| | Class 1 Device Recall Spectrum Central Venous Tray Minocycline/Rifampin Impregnated Triple Lumen Polyurethane Catheter |  |
| Date Initiated by Firm | October 04, 2010 |
| Date Posted | February 23, 2011 |
| Recall Status1 |
Terminated 3 on June 21, 2011 |
| Recall Number | Z-1201-2011 |
| Recall Event ID |
57663 |
| 510(K)Number | K081113 |
| Product Classification |
Peripherally Inserted, Central Venous Catheter Tray - Product Code FOZ
|
| Product | Spectrum Central Venous Tray Minocycline/Rifampin Impregnated Triple Lumen Polyurethane Catheter, 5.0 Fr/ 8 cm, C-UTLMYJC-501J-PED-ABRM-HC-IHI-FST, G51264, Sterile, Cook Medical, Cook Incorporated, Bloomington, IN. |
| Code Information |
Part number: C-UTLMYJC-501J-PED-ABRM-HC-IHI-FST, Global part number G51264. Lot number: F2365655 |
| FEI Number |
1820334
|
Recalling Firm/ Manufacturer |
Cook, Inc. 750 Daniels Way Bloomington IN 47404-9120
|
| For Additional Information Contact | Rita A. Harden 812-339-2235 |
Manufacturer Reason for Recall | Cook, Inc. is recalling Single, Double, Triple, and Five-Lumen Central Venous Catheter Trays AND Single and Double Lumen PICC Peripherally Inserted Central Venous Catheter Trays. These kits and trays contain 0.9% Sodium Chloride Injection USP 5 mL fill in 6 mL flush syringes that have been recalled by Excelsior Medical. During a routine test for syringe sterility, a possible integrity failure wa |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Cook, Inc. sent URGENT DEVICE RECALLS letters to its customers dated 10/4/2010. The letters asked customers to examine their inventory and notify Cook for product return. The letter included a response form. |
| Quantity in Commerce | 51,473 total trays |
| Distribution | Worldwide distribution: USA, Chile, Colombia, Panama, Barbados, Canada, Guatemala, and Saudi Arabia. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = FOZ
|
|
|
|