463190, 469667, 466014, 466015, 466016, 468480, 468481, 465459, 465464, 468479, 465456, 465457, 465460, 465461, 465463, 465465, 468484, 468485, 468487, 468488, 468482, 468483, 469660, 469663, 469664, 466007, 466008, 469665, 469666, 466009, 466010, 466011, 465546, 465544, 465547, 465548, 465545, 454679, 461872, 466808, 466812, 466813, 466809, 466811, 466814, 466815, 461774, 466774, 464213, 464214, 464218, 465541, 465542, 465543, 466773, 464216, 464217, 464219, 464220, 461866, 462937, 461864, 462935, 462936, 461865, 461869, 461871, 462939, 466771, 461766, 461768, 462930, 462933, 462938, 466769, 466770, 461771, 461772, 461773, 461778, 461775, 461776, 461777, 461780, 466767, 466768, 461779, 461782, 461758, 461728, 461755, 461757, 461759, 461730, 461754, 461763, 461756, 460044, 460045, 460046, 460049, 460366, 460377, 460042, 460048, 460375, 460367, 460365, 460368, 461727, 461731, 461732, 460378, 460382, 460383, 461729, 460372, 460373, 461119, 461120, 461121, 461124, 461733, 461122, 461123, 461126, 461127, 460030, 460032, 460034, 459666, 460031, 460036, 460037, 460038, 459668, 460040, 459667, 459671, 459672, 459674, 444540A, 459730, 459572, 459573, 459725, 459726, 459727, 459728, 459729, 459566, 459662, 459663, 459664, 459567, 459568, 459570, 459655, 459660, 459661, 459731, 459732, 459733, 459734, 459657, 459658, 459659, 459571, 459363, 459364, 459365, 459366, 459367, 459370, 459381, 459385, 459386, 459382, 459387, 459368, 459369, 459383, 459384, 459389, 459390, 459195, 459193, 459194, 459196, 459190, 459191, 459208, 457733, 459189, 459192, 459207, 457511, 457734, 457509, 457512, 459205, 459206, 457728, 457729, 457731, 457732, 459202, 459203, 459204, 457508, 457513, 457515, 457727, 459361, 459362, 457503, 457505, 457498, 457499, 457506, 457496, 457497, 457500, 456135, 456134, 456136, 456137, 456139, 456140, 456141, 456206, 456132, 456202, 456203, 456207, 456092, 456095, 456097, 456210, 456211, 456801, 456093, 456094, 456208, 456803, 457396, 457400, 457403, 456069, 456098, 456099, 456100, 456800, 457405, 457404, 455614, 455615, 455616, 455618, 456062, 456063, 456065, 456067, 456068, 456064, 455609, 455610, 442270A, 454308, 455574, 455575, 454307, 443794, 454306, 455570, 455571, 455573, 455577, 454299, 454302, 454673, 455569, 454303, 454304, 454674, 454678, 454680, 454681, 454677, 454682, 451712, 451713, 451715, 451716, 453978, 453982, 453983, 451711, 453976, 453977, 454015, 454019, 454021, 454023, 455325, 455328, 455321, 451717, 455320, 455324, 451665, 451667, 451668, 451669, 451671, 448756, 448762, 451666, 452467, 452470, 452472, 452473, 448753, 448754, 448755, 451673, 448661, 451218, 451219, 452558, 452560, 452561, 451212, 451215, 452476, 448656, 448657, 448658, 448660, 448663, 451210, 448654, 450016, 450017, 450363, 450364, 450355, 450357, 450358, 450360, 450361, 450362, 442215, 442217, 442218, 427706A, 446244, 446245, 446246, 446249, 446549, 446242, 446243, 446540, 446545, 446547, 446548, 446240, 446248, 446542, 446241, 446247, 446920, 446921, 446925, 446926, 446927, 446928, 449911, 449915, 449912, 449909, 449913, 449914, 449918, 446395, 446397, 446398, 446404, 445897, 445898, 445906, 445899, 445900, 445905, 445901, 445902, 445903, 445904, 444833, 444831, 444830, 444775, 444777, 445888, 445890, 445893, 445894, 444769, 444796, 444798, 444793, 444795, 444791, 443496, 442498, 442730, 442731, 443497, 443498, 443499, 443502, 440916, 442493, 442494, 443504, 443505, 443796, 444539, 444543, 442500, 442724, 444541, 444544, 444546, 444548, 442725, 442727, 443590, 443631, 443586, 443592, 443594, 443595, 443627, 443628, 443630, 443633, 443635, 444529, 443800, 444530, 444532, 444533, 444534, 443792, 443795, 443797, 443798, 443799, 444535, 444536, 444531, 444537, 444538, 442276, 442274, 442275, 442281, 442279, 442263, 442273, 442267, 442268, 442271, 440909, 440912, 440914, 440907, 440911, 436350, 437774, 440329, 440330, 440331, 440333, 440334, 436342, 436345, 437775, 437776, 437777, 437779, 437780, 436344, 436347, 436348, 434310, 434308, 434309, 427420, 434212, 434219, 434311, 434312, 434313, 434213, 434214, 434215, 434216, 434217, 434218, 434306, 434307, 434211, 434284, 434286, 434287, 434288, 434289, 434290, 434291, 434292, 434293, 427418, 428696, 432885, 432886, 432891, 432888, 432890, 427411, 427412, 428698, 432887, 432892, 427365, 427409, 427410, 428699, 428702, 427360, 427364, 427366, 428693, 428695, 428357, 428359, 428362, 428364, 428355, 428356, 431686, 431687, 431688, 431689, 431690, 431691, 428935, 428936, 428938, 431692, 428939, 428940, 431107, 431110, 431114, 431113, 431115, 431684, 432880, 432881, 432882, 426557, 431685, 432875, 432876, 432877, 432878, 427423, 427425, 427426, 427428, 427709, 427708, 427710, 427419, 427421, 427422, 427427, 427707, 428928, 428931, 428934, 428926, 428933, 427711, 427712, 427713, 428925, 428333, 428334, 428335, 428338, 428341, 428342, 428336, 428337, 416217A, 426564, 426562, 426565, 426558, 426560, 426563, 426556, 422560, 427107, 427106, 427111, 422557, 422544, 422729, 422736, 422548, 427102, 422552, 422728, 422558, 422559, 427109, 427110, 422545, 422547, 422550, 422551, 422734, 422735, 422546, 422549, 422727, 422730, 425774, 425259, 424879, 424881, 424883, 424884, 425258, 425780, 424548, 424885, 425254, 425257, 424546, 424547, 424887, 424888, 425261, 425850, 425854, 425260, 425771, 425779, 425852, 425855, 425857, 425858, 425853, 423325, 423608, 423609, 423865, 424539, 424545, 410463, 423601, 423602, 423866, 421655, 423606, 423870, 423872, 423874, 424540, 423607, 423868, 423871, 423321, 423322, 423326, 423329, 423330, 423867, 423869, 424541, 424542, 424543, 424544, 421657, 421658, 421659, 421660, 421661, 413759A, 417581, 417582, 417583, 417584, 417586, 417588, 417579, 422155, 422156, 422157, 422554, 422154, 422158, 422159, 422160, 422553, 422152, 422556, 422561, 419372, 421652, 421653, 421654, 422142, 422145, 422146, 419368, 419370, 419377, 419373, 419374, 422147, 422149, 422150, 422148, 422143, 422144, 421168, 418886, 421163, 421165, 421169, 420317, 421160, 421167, 418887, 418888, 418889, 420318, 420319, 421161, 421162, 420313, 420314, 419990, 420311, 420312, 420315, 419989, 419983, 419985, 419986, 420310, 420316, 415264, 415266, 417877, 417878, 415257, 415259, 415261, 415262, 415263, 415260, 418446, 418448, 418883, 416214, 416221, 417882, 417884, 417885, 417886, 418884, 418885, 416215, 416216, 416218, 416220, 416223, 418440, 418441, 418444, 418445, 414298, 414299, 413709, 414296, 414301, 414303, 414985, 414987, 414983, 414984, 414992, 414304, 414305, 414988, 414989, 414990, 406483, 406484, 397113, 413701, 413702, 413703, 413704, 413705, 413706, 412898, 412899, 412901, 412894, 412902, 412903, 413707, 412895, 412896, 412897, 399628, 418124, 418125, 418126, 418127, 418128, 418129, 418131, 418132, 403394, 418133, 415228, 415229, 415234, 415235, 410458A, 412261, 413754, 413756, 413757, 413758, 415227, 413755, 415231, 415232, 415233, 415236, 409387, 412266, 412505, 412509, 412511, 413750, 413752, 413753, 413683, 413685, 410454, 413688, 412259, 412263, 410455, 410456, 411139, 412264, 412503, 412504, 412506, 413686, 413687, 413689, 413680, 413681, 413682, 411142, 411144, 411147, 410457, 410459, 411141, 411143, 410461, 410462, 411140, 409935A, 401564, 411138, 401563, 401566, 410094, 410095, 410096, 410099, 410100, 410102, 411092, 410103, 411093, 411094, 411095, 411096, 401759, 411088, 411089, 411090, 401757, 401758, 411091, 401753, 401754, 401760, 401761, 411097, 401755, 401756, 409940, 409395, 401559, 401560, 401561, 409248, 409389, 409390, 409391, 409394, 409936, 409937, 409941, 409938, 408876, 408878, 409393, 409939, 409944, 402230, 408873, 408874, 409245, 409388, 409392, 408871, 408875, 408879, 409942, 409943, 402225, 402227, 408353, 402229, 408354, 409243, 409247, 409249, 409250, 408576, 409105, 402476, 402483, 409102, 409103, 409104, 409106, 409108, 402475, 402481, 408577, 408580, 408583, 408584, 408585, 409107, 402474, 402479, 402480, 402482, 407024, 408158, 408161, 407019, 407021, 407023, 407025, 408156, 408157, 405865, 405866, 405868, 405869, 405870, 407018, 408164, 405863, 405871, 405872, 408162, 408163, 408165, 408357, 408358, 402224, 402231, 402232, 406491, 408356, 408359, 405060, 405061, 405062, 406489, 406486, 406487, 406490, 406483, 406484, 406482, 403467, 405059, 405055, 405056, 405058, 403464, 403465, 403468, 405054, 403073, 403074, 403075, 403079, 403080, 394644, 394645, 394640, 394641, 394643, 401999, 402000, 403461, 403463, 402002, 402004, 402005, 402007, 403071, 403401, 403460, 403394, 403398, 403399, 403400, 403402, 394646, 394649, 403393, 403397, 401079, 401083. |
| FEI Number |
1220648
|
Recalling Firm/ Manufacturer |
Abiomed, Inc.
24 Cherry Hill Dr
Danvers
MA
01923-2575
|
| For Additional Information Contact | Morgaine Johnson 978-646-1400 |
Manufacturer Reason for Recall | Optical Sensors have been damaged due to physical interaction between the inlet and another device resulting in a Placement Signal Not Reliable (PSNR) alarm and loss of Central Venous (CV) Placement Signal, Pulmonary Artery (PA) Placement Signal & Pulmonary Artery Pulse Index (PAPi) metrics. Abiomed is reinforcing that there is a risk of potential interaction between the tip of guidewires, indwelling central venous lines or devices and inlet of the Impella pumps listed above during the insertion, manipulation, and removal of those devices. The interaction may result in optical sensor damage, temporary pump stop, or permanent pump stop.
|
FDA Determined Cause 2 | Under Investigation by firm |
| Action | On December 13, 2024, URGENT MEDICAL DEVICE CORRECTION (NOTIFICATION) letters were sent to customers. RECOMMENDATIONS:
Product is not being removed and hospital inventory may continue to be used.
It is recommended that users lower the pump flow level to P-2 and use imaging guidance when inserting, manipulating, or removing any guidewires, indwelling venous lines or devices to reduce the potential for interaction of the device tip with the pump inlet.
If optical sensor damage occurs and Placement Signal Not Reliable (PSNR) alarm is displayed no additional actions are required. Hemodynamic support and Impella position are not affected and can be monitored as part of routine patient/pump management. In addition, calculated flow and suction monitoring are also unaffected when an optical sensor PSNR alarm is displayed. In the event of a pump stop, follow the displayed prompts of the Impella Stopped alarm.
Damage to the guidewire, indwelling central venous line or other device involved in the unintended interaction should be evaluated, and users should assess if a new device is required.
ACTIONS TO BE TAKEN BY CUSTOMER/USER:
- Product is NOT being removed from the field and does not need to be returned.
- Review, complete all fields, sign, and return the attached business response form (BRF) (refer to Attachment 2) to Abiomed6329@sedgwick.com.
- Forward this notice to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products).
- If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice.
- Post a copy of this notice in a visible area for awareness. As with any medical device, adverse reactions or quality problems experienced with the use of this product should be reported to the FDA s MedWatch Adverse Event Reporting Program as per below instructions:
- Complete and submit the report online: www.fda.gov/medwatch/report.htm or Regula |
| Quantity in Commerce | 1,560 units |
| Distribution | Domestic: Nationwide Distribution
International: Canada |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| PMA Database | PMAs with Product Code = PYX
|
|
|
|