• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall POWERPORT(R) ClearVUE(R) Slim with Smooth Septum, 8F ChronoFlex(TM), Polyurethane Catheter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall POWERPORT(R) ClearVUE(R) Slim with Smooth Septum, 8F ChronoFlex(TM), Polyurethane Catheter see related information
Date Initiated by Firm February 08, 2018
Create Date May 09, 2018
Recall Status1 Open3, Classified
Recall Number Z-1770-2018
Recall Event ID 79841
510(K)Number K122899  
Product Classification Port & catheter, implanted, subcutaneous, intravascular - Product Code LJT
Product POWERPORT ClearVUE Slim with Smooth Septum, 8F ChronoFlex, Polyurethane Catheter with open suture holes, REF 5678300
Code Information Lot Number Unique Device Identifier (UDI) REAN0061 (01)00801741110818(17)170628(10)REAN0061 REAP0772 (01)00801741110818(17)170728(10)REAP0772 REAT1524 (01)00801741110818(17)171128(10)REAT1524 REAV1479 (01)00801741110818(17)171228(10)REAV1479 REAV1938 (01)00801741110818(17)171028(10)REAV1938 REAW0937 (01)00801741110818(17)171031(10)REAW0937 REAW1186 (01)00801741110818(17)171031(10)REAW1186 REAW1666 (01)00801741110818(17)171031(10)REAW1666 REAY0528 (01)00801741110818(17)171031(10)REAY0528 REAY0795 (01)00801741110818(17)171031(10)REAY0795 REAZ1529 (01)00801741110818(17)171231(10)REAZ1529 REBN0024 (01)00801741110818(17)171231(10)REBN0024 REBN1581 (01)00801741110818(17)180131(10)REBN1581 REBQ0210 (01)00801741110818(17)180131(10)REBQ0210 REBS1277 (01)00801741110818(17)181031(10)REBS1277 REBR1125 (01)00801741110818(17)181031(10)REBR1125 REBS1395 (01)00801741110818(17)181031(10)REBS1395 REBS2412 (01)00801741110818(17)181130(10)REBS2412 REBT1785 (01)00801741110818(17)181231(10)REBT1785 REBT2349 (01)00801741110818(17)181231(10)REBT2349 REBU0563 (01)00801741110818(17)181231(10)REBU0563 REBV0760 (01)00801741110818(17)190228(10)REBV0760 REBW1037 (01)00801741110818(17)190331(10)REBW1037 
Recalling Firm/
Manufacturer
Bard Peripheral Vascular Inc
1625 W 3rd St Bldg 1
Tempe AZ 85281-2438
Manufacturer Reason
for Recall
The Patient ID Card packaged with the POWERPORT ClearVUE Implantable Port with Smooth Septum contains incorrect port flushing instructions.
FDA Determined
Cause 2
Error in labeling
Action The firm initiated the recall by letter on 02/08/2018. The letter explained the problem and the products involved and requested the following actions: "Please check all inventory locations within your institution for affected product code / lot number combinations listed in the field correction notice. If you have further distributed any of the product code / lot numbers, please immediately contact that location, advise them of this field correction notice, and forward these instructions. For any identified product on your shelves, please replace the products Patient ID Card with the corrected Patient ID Card enclosed with this field correction notice. Enclosed with this field correction notice is a Patient Communication Packet that contains one patient letter, one corrected Patient ID Card, and one POWERPORT¿ Patient Guide. Please contact all patients that have received any product listed in this field correction notice and provide them with the Patient Communication Packet."
Quantity in Commerce 1411 units
Distribution Nationwide
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database 510(K)s with Product Code = LJT and Original Applicant = C.R. BARD, INC.
-
-