• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall Transforaminal Lumbar Interbody Fusion (TLIF) pack

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Transforaminal Lumbar Interbody Fusion (TLIF) packsee related information
Date Initiated by FirmOctober 08, 2014
Date PostedDecember 16, 2014
Recall Status1 Terminated 3 on August 14, 2017
Recall NumberZ-0367-2015
Recall Event ID 69448
Product Classification Orthopedic Tray - Product Code OJH
ProductTransforaminal Lumbar Interbody Fusion (TLIF) pack, Catalog # 900-2918 Used by medical staff to perform patient care procedures such as wound healing, suture removals, irrigate wounds, etc.
Code Information Lot numbers 140312147 140613795 140714523 
FEI Number 2648727
Recalling Firm/
Manufacturer
Customed, Inc
Calle Igualdad Final #7
Fajardo PR 00738
For Additional Information Contact
787-622-5151
Manufacturer Reason
for Recall
Significant compromised sterility, not limited to potentially damaged packaging. Serious deficiencies in the manufacturing processes and uncontrolled and inadequate storage conditions increase the risk of contamination and may result in patient infection.
FDA Determined
Cause 2
Packaging
ActionCustomed initiated a recall on October 8th, 2014 of sterile convenience surgical Packs-trays, bags (known as Convenience Packs) and new lots of products added to the initial June 3, 2014 recall of 233 sterile surgical packs. The products are being recalled because of significant compromised sterility, not limited to potentially damaged packaging. Customers were instructed to quarantine and cease use of the affected product identified. Customers are to return the response form included with the letter, to the contact information provided. Upon receipt, the firm will contact customers with further instructions to obtain credit, replacement product, and for return of the recalled product or any other disposition action. If customers have any questions, they should contact the Customed Recall Coordinator at 407-850-5558 extensions 8001 or 8005, or via e-mail: pablo.diaz@customedhealing.com.
Quantity in Commerce51
DistributionUS Distribution to Florida, New York and Puerto Rico..
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-