 Class 2 Device Recall SIGNA Premiersee related information
Date Initiated by FirmDecember 23, 2019
Date PostedJanuary 28, 2020
Recall Status1 Terminated 3 on December 11, 2020
Recall NumberZ-0874-2020
Recall Event ID 84655
510(K)NumberK171128 
Product Classification System, nuclear magnetic resonance imaging - Product Code LNH
ProductSIGNA Premier Nuclear Magnetic Resonance Imaging System - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
Code Information Mfg. Lot or Serial# System ID 33-4-1031 415723SHMR19 To be provided RTD01204 To be provided DPN88672 
Recalling Firm/
Manufacturer
GE Healthcare, LLC
3000 N Grandview Blvd
Waukesha WI 53188-1615
For Additional Information Contact
800-437-1171
Manufacturer Reason
for Recall
It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system's date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images.
FDA Determined
Cause 2
Software design
ActionThis issue will be resolved by sending an Urgent Medical Device Correction letter to Medical Facility staff (Director of Clinical/Radiology, Risk Manager/Hospital Administrator, Director of Biomedical Engineering) with affected systems. This letter will inform the customer of the issue and instruct the user to ensure the displayed system date is correct. The long-term solution is to correct affected systems with updated software.
Quantity in Commerce62 (30 US, 32 OUS) total devices
DistributionWorldwide distribution - US nationwide distribution and countries of Australia, China, France, Germany, India, Italy, Japan, Netherlands, Oman, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = LNH
-
-