Class 2 Device Recall NexGen CR Complete Knee Solution

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall NexGen CR Complete Knee Solutionsee related information
Date Initiated by FirmFebruary 24, 2012
Date PostedMay 14, 2012
Recall Status1 Terminated 3 on April 18, 2014
Recall NumberZ-1573-2012
Recall Event ID 61468
510(K)NumberK003910 K892800 K933785 K951185 
Product Classification Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
Product00-5972-020-17, NexGen CR Complete Knee Solution, Cruciate Retaining (CR) Articular Surface Regular Constraint, Size Purple, AE Micro 17mm Height , Sterile R, Rx ,Zimmer, Inc. Warsaw, IN The NexGen Complete Knee System ,integrated, with an extensive cruciate retaining, cruciate substituting, and fully constrained component configurations, design specific, conforming surfaces, femoral and tibial augmentation. The femoral, patellar, articular surface, and tibial base plate components should be design specific to the condition and availability of the bone stock, and the condition and functionality of soft tissues. This allows the surgeon to select the most appropriate component combination to meet specific patient needs. Articulating surfaces on micro femoral and tibial components have a smaller bearing spacing than those on the standard and macro components.
Code Information Part 00-5972-020-17, all lots
FEI Number 1000220733
Recalling Firm/
Manufacturer
Zimmer Inc.
345 E Main St
Warsaw IN 46580-2746
For Additional Information Contact
574-372-4807
Manufacturer Reason
for Recall
Zimmer has received 5 complaints where a NexGen CR micro articular surface was used with a standard CR femur, even though the compatibility chart and product labeling indicates that these components are NOT compatible. Investigation identified additional instances where a CR micro articular surface and standard CR femur were sold together on the same purchase order. CR micro articular surfaces are
FDA Determined
Cause 2
Labeling design
ActionZimmer sent an Urgent Correction Notice dated March 5, 2012, to all affected customers via telephone, E-mail and letter notifying users of the appropriate compatibility of micro sizes of NexGen CR and CR-Flex Articular Surfaces and informing them of known cases of incompatible devices being implanted, including associated risks. Customers and surgeons were asked to acknowledge receipt by calling 1-888-912-7349. Questions and concerns or help in notiying accounts about this correction were to be addressed to 1-888-548-8514. For questions regarding this recall call 574-372-4807.
Quantity in Commerce83,322 all sizes
DistributionWorldwide Distribution - USA (nationwide) and the countries of Poland, Taiwan, Denmark, Korea, Singapore, China, Canada, Australia, Japan, France, Italy, UK, Germany, Spain, Belgium, Portugal, Switzerland, Sweden, India, Thailand, Peru, The Netherlands, Malaysia, and Austria.
Total Product Life CycleTPLC Device Report