• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Premisa

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Premisasee related information
Date Initiated by FirmSeptember 05, 2008
Date PostedNovember 25, 2008
Recall Status1 Terminated 3 on March 07, 2012
Recall NumberZ-0260-2009
Recall Event ID 49506
510(K)NumberK060472 
Product Classification dental composite - Product Code EBF
ProductPREMISE, PART NO. 32814, 10 PACK UNIDOSE PREMISA A3.5 (INTERNATIONAL ONLY), dental composite
Code Information LOT NUMBER(S): 3001322, and 3018914
FEI Number 2024312
Recalling Firm/
Manufacturer
Kerr Corp
1717 W Collins Ave
Orange CA 92867-5422
Manufacturer Reason
for Recall
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
FDA Determined
Cause 2
Component design/selection
ActionKerr sent an Urgent Medical Device Recall letter to consignees via USPS first class mail the week of September 8, 2008, requesting they return the product for replacement or credit. The recall notification requested the return of the affected products because the material appears to stiffen and become difficult to extrude over time. Customers were instructed if they have any of the affected lots of product in their inventory to immediately return it to Kerr. Any product returned will be replaced at no charge or credited to your account. In addition, customers were instructed to fax the attached Recall Return Form to 714-516-7739. If the customers were an authorized Kerr distributor, they were requested to identify and recover the affected product lots listed on the attached Lot Number Table that may have been shipped to their customers. Also, customers were instructed to contact Kerr Customer Care at 1 -800-537-7123 directly to handle the arrangements of a quick return and replacement
Quantity in Commerce22,478 packs for all affected lots
DistributionWorldwide: USA, Australia, Canada, Hong Kong, Malaysia, Mexico, and Singapore.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = EBF
-
-